K113187 · Corventis, Inc. · MHX · Jan 4, 2012 · Cardiovascular
Device Facts
Record ID
K113187
Device Name
AVIVO MOBILE PATIENT MANAGEMENT
Applicant
Corventis, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jan 4, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The AVIVO Mobile Patient Management System is intended to continuously record, store, and periodically transmit physiological data. The System is indicated for patients: - with fluid management problems . - taking diuretic medication - living with heart failure . - living with end-stage renal disease - suffering from recurrent dehydration - who require monitoring for the detection of non-lethal cardiac arrhythmias such . as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradvarrhythmias and conduction disorders. The AVIVO Mobile Patient Management System also monitors, derives and displays: - . ECG - Heart Rate (including HR variability) . - Activity . - . Posture - . Respiration rate (including RR variability) - Body fluid status .
Device Story
Wearable, wireless system for continuous ECG and physiological monitoring. Components: PiiX (adherent torso-worn sensor) and zLink (handheld gateway). PiiX records ECG, heart rate, activity, posture, respiration, and body fluid status; transmits data via Bluetooth to zLink; zLink transmits to Corventis Server via cellular network. Corventis cardiographic technicians review data and provide reports to prescribing physicians. Used in clinical or home settings for up to 7 days per PiiX application. Enables arrhythmia detection and fluid status monitoring to assist clinical decision-making for patients with heart failure, renal disease, or dehydration. Benefits include remote monitoring of cardiac and physiological status without requiring constant in-clinic presence.
Clinical Evidence
Bench testing only. Device performance verified against AAMI/ANSI EC38 (ambulatory ECG systems) and AAMI/ANSI EC57 (cardiac rhythm and ST-segment measurement algorithms). Compliance with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC) confirmed.
Technological Characteristics
Wearable, wireless system. Components: PiiX (adherent torso-worn sensor) and zLink (gateway). Connectivity: Bluetooth (PiiX to zLink), cellular (zLink to server). Monitors ECG, HR, activity, posture, respiration, and fluid status. Standards: IEC 60601-1, IEC 60601-1-2, AAMI/ANSI EC38, AAMI/ANSI EC57.
Indications for Use
Indicated for patients with fluid management problems, those on diuretics, heart failure patients, end-stage renal disease patients, those with recurrent dehydration, and patients requiring monitoring for non-lethal cardiac arrhythmias (e.g., SVTs, atrial fibrillation, atrial flutter, ventricular ectopy, bradyarrhythmias, conduction disorders).
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
#### JAN - 4 2012
## ECTION 7 - 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
| Submitter Name & Address: | Corventis, Inc.<br>1410 Energy Park Drive, Suite 1<br>St. Paul, MN 55108 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Michele Chin-Purcell, Ph.D.<br>651-925-3803 (phone)<br>651-389-3251 (fax) |
| Trade/Proprietary Name: | AVIVO® Mobile Patient Management System |
| Common/Usual Name: | Mobile Patient Management System |
| Classification Name: | Arrhythmia Detector and Alarm<br>(21 CFR 870.1025, Product Code DSI)<br>Patient Physiological Monitor (with arrhythmia<br>detection)<br>(21 CFR 870.1025, Product Code MHX) |
| Class: | Class II, Special Controls |
| 510(k): | Abbreviated 510(k) |
| Date Prepared: | October 28, 2011 |
#### Predicate Devices:
- 1. AVIVO Mobile Patient Management System, Corventis, Inc. cleared by FDA under 510(k) numbers K083287, K090696, and K091971; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 2. NUVANT® Mobile Cardiac Telemetry System, Corventis, Inc. cleared by FDA under 510(k) number K090696, K091971, and K111917; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 3. ZOE Fluid Status Monitor, Noninvasive Medical Technologies, LLC, cleared by FDA under K042113; 21 CFR 870.2770, DSB "Impedance Plethysmograph"
{1}------------------------------------------------
# BACKGROUND INFORMATION
The following is the subject device of this Abbreviated 510(k):
- . The AVIVO Mobile Patient Management System (hereafter referred to as "subject AVIVO")
This submission proposes to add clarifying language (e.g., listing the type of non-lethal arrhythmias) to the predicate AVIVO indications for use. This clarifying language is similar to that used in the predicate NUVANT device. Additional language is being proposed (e.g.: when fluid status measurements may be useful) to align with the predicate ZOE indications for use. These changes are for clarification and do not change to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device, and do not affect the safety and effectiveness of the device when used as labeled [807.92(a)(5)].
The intent of the subject devices remains the same as that of the predicate devices, which is to monitor ECG, heart rate and other physiological parameters such as body fluid status.
# INDICATION FOR USE STATEMENT
The following are the proposed indications for use in this submission:
The AVIVO Mobile Patient Management System is intended to continuously record, store, and periodically transmit physiological data. The System is indicated for patients:
- with fluid management problems .
- . taking diuretic medication
- living with heart failure .
