K142596 · B.Braun Medical, Inc. · FRN · Jun 18, 2015 · General Hospital
Device Facts
Record ID
K142596
Device Name
Infusomat Space Volumetric Infusion Pump
Applicant
B.Braun Medical, Inc.
Product Code
FRN · General Hospital
Decision Date
Jun 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Infusomat® Space Volumetric Infusion Pump System is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral fluids, medications, blood and blood products through clinically accepted routes of administration. These routes include intravenous, intra-arterial, subcutaneous, and epidural. The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities.
Device Story
Infusomat Space Volumetric Infusion Pump System provides intermittent or continuous delivery of parenteral fluids, medications, and blood products. Device utilizes linear peristaltic pumping mechanism; powered by external supply or rechargeable battery. System includes SpaceStation docking/communication system with integrated SpaceCom module supporting Ethernet, USB, serial, and WLAN interfaces. Data transfer between pumps and station occurs via internal CAN bus. Device supports inbound/outbound wireless communication, including EHR-based autoprogramming and drug library uploads via server-based Drug Library Manager (DLM) and Upload Manager (ULM). Used in healthcare facilities by trained professionals; biomedical technicians perform initial setup. Healthcare providers use device to manage infusion parameters; autoprogramming feature allows remote order receipt from EHR, though infusion start remains manual. System enhances medication safety through standardized drug libraries and patient profiles.
Clinical Evidence
No clinical testing conducted. Substantial equivalence supported by non-clinical performance testing, including software/system verification and validation, delivery accuracy, usability, electromagnetic compatibility, functional/electrical safety, and human factors studies simulating clinical conditions.
Technological Characteristics
Linear peristaltic infusion pump; transportable. Connectivity: Ethernet, USB, serial, WLAN, CAN bus. Power: 115V AC or rechargeable battery. Software: Server-based Drug Library Manager and Upload Manager. System architecture includes SpaceStation docking and SpaceCom communication module. No specific ASTM materials cited.
Indications for Use
Indicated for adults, pediatrics, and neonates requiring intermittent or continuous delivery of parenteral fluids, medications, blood, and blood products via intravenous, intra-arterial, subcutaneous, or epidural routes. Intended for use by trained healthcare professionals in healthcare facilities.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Infusomat® Space Volumetric Infusion Pump System (K083689)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a simple, flowing line style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2015
B. Braun Medical Inc. Mr. Gary Johnson Manager, Regulatory Affairs 901 Marcon Boulevard ALLENTOWN, PA 18109-9341
Re: K142596
Trade/Device Name: Infusomat® Space Volumetric Infusion Pump System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: May 18, 2015 Received: May 20, 2015
Dear Mr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142596
Device Name
Infusomat® Space Volumetric Infusion Pump System
Indications for Use (Describe)
The Infusomat® Space Volumetric Infusion Pump System is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral fluids, medications, blood products through clinically accepted routes of administration. These routes include intravenous, intra-arterial, subcutaneous, and epidural.
The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|---------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5. 510(k) SUMMARY
This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92
| APPLICANT: | B. Braun Medical Inc.<br>901 Marcon Boulevard<br>Allentown, PA 18109-9341<br>Establishment Registration 2523676 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Gary Johnson<br>Manager, Regulatory Affairs<br>Phone: 610-596-2263<br>Fax Number: 610-266-4962<br>Email: gary.johnson@bbraun.com |
| DATE OF PREPARATION: | June 18, 2015 |
| DEVICE NAME: | Infusomat® Space Volumetric Infusion Pump System |
| COMMON OR<br>USUAL NAME: | Infusion Pump |
| DEVICE<br>CLASSIFICATION: | Class II per 21 CFR §880.5725: Infusion Pump, Infusion Pump<br>product code FRN |
| CLASSIFICATION<br>PANEL: | General Hospital |
| PREDICATE<br>DEVICES: | The Infusomat® Space Volumetric Infusion Pump System is<br>comparable to and substantially equivalent to the Infusomat®<br>Space Volumetric Infusion Pump System marketed under<br>cleared 510(k) K083689.<br>The Alaris® Pump system by Care Fusion cleared under 510(k)<br>K072105 is provided as a reference device for the new<br>Autoprogramming feature. The Autoprogramming feature<br>allows the pump to accept orders wirelessly from the facility's<br>electronic health record (EHR) system. Although the pump can<br>accept remote orders, it does not have the ability to start an<br>infusion remotely. |
{4}------------------------------------------------
#### DESCRIPTION:
The Infusomat® Space Volumetric Infusion Pump System includes an external, electronic infusion pump and pump accessories. The pump is transportable within a facility.
The Infusomat® Space pump utilizes a linear peristaltic pumping mechanism and is intended to provide infusions of parenteral fluids. The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities. A trained Biomedical Technician must perform a complete set-up of the pump prior to use in a clinical setting.
The system is intended to provide intermittent or continuous flow of parenteral fluids to the patient. Parenteral fluids may include standard fluids and/or medications indicated for infusion as well as blood and blood products.
The Infusomat® Space is powered by an external power supply or by a rechargeable battery.
Infusomat® Space is capable of wireless communication both inbound and outbound.
