K091496 · Midwest Rf, LLC · OAP · Feb 23, 2010 · Physical Medicine
Device Facts
Record ID
K091496
Device Name
MEP-90 HAIR GROWTH STIMULATION SYSTEM
Applicant
Midwest Rf, LLC
Product Code
OAP · Physical Medicine
Decision Date
Feb 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MEP-90 is a non-heating lamp as described under the provisions of 21 CFR §890.5500 and is indicated for: Medically prescribed use for the treatment of androgenic alopecia in females: The treatment of androgenic alopecia in females by promoting hair growth of females with androgenetic alopecia who have Ludwig and Savin Hair Loss Scale classifications of I to II and who have been determined to have a Fitzpatrick Skin Typing of I to IV.
Device Story
MEP-90 is a stationary low-level laser device for hair growth stimulation in females with androgenic alopecia. Device delivers automated, timed, equal distribution of laser light to 100% of the scalp. Operated by clinicians in a medical setting; device uses biostimulation via measured laser wavelength to promote hair growth. System includes an operating system allowing treatment flexibility. Patients undergo a regimen of 36, 20-minute treatments over 18 weeks. Clinical outcomes include increased hair count and stabilization of hair loss rates. Benefits include improved hair density and reduced visible alopecia areas without reported adverse events.
Clinical Evidence
Multi-phased IRB-approved study; N=subject population (size not specified). Subjects received 36, 20-minute treatments over 18 weeks. Results at 10 weeks: 92% had ≥10% hair count increase; 98% showed stabilized hair loss. Results at 18 weeks: 97% had ≥20% hair count increase; 89% had ≥30% increase; 57% had ≥50% increase. 65% reported smaller bald spots. No adverse events reported.
Technological Characteristics
Stationary low-level laser device; multiple lasers; visible laser wavelength; wavelength tolerance ≤±0.76%; automated/timed distribution system; operating system control.
Indications for Use
Indicated for females with androgenic alopecia, Ludwig and Savin scale I-II, and Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
idwestrf
as required per 21CFR; §807.92)
FEB 2 8 2010
# MEP-90 Hair Growth Stimulation System
| I. Applicant | Midwest RF LLC |
|--------------|-------------------------|
| | 1050 Walnut Ridge Drive |
| | Hartland, WI 53029 USA |
Phone: (262) 867-8254 Fax: (262) 867-8554
| II. Contact Name | Helmut Keidl, President |
|------------------|-----------------------------|
| | helmut@midwestcomposite.com |
# III. Device Name
| Proprietary Name................................................................ | MEP-90 Hair Growth Stimulation System |
|----------------------------------------------------------------------------------------|---------------------------------------|
| Common/Usual Name(s) ............................................................ | Light Therapy Hair System |
| Classification Name ................................................................ | Infrared Lamp per 21CFR 890.5500 |
| Product Code(s) ...................................................................... | OAP; NHN |
## IV. Predicate Devices
| 510(k) Number | Device | Manufacturer |
|---------------|------------------------------|-----------------------------|
| K060305 | Hairmax Lasercomb | Lexington International LLC |
| K032816 | Quantum Light Therapy System | Stargate International |
{1}------------------------------------------------
idwestrf
#### VI. Indications For Use
The MEP-90 is a non-heating lamp as described under the provisions of 21 CFR §890.5500 and is indicated for:
Medically prescribed use for the treatment of androgenic alopecia in females:
The treatment of androgenic alopecia in females by promoting hair growth of females with androgenetic alopecia who have Ludwig and Savin Hair Loss Scale classifications of I to II and who have been determined to have a Fitzpatrick Skin Typing of I to IV.
### VII. Technological Characteristics
The MEP-90 is a stationary low-level laser device that promotes hair growth and provides treatment for androgenic (androgenetic) alopecia in females. The device provides automated and timed equal distribution of laser light to 100% of the scalp.
The MEP-90 operation is controlled by an operating system that affords the user maximum flexibility for individual treatments. The device applies a measured very high tolerance (≤+.76%) wavelength (λ) to the scalp stimulating hair growth by the proven concept of biostimulation.
#### VII. Performance Data And Clinical Efficacy
A multi-phased experimental study was performed with Institutional Review Board (IRB) preapproval and oversight, in accordance with all applicable references of the Food and Drug Cosmetic Act and Title 21; Code of Federal Regulations.
Androgenic alopecia in women is a chronic medical condition requiring diagnosis, treatment, and monitoring by a licensed medical physician. The condition in women demonstrates both physical and emotional symptoms, which requires addressing by a licensed medical professional.
For the MEP-90 efficacy determination, each subject received a total of 36 each, 20-minute treatments with the MEP-90, over a period of 18 weeks. Results were reviewed at the 10-Week (20 treatment) and 18-Week (36 treatment) levels.
