ALARIS SYSTEM WITH MMS, MODEL 8XXX

K072105 · Cardinal Health 303,Inc. · FRN · Dec 28, 2007 · General Hospital

Device Facts

Record IDK072105
Device NameALARIS SYSTEM WITH MMS, MODEL 8XXX
ApplicantCardinal Health 303,Inc.
Product CodeFRN · General Hospital
Decision DateDec 28, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Intended Use

The incorporation of the Medication Management System (MMS) with the Alaris® System (a.k.a. Medley™ System) provides wired or wireless communication between the Alaris® System and external devices. This is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps to enter infusion data. All data entry and validation of infusion parameters using MMS is performed by the trained healthcare professional. The Alaris System with MMS is integrated into an existing hospital network infrastructure and allows communications to and from external devices, including personal computers, Personal Digital Assistants (PDAs), hospital monitoring systems and Hospital Information Management Systems (HIMS). Bi-directional communication of data includes infusion parameters, system configuration, history, events, trending, alarms and status. In addition the Alaris System with MMS has the capability to transmit, receive and/or store: features, calibration data, datasets, and libraries from external components or devices to the pump.

Device Story

Alaris System with MMS facilitates wireless/wired communication between infusion pumps and hospital network infrastructure (servers, PCs, PDAs, HIMS). Device acts as a bedside safety net by automating infusion parameter programming, reducing manual data entry steps. Trained healthcare professionals operate the system, performing all data entry and validation. System enables bi-directional data exchange, including infusion parameters, system configuration, history, events, trending, alarms, and status. It also transmits/receives/stores calibration data, datasets, and libraries. By automating programming, the device aims to reduce manual programming errors, potentially improving patient safety during fluid, medication, and blood product delivery.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Infusion pump system with integrated wired/wireless communication modules. Connectivity includes integration with hospital network infrastructure (HIMS, PCs, PDAs). System supports bi-directional data transfer for infusion parameters, configuration, and device libraries. Fundamental scientific technology and materials are identical to the predicate device.

Indications for Use

Indicated for use in professional healthcare facilities for the delivery of fluids, medications, blood, and blood products via infusion devices. Intended for operation by trained healthcare professionals to automate infusion parameter programming based on physician orders.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ CARDINAL HEALTH SPECIAL 510(K) ALARIS® SYSTEM WITH MMS # ATTACHMENT D K072105 # SUMMARY OF SAFETY AND EFFECTIVENESS Cardinal Health 303, Inc. Alaris® System with MMS | SUBMITTER'S NAME: | Cardinal Health 303, Inc.<br>d.b.a. Cardinal Health<br>10221 Wateridge Circle<br>San Diego, CA 92121-2772<br>(858) 458-7830<br>(858) 458-6114 FAX | DEC 2 8 2007 | |-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | CONTACT PERSON: | Stacy L. Lewis<br>Principal Regulatory Affairs Specialist | | | DATE PREPARED: | July 27, 2007 | | | DEVICE NAME: | Proprietary Name<br>Alaris® System with MMS | | | | Common Name<br>Infusion Pump | | | | Classification Name<br>Pump, Infusion and accessories, FRN (880.5725) | | | PREDICATE DEVICE: | Alaris® System with MMS (a.k.a. Medley™ System with<br>MMS) K030459 | | #### DEVICE DESCRIPTION As with the predicate device, the Alaris® System with MMS assists our customers in reducing the number of manual steps needed to program an infusion by allowing wireless communication capability to a server and an existing infusion device (the Alaris® System). The wireless communication capability provides a "safety net" at the bedside to help reduce the number of programming errors at the point of care. It allows the Alaris® System to transmit and receive messages with the Alaris® Server which in turn allows communication capability with external devices, including personal computers, Personal Digital Assistants (PDA;s), hospital monitoring systems and Hospital Information Management Systems (HIMS). {1}------------------------------------------------ # ALARIS® SYSTEM WITH MMS SUMMARY OF SAFETY AND EFFECTIVENESS Page 2 of 2 #### SUBSTANTIAL EQUIVALENCE The Alaris® System with MMS is essentially the same as the predicate device in that they have the same intended use, operating principles, method of operation, technology, materials and manufacturing processes. The changes as described in this Special 510(k) pose no new issues of safety or efficacy. The Alaris® System with MMS as described in this submission is substantially equivalent to the predicate device. #### INTENDED USE The incorporation of the Medication Management System (MMS) with the Alaris® System (a.k.a. Medley™ System) provides wired or wireless communication between the Alaris® System and external devices. This is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps to enter infusion data. All data entry and validation of infusion parameters using MMS is performed by the trained healthcare professional. The Alaris System with MMS is integrated into an existing hospital network infrastructure and allows communications to and from external devices, including personal computers, Personal Digital Assistants (PDAs), hospital monitoring systems and Hospital Information Management Systems (HIMS). Bi-directional communication of data includes infusion parameters, system configuration, history, events, trending, alarms and status. In addition the Alaris System with MMS has the capability to transmit, receive and/or store: features, calibration data, datasets, and libraries from external components or devices to the pump. ## TECHNOLOGICAL CHARACTERISTICS A comparison of the technological characteristics of the Alaris® System with MMS and the predicate device has been performed. The results of this comparison demonstrate that the Alaris® System with MMS is equivalent in technological characteristics and the fundamental scientific technology of the predicate device has not been altered. $$ \begin{array}{ccc} \mathsf{k}\mathsf{o}\mathsf{7}\mathsf{z}\mathsf{o}\mathsf{6} & \mathsf{p}\mathsf{4} & \mathsf{z}\mathsf{o}\mathsf{z} \\ \mathsf{k}\mathsf{z}\mathsf{o}\mathsf{z} & \mathsf{p}\mathsf{z} & \mathsf{z}\mathsf{o}\mathsf{z} \\ \end{array} $$ 00768 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines representing the serpent winding around the staff. DEC 2 8 2007 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Stacey L. Lewis Principal Regulatory Affairs Specialist Cardinal Health 303, Incorporated 10221 Wateridge Circle San Diego, California 92121 Re: K072105 Trade/Device Name: Alaris System with MMS Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: December 3, 2006 Received: December 4, 2006 Dear Ms. Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Ms. Lewis Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Shiela A Murphy, MD for Chu Lin, Ph D Chu Lin, Ph D Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ CARDINAL HEALTH # Attachment B # INDICATIONS FOR USE 510(k) Number: 12) 05 (To Be Assigned By FDA) Device Trade Name: Alaris® System with MMS ### Indications For Use: The Alaris® System with Medication Management System (also known as the Medley™ System with MMS and the Alaris® System with MMS) is intended for use in today's growing professional healthcare environment for facilities that utilize infusion devices for the delivery of fluids, medications, blood and blood products. The Alaris® System with MMS is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician's order. Prescription Use X Over-Th Over-The-Counter Use PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) OR Concurrence of CDRH, Office of Device Evaluation (ODE) (Per 21 CFR 801.109) Clint Damer (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K472145 00763
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...