K142824 · Hairlabs International, Inc. · OAP · Apr 14, 2015 · Physical Medicine
Device Facts
Record ID
K142824
Device Name
LX-100 Hair Growth Stimulation System
Applicant
Hairlabs International, Inc.
Product Code
OAP · Physical Medicine
Decision Date
Apr 14, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LX-100 is a non-heating lamp as described under the provisions of 21 CFR §890.5500 and is indicated for: Adjunctive use for the treatment of androgenic alopecia in females; the LX-100 is indicated to promote hair growth of females with androgenetic alopecia who have Ludwig and Savin Hair Loss Scale classifications of I to II and who have been determined to have a Fitzpatrick Skin Typing of I to IV.
Device Story
LX-100 is a stationary, spherical laser helmet for hair growth stimulation; uses 90 lasers at 650nm wavelength to provide biostimulation to the scalp. Operated by a physician or trained technician in a clinical setting; requires prescription. Device features computer-controlled treatment protocols, including adjustable time and dosage settings. Safety features include key lock, warning labels, on-screen prompts for safety glasses, and beam interruption sensors that disable lasers if the patient moves or tilts their head. Output is continuous wave laser light. Benefits include non-invasive promotion of hair growth for specified alopecia patients.
Clinical Evidence
Bench testing only. The device met all design verification and validation criteria. Equivalence is based on technological characteristics (wavelength, power, dosage) and comparison to previously cleared devices (K091496, K122950) with established clinical efficacy for the same indications.
Technological Characteristics
Stationary laser helmet; 90 lasers; 650nm (±1%) wavelength; ≤4.5mW per diode; 3V DC power; injection-molded polycarbonate/polystyrene and fiberglass construction; two-part epoxy paint. Features computer-controlled treatment timing and dosage. Safety: key lock, beam interruption sensors, head proximity circuitry. No sterilization required.
Indications for Use
Indicated for females with androgenetic alopecia, Ludwig and Savin scale I-II, and Fitzpatrick skin types I-IV, to promote hair growth as an adjunctive treatment.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting movement or connection. The profiles are arranged in a way that creates a sense of depth and unity. Encircling the image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2015
HairLabs International Incorporated Mr. James Britt President 742 Harpeth Knoll Road Nashville, Tennessee 37221
Re: K142824
Trade Name: LX-100 Hair Growth Stimulation System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP, NHN Dated: September 25, 2014 Received: September 30, 2014
Dear Mr. Britt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Concurrence & Template History Page [THIS PAGE IS INCLUDED IN IMAGE COPY ONLY]
Full Submission Number:
For Office of Compliance Contact Information:
http://insideportlets.fda.gov:9010/portal/page?_pageid=197,415881&_dad=portal&_schema=PORTAL&org=318
For Office of Surveillance and Biometrics Contact Information:
http://insideportlets.fda.gov:9010/portal/page? pageid=197,415881& dad=portal& schema=PORTAL&org=423
| Digital Signature Concurrence Table | |
|-------------------------------------|--|
| Reviewer Sign-Off | |
| Branch Chief Sign-Off | |
| Division Sign-Off | |
Template Name: K1(A) – SE after 1996
Template History:
| Date of Update | By | Description of Update |
|----------------|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 7/27/09 | Brandi Stuart | Added Updates to Boiler Table |
| 8/7/09 | Brandi Stuart | Updated HFZ Table |
| 1/11/10 | Diane Garcia | Liability/Warranty sentence added at bottom of 1st page |
| 10/4/11 | M. McCabe Janicki | Removed IFU sheet and placed in Forms |
| 9/25/12 | Edwena Jones | Added digital signature format |
| 12/12/12 | M. McCabe Janicki | Added an extra line between letter signature block and the word<br>"Enclosure". Also, added a missing digit in 4-digit extension on<br>letterhead zip code: "002" should be "0002". |
| 4/2/2013 | M. McCabe Janicki | Edited sentence that starts "If you desire specific advice for your<br>device on our labeling regulation (21 CFR Part 801)." Replaced<br>broken Compliance link with general link to DSMICA. |
| 4/12/2013 | Margaret McCabe<br>Janicki | Fixed a typo: Paragraph 1, final sentence, "We remind you,<br>however; that device labeling must be truthful." Replaced<br>incorrect semicolon with a comma. |
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142824
### Device Name
LX-100 Hair Growth Stimulation System
Indications for Use (Describe)
The LX-100 is a non-heating lamp as described under the provisions of 21 CFR §890.5500 and is indicated for:
Adjunctive use for the treatment of androgenic alopecia in females; the LX-100 is indicated to promote hair growth of females with androgenetic alopecia who have Ludwig and Savin Hair Loss Scale classifications of I to II and who have been determined to have a Fitzpatrick Skin Typing of I to IV.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# HairLabs International Inc.
