K060635 · Hayes Medical, Inc. · LPH · Apr 28, 2006 · Orthopedic
Device Facts
Record ID
K060635
Device Name
CONSENSUS ACETABULAR SHELL SYSTEM
Applicant
Hayes Medical, Inc.
Product Code
LPH · Orthopedic
Decision Date
Apr 28, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. Proximal femoral fractures Avascular necrosis of the femoral head Non-union of proximal femoral neck fractures Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomaliacia, osteoporosis, pseudarthrosis conversion and structural abnormalities.
Device Story
Consensus Acetabular Shell System is a modular, cementless acetabular component for total hip arthroplasty. Device consists of titanium alloy (ASTM F136) shells with porous coating of irregular sintered CP titanium beads (ASTM F67) for press-fit fixation. Available in hemispherical and rim-flare profiles, with or without screw holes. Features include threaded apical dome hole for impaction tool and optional threaded titanium plug to prevent particle migration. Device mates with existing polyethylene (UHMWPE) inserts. Used by orthopedic surgeons in clinical settings to replace acetabular surface, aiming to restore hip function and reduce pain in patients with degenerative or traumatic joint conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by design comparisons and material data.
Technological Characteristics
Substrate: Titanium alloy (ASTM F136, Ti-6Al-4V). Porous coating: Irregular sintered CP titanium beads (ASTM F67). Fixation: Cementless press-fit. Features: Threaded apical dome hole with titanium plug. Locking mechanism: Identical to K922561 for UHMWPE liners. Modular options: Hemispherical and rim-flare profiles.
Indications for Use
Indicated for skeletally mature patients undergoing total hip arthroplasty for pain relief or improved function due to rheumatoid, osteoarthritis, post-traumatic arthritis, failed prior hip procedures, proximal femoral fractures, avascular necrosis, non-union of femoral neck fractures, congenital dysplasia, or other metabolic/structural abnormalities.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Hayes Medical, Consensus Acetabular Shell (K922561)
Orthopaedic Source, Avalon Acetabular Cup (K022711)
Encore, Foundation Porous Acetabular System (K974095)
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of 2
# Section 9 510(k) Summary of Safety and Effectiveness
Defined in 21 CFR 807 In accordance with 21 CFR 807.92 (Summary)
Fir 28 2003
| Applicant's Name: | Hayes Medical, Inc.<br>1115 Windfield Way, Suite 100<br>El Dorado Hills, CA 95682 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Luke Rose |
| Trade Name: | Consensus Acetabular Shell System |
| Common Name: | Acetabular Shell, porous, uncemented |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis (21 CFR 888.3358, Product Code LPH) |
| | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis (21 CFR 888.3353, Product<br>Code LZO) |
| Proposed Regulatory Class: | Class II |
| Device Classification Panel: | Orthopaedic |
| Substantially Equivalent To: | Hayes Medical, Consensus Acetabular Shell (K922561)<br>Wright Medical, Lineage Cup (K002149)<br>Apex Surgical, Modular Acetabular Cup (K031110)<br>Orthopaedic Source, Avalon Acetabular Cup (K022711)<br>Encore, Foundation Porous Acetabular System (K974095) |
# Intended Use:
Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. .
- Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. .
- . Proximal femoral fractures
- . Avascular necrosis of the femoral head
- . Non-union of proximal femoral neck fractures
- Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa ● vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomaliacia, osteoporosis, pseudarthrosis conversion and structural abnormalities.
### Device Description:
The Consensus Acetabular Shell System consists of four modular shell options which mate with currently marketed polyethelyne insert components for replacement of the acetabular surface during total hip arthroplasty. The shells are designed to be implanted without bone cement. Cementless (press-fit) fixation is achieved by coating the external surface of the cup with commercially pure (CP) titanium sintered beads.
All of the shell options have the following common features:
- The substrate material of the Consensus Shells is titanium alloy (ASTM F136, Ti-6AI-4V) ●
{1}------------------------------------------------
K060635
Page 2 of 2
- · The shell is porous coated with commercially pure (CP) titanium sintered beads (ASTM F67, irregular bead geometry).
- · The shells have the same internal locking mechanism as the previously cleared Consensus Acetabular Shells (K922561).
- · The apical dome hole of the shells is threaded to accept an impaction tool for implanting the component. After implantation and the impaction tool is removed, a threaded titanium (ASTM F67 or F136) hole plug can be inserted into the threaded hole to prevent migration of unwanted particles through the dome hole.
The following shell options are available:
- · Total Hemispherical in no screw holes or three (3) quadrant screw holes options.
- · Hemispherical with rim flare in no screw holes or three (3) quadrant screw holes options.
#### Comparison to Cleared Device
The following changes have been made to the previously cleared Consensus Acetabular Shell (K922561):
- Addition of the Hemisphical with rim flare outer profile .
- Removal of the two inferior screw holes. Holed cups now have three instead of five screw holes. .
- The apical dome hole has been threaded. .
- . A threaded apical dome hole plug is now available.
- . The porous coating is being modified from a spherical sintered CP titanium bead to an irregular sintered CP titanium bead.
The locking mechanism between the polyethylene (UHMWPE) liner and the acetabular shell is identical to the previously cleared Consensus Acetabular Shell (K922561). The safety and effectiveness of this locking detail is described in testing contained within the original submission.
# Substantial Equivalence Information
The intended use, material, design features and type of interface of the Consensus Acetabular Shell System are substantially equivalent to competitive devices previously cleared for market. The safety and effectiveness of the Consensus Acetabular Shell System are adequately supported by the substantial equivalence information and materials data provided within this Special 510(k) submission.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the top half of the circle. Inside the circle is an abstract image of a bird-like figure with three wing-like shapes above a wavy line.
APR 2 8 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hayes Medical, Inc. c/o Mr. Luke Rose Director, Quality Systems and Regulatory Affairs 1115 Windfield Way - Suite 100 El Dorado Hills, California 95762
Re: K060635
Trade/Device Name: Consensus Total Hip System, Acetabular Cup Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Codes: LPH and LZO Dated: March 30, 2006 Received: March 31, 2006
Dear Mr. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. În addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act`s requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if
{3}------------------------------------------------
Page 2 - Mr. Luke Rose
applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Hicker Lerner (o)
Mark N. Melkerson, M.S. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Consensus Acetabular Cup System Device Name:
Indications For Use:
Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C. Proximal femoral fractures
- D. Avascular necrosis of the femoral head
- E. Non-union of proximal femoral neck fractures
- F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion and structural abnormalities.
Prescription Use __ X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH/Office of Device Evaluation (ODE)
f CERTAIN STATES OF AMERICA
Division of General. Restorative. and Neurological Devices
| Onno<br>at | of | |
|------------|----|-----------|
| 4 | | -4-4-46-4 |
**510(k) Number** K060635
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.