- . living with end-stage renal disease
- . suffering from recurrent dehydration
- who require monitoring for the detection of non-lethal cardiac arrhythmias such . as, but not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial
{2}------------------------------------------------
flutter, paroxysmal SVTs), ventricular ectopy, bradvarrhythmias and conduction disorders.
The AVIVO Mobile Patient Management System also monitors, derives and displays:
- . ECG
- Heart Rate (including HR variability) .
- Activity .
- . Posture
- . Respiration rate (including RR variability)
- Body fluid status .
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The AVIVO System is wearable, wireless arrhythmia detection system that is used by clinicians to identify suspected cardiac arrhythmias and monitor physiologic signals. In combination with interpretation services provided by learned intermediaries in the Corventis Monitoring Center as well as online review of data (for prescribing physicians only), AVIVO enables arrhythmia detection and other physiological data monitoring for up to 7 days for each PiiX application.
The AVIVO system components are:
- . PiiX® (aka: Adherent Device) - a patient-worn device which is applied to the patient's torso. It contains the ECG electrodes for recording ECG, heart rate and other physiological data.
- . zLink® (aka: Gateway) – hand-held device that receives information from the PiiX and transmits it to the Corventis Server via cellular technology.
- Server The Server receives sensor data from the PiiX via zLink. ECG, heart . rate and other physiological parameters are presented to learned intermediaries, Corventis cardiographic technicians, who prepare and deliver the information to prescribing physicians
The communication between the PiiX and the zLink is enabled via BlueTooth Technology. Sensor data and ECGs collected by the PiiX are transmitted to the Server via zLink.
The AVIVO and NUVANT system components are identical expect for the use of a Patient Trigger Magnet, which is available for the patient to manually trigger the ECG
{3}------------------------------------------------
collection when he/she experiences symptoms when using the NUVANT system. The system components of the subject AVIVO are identical to the system components of the predicate AVIVO and predicate NUVANT.
## PREDICATE DEVICES
Three (3) predicate devices have been identified:
- 1. AVIVO Mobile Patient Management System, Corventis, Inc. cleared by FDA under 510(k) numbers K083287, K090696, and K091971; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 2. NUVANT® Mobile Cardiac Telemetry System, Corventis, Inc. cleared by FDA under 510(k) number K090696, K091971, and K111917; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 3. ZOE Fluid Status Monitor, Noninvasive Medical Technologies, LLC, cleared by FDA under K042113; 21 CFR 870.2770, DSB "Impedance Plethysmograph"
The following table summarizes the compared features and the corresponding predicate devices.
| Features being compared | Predicate Devices |
|-------------------------|----------------------------------------------|
| All technical features | Predicate AVIVO (K083287, K090696, K091971) |
| All technical features | Predicate NUVANT (K090696, K091971, K111917) |
| Body Fluid Status | Predicate Zoe (K042113) |
| Indications for Use | Predicate NUVANT (K111917) |
| Indications for Use | Predicate ZOE (K042113) |
## Features comparison and corresponding predicate devices
## SUMMARY OF PERFORMACE TESTING
The subject and predicate AVIVO and predicate NUVANT meet the requirements of following performance standards in accordance with FDA Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm.
{4}------------------------------------------------
#### IEC 60601-1
Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995
## IEC 60601-1-2
Medical Electrical Equipment - Part 1-2: General Requirements for Safety -Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004))
## AAMI/ANSI EC38
Medical electrical equipment - Part 2-47: Particular requirements for the safety, including essential performance, of ambulatory electrocardiographic systems, 2007.
## AAMI/ANSI EC57
Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms, 1998/(R) 2008
# CONCLUSION
The modifications proposed in the subject AVIVO System make clarifications to the indications for use and do not change the fundamental scientific technology or use of the devices. As supported by the descriptive information and the bench tests, it is concluded that the AVIVO Systems are as safe and as effective as the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN - 4 2012
Corventis, Inc. c/o Kathleen Lundberg VP of Ouality, Regulatory, and Clinical Affairs 1410 Energy Park Drive, Suite #1 St. Paul, MN 55108
Re: K113187
Trade/Device Name: AVIVO Mobile Patient Management System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX, DSB, DSI Dated: October 28, 2011 Received: October 31, 2011
Dear Ms. Lundberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 -- Ms. Kathleen Lundberg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram Zuckerman, MD
Bram Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation -Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use
510(k) Number (if known): NA K || 3 | 8 7
Device Name: AVIVO® Mobile Patient Management
Indications for Use:
The AVIVO Mobile Patient Management System is intended to continuously record, store, and periodically transmit physiological data. The System is indicated for patients:
- with fluid management problems .
- taking diuretic medication .
- living with heart failure .
- living with end-stage renal disease .
- suffering from recurrent dehydration .
- who require monitoring for the detection of non-lethal cardiac arrhythmias such as, but . not limited to, supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, paroxysmal SVTs), ventricular ectopy, bradyarrhythmias and conduction disorders.
The AVIVO Mobile Patient Management System also monitors, derives and displays:
- ECG .
- Heart Rate (including HR variability) .
- . Activity
- . Posture
- Respiration rate (including RR variability) �
- Body fluid status �
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Page 1 of 1
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K113187 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.