# 510(k) Summary
(as required per 21 CFR: §807.92)
## LX-100 Hair Growth Stimulation System
- l. Applicant: Hairlabs International, Incorporated 742 Harpeth Knoll Road Nashville, TN 37221
Phone: (615)-646-9537
- II. Contact Name: James Britt, President james@hairlabs.com
### III. Device Name:
| Proprietary Name | LX-100 Hair Growth Stimulation System |
|---------------------|---------------------------------------|
| Common/Usual Name | Light Therapy Hair System |
| Classification Name | Infrared Lamp per 21CFR 890.5500 |
Product Code(s)............................................................... OAP
### IV. Predicate Devices And Comparison
| Predicate Device<br>Comparative Item | Hairlabs<br>International<br>LX-100 | Midwest RF<br>MEP-90 | Theradome Inc.<br>LH80 Pro |
|-----------------------------------------|---------------------------------------------|---------------------------------------------|----------------------------|
| 501(k) Number | K142824 | K091496 | K122950 |
| Device Name | LX-100 Hair<br>Growth Stimulation<br>System | MEP-90 Hair<br>Growth Stimulation<br>System | Laser Helmet LH80<br>PRO |
| Manufacture | Hairlabs<br>International, Inc. | Midwest RF, LLC | Theradome, Inc. |
| Establishment<br>Registration<br>Number | N/A | 2134565 | Unknown |
| Device<br>Regulation<br>Description | Infrared Lamp | Infrared Lamp | Infrared Lamp |
| Device<br>Regulation<br>Number | 21CFR; §890.5500 | 21CFR; §890.5500 | 21CFR; §890.5500 |
| Device | A device that emits | A device that emits | A device that emits |
{5}------------------------------------------------
| Regulation<br>Identification/<br>Classification | energy at infrared<br>frequencies<br>(approximately 700<br>nanometers to<br>50,000 nanometers<br>to provide topical<br>heating. | energy at infrared<br>frequencies<br>(approximately 700<br>nanometers to<br>50,000 nanometers<br>to provide topical<br>heating. | energy at infrared<br>frequencies<br>(approximately 700<br>nanometers to<br>50,000 nanometers<br>to provide topical<br>heating. |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Physical State | Light emitting<br>stimulator | Light emitting<br>stimulator | Light emitting<br>stimulator |
| Product<br>Nomenclature | Lamp, infrared | Lamp, infrared | Lamp, infrared |
| Product Code | OAP | OAP | OAP |
| Device Class | Class 2 | Class 2 | Class 2 |
| 21CFR Part 1010<br>Laser<br>Classification/<br>Compliance | Illa; IIIr<br>Full Compliance | Illa; IIIr<br>Full Compliance | Illa; IIIr |
| FDA Device Risk<br>Classification | Non-significant<br>(NSR) | Non-significant<br>(NSR) | Non-significant<br>(NSR) |
| Wavelength (λ) | 650nm (+/-1%)<br>644nm to 656nm<br>Measured | 650nm (+/-1%)<br>644nm to 656nm<br>Measured | 678nm center<br>wavelength<br>Published |
| Output Per Diode<br>in mw | ≤4.5mW Measured | ≤4.5mW Measured | ≤5mW<br>Unpublished |
| Treatment Area in<br>cm2 | 1065.10 cm2<br>Mathematically<br>derived | 990.80 cm2<br>Mathematically<br>derived | 990.80 cm2<br>Mathematically<br>derived |
| Total Output for<br>Treatment Area,<br>Dosage mW/cm2 | .3802 mW/cm2<br>Mathematically<br>Derived | .3724 mW/cm2<br>Mathematically<br>Derived | .4037 mW/cm2<br>Mathematically<br>Derived |
| Recommended<br>Treatment<br>Regimen | 2 - 20 Minute<br>Treatments Per<br>Week, On Non-<br>consecutive Days<br>for 26 Weeks for a<br>Total of 52<br>Treatments | 2 - 20 Minute<br>Treatments Per<br>Week, On Non-<br>consecutive Days<br>for 18 Weeks for a<br>Total of 36<br>Treatments | 2 - 20 Minute<br>Treatments Per<br>Week, On Non-<br>consecutive Days<br>for 26 Weeks for a<br>Total of 52<br>Treatments |
| Fluence, 20<br>Minute Treatment<br>J/cm2 | .45629 J/cm2<br>Mathematically<br>derived | .44691 J/cm2<br>Mathematically<br>derived | .48446 J/cm2<br>Mathematically<br>Derived |
| Number of Lasers | 90 | 82 | 80 |
| Laser Pulse Rate | Continuous | Continuous | Continuous |
| Laser Pulse<br>Duration | Continuous | Continuous | Continuous |
| Power | 3Volts DC;<br>110 Volts AC<br>converted to<br>24 Volts DC | 3 Volts DC:<br>110 Volts AC<br>converted to<br>24 Volts DC | Unknown<br>proprietary |
{6}------------------------------------------------
