K141919 · Ad Rem Technology Sarl · NGX · Jun 16, 2015 · Physical Medicine
Device Facts
Record ID
K141919
Device Name
VEINOPLUS SPORT NEUROMUSCULAR STIMULATOR
Applicant
Ad Rem Technology Sarl
Product Code
NGX · Physical Medicine
Decision Date
Jun 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VEINOPLUS® SPORT is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
Portable, battery-powered neuromuscular stimulator; delivers low-voltage, low-frequency electrical pulses via self-adhesive electrodes to calf muscles or other body parts (arms, buttocks, back, shoulder, abdomen). Device features single operating mode stored in internal memory; operated by user in home, gym, or clinical settings. Output is monophasic rectangular asymmetrical waveform. User controls intensity; device provides 30-minute stimulation session. Output facilitates muscle contraction; intended to improve muscle performance. Benefits include non-invasive muscle stimulation for healthy users.
Clinical Evidence
Bench testing only. Includes electrical safety testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10), biocompatibility testing of electrode hydrogel (ISO 10993-5, ISO 10993-10), and software verification per FDA guidance. A usability study was conducted to confirm safe and effective operation by the user.
Technological Characteristics
Powered by 9V alkaline battery. Monophasic rectangular asymmetrical waveform; pulse duration 25-240 µs; max output 50V/100mA at 500 Ohm. Single-channel output. Plastic enclosure. Connectivity: None (standalone). Sterilization: Not applicable (electrodes are self-adhesive). Software: Pre-set program encoded in microchip. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and ISO 14971.
Indications for Use
Indicated for healthy individuals to stimulate muscles to improve and facilitate muscle performance.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three faces in profile, suggesting care and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 15, 2015
Ad Rem Technology SARL c/o Norman F. Estrin, PhD Estrin Consulting Group, LLC 9109 Copenhaver Drive Potomac, MD 20854
Re: K141919
Trade/Device Name: Veinoplus® Sport Neuromuscular Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX, GXY Dated: May 18, 2015 Received: May 19, 2015
Dear Dr. Estrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K141919
Device Name
VEINOPLUS® SPORT Neuromuscular Stimulator
Indications for Use (Describe) The VEINOPLUS® SPORT is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo and contact information for Ad Rem Technology. The logo features the company name in a stylized font, with the words "Ad Rem" in a larger font size and a blue color, and the word "TECHNOLOGY" in a smaller font size. Below the logo is the text "ISO 13485:2003 and ISO 9001:2008", followed by the address "162, rue du Faubourg Saint-Honoré, 75008 Paris - France", telephone number "+33 (0)1 42 60 00 22", fax number "+33 (0)1 42 60 00 63", email address "info@adremtechnology.com", and website address "www.adremtechnology.fr".
# 510(k) SUMMARY (Clean) Ad Rem Technology Veinoplus Sport Neuromuscular Stimulator
| Sponsor identification | Ad Rem Technology SARL<br>162 Rue du Faubourg Saint- Honoré<br>75008 Paris, France<br>Ms. Noémie Dumérain, Quality and Regulatory Affairs Manager,<br>nd@veinoplus.com<br>Tel: + 33 (0)1.42.60.00.22; Fax: + 33 (0)1.42.60.00.63 |
|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>registration number: | 3005671018 |
| Date of preparation | June 13, 2014 |
| Contact person | Norman F. Estrin, Ph.D.<br>Managing Partner<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Email: estrin@yourFDAconsultant.com |
| Authorized Agent in<br>the United States | Norman F. Estrin, Ph.D.<br>Managing Partner<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Email: estrin@yourFDAconsultant.com |
| Proprietary Name | Veinoplus® Sport Neuromuscular Stimulator |
| Trade name | VEINOPLUS® SPORT |
| Device Classification<br>Name, Product Code,<br>Reviewing Panel,<br>Regulatory Class,<br>(Regulation) | Device, Stimulator, Muscle, Powered, For Muscle Conditioning<br>NGX Physical Medicine, Class II (21 CFR 890.5850)<br>Electrode, cutaneous<br>GXY Neurology, Class II (21 CFR 882.1320) |
| Indications for<br>use: | The VEINOPLUS® SPORT is intended to stimulate healthy<br>muscles in order to improve and facilitate muscle performance. |
{4}------------------------------------------------
| Device Description | Design and features: |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The VEINOPLUS® SPORT, like a number of legally marketed predicate devices is a traditional battery powered muscle stimulator. It is portable, hand-held device, equipped with one single channel intended to stimulate the calf muscles or other muscles of the body (for example thighs, arms, buttock, back, shoulder or abdomen). The VEINOPLUS® SPORT device has only one operating mode, called program, stored in the internal memory.<br><br>The output stimulus is low-voltage, low frequency with rectangular voltage waveform. The device is housed in a plastic enclosure, and it is powered by an alkaline 9V primary battery. The accessories include the output cables, battery, user manual and electrode pads. No provisions are made for AC adapters, so the device cannot be connected to line under any circumstances.<br><br>The VEINOPLUS® SPORT device is supplied with all accessories (as listed below) necessary for a safe and effective session of stimulation. Main unit, cable, the pair of electrodes, battery, lanyard and user manual are packaged in carrying case then packaged in a carton box with printed messages. Refer to the picture in Section 12 Device Description.<br><br>List of accessories:<br>- 1pair of electrodes VEINOPLUS® Pack. Self-adhesive. Oval shape. Size 8x13 cm. Low resistance. This includes a Hydrogel for the electrodes (Refer to Sections 19 and 23F).<br>- 1 conductor cable for connecting the electrodes. Approximate length (w/o plug) 135 cm.<br>- 1 instruction manual<br>- 1 9V alkaline battery<br>- 1 lanyard<br>- 1 carrying case: L22.5*W4.5*H17cm. |
| Predicate Devices | Veinoplus® (K072252)<br>Compex® Sport Plus (K083140) |
## Comparison of the indications for use with the predicate devices:
The proposed device, VEINOPLUS® SPORT, is substantially equivalent to both of its legally marketed Class II predicate devices:
- 1. The mode 1 (powered muscle stimulator, IPF) of the Veinoplus® with 510k K072252 (except for the indications for use).
