K123830 · Dalian Labtek Science & Development Co., Ltd. · JOW · Aug 16, 2013 · Cardiovascular
Device Facts
Record ID
K123830
Device Name
INTERMITTENT PNEUMATIC COMPRESSION DEVICE
Applicant
Dalian Labtek Science & Development Co., Ltd.
Product Code
JOW · Cardiovascular
Decision Date
Aug 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Veinoflow SCD system, Model LBTK-M-I 5001 is a system to prevent DVT (Deep Vein Thrombosis) by improving the blood velocity of patients. LBTK-M-1 5001 is indicated for Circulation Enhancement, Deep Vein Thrombosis Prophylaxis, Edema -Acute, Edema - Chronic, Extremity Pain Incident to Trauma or Surgery, Leg ulcers, Venous Stasis / Venous Insufficiency.
Device Story
Veinoflow SCD (Model LBTK-M-I 5001) is a portable intermittent pneumatic compression system. It consists of a pump controller, air supply tubes, and inflatable garments (thigh-calf, calf, or foot cuffs). The system operates by supplying compressed air to garments, providing sequential or uniform compression to lower limbs to enhance blood velocity and circulation. The controller automatically detects the connected garment type and applies preset, adjustable pressure and inflation intervals. It includes a self-test system with visual and audible alarms for errors (e.g., pressure, connection, system, battery). Used in clinical settings, the device is operated by healthcare professionals to prevent DVT and manage venous conditions. Output is mechanical compression; clinical benefit is improved venous blood flow and reduced risk of thrombosis.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1, -5, -10), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and mechanical performance (pressure testing, inflation/deflation timing, air leakage, burst testing, bladder fatiguing, and Velcro adhesion).
Technological Characteristics
Portable pneumatic pump controller with optional 14.8V Li-ion battery. Garments: thigh-calf, calf, foot cuffs. Materials: biocompatible per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2. Compression: sequential (thigh/calf) or uniform (foot). Connectivity: standalone. Sterilization: non-sterile.
Indications for Use
Indicated for patients requiring circulation enhancement, DVT prophylaxis, management of acute/chronic edema, relief of extremity pain post-trauma/surgery, treatment of leg ulcers, and venous stasis/insufficiency.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Dalian Labtek Science & Development Co., Ltd. 2-2-2-2, No.4 Zhengren Street, Shahekou District, Dalian, China 116021
#### Section 5
## 510(k) Summary
[As required by 21 CFR 807.92]
#### 1. Submission Information: AUG 1 6 2013 510(k) Number: K123830 June 26th, 2013 Date: Type of 510(k) Submission: Traditional Basis for 510(k) Submission: New device Submitter/Manufacturer: Dalian Labtek Science & Development Co., Ltd. 2-2-2-2, No.4 Zhengren Street, Shahekou District, Dalian, China 116021 Contactor: Doris Dong, Consultant Shanghai CV Technology Co., Ltd. Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, China 201600 E-mail: doris d@126.com / Url: www.ceve.org.cn Tel: 86 21-31261348 / Fax: 86 21-37824346 2. Device Description: Proprietary Name: Veinoflow SCD, Model LBTK-M-I 5001
| Common Name: | Intermittent Pneumatic Compression Device |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Compressible limb sleeve |
| Regulation Number: | 21 CFR 870.5800 |
| Product Code: | JOW |
| Device Class: | II |
| Review Panel: | Cardiovascular |
| Indications for use: | Veinoflow SCD system, Model LBTK-M-I 5001 is a system to prevent<br>DVT (Deep Vein Thrombosis) by improving the blood velocity of patients.<br>LBTK-M-1 5001 is indicated for Circulation Enhancement, Deep Vein<br>Thrombosis Prophylaxis, Edema -Acute, Edema - Chronic, Extremity Pain<br>Incident to Trauma or Surgery, Leg ulcers, Venous Stasis / Venous<br>Insufficiency. |
| Device Description: | 1) Description of the compression system<br>Veinoflow SCD, Model LBTK-M-I 5001, is a pneumatic pump system that<br>supplies compressed air to inflate garments that are attached to a patent's<br>lower limbs. It consists of a pump controller, specially designed inflation<br>and deflation garments for feet and legs, air supply tubes, and power line.<br>The inflation and deflation garments have 3 types: ① thigh-calf garments,<br>② calf garments, and ③ foot cuffs.<br>The system offers sequential inflation and propels the vein blood from<br>limb to heart, therefore enhance the blood circulation. The controller can<br>automatically detect the external garment type, and provides pressure<br>correspondingly. The pressure value and inflatable interval time are preset<br>and adjustable. |
The controller has a self test system, when there is any error, it will alarm both in visual and audio.
