K142087 · Tosense, Inc. · MWI · May 1, 2015 · Cardiovascular
Device Facts
Record ID
K142087
Device Name
COVA MONITORING SYSTEM
Applicant
Tosense, Inc.
Product Code
MWI · Cardiovascular
Decision Date
May 1, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The CoVa™ Monitoring System is intended for use under the direction of a licensed medical professional by adult patients at home to record, store, and transmit the following physiological data: i) Heart Rate including Heart Rate Variability; ii) Thoracic Impedance; iii) Respiration Rate; and iv) Posture. The CoVat™ Monitoring System only displays these physiological data to licensed medical professionals.
Device Story
Body-worn necklace-style sensor measures single-lead ECG, thoracic bioimpedance (TBI), skin temperature, and posture via accelerometer. Sensor connects to disposable electrodes via magnetic interface. Data transmitted via Bluetooth to Android-based gateway (tablet/phone), then forwarded via cellular/Wi-Fi to web-based system for clinician review. Used at home by patients; operated under direction of licensed medical professional. Clinicians access data via web interface to monitor fluid status, heart rate, and respiration. Benefits include remote physiological monitoring for patients with heart failure, renal disease, or fluid-management issues, facilitating clinical decision-making regarding medication or treatment adjustments.
Clinical Evidence
Clinical data included three studies (N=33, N=23, N=19) and bench testing (N=88, 15, 67, 15, 40 measurements). Study 1 (hemodialysis patients) showed TI correlation with fluid removal (r=0.93, p<0.0001). Studies 1-3 compared TI and RR against FDA-cleared reference devices using rank-regression; results showed slope/intercept consistency. Bench studies (4a-e) confirmed accuracy for TI, RR, HR, HRV, and temperature using calibrated simulators. No clinical data on long-term outcomes provided; focus on measurement equivalence.
Technological Characteristics
Necklace-form factor sensor; single-lead ECG; thoracic bioimpedance; tri-axial accelerometer; thermal sensor. Connectivity: Bluetooth (sensor-to-gateway), cellular/802.11 (gateway-to-web). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366, AAMI/ANSI EC13, ISO 80601-2-56. Disposable electrodes with magnetic interface.
Indications for Use
Indicated for adult patients at home with fluid-management problems, taking diuretic medication, living with heart failure, living with end-stage renal disease, recovering from a coronary artery disease-related event, or suffering from recurrent dehydration. Contraindicated for patients with implanted cardiac devices (pacemakers/ICDs), patients requiring external defibrillation, and patients undergoing MRI procedures.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 1, 2015
toSense, Inc. Nandini Murthy Regulatory Consultant to toSense, Inc 4225 Executive Square, Suite 570 La Jolla, California 92037
> K142087 Trade/Device Name: Cova Monitoring System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, DSB, FLL Dated: April 1, 2015 Received: April 2, 2015
Dear Nandini Mirthy,
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours,
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Section 4
### Indications for Use Form
510(k) Number (if known): K142087
Device Name:
#### Indications for Use:
The CoVa™ Monitoring System is intended for use under the direction of a licensed medical professional by adult patients at home to record, store, and transmit the following physiological data: i) Heart Rate including Heart Rate Variability; ii) Thoracic Impedance; iii) Respiration Rate; and iv) Posture. The CoVat™ Monitoring System only displays these physiological data to licensed medical professionals.