| Aiming Beam | No lens; diffused beam, fixed coverage | No lens; diffused beam, fixed coverage | No lens, diffused beam, fixed coverage |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Laser Beam Scattering Outside Aperture | None - Fixed angulation and required beam interruption prevent beam scattering outside of Helmet assembly | None - Fixed angulation and required beam interruption prevent beam scattering outside of Helmet assembly | None when patient is in helmet. |
| Output Mode | Direct Light | Direct Light | Direct Light |
| Sterilization | Basic cleaning, Instructions provided; No sterilization claimed, called for or possible | Basic cleaning, Instructions provided; No sterilization claimed, called for or possible | No sterilization claimed or called for in published materials |
| Device Type | Free standing, movable laser helmut, spherical laser helmet | Free standing, movable laser helmet, spherical laser helmet | Portable helmet only, sits on patients head, spherical laser helmet |
| Accessories | None; all items described are necessary for basic operation including safety keys for key lock, 10' medical grade power cord, operation manual and 2 pair laser safety glasses | None; all items described are necessary for basic operation including mouse, keyboard, monitor, safety keys for key lock, 10' medical grade power cord, operation manual and 2 pair laser safety glasses | None; all items described are necessary for basic operation including wall charger |
{7}------------------------------------------------
| Materials | Injection molded<br>and painted<br>polycarbonate and<br>polystyrene,<br>Molded and<br>painted fiberglass.<br>All paint is two part<br>epoxy based. | Injection molded<br>and painted<br>polycarbonate and<br>polystyrene,<br>Molded and<br>painted fiberglass.<br>All paint is two part<br>epoxy based. | Unkown and would<br>be proprietary<br>information under<br>18U.S.C.§1832-<br>Assumed no<br>issues due to<br>issuance of 510(k) | Unknown and<br>would be<br>proprietary<br>information under<br>18U.S.C§1832-<br>Assumed no<br>issues due to<br>issuance of 510(k) |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | There are no new<br>materials in use on<br>this device that are<br>not in routine use<br>on other devices.<br>The<br>biocompatibility is<br>comparable to any<br>of the legally<br>marketed devices<br>listed as follows:<br>Midwest RF<br>K091496<br>Model MEP-90<br>Sunetics<br>International<br>K132646<br>Model Clinical<br>Laser G<br>Theradome, Inc<br>K122950<br>Model LH80 PRO | Assumed no<br>issues due to<br>issuance of 510(k) | Assumed no<br>issues due to<br>issuance of 510(k) | Assumed no<br>issues due to<br>issuance of 510(k) |
| Indications of Use | LX-100 is for<br>adjunctive use for<br>the treatment of<br>androgenic<br>(androgenetic)<br>alopecia in females<br>and is indicated to<br>promote hair<br>growth of females<br>with androgenic<br>(androgenetic)<br>alopecia who have<br>Ludwig and Savin<br>Hair Loss Scale<br>classifications of I<br>to II and who have<br>been determined to<br>have a Fitzpatrick<br>Skin Typing of I to<br>IV | MEP-90 is for<br>adjunctive use for<br>the treatment of<br>androgenic<br>(androgenetic)<br>alopecia in females<br>and is indicated to<br>promote hair<br>growth of females<br>with androgenic<br>(androgenetic)<br>alopecia who have<br>Ludwig and Savin<br>Hair Loss Scale<br>classifications of I<br>to II and who have<br>been determined to<br>have a Fitzpatrick<br>Skin Typing of I to<br>IV | Laser Helmet LH80<br>PRO is an over the<br>counter (OTC)<br>device indicated to<br>treat androgenic<br>alopecia, to<br>promote hair<br>growth in females<br>with female pattern<br>hair loss (FPHL) on<br>the Ludwig and<br>Savin Hair Loss<br>Scale I-II,<br>Fitzpatrick Skin-<br>Types I to IV | |
| Indications of Use<br>Source | Prescription | Prescription | Over The Counter | |
| Indications of Use<br>Sale and Usage<br>Restrictions | Direction of<br>Licensed Physician<br>only | Direction of<br>Licensed Physician<br>only | Over The Counter | |
| Indications of Use<br>Installation | Certified on site | Certified on site | Drop shipped to<br>user | |
| Indications of Use<br>Operator and<br>User Training /<br>Education | Factory and/or on<br>site training at<br>installation;<br>Internet access for<br>Operations Manual | Factory and/or on<br>site training (User<br>option) of<br>approximately 6-8<br>hours at<br>installation;<br>Internet access for<br>operational<br>updates; Operation<br>Manual; Continuing<br>education program<br>TBD | Operation manual | |
