K141403 · Quanta System Spa · GEX · Aug 7, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141403
Device Name
LITHO DK30
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 7, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device Litho DK30 including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones.
Device Story
Litho DK30 is a desktop-version pulsed Holmium:YAG laser surgical instrument (30 W, 2.1 µm). It delivers laser energy via a fiber optic delivery system to perform soft tissue surgery and lithotripsy. Operated by clinicians in surgical settings (OR, clinic), the device features a touch-screen interface and an RFID system for automatic fiber diameter recognition. Modifications from the predicate (Litho K091909) include a new mechanical structure, compact power supply, improved cooling (pump and heat-exchanger), updated focusing optics, and revised control hardware/software. The device assists surgeons in precise tissue management and stone fragmentation, providing a more compact form factor for clinical use.
Clinical Evidence
Bench testing only. Verification and validation activities were conducted to demonstrate that the modified device meets the requirements of the third edition of IEC 60601-1 and its collateral standards. No clinical data was required or provided.
Technological Characteristics
Pulsed Holmium:YAG laser; 30 W max output power; 2.1 µm wavelength. Desktop form factor. Features: RFID fiber recognition system, touch-screen GUI, water-to-air heat exchanger, and water pump cooling. Complies with IEC 60601-1 (3rd ed).
Indications for Use
Indicated for patients requiring surgical incision, excision, resection, ablation, vaporization, coagulation, or hemostasis of soft tissue, or lithotripsy of stones. Applicable across medical specialties including Urology, Gastroenterology, Arthroscopy, Neurosurgery, Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery, and General Surgery. Used in open, laparoscopic, and endoscopic procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# 7. Special 510(K) SUMMARY - Device Modifications
## Introduction:
This document contains the 510(k) Summary for the device Litho DK30. The basis of this submission is Modifications to Device already cleared. The content of this summary is based on the requirements of 21 CFR 807.92(c).
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via IV Novembre, 116<br>Solbiate Olona (VA)<br>Italy, 21058 |
|--------------------------------------------------|--------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Maurizio Bianchi<br>Regulatory Affairs Manager<br>Quanta System SPA |
| | Email: maurizio.bianchi@quantasystem.com<br>Phone: +39 0331 376797<br>Fax: +39 0331 367815 |
| Date Prepared: | May 24th, 2014 |
| Device Name: | Litho DK30 |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology. |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Basis for Submission: | Device modifications |
| Legally Marketed Device | Litho (K091909) |
The modified device Litho DK30 is claimed to be derived from the legally marketed (unmodified) device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).
#### Performance Standards:
There are no mandatory performance standards for this device.
{1}------------------------------------------------
#### General Device Description:
The modified device Litho DK30 is a Laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).
This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909): Litho DK30 is the desktop version of the already cleared device Litho (K09109).
The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
Both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 um.
The sum of the incremental changes from the original clearance K091009 has been taken into account and all the occurred modifications will be listed and described within this submission.
The device Litho including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones.
The device Litho including a fiber optic delivery system is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones in medical specialties including, but not limited to: Urology, Gastroenterology, Arthroscopy, Neurosurgery, Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery.
The modified device Litho DK30 is claimed to be derived from the legally marketed (unmodified) device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).
#### Description of the modifications:
The modified device Litho DK30 is a Laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).
This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).
The modified device Litho DK30 share the same architecture and the same laser source of the unmodified device Litho: both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 µm, with no change in the fundamental scientific technology of the device.
The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
{2}------------------------------------------------
The following modifications have been implemented on the cleared device Litho (K091909) in order to get the desktop version named Litho DK30:
- A new mechanical structure and related covers have been developed
- A more compact and lightweight power-supply with extended working range has been qualified.
- The water pump of the cooling systern has been changed to improve the cooling efficiency. -
- The water-to-air heat-exchanger has been changed to improve the cooling efficiency.
- i The technical specifications of lens focusing the laser beam into the optical fiber has been changed to improve the incoupling.
- An RFID recognition system has been added to read the fiber diameter directly from the n fiber equipped with the suitable transponder.
- Review of the combinations of the laser parameters (energy and frequency) achievable within 30 W maximum power.
