K141390 · Wuhan Landing Medical Hi-Tech, Ltd. · JOW · Nov 18, 2014 · Cardiovascular
Device Facts
Record ID
K141390
Device Name
INTERMITTENT PNEUMATIC COMPRESSION DEVICE
Applicant
Wuhan Landing Medical Hi-Tech, Ltd.
Product Code
JOW · Cardiovascular
Decision Date
Nov 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LD Q-IPC III is a system to prevent DVT(Deep Vein Thrombosis) by improving the blood velocity of patients. Leg Compression The use of the product with Calf Cuff and Thigh Cuff is indicated for: 1. Deep vein thrombosis and pulmonary embolism prophylaxis. Foot Compression The use of the product with Foot Cuff is indicated for: 1. Circulation enhancement. 2. Deep vein thrombosis prophylaxis. 3. Edema-Acute. 4. Edema-Chronic. 5. Extremity pain incident to trauma or surgery. 6. Leg Ulcers. 7. Venous stasis /Venous insufficiency.
Device Story
LD Q-IPC III is an intermittent pneumatic compression system; prevents DVT by increasing blood velocity. System comprises controller, tubing, and single-patient-use cuffs (calf, thigh, foot). Controller delivers air to cuffs; sequentially pressurizes from distal to proximal chambers; provides interval for vein refilling. Used in clinical settings; operated by healthcare providers. Output is mechanical compression of limbs; affects venous return and blood flow; benefits patients by reducing DVT incidence. Controller features audible/visual alarms for pressure/system errors.
Clinical Evidence
Bench testing only. Device performance verified through compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 10993 (biocompatibility: cytotoxicity, irritation, sensitization). Software verification conducted per FDA guidance.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2014
Wuhan Landing Medical Hi-Tech Co., Ltd Yan Liu Quality Manager 818 Gaoxin Rd East Lake High Tech Development Zone Floors 1 & 2, Units B & C & D, Building No. B7 Med Equipment Park Wuham, CH 430206
Re: K141390
> Trade/Device Name: Intermittent pneumatic compression device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: Undated Received: September 15, 2014
Dear Yan Liu,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hilleman
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K141390
Device Name
Intermittent pneumatic compression device
#### Indications for Use (Describe)
The LD Q-IPC III is a system to prevent DVT(Deep Vein Thrombosis) by improving the blood velocity of patients.
Leg Compression
- The use of the product with Calf Cuff and Thigh Cuff is indicated for:
1. Deep vein thrombosis and pulmonary embolism prophylaxis.
Foot Compression
The use of the product with Foot Cuff is indicated for:
- 1. Circulation enhancement.
- 2. Deep vein thrombosis prophylaxis.
- 3. Edema-Acute.
- 4. Edema-Chronic.
- 5. Extremity pain incident to trauma or surgery.
- 6. Leg Ulcers.
- 7. Venous stasis /Venous insufficiency.
## Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Section 5 - 510(k) Summary
Date of Summary Preparation: 9/9/2014
#### 1. Submitter's Identifications
Submitter's Name: Wuhan Landing Medical Hi-Tech Co., LTD Address: Floors 1&2,Units B&C&D,Building No.B7 Medical Equipment Park,818 Gaoxin Road East Lake High Tech Development Zone 430206 Wuhan, Hubei,China Contact Person: Yan Liu Contact Email Address: fdalanding@sohu.com Telephone: +86-27-59263677 Fax: +86-27-59368637
#### 2. Correspondent's Identifications
Correspondent's Name: Wuhan Landing Medical Hi-Tech Co.,LTD Address: Floors 1&2,Units B&C&D,Building No.B7 Medical Equipment Park,818 Gaoxin Road East Lake High Tech Development Zone 430206 Wuhan, Hubei,China Contact Person: Yan Liu Contact Email Address: fdalanding@sohu.com Telephone: +86-27-59263677 Fax: +86-27-59368637
#### 3. Name of the Device
Device Classification Name: sleeve, limb, compressible Product Name: Compressible limb sleeve Trade Name: Intermittent pneumatic compression device Model: LD Q-IPC III Classification Panel: Cardiovascular Product Code: JOW Device Classification: Class II
#### 4. The Predicate Devices
Veinoflow SCD, LBTK-M-I 5001 K123830
#### 5. Device Description
The LD Q-IPC III Intermittent pneumatic compression device is a pneumatic compression device that noninvasively helps reduce the incidence of deep vein thrombosis, a potentially life threatening condition.
