ALIANS ULNA LOCKING PLATING SYSTEM

K142502 · Newclip Technics · HRS · Dec 23, 2014 · Orthopedic

Device Facts

Record IDK142502
Device NameALIANS ULNA LOCKING PLATING SYSTEM
ApplicantNewclip Technics
Product CodeHRS · Orthopedic
Decision DateDec 23, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Alians Ulna Locking Plating System is indicated for ulnar shortening osteotomy.

Device Story

The Alians Ulna Locking Plating System consists of titanium alloy plates and screws designed for ulnar shortening osteotomy. The device is provided non-sterile and requires sterilization by healthcare professionals prior to use. It is intended for surgical fixation of the ulna. The system functions as a bone fixation appliance to stabilize the bone during the healing process following an osteotomy procedure. It is used in a clinical/surgical setting by orthopedic surgeons.

Clinical Evidence

No clinical studies were performed. Evidence is based on non-clinical bench testing, including comparative fatigue compression tests, static tests on plates, and pullout/torsional strength tests on screws.

Technological Characteristics

Materials: Titanium alloy Ti-6Al-4V ELI (ASTM F 136-12a / ISO 5832-3). Form factor: Plates with multiple screw holes/slots. Sterilization: Provided non-sterile for healthcare professional sterilization. Mechanical fixation principle.

Indications for Use

Indicated for patients requiring ulnar shortening osteotomy.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 23, 2014 NEWCLIP TECHNICS % Mr. J.D. Webb Official Correspondent The OrthoMedix Group, Incorporated 1001 Oakwood Boulevard Round Rock, Texas 78681 Re: K142502 Trade/Device Name: Alians Ulna Locking Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: October 3, 2014 Received: October 6, 2014 Dear Mr. Webb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K142502 Device Name ALIANS ULNA LOCKING PLATING SYSTEM Indications for Use (Describe) The Alians Ulna Locking Plating System is intended for ulnar shortening osteotomy. Type of Use (Select one or both, as applicable) X | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for a company called NEWCLIP-TECHNICS. The logo consists of a circle with the letters "nct" inside. The "t" in "nct" has an arrow pointing upwards. The words "NEWCLIP-TECHNICS" are written in a sans-serif font below the circle. #### 4. 510 (k) Summary for the Alians Ulna Locking Plating System In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following 510(k) summary is submitted for the Alians Ulna Locking Plating System. Contact Person: Summary preparation date: July 17, 2014 ### 1. Submitter: #### NEWCLIP TECHNICS J.D. Webb The OrthoMedix Group, Inc. P.A. de la Lande Saint Martin 45 rue des Garottières 1001 Oakwood Blvd F-44115 Haute-Goulaine - France Round Rock, TX 78681 Telephone: (33) 2 28 21 37 12 Telephone: 512-388-0199 | 2. Trade name: | Alians Ulna Locking Plating System | |----------------------|------------------------------------------------------------------------------------------------------| | Common Name: | Plates for ulna shortening osteotomy<br>Screws associated | | Product code: | HRS/HWC | | Classification Name: | Plate, Fixation, Bone<br>(21 CFR part. 888.3030)<br>Screw, Fixation, Bone<br>(21 CFR part. 888.3040) | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for NEWCLIP-TECHNICS. The logo features the letters "nct" inside of a circle. The "t" in "nct" has a red arrow pointing upwards. Below the circle, the words "NEWCLIP-TECHNICS" are written in all caps. ### 3. Predicate or legally marketed devices which are substantially equivalent: - The Synthes One-Third Tubular DCL Plate (K011335). . - Newclip Radius Plates (K061917) . - Newclip Clavicle Plate System (K100944) ● - Newclip Distal Radius Plate (K130774) . | 4. Description of the device: | The Alians Ulna Locking Plating system<br>consists of a plate with many screws as<br>holes (slots) designed for ulnar shortening<br>osteotomy. The plates and screws are<br>manufactured from titanium alloy and<br>color anodized. | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The Alians Ulna Locking Plating system<br>will be provided non-sterile for<br>sterilization by health care professional's<br>prior use. | | Materials: | Titanium alloy Ti-6Al-4V ELI (conform to<br>ASTM F 136-12a and/or ISO 5832-3). | | Function: | The implants of the Alians Ulna Locking<br>Plating System are indicated for fixation of<br>ulnar shortening osteotomy. | ## 5. Substantial equivalence claimed to predicate devices: The Alians Ulna Locking Plating System is substantially equivalent to Synthes One-Third Tubular DCL Plate (K011335), to the Newclip Radius Plates (K061917), to the Newclip Clavicle Plate System (K100944) and to the Newclip Distal Radius Plate (K130774) in terms of intended use, design, materials used, mechanical safety and/or performance. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Newclip-Technics. The logo features the letters 'nct' in a circle, with a red arrow pointing upwards above the 't'. Below the circle, the words 'NEWCLIP-TECHNICS' are written in a sans-serif font. ## 6. Intended use: The Alians Ulna Locking Plating System is indicated for ulnar shortening osteotomy. ### 7. Non-clinical Test Summary: The following tests were conducted: - Comparative fatigue compression tests on plates. . - Comparative static tests on plates. . - Pullout and torsional strength tests on screws. . ## 8. Clinical Test Summary: No clinical studies were performed. ### 9. Conclusions Non-clinical and Clinical: The Alians Ulna Locking Plating System is substantially equivalent to the predicate devices in terms of indications for use, design, material, and function.