{1}------------------------------------------------
### 3. Substantial Equivalence:
.
Detailed comparison data is included in the section 9 of "Substantial Equivalence Discussion" of this 510(k) submission.
{2}------------------------------------------------
Dalian Labtek Science & Development Co., Ltd.
2-2-2-2, No.4 Zheneren Street Shahekou District Dalian, China 116021
| | New Device | Predicate Device |
|----------------------------------|-------------------------------------------------------------------------------|------------------------------------------------------------------------|
| 510(k) Number: | K123830 | K112677 |
| Product Code: | JOW | JOW |
| Proprietary Name: | Veinoflow SCD, LBTK-M-I 5001 | The Venous Assist System DVT-2600 |
| Manufacturer: | Dalian Labtek Science & Development Co., Ltd. | DAESUNG MAREF CO LTD |
| Indications for use: | Veinoflow SCD system, Model LBTK-M-I 5001 is a system to prevent DVT | DVT-2600 is a system to prevent DVT (Deep Vein Thrombosis) by |
| | (Deep Vein Thrombosis) by improving the blood velocity of patients. | improving the blood velocity of patients. DVT-2600 is indicated for |
| | LBTK-M-I 5001 is indicated for Circulation Enhancement, Deep Vein | Circulation Enhancement, Deep Vein Thrombosis Prophylaxis, |
| | Thrombosis Prophylaxis, Edema -Acute, Edema - Chronic, Extremity Pain | Edema -Acute, Edema - Chronic, Extremity Pain Incident to Trauma |
| | Incident to Trauma or Surgery, Leg ulcers, Venous Stasis / Venous | or Surgery, Leg ulcers, Venous Stasis / Venous Insufficiency. |
| | Insufficiency. | |
| Components: | pump controller, multi-cavity thigh-calf garments, calf garments, foot cuffs, | pump controller, multi-cavity thigh-calf garments, calf garments, foot |
| | battery, air supply tubes, power line | cuffs, battery, air supply tubes, power line |
| Ingress of Water Protection | Ordinary | Ordinary |
| Compression Type | Thigh-calf garments & calf garments: Sequential, Gradient | Thigh-calf garments & calf garments: Sequential, Gradient |
| | Foot Cuffs: Uniform | Foot Cuffs: Uniform |
| Compression time | Single thigh-calf garment: 5.5s Compression | Single thigh-calf garment: 5.5s Compression |
| | Dual thigh-calf garments: 11s Compression | Dual thigh-calf garments: 11s Compression |
| | Single calf garment: 4s Compression | Single calf garment: 4s Compression |
| | Dual calf garments: 8s Compression | Dual calf garments: 8s Compression |
| | Single foot cuff: 2.5s Compression | Single foot cuff: 2.5s Compression |
| | Dual foot cuffs: 5s Compression | Dual foot cuffs: 5s Compression |
| Deflation time | 2~3s | 2~3s |
| Default inflatable interval time | 48s | 48s |
| Adjustable inflatable interval | 24s, 48s, 60s | 24s, 48s, 60s |
| time | | |
| Default Pressure | Thigh-calf garments & calf garments: 40 mmHg | Thigh-calf garments & calf garments: 40 mmHg |
| | Foot Cuffs: 130 mmHg | Foot Cuffs: 120 mmHg |
| Adjustable pressure | Thigh-calf garments & calf garments: 30-60mmHg | Thigh-calf garments & calf garments: 20-60mmHg |
:
:
:
.