- A larger touch-screen has been adopted
- The control hardware (boards) and related software have been reviewed
- The GUI (Graphic User Interface) has been reviewed to improve the readability i
All the modifications implemented have been evaluated following the requirements of third edition of IEC 60601-1 and to its collateral standards. The introduction of the third edition of the IEC 60601-1 and to its collateral standards has been taken into account: it leaded to a basic change in the (regulatory) design input requirements with the subsequent revision of risk analysis and specific verification and validation activities to demonstrate that modified device meet the new requirements.
Picture 7-1 shows the the external appearance of the modified device and the unmodified device.
Image /page/2/Picture/13 description: The image contains two medical devices. The device on the left is a larger, wheeled machine with a display screen and various controls. It has a light-colored body with a darker top section and a foot pedal. The device on the right is smaller and more compact, labeled "Litho DK". It also features a display screen and several buttons, indicating it is a medical device for lithotripsy.
Picture 7-1 - device Litho (on left) and Litho DK30 (on right) 7-3
{3}------------------------------------------------
## Intended Use/Indications for Use
The modified device Litho DK30 is a Laser surgical instrument for use in general and plastic surgery and in dermatology (GEX).
This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909) because Litho DK30 is the desktop version of the already cleared device Litho (K09109).
The modified device Litho DK30 share the same architecture and the same laser source of the unmodified device Litho: both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 um, with no change in the fundamental scientific technology of the device.
The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
The Intended Use/Indication for use statement of the unmodified device (Litho - K091009) has been analyzed. In the original statement there is not a clear definition (separation) of the Intended use and of the Indications for use and thus a revised statement is proposed.
The revised version of the statement has separated sections for the Intended Use and for the Indications for use.
Moreover, in order to have a more understandable definition of the Indications for use, it has been removed the (long) list of diseases/surgical operations given in K091909 because this list, as said by several physicians, does not add useful information.
#### Thus the Intended Use is:
The device Litho including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic, to perform incision, excision, excision, resection, ablation, vaporization, coaqulation and hemostasis of soft tissue and in Lithotripsy of stones.
#### And the Indications for use are:
The device Litho including a fiber optic delivery system is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones in medical specialties including, but not limited to: Urology, Gastroenterology, Arthroscopy, Neurosurgery, Pulmonary, Gynecology, ENT, Dermatology , Plastic Surgery and General Surgery.
#### The modified device Litho DK30 has the same intended use of the unmodified device.
Moreover the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
{4}------------------------------------------------
## Substantial Equivalence:
This Special 510(k) of the modified device Litho DK30 is submitted due to Device Modifications of the already cleared device Litho (K091909) because because Litho DK30 is the desktop version of the already cleared device Litho (K09109).
The modified device Litho DK30 share the same architecture and the same laser source of the unmodified device Litho: both devices are Pulsed Holmium:YAG laser with a maximum output power of 30 W @ 2.1 µm, with no change in the fundamental scientific technology of the device.
The modified device Litho DK30 has the same intended use of the unmodified device. Moreover the the intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.
Thus the modified device Litho DK30 is substantially equivalent to the previously legally marketed device Litho (K091909).
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined strands. The caduceus is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is presented in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2014
Quanta System SPA Mr. Maurizio Bianchi Regulatory Affairs Manager Via IV Novembre, 116 Solbiate Olona (VA) Italy, 21058
Re: K141403
Trade/Device Name: Litho DK30 (and its accessories) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 7, 2014 Received: July 8, 2014
Dear Mr. Bianchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices . . Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K141403
Device Name Litho DK30
#### Indications for Use (Describe)
#### INTENDED USE
The device Litho DK30 including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones.
#### INDICATIONS FOR USE
The device Litho DK30 including a fiber optic delivery system is insurgical procedures such as open, laparoscopic and endoscopic, to perform incision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones in medical specialties including, but not limited to:
- Urology
- · Gastroenterology
- Arthroscopy
- Neurosurgery
- Pulmonary
- Gynecology
- ENT
- Dermatology
- Plastic Surgery
- General Surgery
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (2) CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
David Krause -S 2014.08.18 13:46:18 -04'00'
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.