- 2. The mode (program) "active recovery" of the Compex® Sport Plus with 510k K083140.
{5}------------------------------------------------
### Summary of technological characteristics of VEINOPLUS® SPORT with its two predicates Veinoplus® and Compex Sport Plus:
| Device name: | VEINOPLUS® SPORT | Veinoplus® (mode 1 of stimulation) | Compex® Sport Plus (Program Active Recovery) |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | K141919 | K072252 | K083140 |
| Intended use: | Muscle electrical stimulation | Muscle electrical stimulation | Muscle electrical stimulation |
| Indications for use: | The VEINOPLUS® SPORT is intended to stimulate healthy muscles in order to improve and facilitate muscle performance. | Increasing local blood circulation (common claim with predicates). Relaxation of muscle spasm/ Prevention or retardation of muscle disuse atrophy/ Muscle reeducation/ Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis/Maintaining or increasing range of motion. | Stimulate healthy muscles in order to improve or facilitate muscle performance. |
| Mode of Action: | The VEINOPLUS® SPORT contracts the muscles by electrical stimulation. | The Veinoplus stimulates the muscles, resulting in rhythmic muscles contraction, which then increases the blood flow. | The Active Recovery program produces muscle twitches at a very low frequency. Those twitches act like a massage and induce an increase of the blood flow and a faster reduction of the lactic acid blood level. It facilitates recovery of the stimulated muscles after active muscle training or competition. |
| Purpose: | Non-Medical | Medical | Non-Medical |
| Target Population: | Healthy user | Medical patient | Healthy user |
| Parts to be stimulated: | Calf muscles and other body parts (arms, buttocks, back (neck/shoulders). | Calf muscles | Legs/feet, arms/hands, buttock, back, shoulder or abdomen) |
| Where used: | Home, hospital, gym, etc. | Home, hospital, gym, etc. | Data unknown |
| Device name: | VEINOPLUS® SPORT | Veinoplus® (mode 1 of<br>stimulation) | Compex® Sport Plus<br>(Program Active Recovery) |
| OTC/Rx Only<br>information : | OTC use | Prescription use | OTC use |
| Energy<br>delivered: | Electrical | Electrical | Electrical |
| Technical information: | | | |
| Power Source: | One Internal 9 V alkaline<br>battery.<br>Types: 522, 6LR61,<br>MN1604<br>Shape: rectangular<br>Size: 48x25x16 mm;<br>Volume: 20 cm³ | One Internal 9 V alkaline<br>battery<br>Types: 522, 6LR61,<br>MN1604<br>Shape: rectangular<br>Size: 48x25x16 mm;<br>Volume: 20 cm³ | One NIMH rechargeable<br>battery ( 4.8V ≥1200 mA/h )<br>(4 cells pack of 1.2V)<br>Type: 4H-AA1500<br>Shape: cylindrical<br>Size: 1 cell = 14 mm x<br>47mm , diameter 14 mm;<br>Volume: 7.3 cm³ |
| Number of<br>output modes | 1 mode/program | 2 modes/programs :<br>Mode 1 : EMS<br>Mode 2 : TENS | 9 modes/programs :<br>Potentiation<br>Endurance<br>Strength<br>Resistance<br>Active recovery<br>Recovery plus<br>Explosive strength<br>Pre warm-up<br>Massage |
| Timer range | 30 minutes | Mode 1: 20 minutes<br>Mode 2: 20 minutes | Potentiation : 3' 27''<br>Endurance: 50' 23''+ 5'03'' of<br>optional warm up<br>Strength: 22'07''23''+ 5'03'' of<br>optional warm up<br>Resistance: 27'33''23''+ 5'03''<br>of optional warm up<br>Active recovery: 26'40''23''+<br>5'03'' of optional warm up<br>Recovery plus: 24'20''<br>Explosive strength: 25'18''<br>Pre warm up: 25'03''<br>Massage: 20'03'' |
| Number of<br>output<br>channels: | 1 | 1 | 4 |
| Automatic<br>overload trip | NO | NO | NO |
| Automatic no<br>load trip | NO | NO | YES |
| Accessories: | Electrodes, cable, battery | Electrodes, cable, battery | Electrodes, cable, Battery |
| Device name: | VEINOPLUS® SPORT | Veinoplus® (mode 1 of stimulation) | Compex® Sport Plus (Program Active Recovery) |
| | | (same model and manufacture as Veinoplus Sport) | charger. |
{6}------------------------------------------------
{7}------------------------------------------------
| Waveform | Monophasic rectangular<br>asymmetrical | Monophasic rectangular<br>asymmetrical | Biphasic rectangular<br>symmetrical |
