K091909 · Quanta System Spa · GEX · Jul 9, 2009 · General, Plastic Surgery
Device Facts
Record ID
K091909
Device Name
LITHO LASER SYSTEM
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 9, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The LITHO laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotnpsy. Gasteroenterology, Arthroscopy, Discetomy Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Device Story
LITHO Laser System is a surgical Ho:YAG laser (2.1um wavelength) used for soft tissue surgery and lithotripsy. Device comprises high-voltage power supply, internal water-to-air cooling system, laser rod, optical fiber delivery system, and control electronics. Operated by physicians in clinical settings (OR, endoscopic suites). User selects parameters via control panel; laser energy delivered via fiber optic to target tissue for incision, ablation, or coagulation. Safety features include remote interlock, emergency stop, and footswitch. Provides precise tissue interaction for various specialties, including urology (e.g., prostatectomy, stone fragmentation), gastroenterology, orthopedics, and dermatology. Benefits include minimally invasive access and effective hemostasis.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Ho:YAG laser; 2.1um wavelength; up to 20 Hz frequency. Five subsystems: high-voltage power supply, internal water-to-air cooling, laser rod, optical fiber delivery, and display/control electronics. Safety features: remote interlock, emergency stop, footswitch, key switch.
Indications for Use
Indicated for patients requiring surgical incision, excision, resection, ablation, vaporization, coagulation, or hemostasis of soft, cartilaginous, or bony tissue across urological, gastroenterological, orthopedic, pulmonary, gynecological, ENT, dermatological, and general surgical specialties. Includes lithotripsy for urinary calculi and spinal disc decompression.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K091909
Title:
# 510(k) SUMMARY Quanta System LITHO Laser System
JUL - 9 2009
Quanta System SpA via IV Novembre, 116 21058 Solbiate Olona VA / Italy
Contact:
Submitter:
Dr. Isabella Carrer Medical Division Manager
Quanta System LITHO Laser System
Instrument, surgical, powered, laser
Date Prepared: April 10, 2009
Device Trade Name:
Common Name:
Laser surgical instrument for use in general surgery and dermatology
Classification Name:
Predicate Devices:
- AllMed System Sphinx Laser System (K033437):
Intended Use / Indications for Use:
# 2.1 um Applications:
The LITHO laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotnpsy. Gasteroenterology, Arthroscopy, Discetomy Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery.
### Urology
Open and endoscopic surgery (incision, excision, resection, ablation,
vaporization, coagulation and hemostasis) including:
- Urethral Strictures .
- Bladder Neck Incisions (BNI)
- Ablation and resection of Bladder Tumors, ● Uretheral Tumors and Ureteral
{1}------------------------------------------------
- Tumors,
- Ablation of Benign Prostatic Hypertrophy (BHP),
- Transurethral incision of the prostate (TUIP)
- Holmiurn Laser Resection of the Prostrate (HoLRP)
- Holmium Laser Enuculeation of the Prostate (HoLEP)
- Holmiurm laser Ablation of the Prostate (HoLAP)
- Condylomas
- Lesions of external genitalia
# Lithotripsy and Percutaneous Urinary Lithotripsy
- Endoscopic fragmentation of urethral, ureteral, . bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate
- dehydrate stones.
- Endoscopic fragmentation of kidney calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
## Gasteroenterlogy
Open and endoscopic gasteroenterlogy surgery (incision, excision, resection,
ablation, vaporization, coagulation and hemostasis ) including:
- Appendectomy .
- Polyps ●
- Biopsv .
- Gall Bladder calculi .
- Biliary/Bile duct calculi .
- Ulcers ●
- Gastric ulcers .
- Duodenal ulcers .
- Non Bleeding Ulcers ●
- Pancreatitas .
- Hemorrhoids
- Cholecystectomy .
- � Benign and Malignant Neoplasm
- Angiodysplasia
- Colorectal cancer
- Telangiectasias
- Telangiectasias of the Osler-Weber-Renu disease
{2}------------------------------------------------
- Vascular Malformation
- . Gastritis
- Esophagitis .
- Esophageal ulcers .
- Varices �
- Colitis .
- Mallory-Weiss tear ●
- Gastric Erosions .
# Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including:
- Ligament and tendon Release
- Countouring and sculpting of articular surfaces ●
- Capsulectomy in the Knee .
- � Chondreplasty in the Knee
- Debridement of inflamed synovial tissue .
- Chondromalacia Ablation .
- Chondromalacia and tears .
- Plica Removal .
- Meniscectomy ◆
- Loose Body Debuidement ◆
- Lateral retinecular release
Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including
- Percutaneous Laser Disc . Decompression/Discectomy of the L4-5 and L5-SI lumbar discs, including Foraminoplasty Percutaneous Cervical Disc Decompression/Discectomy Percutaneous Thoracic Disc Decompression/Discectomy
# Pulmonary
Open and endoscopic pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue)
#### Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
{3}------------------------------------------------
# ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and cartilage) including:
- Endonasal/sinus Surgery .
- Partial turbinectomy
- Polypectomy
- Dacryocystorhinostomy
- Frontal Sinusotomy
- Ethmoidectomy
- Maxillary antrostomy
- Functional endoscopic sinus surgery
### Dermatology and Plastic Surgery
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- Basal Cell Carcinomas .
- Lesions of skin and subcutaneous tissue
- Skin tags
- Plantar warts
- Lesions of skin and subcutaneous tissue
- Port Wine Stains
- Papillomas
# General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy .
- Skin incision .
- Excision of external and internal lesions .
- Complete of partial resection of internal organs, tumors and lesions
- Biopsy .