The CoVaTM Monitoring System is indicated for patients: i) with fluid-management problems; ii) taking diuretic medication; iii) living with heart failure; iv) living with end-stage renal disease; v) recovering from a coronary artery disease-related event; and/or vi) suffering from recurrent dehydration.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) Summary
| Date<br>prepared | May 1, 2015 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | toSense, Inc.<br>4225 Executive Square<br>Suite 570<br>La Jolla, CA 92037<br>(619) 733-4968 |
| Contact<br>Person | Matthew Banet, Ph.D.<br>Chief Technology Officer |
| Trade Name | CoVaTM Monitoring System |
| Common<br>Name | Mobile patient monitoring system, cardiac monitor |
| Classification<br>Name | Patient Physiological Monitor (without arrhythmia detection or alarm); Impedance Plethysmograph |
| Classification<br>Regulations | 21 CFR 870.2300 (MWI), 21 CFR 870.2770 (DSB), 21 CFR 880.2910 (FLL) |
| Predicate<br>Devices | Zoe Fluid Status Monitor (referred to herein as the 'Zoe monitor')<br>Manufacturer: Non-Invasive Technologies, Inc.<br>Number: K042113<br>Clearance Date: September 14, 2004<br><br>Avivo Mobile Patient Management System (referred to herein as the 'Corventis monitor')<br>Manufacturer: Corventis<br>Number: K113187<br>Clearance Date: January 4, 2012<br><br>ViSi Mobile Monitoring System (referred to herein as the 'ViSi monitor')<br>Manufacturer: Sotera Wireless<br>Number: K112478<br>Clearance Date: March 22, 2012 |
| Description | toSense's CoVaTM Monitoring System features a body-worn Sensor, Gateway, and Web-based System.<br><br>The Sensor non-invasively measures heart rate (HR), heart rate variability (HRV), respiration rate (RR), and thoracic index (TI). To determine these parameters, the Sensor measures and processes single-lead electrocardiogram (ECG) and thoracic bioimpedance (TBI) waveforms. Additionally, the Sensor measures skin temperature (TEMP) and posture using, respectively, a temperature sensor and accelerometer.<br><br>The Sensor has a form factor similar to a conventional necklace, with all measurement electronics built into its strands and base. A pair of customized disposable Electrodes, each featuring two electrode regions, snaps into a magnetic interface on the backside of the base, and then attaches to the patient's chest to make a measurement. The Sensor is designed for measurement periods less |
{4}------------------------------------------------
than about 5 minutes.
Using a Bluetooth transceiver, the Sensor wirelessly transmits information it measures from the patient to a Gateway, which can be a tablet computer or mobile phone running the Android operating system. Each of these systems receives information from the Sensor, and then forwards it to a Web-based System through either a local-area network (e.g., network based on 802.11), or a wide-area cellular network (e.g. AT&T). The Web-based System displays information, and can also forward it to a third-party system through a web-services interface.
Indications for Use
The CoVa" Monitoring System is intended for use under the direction of a licensed medical professional by adult patients at home to record, store, and transmit the following physiological data: i) Heart rate including heart rate variability; ii) Respiration rate; iii) Thoracic impedance; and iv) Posture. The CoVa"M Monitoring System only displays these physiological data to licensed medical professionals.
The CoVa™ Monitoring System is indicated for patients: i) with fluid-management problems; ii) taking diuretic medication; iii) living with heart failure; iv) living with end-stage renal disease; v) recovering from coronary artery disease-related event; and/or vi) suffering from recurrent dehydration.
The following are the contraindications for the CoVa™ Monitoring System:
- . The System is not to be used on patients with implanted cardiac devices, such as pacemakers and implanted cardio-defibrillators (ICDs).
- . The System is not defibrillator-proof. Thus the sensor should be removed from a patient before using an external defibrillator.
- . The System includes magnetically active materials, and thus should not be used by patients undergoing a procedure involving magnetic resonance imaging (MRI).
Summary of Each indication for use and measurement assigned to the CoVa™ Monitoring System is also Substantial assigned to one or more of the Corventis, Zoe, or ViSi monitors. The System's proposed intended-Equivalence use population is also aligned with that of one or more of the predicate monitors.
> The Sensor, along with the Zoe and Corventis monitors, all use an impedance-based technology to measure TI or an equivalent thereof. Our validations efforts, which compare the Sensor's measurements of TI to those made by a FDA-cleared reference device, indicate that measurements of TI made by these systems are substantially equivalent.
> Both the Sensor and the Corventis monitors use the same impedance-based technology to measure RR. Our validations efforts, which compare the Sensor's measurements of RR to those made by a FDA-cleared reference device, indicate that measurements of RR made by these systems are substantially equivalent.