|--------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Max Output<br>Voltage | 50 V@ 500 Ohm | 50 V@ 500 Ohm | 60 V@ 500 Ohm (each<br>phase) |
| Max Output<br>Current | 100 mA@ 500 Ohm | 100 mA@ 500 Ohm | 120 mA @ 500 Ohm (each<br>phase) |
| Max Phase<br>Charge (µC) | 24μC @500Ω | 24μC @500Ω | 48µC (each phase)<br>@500Ω |
| Maximum<br>current density<br>(mA/cm2, r.m.s.) | <14 μΑ (rms) /cm2<br>@500Ω | <13 μΑ (rms)/cm2 @500Ω | Not available |
| Maximum<br>Average power<br>density (W/cm2,<br>r.m.s.) (using<br>smallest electrode<br>conductive<br>surface area) | <500 μW (r.m.s.) /cm2<br>(microwatt) @500 Ω<br>Only one size of<br>electrodes is permitted<br>and supplied by the<br>manufacturer. | <10 μW/cm2 (microwatt)<br>@500 Ω<br>Only one size of electrodes<br>is permitted and supplied by<br>the manufacturer. | <12 mW/cm2 (microwatt)<br>@500 Ω |
| Pulse duration | from 25 to 240 µs | from 25 to 240 µs | 400 µs (each phase) |
| Session duration: | 30 mn | 20 mn | 29 mn including warm-up |
| Automatic shut-<br>off | Yes at 30 mn | Yes at 20 mn | Yes, at the end of the<br>session |
| User override<br>Control | Yes | Yes | Yes |
| Software | Stimulation pre-set<br>program encoded in a<br>microchip | Stimulation pre-set program<br>encoded in a microchip | Stimulation pre-set<br>program encoded in a<br>microchip |
| Controls | 3 controls | 3 controls | Similar, 6 controls |
| Display/interface | LCD Screen and data<br>display | LCD Screen and data<br>display | LCD screen and similar<br>data. |
| Electrode<br>size/shape | 8x13 cm OVAL | 8x13 cm OVAL | 5x5 cm square and 5x10<br>cm rectangular |
{8}------------------------------------------------
| Compliance with safety and performance Standards: | | | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Biocompatibility | Electrodes material in<br>contact with skin user are<br>compliant with ISO<br>10993-5 and ISO 109993-<br>10 standards requirements | Electrodes material in<br>contact with skin user are<br>compliant with ISO<br>10993-5 and ISO 109993-<br>10 standards requirements | Data unknown |
| Electrical safety | Comply with IEC 60601-<br>1, IEC 60601-1-2 (EMC),<br>IEC 60601-2-10 | Comply with IEC 60601-<br>1, IEC 60601-1-2 (EMC),<br>IEC 60601-2-10 | Comply with IEC 60601-1,<br>IEC 60601-1-2 (EMC), IEC<br>60601-2-10 |
VEINOPLUS® SPORT is substantially equivalent to Veinoplus® in the Substantial Equivalence perspective of technological characteristics and general intended uses. Both summary: devices are intended for electrical stimulation of muscles. VEINOPLUS® SPORT was developed based on Veinoplus® device and both devices are technically very close. The indications for use of VEINOPLUS®, which is intended to be sold under prescription and for medical purpose, are not used for demonstration of substantial equivalence of VEINOPLUS® SPORT. VEINOPLUS® SPORT is substantial equivalent to Compex® Sport Plus in term of intended use, indications for use and technological characteristics. The technical differences between VEINOPLUS® SPORT and Compex® Sport Plus (power supply, waveform shape ...) do not raise additional issues of safety Compliance to applicable voluntary standards includes IEC 60601-1, IEC 60601-Summary of Non-1-2, IEC 60601-2-10 and ISO 14971. clinical The gel of electrodes was tested according ISO 10993-5 and ISO 10993-10 performance testing: Biocompatibility standards. In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA guidance for the content of premarket submissions for software contained in Medical Devices. A usability study support as well that the user can operate the device in a safe and effective manner with labeling supplied with the device. Conclusion: The safety and performance of the device is supported by testing with safety and performance standards IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-10 standards. The differences between VEINOPLUS® SPORT® and the predicate devices do not raise new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.