The LD Q-IPC III Intermittent pneumatic compression device consists of the controller, the Tubing Sets and single-patient used cuff. The cuff includes calf cuffs, thigh cuffs and foot cuffs.
The controller will deliver air to the cuffs and sequentially pressurize them. Cuffs will be
{5}------------------------------------------------
# Wuhan Landing Medical Hi-Tech Co,LTD
pressurized from the distal air chamber to the proximal air chamber. After the compression, the controller will allow an interval time for vein refilling, before the controller starts to pressurize the cuffs again. The cycles will be repeated throughout the time the unit is in use or for the duration of the default operating time.
#### 6. Intended Use of Device
The LD Q-IPC III is a system to prevent DVT(Deep Vein Thrombosis) by improving the blood velocity of patients.
Leg Compression
The use of the product with Calf Cuff and Thigh Cuff is indicated for:
- 1. Deep vein thrombosis and pulmonary embolism prophylaxis.
#### Foot Compression
The use of the product with Foot Cuff is indicated for:
- 1. Circulation enhancement.
- 2. Deep vein thrombosis prophylaxis.
- 3. Edema-Acute.
- 4. Edema-Chronic.
- 5. Extremity pain incident to trauma or surgery.
- 6. Leg Ulcers.
- 7. Venous stasis /Venous insufficiency.
#### 7. Summary of Substantial Equivalence
Table : The difference between LD Q-IPC III and Predicate Veinoflow SCD (K123830)
| | New Device | Predicate Device |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K 141390 | K123830 |
| Product Code: | JOW | JOW |
| Proprietary Name: | Intermittent Pneumatic Compression Device | Veinoflow SCD, LBTK-M-I 5001 |
| Manufacturer: | Wuhan Landing Medical Hi-Tech Co., Ltd | Dalian Labtek Science & Development Co., Ltd |
| Indication for use: | The LD Q-IPC III is a system to prevent | Veinoflow SCD system, Model LBTK-M-I 5001 |
| | DVT(Deep Vein Thrombosis) by improving | is a system to prevent DVT (Deep Vein |
| | the blood velocity of patients. LD Q-IPC III | Thrombosis) by improving the blood velocity of |
| | is indicated for Circulation enhancement, | patients. LBTK-M-I 5001 is indicated for |
| | Deep vein thrombosis and pulmonary | Circulation enhancement, Deep vein thrombosis |
| | embolism prophylaxis, Edema-Acute, | and pulmonary embolism prophylaxis, |
| | Edema-Chronic, Extremity pain incident to | Edema-Acute, Edema-Chronic, Extremity pain |
| | trauma or surgery, Leg Ulcers, Venous | incident to trauma or surgery, Leg Ulcers, Venous |
| | stasis /Venous insufficiency | stasis /Venous insufficiency |
| Components | Pump controller, thigh cuffs, calf cuffs, foot<br>cuffs, air supply tubes, power line | Pump controller, thigh cuffs, calf cuffs, foot cuffs,<br>battery, air supply tubes, power line |
| Ingress of Water | Ordinary | Ordinary |
| protection | | |
| Compression type | thigh cuffs & calf cuffs: Sequential, Gradient<br>foot cuffs: Uniform | thigh cuffs & calf cuffs: Sequential, Gradient<br>foot cuffs: Uniform |
| Compression time | Single thigh cuffs: 6s Compression<br>Dual thigh cuffs: 12s Compression<br>Single calf cuffs: 4s Compression<br>Dual calf cuffs: 8s Compression<br>Single foot cuffs: 2.5s Compression<br>Dual foot cuffs: 5s Compression | Single thigh cuffs: 5.5s Compression<br>Dual thigh cuffs: 11s Compression<br>Single calf cuffs: 4s Compression<br>Dual calf cuffs: 8s Compression<br>Single foot cuffs: 2.5s Compression<br>Dual foot cuffs: 5s Compression |