#### SpaceStation with SpaceCom
The B. Braun Space Station is an 115V AC powered flexible docking and communication system for use in a medical facility. It is designed to accommodate multiple Infusomat® Space volumetric infusion pumps.
SpaceCom is a communication device that has been integrated into the SpaceStation. SpaceCom supports different interfaces such as Ethernet, PS2-Keyboard, Serial, USB ports and WLAN network card. Data transfer with the pumps is provided via an internal CAN bus. For barcoding, a barcode image reader can be connected to the PS2-Keyboard or USB interface. The pumps are coupled together with connectors on the inner backside of the SpaceStation. These connectors provide the voltage supply, distribute the information in the Space system via a serial interface, transfer data via a bus system (CAN bus) and transmit a staff call, which may be pending.
The outbound data communication through SpaceCom transfers status data of the infusion pumps to a hospital server and was cleared through the Infusomat Space Volumetric Infusion Pump System 510(k) K083689. The software in SpaceCom is being modified to allow inbound wired or wireless data communication to be uploaded. This includes data such as but not limited to drug libraries created by the DrugLibraryManager (DLM) to the infusion pumps.
{5}------------------------------------------------
#### Drug Library Manager and Drug Upload Manager
The Space OnlineSuite is a server based software system which provides the following applications:
### Space Server Core
Space Server Core is the basic server framework for Space applications. This application framework provides basic server functions like User Management, License Management, Data Management, Communication Service, Security and Maintenance Functions. These functions are used by the administrator of the Space OnlineSuite. The applications, Drug Library manager and Upload Manager also use the basic functions of SpaceServer Core.
## Drug Library Manager (DLM)
The Drug Library Manager allows the creation of unit and ward specific drug libraries and patient profiles. These features are designed to enhance medication safety and reduce medication errors.
## Upload Manager (ULM)
The server-based Upload Manager can be used to manage the upload of drug libraries to any single B. Braun Space infusion pump, multiple Space pumps within a facility or to all Space pumps on the System. The drug libraries will be uploaded to SpaceCom residing in SpaceStation and then transferred via CAN bus (Controller Area Network) to the pumps or to a single Infusomat® Space Volumetric Infusion Pump directly.
#### The Infusomat® Space Volumetric Infusion Pump System is INTENDED USE: intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral fluids, medications, blood and blood products through clinically accepted routes of administration. These routes include intravenous, intraarterial, subcutaneous, and epidural. The Infusomat® Space Volumetric Infusion Pump System is intended to be used by trained healthcare professionals in healthcare facilities.
The predicate device, the Infusomat® Space Volumetric Infusion Pump System (cleared in 510(k) K083689) is also intended for use on adults, pediatrics and neonates for infusion of parenteral fluids, medications, blood and blood products through FDA approved routes of administration. The patient population and infusion solutions are the same for the subject and predicate devices. The difference in the indication for the subject device is clarification of the routes of administration; intravenous, intra-arterial, subcutaneous and epidural. The routes of administration for the subject device are also FDA approved
{6}------------------------------------------------
routes of administration. Testing was conducted with the subject device to demonstrate that the differences in routes of administration do not affect the performance of the subject device. Both the subject and predicate devices are intended for prescription use as provided in the indications for use. Therefore, use by trained professionals in healthcare facilities was added to the indications for the subject device to provide further details. This does not affect the intended use since both the subject and predicate devices are intended for use by trained professionals in healthcare facilities.
SUBSTANTIAL The subject device, the B. Braun Infusomat Space Volumetric EQUIVALENCE: Infusion Pump System is substantially equivalent to the Infusomat® Space Volumetric Infusion Pump System marketed under cleared 510(k) K083689. The intended use, operation and function of the subject device are similar to the identified predicate devices.
> The similarities and differences between the subject device and predicate devices are identified in this submission. Performance testing conducted with the subject device demonstrates the device will function as intended. No new questions or issues of safety and effectiveness have been raised during verification and validation performed with the subject device.
> Based on the information provided in this submission, the Infusomat Space Volumetric Infusion Pump System is substantially equivalent to the legally marketed predicate devices when used as instructed by trained healthcare professionals.
#### PERFORMANCE DATA NON-CLINICAL
TESTING:
Performance data is provided for the subject device to support substantial equivalence. Non clinical testing was conducted with the subject device to demonstrate substantial equivalence to the predicate device. To verify that the subject device design is appropriate for the intended use and meets the user requirements, verification and validation testing were conducted. Test results demonstrate that the subject device functions as intended and is substantially equivalent to predicate device. Testing conducted included software and system verification and validation, delivery accuracy, usability, electromagnetic compatibility, functional and electrical safety. Human factors studies to validate the use safety and usability of the subject device were performed. The human factors studies were conducted with the intended user population, use
{7}------------------------------------------------
environment, and use scenarios to simulate clinical conditions of use. Results of the human factors testing demonstrate the device functions as intended under simulated conditions of use by the intended user population.
## PERFORMANCE DATA CLINICAL TESTING:
Clinical testing was not conducted with the subject device.
CONCLUSION: The nonclinical and simulated clinical use testing conducted using the subject device summarized above demonstrates that the subject device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.