Bonding of light cured composites and acid modified composites (compomers) to tooth structure. Bonding of dual cured luting and core build up composites to tooth structure as long as these materials are light cured. Intraoral repairs of porcelain fused to metal crowns and composite veneer crowns with metal backing. Intraoral repairs of all ceramic crowns (except zirconia and alumina), hybrid resin jacket crowns, CAD/CAM hybrid resin crowns and composites. Intraoral repairs of porcelain fused to zirconia crowns, porcelain fused to alumina crowns and full zirconia crowns. Treatment of hypersensitive teeth.
Device Story
G-Premio BOND is a one-component, light-cured dental adhesive; supplied in 5 mL bottles or unit doses. Applied by dental professionals to tooth structure or restorative surfaces (metals, ceramics, composites); functions via chemical and mechanical bonding through polymerization of methacrylate ester monomers. Used for bonding restorative materials and treating tooth hypersensitivity. Output is a cured adhesive film providing structural bond; clinical benefit includes restoration retention and reduction of dentin hypersensitivity.
Clinical Evidence
Bench testing only. Performance evaluated against company specification AB-15-Q-301-631(1). Metrics included bond strength (>10MPa for enamel, dentin, metal, porcelain, composite, and zirconia), appearance, curing property, application characteristics, and SEM observation of dentin tubule sealing. Results demonstrated equivalence to predicate device.
Technological Characteristics
One-component, light-cured resin bonding agent. Composition includes methacrylate ester monomers. Polymerization-based bonding mechanism. Form factor: 5 mL liquid bottle or unit dose. Shelf life: 2 years. Storage: 1-25°C. No software or electronic components.
Indications for Use
Indicated for patients requiring dental restorative procedures, including bonding of light-cured/dual-cured composites to tooth structure, intraoral repair of various crown/veneer types (metal, ceramic, hybrid resin), and treatment of hypersensitive teeth.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with a flowing, wave-like design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 20, 2015
GC America Inc. Mark Heiss, D.D.S Director Regulatory Affairs 3737 W. 127th Street Alsip. IL 60803
Re: K143140
Trade/Device Name: G-Premio BOND Regulation Number: 21 CFR 872.3200 Regulation Name: Resin, tooth bonding agent Regulatory Class: II Product Code: KLE Dated: December 15, 2014 Received: January 20, 2015
Dear Dr. Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Section 4 - Indications for Use Statement
## Indications for Use
510(k) Number (if known): K143140
Device Name: G-Premio BOND
Indications for Use:
- 1. Bonding of light cured composites and acid modified composites (compomers) to tooth structure.
- Bonding of dual cured luting and core build up composites to tooth structure as long as these materials 2. are light cured.
- 3. Intraoral repairs of porcelain fused to metal crowns and composite veneer crowns with metal backing.
- Intraoral repairs of all ceramic crowns (except zirconia and alumina), hybrid resin jacket crowns, 4. CAD/CAM hybrid resin crowns and composites.
- 5. Intraoral repairs of porcelain fused to zirconia crowns, porcelain fused to alumina crowns and full zirconia crowns.
- Treatment of hypersensitive teeth. 6.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
- 1. Submitter Information:
GC America Inc. 3737 W. 127th Street Alsip, IL 60803
| Contact Person: | Mark Heiss, D.D.S. |
|-----------------|--------------------|
| Phone: | (708) 926-3090 |
| Fax: | (708) 926-9100 |
Date Prepared: October 30, 2014
#### 2 Device Name:
| 510(k) Number: | K143140 |
|------------------------|---------------------------|
| Proprietary Name: | G-Premio BOND |
| Classification Name: | Resin tooth bonding agent |
| Device Classification: | Class II, 872.3200 |
| Product Code: | KLE |
#### Predicate Devices: 1.
| Company | Device | 510(k) No. | Date Cleared |
|-----------------|--------------|------------|--------------|
| GC America Inc. | G-ænial Bond | K082768 | 2008/10/21 |
#### 2. Description of Device:
G-Premio BOND is a one component, light-cured bonding agent available in a 5 mL liquid bottle or unit dose.
#### 3. Indications for Use:
- Bonding of light cured composites and acid modified composites (compomers) to tooth a. structure.
- Bonding of dual cured luting and core build up composites to tooth structure as long as these b. materials are light cured.
- Intraoral repairs of porcelain fused to metal crowns and composite veneer crowns with metal C. backing.
