OMNI Interface Acetabular System

K142201 · Omnlife Science · LPH · Jan 26, 2015 · Orthopedic

Device Facts

Record IDK142201
Device NameOMNI Interface Acetabular System
ApplicantOmnlife Science
Product CodeLPH · Orthopedic
Decision DateJan 26, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Intended Use

The OMNI Interface™ Acetabular System is designed to be used in hip arthroplasty. The specific indications for use include - . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - Rheumatoid arthritis; - Correction of functional deformity; - Congenital dislocation; - Revision procedures where other treatments or devices have failed; - Femoral neck and trochanteric fractures of the proximal femur. The OMNI Interface Acetabular System is intended for cementless and single use implantation only.

Device Story

OMNI Interface Acetabular System is a cementless, single-use hip prosthesis for primary and revision total hip replacement. System components include titanium alloy acetabular shells (no-hole or 3-hole configurations), ultra-high molecular weight polyethylene (UHMWPE) shell inserts, and femoral heads. Device is implanted by orthopedic surgeons during hip arthroplasty procedures. System is designed for compatibility with femoral hip stems, including the Paragon stem. Output is a reconstructed hip joint interface; clinical benefit is restoration of joint function and mobility. Compatibility with the Paragon stem was verified via range of motion testing.

Clinical Evidence

Bench testing only. Performance requirements for range of motion were evaluated per ISO 21535 to confirm compatibility between the OMNI Interface Acetabular System and the Paragon stem. No clinical data was provided.

Technological Characteristics

Materials: Titanium alloy (shells), standard and crosslinked ultra-high molecular weight polyethylene (inserts). Design: Semi-constrained, porous-coated, uncemented acetabular prosthesis. Configurations: No-hole and 3-hole shells, sizes 46-76 mm and X-size (48x-58xmm). Standards: ISO 21535 (range of motion).

Indications for Use

Indicated for patients requiring hip arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, congenital dislocation, revision procedures, or femoral neck/trochanteric fractures of the proximal femur.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes an abstract image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the seal. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 26, 2015 OMNIlife Science, Inc. Ms. Vani Sindwani Regulatory Affairs Specialist 50 O'Connell Way, Suite 10 East Taunton, Massachusetts 02718 Re: K142201 Trade/Device Name: OMNI Interface™ Acetabular System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, MEH Dated: November 25, 2014 Received: November 28, 2014 Dear Ms. Sindwani: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {1}------------------------------------------------ ## Page 2 - Ms. Vani Sindwani (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known): K142201 Device Name: OMNI (previously Apex) Interface™ Acetabular System Indications for Use: The OMNI Interface™ Acetabular System is designed to be used in hip arthroplasty. The specific indications for use include - - . Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - Rheumatoid arthritis; - Correction of functional deformity; - Congenital dislocation; - Revision procedures where other treatments or devices have failed; - Femoral neck and trochanteric fractures of the proximal femur. The OMNI Interface Acetabular System is intended for cementless and single use implantation only. X AND/OR Over-The-Counter Use Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ ## 510(k) Summary | TABLE 1: 510(k) SUMMARY | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Summary Prepared | 08/08/2014 | | Manufacturer/Distributor/Sponsor | OMNIlife Science, Inc.<br>50 O'Connell Way<br>Suite 10<br>East Taunton, MA 02718 | | 510(k) Contact | Vani Sindwani<br>Regulatory Affairs Specialist<br>OMNIlife Science<br>50 O' Connell Way<br>Suite 10<br>East Taunton, MA 02718<br>Telephone: 774-226-1871<br>Fax: 508-822-6030<br>Email: vsindwani@omnils.com | | Trade Name | OMNI Interface™ Acetabular System | | Common Name | Prosthesis, hip, semi-constrained, metal/polymer,<br>porous uncemented | | Classification | 21 CFR 888.3358 Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented prosthesis. LPH | | Predicate Device | K112779 Apex Interface Acetabular System<br>K060635 Consensus Acetabular Shell System<br>K123782 Paragon Hip System<br><br>Reference Predicates:<br>K031110 Apex Acetabular Shell | | Purpose of Submission | This traditional 510(k) premarket notification is being<br>submitted to expand the indications for OMNI Interface<br>Acetabular System. Through this premarket notification<br>OMNI intends to obtain clearance for the use of OMNI<br>Interface Acetabular System with Global's Paragon<br>stem cleared via K123782. | | Device Description and Intended Use | The OMNI Interface Acetabular System consists of<br>acetabular cups/shells that are made of titanium alloy,<br>standard and crosslinked ultra high molecular weight<br>polyethylene shell inserts and femoral heads that are<br>used with compatible femoral hip stems for primary and | | | revision total hip replacement. The shells are available<br>in “no hole” and “3-hole” configuration and a size range<br>from 46-76 mm and X size (48x-58xmm). | | | The OMNI Interface™ Acetabular System is designed<br>to be used in hip arthroplasty. The specific indications<br>for use include -<br>Non-inflammatory degenerative joint disease,<br>including osteoarthritis and avascular necrosis;Rheumatoid arthritis;Correction of functional deformity;Congenital dislocation;Revision procedures where other treatments or<br>devices have failed;Femoral neck and trochanteric fractures of the<br>proximal femur. | | | The OMNI Interface Acetabular System is intended for<br>cementless and single use implantation only. | | Substantial Equivalence Summary | The OMNI Interface Acetabular System is substantially<br>equivalent to the predicates as the basic design,<br>interface, fundamental technology, materials and<br>intended use are the same. | | | Paragon stem (K123782) is cleared to be used in<br>conjunction with a compatible femoral head and<br>acetabular component and is used with Consensus CS2<br>Acetabular Shell System (K060635). OMNIlife science<br>has determined OMNI Interface Acetabular System to<br>be substantially equivalent to CS2 Acetabular Shell<br>System thereby supporting its use with the Paragon<br>stem. | | | The use of OMNI Interface Acetabular System with the<br>Paragon stem is determined to be as safe and effective<br>as the predicate and does not raise any new safety or<br>effectiveness concerns. The results of testing for the<br>proposed use of OMNI Interface Acetabular System<br>with the Paragon Stem met the performance<br>requirements for range of motion per ISO 21535 | | The comparison between OMNI hip stems and the<br>Paragon stem, and OMNI and Global femoral heads<br>have been included in this submission that further<br>supports the compatibility of OMNI Interface<br>Acetabular System and Paragon Stem. | | | Based on the identical material, the design and<br>characterization data, compatibility and mechanical<br>testing, and technological characteristics, OMNIlife<br>science believes the proposed OMNI Interface<br>Acetabular System to be substantially equivalent to<br>legally marketed predicates and can be used with<br>Paragon stem for primary and revision hip arthroplasty. | | {4}------------------------------------------------ {5}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...