NUVANT CARDIAC EVENT MONITORING SYSTEM; NUVANT MOBILE CARDIAC TELEMETRY SYSTEM
K090696 · Corventis, Inc. · QYX · Jun 19, 2009 · Cardiovascular
Device Facts
Record ID
K090696
Device Name
NUVANT CARDIAC EVENT MONITORING SYSTEM; NUVANT MOBILE CARDIAC TELEMETRY SYSTEM
Applicant
Corventis, Inc.
Product Code
QYX · Cardiovascular
Decision Date
Jun 19, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The NUVANT™ Cardiac Event Monitor (CEM) System and the NUVANT™ Mobile Cardiac Telemetry (MCT) System are intended to continuously measure, record, and periodically transmit physiological data. The Systems are indicated for those patients who require monitoring for the detection of non-lethal cardiac arrhythmias. Both NUVANT system models monitor, derive and display: ECG Heart rate The Systems may also monitor, derive and display: Activity Posture Respiration rate (including RR Variability) Body fluid status Heart rate variability
Device Story
System consists of PiiX (patient-worn adherent device), zLink (gateway), and secure server. PiiX monitors ECG, activity, posture, respiration, fluid status, and heart rate variability. ECG data recorded via baseline collection, detection algorithms, or patient-activated magnet (45s total: 15s pre-event, 30s post-event). PiiX transmits data to zLink via Bluetooth; zLink transmits to server via cellular network. Server derives physiological parameters, displays trends/ECG waveforms, and provides visual notifications for threshold breaches. Healthcare professionals access data via standard browsers to acknowledge/dismiss events and select CEM vs. MCT modes. System aids clinical decision-making by providing continuous monitoring and arrhythmia detection for patients requiring cardiac surveillance.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
System components: PiiX (adherent torso-worn device), zLink (gateway), and server. Sensing: ECG, activity, posture, respiration, fluid status. Connectivity: Bluetooth (PiiX to zLink), cellular (zLink to server). Data access: Web-based browser. Software: Server-side processing for parameter derivation and threshold-based notifications.
Indications for Use
Indicated for patients requiring monitoring for the detection of non-lethal cardiac arrhythmias.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
SJM Confirm Model DM2100 Implantable Cardiac Monitor and Model DM 2100A Patient Activator (K081365)
Cardionet Ambulatory ECG Monitor with Arrhythmia Detections, Model CN1004 (K063222)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and consists of a stylized eagle. The FDA logo is on the right and consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 6, 2023
Corventis, Inc. Dawn Chang Senior Regulatory Affairs Associate 2226 N. First Street San Jose, California 95131
## Re: K090696
Trade/Device Name: Nuvant Cardiac Event Monitoring System; Nuvant Mobile Cardiac Telemetry System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: QYX, DSI
Dear Dawn Chang:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 19, 2009. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, Jennifer.Kozen@fda.hhs.gov.
Sincerely.
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows a logo for the U.S. Department of Health and Human Services. The logo features a stylized human figure with three flowing lines extending from it. The text "HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the logo. The logo is black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
19 2009 . IIIN
Corventis, Inc. c/o Ms. Dawn Chang Senior Regulatory Affairs Associate 2226 N. First Street San Jose, CA 95131
Re: K090696
> Trade/Device Name: NUVANTTM Cardiac Event Monitor System, and NUVANT™ Mobile Cardiac Telemetry System
Regulation Number: 21 CFR 870:1025
Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm)
Regulatory Class: Class II (Two)
Product Code: DSI
Dated: May 19, 2009
Received: May 20, 2009
## Dear Ms. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. I i
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Ms. Dawn Chang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240). 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
er R. haines
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{4}------------------------------------------------
# KO90696
## Indications for Use
## 510(k) Number (if known):
Device Names:
NUVANT™ Cardiac Event Monitor System NUVANT™ Mobile Cardiac Telemetry System
## Indications for Use:
The NUVANT" Cardiac Event Monitor (CEM) System and the NUVANT "Mobile Cardiac Telemetry (MCT) System are intended to continuously measure, record, and periodically transmit physiological data. The Systems are indicated for those patients who require monitoring for the detection of non-lethal cardiac arrhythmias. Both NUVANT system models monitor, derive and display:
- ECG
- Heart rate
The Systems may also monitor, derive and display:
- Activity
- Posture
- Respiration rate (including RR Variability)
- Body fluid status
- Heart rate variability
Prescription Use X (Part 21 CFR 801 Subpart D) :
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page / of
uriel D. Vorhies
(Division Sign-Off) Division of Cardrovascular Devices
510(k) Number_K090696
{5}------------------------------------------------
## 510(k) Summary of Safety & Effectiveness
50000
| Submitter Name & Address: | Corventis, Inc.<br>2226 N. First Street<br>San Jose, CA 95131 | JUN 19 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Contact Person: | Madhuri Bhat, VP, Clinical & Regulatory Affairs<br>408-790-9306 (phone)<br>408-790-9350 (fax) | |
| Trade/Proprietary Name: | NUVANT™ Cardiac Event Monitor System<br>NUVANT™ Mobile Cardiac Telemetry System | |
| Common/Usual Name: | Cardiac Event Monitor<br>Mobile Cardiac Telemetry | |
| Classification Name: | Arrhythmia Detector and Alarm<br>(21 CFR 870.1025, Product Code DSI)<br>Patient Physiological Monitor (with arrhythmia<br>detection)<br>(21 CFR 870.1025, Product Code MHX) | |
| Class: | Class II, Special Controls | |
| 510(k): | Special 510(k): Device Modification | |
| | | |
Predicate Devices:
- 1. AVIVO Mobile Patient Management System, Corventis, Inc., cleared by FDA under 510(k) number K083287; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 2. E-Tac EX-1000 electrocardiographic Event Recorder, Datrix, Inc., cleared by FDA under 510(k) number K042022; 21 CFR 870.2920, DXH "Telephone electrocardiographic transmitter and receiver"
- 3. SJM Confirm Model DM2100 Implantable Cardiac Monitor and Model DM 2100A Patient Activator, St. Jude Medical, CRMD, cleared by FDA under 510(k) number K081365; 21 CFR 870.2800, MXC "Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection)"
- 4. Cardionet Ambulatory ECG Monitor with Arrhythmia Detections, Model CN1004, Cardionet, Inc., cleared by FDA under 510(k) number K063222: CFR 870.1025, DSI "Arrhythmia Detector and Alarm"
Page 1 of 4
{6}------------------------------------------------
## Indication for Use Statement:
The NUVANT™ Cardiac Event Monitor (CEM) System and the NUVANT™ Mobile Cardiac Telemetry (MCT) System are intended to continuously measure, record and periodically transmit physiological data. The Systems are indicated for those patients who require monitoring for the detection of non-lethal cardiac arrhythmias. Both NUVANT system models monitor, derive and display:
- . ECG
- Heart Rate ●
The Systems may also monitor, derive and display:
- Activity ●
- . Posture
- Respiration rate (including RR variability) .