- d. Intraoral repairs of all ceramic crowns (except zirconia and alumina), hybrid resin jacket crowns, CAD/CAM hybrid resin crowns and composites.
- Intraoral repairs of porcelain fused to zirconia crowns, porcelain fused to alumina crowns and e. full zirconia crowns.
- Treatment of hypersensitive teeth. f.
- 4. Technological characteristics:
All the compounds found in the applicant device, G-Premio BOND, have already been used in the predicate devices. The bonding mechanism of the predicate device is applying it to tooth structure, then bonding chemically and mechanically by polymerization of uncured methacrylate ester monomers. The chemical composition of the subject device is such that it can also bond to tooth surfaces (like the predicate) as well as composites, hybrid ceramics, and metals.
- న్. Substantial equivalence:
The subject and predicate devices contain the same compounds, have similar indications, and are both light cured.
{5}------------------------------------------------
| | Applicant device | Comparative device |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Product category | Resin tooth bonding agent, Class II | Resin tooth bonding agent, Class II |
| Trade name | G-Premio BOND | G-ænial Bond |
| Manufacturer | GC Corporation | GC Corporation |
| Intended use | For bonding light-cured composite resin to tooth structure, hybrid ceramics, composites, ceramics (porcelain, lithium disilicate, zirconia and alumina) and metals (precious and non-precious metals) surfaces. For treatment of hypersensitive teeth. | For bonding light-cured composite resin to tooth structure. |
| Product description | G-Premio BOND is a one component, light-cured bonding agent available in a 5 mL liquid bottle or unit dose. | G-aenial Bond is a one component, light-cured bonding agent that allows bonding light-cured composite resins to tooth structure. |
The applicant device complies with all the requirements of Company specification AB-15-Q-301-631(1) (see table below).
| | Property | Requirements |
|---|----------------------------------------------|---------------------------------------------------------------------|
| 1 | Appearance | Should be homogenous and free from foreign<br>matters |
| 2 | Curing property | Should be cured and formed film |
| 3 | Bond strength to<br>tooth structure | >10MPa for enamel<br>>10MPa for dentin |
| 4 | Bond strength to<br>metal | >10MPa for precious metal<br>>10MPa for non-precious metal |
| 5 | Bond strength to<br>ceramic and<br>composite | >10MPa for porcelain<br>>10MPa for composite<br>>10MPa for zirconia |
| 6 | Application<br>characteristics | Should be formed even and homogenous<br>coat. |
| 7 | Sealing property<br>of dentin tubules | Should be sealed dentin tubules when<br>observed using SEM |
### 6. Differences
The following differences may be noted between the predicate device and G-Premio BOND.
- G-Premio BOND is available for bonding light-cured composite resin to composites, hybrid ● ceramics, ceramics and metals surfaces as well as tooth structure.
- . G-Premio BOND is available for treatment of hypersensitive teeth.
The specifications that are important to determine substantial equivalency of a dental adhesive system are bond strengths associated to enamel and dentin (tooth structure). The bond strengths as noted above compare the predicate device (G-aenial Bond, K082768) to the subject device. In both enamel and dentin bonding, the results were equivalent.
{6}------------------------------------------------
In addition, bond strength of subject device to other substrates was equivalent to the predicate bond strengths to enamel and dentin. SEM images of sealed bovine dentin tubules also support indications for treatment of hypersensitive teeth.
Therefore, the subject device has been shown to be substantially equivalent to the predicate device.
#### 7. Performance Bench Tests
It is confirmed that the device conforms to the required specifications and is suitable for its intended use. Performance testing includes:
- · Appearance
- · Curing property
- · Application characteristics
- Color
- pH
- Filler loading
- · Refraction index
#### 8. Packaging
- G-Premio BOND Bottle:
- 5 mL liquid (1), disposable dispensing dish (20), disposable applicator (50) .
- G-Premio BOND Unit Dose:
- 0.1mL liquid (50), disposable applicator (50) .
- 9. Shelf Life Evaluation and Storage Conditions:
- Shelf Life 2 years
- Recommended for optimal performance, store in a cool and dark place. 1-25°C (33.8 77.0°F)
## 10. Conclusion
Based on similarities in intended use, mode of action, chemical composition, and performance testing, G-Premio Bond is substantially equivalent to the selected predicate (K082768).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.