> Both the Sensor and the Corventis monitor use standard technologies to measure ECG waveforms and, from these, calculate HR and HRV. The Sensor and the Corventis monitor also both use conventional tri-axial accelerometer sensors to measure motion-related properties of the patient, such as posture. The Sensor and the Corventis monitor both use a conventional Bluetooth wireless transceiver to send numerical and waveform information to a gateway device. And our Gateway and that used with the Corventis monitor both use a cellular transmitter or, alternatively, 802.11 to send information to a web-based system. Based on this, we do not believe our core technologies outside of those used for TI and RR raise any new questions of safety or effectiveness.
> The Sensor measures TEMP with a thermal sensor in electrical contact with one of the Electrodes. During a measurement, the Electrode adheres to the patient's underlying skin, and within a few
{5}------------------------------------------------
minutes these components reach a thermal equilibrium. After this point the thermal sensor measures TEMP. The ViSi monitor measures skin temperature in a nearly identical manner using a small plastic component in its chest-worn sensor. Based on the commonality of these approaches, we do not believe our core technology for measuring TEMP raises any new questions of safety or effectiveness.
The Sensor, when compared to both the Corventis and Zoe monitors, has a different form factor and electrode configuration, and uses a different mechanism for securing its Electrodes. The Corventis monitor features a single, integrated patch with four electrode regions. The Zoe monitor features a remote, stand-alone console that connects to a pair of dual-region, body-worn electrodes through a pair of cables. The cables feature standard mechanical snaps to connect the electrodes. The console contains all measurement and data-processing electronics. The Zoe monitor is typically used daily and resides with the patient.
In contrast, the Sensor is shaped like a conventional necklace, and drapes around the patient's neck during a measurement. It is designed for one-time, daily use, and typically resides permanently with the patient. 2 disposable Electrodes, each containing 2 electrode regions, connect through a magnetic interface to the Sensor's base, and then attach to the patient. The Sensor's necklaceshaped form factor ensures that, on a daily basis, the Electrodes are positioned consistently on the patient.
Thus, we believe the mechanical differences between the Sensor and its predicates -i.e. the necklace-shaped form factor and the mechanism in which it supports electrodes- do not raise any new questions of safety.
#### Clinical and Non-Clinical Testing
toSense studied the CoVa™ Monitoring System (the test device) and its measurement of thoracic impedance with a clinical trial featuring 33 subjects receiving chronic maintenance hemodialysis and, in some cases, suffering from congestive heart failure. Table 1, below, refers to this study as 'Clinical Study 1'. Measurements were made on all subjects while they were receiving hemodialysis. The trial compared thoracic impedance measured with the test device to a similar impedance parameter measured simultaneously by a FDA-cleared reference device. Trained personnel applied both test and reference devices to subjects. Table 1 summarizes the comparison of the test and reference devices' impedance values using the rank-regression method, as investigated during Clinical Study 1. Additional analysis demonstrated an average Pearson's correlation between the test device's measurement of impedance and fluid removed by a machine performing the hemodialysis of r = 0.93 (p < 0.0001), and that the test device's measurement sensitivity was 1.72 Ohms/Liter. Studies estimate the test device's measurement precision for thoracic impedance when used for long-term monitoring, as calculated from the standard deviation of repeated measurements, may be in the order of magnitude of 1~2 Ohms. Separate studies estimate the test device's measurement precision for thoracic impedance when used for short-term monitoring, calculated in a similar manner, may be less than 1 Ohm.
toSense also studied the test device's measurement of thoracic impedance with a clinical trial featuring 23 subjects, all having fluid-management issues, such as congestive heart failure. Table 1 refers to this study as 'Clinical Study 2'. Here, both test and reference devices measured impedance values daily in the home environment over an extended period of time. Trained personnel applied the reference device to all subjects. In some cases, trained personnel also applied the test device; in others, the subjects applied the test device. Table 1 summarizes the comparison of the test and reference devices' impedance values using the rank-regression method, as investigated during Clinical Study 2.