- Body fluid status .
- Heart rate variability .
## Technological Characteristics and Substantial Equivalence
The NUVANT™ Cardiac Event Monitor System and the NUVANT™ Mobile Cardiac Telemetry System include the following components:
- PiiX™ (a patient worn adherent device) with Magnet .
- zLink .
- Server �
The PiiX is a patient-worn device which is applied to the patient's torso. The Adherent Device monitors the patient's ECG in a looping mode. There are three circumstances under which the ECG data will be recorded: (i) when the device is first applied and activated, a baseline ECG is collected, (ii) when the detection algorithms residing in the Adherent Device trigger the collection, or (iii) when the patient activates the collection using the magnet. When the ECG collection is triggered, a total of 45 seconds of ECG signals will be collected, with 15 seconds of pre-event and 30 seconds post-event. Same as the AVIVO™ Mobile Patient Management System (predicate device), the PiiX has additional sensors which monitor other physiological parameters described in the Indications for Use Statement. The collected ECG signals and the other sensor data will be transmitted to the Server via the zLink.
The zLink receives information from the PiiX and transmits them to the Corventis Server. It also interacts with the Corventis Server to receive configuration updates and other relevant hardware diagnostic information.
{7}------------------------------------------------
The Server of the CEM/MCT receives information from the PiiX via zLink. The secure server performs the following functions:
- Derive physiological parameters using the raw data collected by the Adherent . Device.
- Display the physiological parameters in trend graphs format. .
- . Display ECG waveform when the heart rates are beyond the specified threshold.
- . Provide visual notifications when healthcare professionals need to be aware of heart rates that are beyond the specified threshold.
- . Provide the users the ability to select CEM vs. MCT prescription.
- Provide the users the ability to acknowledge or dismiss events. .
The communication between the PiiX and the zLink is enabled via the BlueTooth™ Technology. The zLink transmits the data to the Server via cellular technology, where healthcare professionals can access with standard browsers.
Four (4) predicate devices have been identified for the various aspects of the CEM System and MCT System. They are:
- l . AVIVO Mobile Patient Management System, Corventis, Inc., cleared by FDA under 510(k) number K083287; 21 CFR 870.1025, DSI "Arrhythmia Detector and Alarm", and 21 CFR 870.1025, MHX "Patient Physiological Monitor (with arrhythmia detection)"
- 2. E-Tac EX-1000 electrocardiographic Event Recorder, Datrix, Inc., cleared by FDA under 510(k) number K042022: 21 CFR 870.2920. DXH "Telephone electrocardiographic transmitter and receiver"
- 3. SJM Confirm Model DM2100 Implantable Cardiac Monitor and Model DM 2100A Patient Activator, St. Jude Medical, CRMD, cleared by FDA under 510(k) number K081365; 21 CFR 870.2800, MXC "Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection)"
- 4. Cardionet Ambulatory ECG Monitor with Arrhythmia Detections, Model CN1004, Cardionet, Inc., cleared by FDA under 510(k) number K063222; CFR 870.1025, DSI "Arrhythmia Detector and Alarm"
{8}------------------------------------------------
| Comparables | Predicate Devices for CEM | Predicate Devices for MCT |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Patient Trigger<br>Feature | E-Tac EX-1000 electrocardiographic Event Recorder<br>(K042022) | |
| Mechanism of the<br>Patient Trigger<br>Feature | SJM Confirm Model DM2100 Implantable Cardiac Monitor<br>and Model DM 2100A Patient Activator (K081365) | |
| All Other Features | AVIVO Mobile Patient Management System (K083287) | |
| Overall Product | E-Tac EX-1000<br>electrocardiographic Event<br>Recorder (K042022) | Cardionet Ambulatory ECG<br>Monitor with Arrhythmia<br>Detections, Model CN1004<br>(K063222) |
## Conclusions
The NUVANT™ Cardiac Event Monitor (CEM) System and the NUVANT™ Mobile Cardiac Telemetry (MCT) System models have the same intended use, similar operating principles and technological characteristics as their predicate devices. As supported by the descriptive information and the design verification tests, it is concluded that the NUVANT™ Cardiac Event Monitor (CEM) System and the NUVANT™ Mobile Cardiac Telemetry (MCT) System are as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.