| Deflation time | 2-3s | 2-3s |
| Default inflatable<br>Interval time | 48s | 48s |
| Adjustable<br>inflatable interval<br>time | 24s, 48s, 60s | 24s, 48s, 60s |
| Default Pressure | thigh cuffs & calf cuffs: 45 mmHg<br>foot cuffs: 130 mmHg | thigh cuffs & calf cuffs: 40 mmHg<br>foot cuffs: 130 mmHg |
| Adjustable<br>Pressure | thigh cuffs & calf cuffs: 30-60 mmHg<br>foot cuffs: 120-140 mmHg | thigh cuffs & calf cuffs: 30-60 mmHg<br>foot cuffs: 120-140 mmHg |
| Mode of operation | Continuous | Continuous |
| Application mode | Single thigh cuffs Compression<br>Dual thigh cuffs Compression<br>Single calf cuffs Compression<br>Dual calf cuffs Compression<br>Single foot cuffs Compression<br>Dual foot cuffs Compression | Single thigh cuffs Compression<br>Dual thigh cuffs Compression<br>Single calf cuffs Compression<br>Dual calf cuffs Compression<br>Single foot cuffs Compression<br>Dual foot cuffs Compression<br>Simultanous single thigh cuffs and single calf<br>cuffs Compression |
| Bed Hook | Yes | Yes |
| Power cord<br>storage | Yes | Yes |
| Audible/Visual<br>Alarms | Low Pressure, High Pressure, Pump Error,<br>Valve Error, System Error, | No garment, Pump Error, Valve Error,<br>Temperature Error, Software Error, System Error,<br>High Pressure, Low Pressure, Low Battery |
| Controll<br>Dimensions | Length: 280mm; Width: 145mm; Height:<br>250mm | Length: 240mm; Width: 140mm; Height: 263mm |
| Controll Weight | 3kg | 3.5kg |
| Power<br>Requirement | AC 100-240V, 60 VA, 50-60Hz | AC 100-240V, 50 VA, 50-60Hz |
| Battery | No | Yes |
| Shipping Unit | Each | Each |
| Standard | IEC60601-1 | IEC60601-1 |
| | ISO10993-1 | ISO10993-1 |
| | ISO10993-5 | ISO10993-5 |
| | ISO10993-10 | ISO10993-10 |
| Non-sterile | Non-sterile | Non-sterile |
| Microprocessor<br>Control | Yes | Yes |
{6}------------------------------------------------
# Wuhan Landing Medical Hi-Tech Co,LTD
{7}------------------------------------------------
# Wuhan Landing Medical Hi-Tech Co,LTD
### 8. Substantial Equivalence:
The proposed devices of LD Q-IPC III have the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance effectiveness, performance safety as the predicate device. The differences only exist in battery, default pressure, and audible/visual alarms. These differences are slight and do not influence the effectiveness and safety of the device.
### 9. Non-Clinical Tests Performed:
The following testing was performed on the LD Q-IPC III in accordance with the requirements of the design control regulations and established quality assurance procedures.
IEC60601-1:2005+CORR.1(2006)+CORR.2(2007),Medical electrical equipment-Part 1:General requirements for basic safety, and essential performance.
IEC60601-1-2:2007, Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility-Requirements and tests.
ISO10993-1: 2009, Biological evaluation of medical devices-Part 1:Evaluation and testing within a risk management process
ISO10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity. ISO10993-10:2010, Biological evaluation of medical devices-Part10: Tests for irritation and skin sensitization
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
#### 10. Conclusion:
The proposed device LD Q-IPC III are determined to be Substantially Equivalent (SE) to the predicate device, Veinoflow SCD, LBTK-M-I 5001 in respect of safety and effectiveness.
--- End of this section ---
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.