toSense also studied the test device and its measurement of respiration rate with a clinical trial featuring 19 adult subjects undergoing breathing patterns with variable rates and tidal volumes that
{6}------------------------------------------------
Page 4 of 5 simulate those encountered by intended-use populations. Table 1 refers to this study as 'Clinical Study 3'. The trial compared respiration rate values measured simultaneously with: 1) the test device; 2) a FDA-cleared monitor based on impedance pneumography (acceptance-standard device); and 3) a FDA-cleared monitor that measures end-tidal carbon dioxide and spriometry (reference device). Table 1 summarizes the comparison of the test and reference devices' impedance values using the rank-regression method, as investigated during Clinical Study 3.
K142087
Finally, toSense studied the test device's measurement accuracy of thoracic impedance, respiration rate, heart rate, heart rate variability, and skin temperature using bench studies featuring calibrated simulators. Table 1 refers to these studies as 'Clinical Studies 4a-e', respectively. Analysis of results from these studies, as determined using a standard regression model, are included in Table 1.
| Clinical<br>Study | Parameter<br>Evaluated | N | Subject Type | Slope from<br>Regression<br>Model | Intercept<br>from<br>Regression<br>Model |
|-------------------|--------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------------------|
| 1 | TI (compared to<br>reference device) | 33 subjects, 13<br>test devices,<br>500<br>measurements | undergoing<br>chronic<br>hemodialysis; may<br>have CHF | 0.86 | 35.8 |
| 2 | TI (compared to<br>reference device) | 23 subjects, 2<br>test devices,<br>560<br>measurements | fluid-management<br>issues | 0.65 | 117 |
| 3 | RR (compared to<br>reference device) | 19 subjects, 2<br>test devices,<br>682<br>measurements | healthy volunteers<br>using breathing<br>patterns<br>encountered by<br>intended-use<br>populations | 1.00 | 1.62 |
| 4a | TI (absolute<br>accuracy) | 8 devices, 88<br>measurements | simulator | 1.00 | 0.09 |
| 4b | RR (absolute<br>accuracy) | 1 device, 15<br>measurements | simulator | 1.02 | -0.60 |
| 4c | HR (absolute<br>accuracy) | 1 device, 67<br>measurements | simulator | 1.00 | -0.41 |
| 4d | HRV (absolute<br>accuracy) | 1 device, 15<br>measurements | simulator | 1.02 | -2.53 |
| 4e | TEMP<br>(absolute<br>accuracy) | 1 device, 40<br>measurements | simulator | 0.99 | 0.35 |
Table 1: Summary of regression results for CoVa's measurements, as determined by Clinical Studies 1-4.
Performance
Components of the CoVa" Monitoring System have been tested and meet the requirements of the
{7}------------------------------------------------
| | K142087<br>Page 5 of 5 |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Testing | relevant sections of the following performance standards: |
| | IEC 60601-1 Standard: Medical Electrical Equipment - Part 1: General Requirements for<br>Basic Safety and Essential Performance (2005) |
| | IEC 60601-1-2 Standard: Medical Electrical Equipment - Part 1-2: General Requirements<br>for Basic Safety and Essential Performance, Collateral Standard: Electromagnetic<br>Compatibility (2007) |
| | 60601-1-6 Standard: Medical Electrical Equipment - Part 1-6: General Requirements for<br>Basic Safety and Essential Performance, Collateral Standard: Usability; and IEC 62366<br>Standard: Medical Devices – Application of Usability Engineering to Medical Device |
| | AAMI/ANSI EC13: Cardiac monitors, heart rate meters, and alarms (2002) |
| | ISO 80601-2-56 Medical electrical equipment - Part 2-56: Particular requirements for basic safet<br>and essential performance of clinical thermometers for body temperature measurements |
| Conclusion | The CoVaTM Monitoring System has been tested and found to comply with recognized<br>performance, safety, and electromagnetic compatibility standards for medical devices. Clinical<br>studies indicate its measurements are substantially equivalent to those made by its predicate<br>devices. |
| | Based on the above, we believe the CoVaTM Monitoring System is as safe and effective as its<br>predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.