| Indications of Use<br>Operation Control<br>and Length of<br>Treatment | Default settings for<br>recommended<br>treatment protocol;<br>Operator resets of<br>time and dosage<br>which is controlled<br>by computer | Default settings for<br>recommended<br>treatment protocol;<br>Operator resets of<br>time and dosage<br>which is controlled<br>by computer | Single button<br>control, button<br>push starts one 20<br>minute cycle, unit<br>turns off when<br>cycle ends | |
| Indications of Use<br>Safety In<br>Operation | Warning labels on<br>device; key lock<br>with on screen<br>warning; default<br>treatment settings;<br>fixed maximum<br>power output<br>regardless of<br>settings; constant<br>thru beam interrupt<br>by patient required<br>for laser operation;<br>warning on screen<br>to insure operator<br>and patient are<br>wearing safty<br>glasses before<br>lasers will operate;<br>no beam scatter<br>outside of helmet<br>assembly; tilting of<br>head no ≥3/8"<br>interrupt | Warning labels on<br>device; key lock<br>with on screen<br>warning; default<br>treatment settings;<br>fixed maximum<br>power output<br>regardless of<br>settings; constant<br>thru beam interrupt<br>by patient required<br>for laser operation;<br>warning on screen<br>to insure operator<br>and patient are<br>wearing safty<br>glasses before<br>lasers will operate;<br>no beam scatter<br>outside of helmet<br>assembly; tilting of<br>head no ≥3/8"<br>interrupt; head<br>proximity safety<br>circuitry | Unknown -<br>proprietary | |
{8}------------------------------------------------
{9}------------------------------------------------
### V. Indications For Use
The LX-100 is a non-heating lamp as described under the provisions of 21 CFR §890.5500 and is indicated for:
Adjunctive use for the treatment of androgenic alopecia in females; the LX-100 is indicated to promote hair growth of females with androgenetic alopecia who have Ludwig and Savin Hair Loss Scale classifications of I to II and who have been determined to have a Fitzpatrick Skin Typing of I to IV
### VI. Technological Characteristics
The LX-100 is a stationary low-level laser device that promotes hair growth and provides adjunctive treatment for androgenic (androgenetic) alopecia in females. The device provides automated and timed equal distribution of laser light to 100% of the scalp.
The LX-100 operation is controlled by an operating system that affords the user maximum flexibility for individual treatments. The device applies a measured very high tolerance (≤+1%) wavelength (λ) to the scalp stimulating hair growth by the proven concept of biostimulation.
{10}------------------------------------------------
## Performance Data and Clinical Efficacy
The LX-100 Hair Growth Stimulation System met all of the specified acceptance criteria of the design verification and validation testing performed under design control and is found to be substantially equivalent to the currently marketed and approved predicate devices.
#### VII. Conclusion
The LX-100 is substantially equivalent to other pulsed therapeutic light therapy svstems currently in commercial distribution. The LX-100 has the same intended use and technological and safety characteristics to the predicate devices approved for commercial distribution under 510(k) number K091496, and K122950.
It equals the clinically accepted therapeutic results standards of FDA 510(k) K091496, and K122950 previously approved light therapy systems.
The technological equivalence to the predicate devices is substantiated by the wavelength, power output and dosage generated by the LX-100. The LX-100 provides expanded treatment benefits and regimens for clinical presentations already approved by the Food and Drug Administration for the predicate device.
The LX-100 is as safe and effective as a combination of the predicate devices listed and numerous others. It has the same intended use of affecting hair growth as the hair growth predicate device (K091496, and K122950). In addition, the LX-100 has the same general indications, i.e., treating androgenic alopecia, and the same specific indication of promoting hair growth as the predicate device.
The LX-100 also has many of the same or similar technological characteristics as a combination of its predicate devices. These include multiple lasers and visible laser wavelength.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.