The LITHO Laser Systemis a surgical laser instrument for use in general surgery and dermatology. The LITHO Laser Systemi includes 1 model:
| <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Sheep of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattle of Cattl<br>100 AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AN | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br> | AND AND AND A CLAND C A color Company CO a<br>Acres And and a<br>10-4-4-4-400-4-4<br>A Company<br>AND A FINE A FINE A CHINA A<br>CAP<br>1 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| 24.00 m<br>and proper productions of the control of<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | -----<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>No. of Acres<br>A CONSUL CONSULTION OF COLLECTION OF CLAIM OF CHILIP OF CLAIM OF CHEAR<br>3 | <br>.<br>1<br>AATTT<br>C<br>1 |
Technological Characteristics:
{4}------------------------------------------------
The LITHO laser system is designed with five major sub-systems, including: (1) Power Supply: a high voltage power supply, which converts and rectifies the AC mains current to provide regulated power for the flash lamp simmer current and main triggering pulse; (2) Cooling System: a cooling system consisting of an internal water flow circuit together with water-to-air heat exchanger; (3) Laser System: an Ho:YAG laser rod, capable of generating laser pulses at 2.1um wavelength and a frequency up to 20 Hz; (4) Delivery System: an optical delivery system, interfacing the energy from the laser to the patient via an optical fiber and focusing hand piece; (5) Display and Control Electronics: a display with control electronics, which regulates the functions of the laser and allows parameter selection by the user.
In addition to the five subsystems, the LITHO laser system incorporates several safety features, including a remote interlock, an emergency red push button, a footswitch and a key switch.
# Performance Data
None
Substantial Equivalence: The Quanta System LITHO Laser System is as safe and effective as the predicate device. The LITHO Laser System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the LITHO Laser System and its predicate devices raise no new issues of safety or effectiveness. Thus, the LITHO Laser System is substantially equivalent to the predicate device.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/14 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Quanta System, S.P.A. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25th Street, Northwest Buffalo, Minnesota 55313
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### Re: K091909
Trade/Device Name: LITHO Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery
JUL - 9 2009
and in dermatology
Regulatory Class: II Product Code: GEX Dated: June 24, 2009 Received: June 25, 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
### Page 2 - Mr. Mark Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Millerson
Mark N.Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known):_K 09 | 9 0 9
#### LITHO Laser System Device Name:
Indications for Use:
# 2.1 um Applications:
The LITHO laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotnpsv. Gasteroenterology, Arthroscopy, Discetomy Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery.
#### Ürology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral Strictures t
- Bladder Neck Incisions (BNI)
- Ablation and resection of Bladder Tumors. Uretheral Tumors and Ureteral
- Tumors,
- Ablation of Benign Prostatic Hypertrophy (BHP), ( ßPH)
- Transurethral incision of the prostate (TUIP)
- Holmiurn Laser Resection of the Prostrate (HoLRP)
- Holmium Laser Enuculeation of the Prostate (HoLEP)
- Holmiurm laser Ablation of the Prostate (HoLAP)
- Condylomas
- Lesions of external genitalia
#### Lithotripsy and Percutaneous Urinary Lithotripsy
- . Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate
- dehydrate stones.
- Endoscopic fragmentation of kidney calculi
- Treatment of distal impacted fragments of steinstrasse, when g 800 be passed.
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
### Gasteroenterlogy
510(k) Number.
Open and endoscopic gasteroenterlogy surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis ) including:
{8}------------------------------------------------
- Appendectomy
- Polyps
- Biopsv
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non Bleeding Ulcers
- Pancreatitas
- Hemorrhoids
- Cholecystectomy
- Benign and Malignant Neoplasm
- Angiodysplasia
- Colorectal cancer
- Telangiectasias
- Telangiectasias of the Osler-Weber-Renu disease
- Vascular Malformation
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including:
- Ligament and tendon Release
- Countouring and sculpting of articular surfaces
- Capsulectomy in the Knee
- Chondreplasty in the Knee
- Debridement of inflamed synovial tissue
- Chondromalacia Ablation
- Chondromalacia and tears
- Plica Removal
- Meniscectomy
- Loose Body Debuidement
- Lateral retinecular release
N.A. Rf. Ogle for mxm
(Division Sign-Off)
Division of Surgical, Orthonedic, and Restorative Devices
510(k) Number K091909
Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including
- Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI . lumbar discs, including Foraminoplasty Percutaneous Cervical Disc
{9}------------------------------------------------
Decompression/Discectomy Percutaneous Thoracic Disc Decompression/Discectomy
### Pulmonary
Open and endoscopic pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue)
# Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and cartilage) including:
- Endonasal/sinus Surgery
- Partial turbinectomy
- Polypectomy
- Dacrvocystorhinostomy
- · Frontal Sinusotomy
- Ethmoidectomy
- Maxillary antrostomy
- Functional endoscopic sinus surgery
### Dermatology and Plastic Surgery
Incision, excision, resection, ablation, vàporization, coagulation and hemostasis of soft mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- Basal Cell Carcinomas
- Lesions of skin and subcutaneous tissue
- Skin tags
- Plantar warts
- Lesions of skin and subcutaneous tissue
- Port Wine Stains
- Papillomas
Nail R. Dale
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091909
### General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Appendectomy .
- Skin incision
- Excision of external and internal lesions
- Complete of partial resection of internal organs, tumors and lesions
- Biopsy
{10}------------------------------------------------
# Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 C.F.R. 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil ke Oglin for nkn
(Division Sign-Off)
Division of Surpical, Orihopedic, and Restorative Devices
Kog1909 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.