After 20 treatments (10-Weeks), 92% of the subjects demonstrated an increased hair count of ≥10% with 57% demonstrating an increase of ≥30%. 98% of the subjects indicated a medically significant stabilization of their rate of hair loss.
After the 36th treatment, 97% of the subject population demonstrated an increased hair count of ≥20%. A total of 89% of all subjects demonstrated an increased hair count of ≥30%, with 57% demonstrating an increased hair count of ≥50%.
{2}------------------------------------------------
idwestrf
87% of the subjects indicated the treatments have helped their condition, with 60% reporting their loss rate has further slowed down from the 10-week period, and 65% reported their visible area of the alopecia (bald spot) had gotten smaller.
100% of the linear trend plotting for all subjects of their Initial, 10-Week, and 18-Week hair counts demonstrated a historical rate of increased hair growth.
No subject experienced any adverse event and/or effect from the treatments.
# VI. Substantial Equivalency
The MEP-90 is substantially equivalent to other pulsed therapeutic light therapy systems currently in commercial distribution. The MEP-90 has the same intended use to the predicate device approved for commercial distribution under 510(k) number K060305 and technological and safety characteristics to the predicate device approved for commercial distribution under 510(k) number K032816.
It exceeds the clinically accepted therapeutic results standards of FDA 510(k) K060305 previously approved light therapy system into a system which provides a more controlled application and larger treatment coverage area at no increased risk to the patient.
The technological equivalence to the predicate devices is substantiated by the wavelength and power output generated by the MEP-90. The MEP-90 provides expanded treatment benefits and regimens for clinical presentations already approved by the Food and Drug Administration for the predicate device.
The MEP-90 is as safe and effective as a combination of the predicate devices listed and numerous others. It has the same intended use of affecting hair growth as the hair growth predicate device (K060305). In addition, the MEP-90 has the same general indications, i.e., treating androgenic alopecia, and the same specific indication of promoting hair growth as the predicate device.
The MEP-90 also has many of the same or similar technological characteristics as a combination of its predicate devices. These include multiple lasers and visible laser wavelength.
The technological differences between the MEP-90 and its predicate devices, specifically the use of red laser to treat androgenic alopecia in females, does not raise new questions of safety or effectiveness for several reasons:
First, the safety and effectiveness profile of the type, wavelength, and power output of this type of laser is well established and previously cleared by the FDA.
Second, FDA's clearance of the predicate device with a much wider wavelength tolerances then the MEP-90's, confirms the favorable risk benefit ratio of visible lasers.
{3}------------------------------------------------
idwestrf
Third, the clinical data acquired confirms both the safety and effectiveness of the MEP-90 for prescription use in promoting hair growth in the intended patient population, despite the difference in technological characteristics between the MEP-90 and K060305. The data demonstrates clear statistical significance of the treatment results obtained and provide mathematical certainty that the results attained did not occur by chance.
These facts exceed FDA's substantial equivalence requirements with respect to the intended use, clinical efficacy, and technological characteristics of the MEP-90.
While there are some technological differences between the MEP-90 and its predicate devices, Midwest conducted an Institutional Review Board approved and monitored clinical study, with the MEP-90, to show that the device functions as intended for its proposed indication for use without any serious side effects or risks.
The clinical and effectiveness data demonstrates that the MEP-90 is effective in promoting hair growth, does not present any safety issues, is classified by the FDA as a non-significant risk (NSR) device, therefore the FDA should approve the medical device by approval of the 510(k).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird with outstretched wings.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
FEB 2 3 2010
Midwest RF. LLC % Mr. Helmut Keidl President 1050 Walnut Ridge Drive Hartland, Wisconsin 53029
Re: K091496
Trade/Device Name: MEP-90 Hair Growth Stimulation System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: January 15, 2010 Received: January 20, 2010
Dear Mr. Keidl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Helmut Keidl
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
idwestrf
# INDICATIONS FOR USE
## 510(k) Number: K091496
Device Name: MEP-90 Hair Growth Stimulation System
Indications For Use: The MEP-90 is a non-heating lamp as described under the provisions of 21 CFR §890.5500 and is indicated for:
> The treatment of androgenic alopecia in females by promoting hair growth of females with androgenetic alopecia who have Ludwig and Savin Hair Loss Scale classifications of I to II and who have been determined to have a Fitzpatrick Skin Typing of I to IV.
Prescription Use: AND/OR (Part 21 CFR 801 Subpart D)
Over The Counter Use:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dhtro
(Division Sign-Off) Division of Surgical, Orthepedic, and Restorative Devices
510(k) Number
16091496
Revised page 10 of 102 Pages - MEP-90 510(k) K091496 Application Dated May 15, 2009
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.