:
{3}------------------------------------------------
| Dalian Labtek Science & Development Co., Ltd. | | |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| | 2-2-2-2, No.4 Zhengren Street, Shahekou District, Dalian, China 116021 | |
| | Foot Cuffs: 120~140mmHg | Foot Cuffs: 120-140mmHg |
| Mode of Operation | Continuous | Continuous |
| Application mode | ① Single thigh-calf garment Compression, | ① Single thigh-calf garment Compression; |
| | ② Dual thigh-calf garments Compression; | ② Dual thigh-calf garments Compression; |
| | ③ Single calf garment Compression; | ③ Single calf garment Compression; |
| | ④ Dual calf garments Compression; | ④ Dual calf garments Compression; |
| | ⑤ Single Foot cuff Compression; | ⑤ Single Foot cuff Compression; |
| | ⑥ Dual Foot cuffs Compression; | ⑥ Dual Foot cuffs Compression; |
| | ⑦ Simultaneous single thigh-calf garment and single calf garment | ⑦ Simultaneous single thigh-calf garment and single calf garment |
| | Compression | Compression; |
| | | ⑧ Simultaneous single foot cuff and single thigh-calf garment |
| | | Compression; |
| | | ⑨ Simultaneous single foot cuff and single calf garment |
| | | Compression |
| Bed Hook | Yes | Yes |
| Power Cord Storage | Yes | Yes |
| Audible/Visual Alarms | No garment, Pump Error, Valve Error, Temperature Error, Software Error,<br>System Error, High Pressure, Low Pressure, Low Battery | Cuff connection Error, Pressure Error, System Error, Power Error |
| Controller Dimensions | Length: 240mm; Width: 140mm; Height: 263mm | Length: 205mm; Width: 155mm; Height: 195mm |
| Controller Weight | 3.5kg | 1.9kg |
| Power Requirement | AC 100-240V, 50VA, 50/60 Hz | AC 100-240V, 35VA, 50/60 Hz |
| Battery | 14.8V, 3100mAhr, Lithium Ion (Optional) | 11.1V, 2500mAhr, Lithium Ion (Optional) |
| | Run Time: 3-4 hours | Run Time: 6-8 hours |
| | Charge Time: 4~5 hours (charging only) | Charge Time: 4 hours (charging only) |
| Shoulder Strap | Yes (optional) | Yes |
| Shipping Unit | Each | Each |
| Standards: | ISO 10993-1, ISO 10993-5, ISO 10993-10,<br>IEC 60601-1, and IEC 60601-1-2 | ISO 10993-1, ISO 10993-5, ISO 10993-10,<br>IEC 60601-1, and IEC 60601-1-2 |
・
4-6
:
.
{4}------------------------------------------------
Dalian Labtek Science & Development Co., Ltd.
.
| 2-2-2-2, No 4 Zhengren Street, Shahekou District, Dalian, China 116021 | 2) Electromagnetic compatibility test | |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 3) Anti-stretch test of garment materials | 3) Electrical Safety test | |
| 4) Anti-stretch test of air chamber welding point | | |
| 5) Air leakage test; 6) Burst testing of garment; 7) Bladder fatiguing test; | | |
| 8) Velcro adhesion evaluation test; 9) Biological compatibility test; | | |
| 10) Electromagnetic compatibility test; 11) Electrical Safety test | | |
| Non-sterile: | Non-sterile | Non-sterile |
| Microprocessor Control? | Yes | Yes |
| Differences: | The new device and the predicate device have different Audible/Visual Alarms, dimensions and weight, battery capacity, and LCD display panel. The<br>new device has more safety testing on thigh-calf garments, calf garments and foot cuffs. | |
| Similarities: | The new device and the predicate device, both are portable, equipped with battery, have same intended use, components, function, working principle, operation mode, compression type, power supply, same compression cycle and similar preset pressure values, and similar conformity standards. | |
| Conclusion: | Since the new device has same components, working principle, intended use, and safety features with the predicate device, they are substantial equivalent.<br>Though the two devices have slightly differences, for example, different default pressure for foot cuffs, however, the new device can be adjusted to the same pressure as the predicate device. And the adjustable pressure range for foot cuffs is same.<br>In addition, the adjustable pressure range for thigh-calf garments/calf garments of the new device is different from the predicate device, however, it is covered by the adjustable pressure range of the predicate device.<br>Except that single foot cuff can not be used with single thigh-calf garments/calf garment simultaneously, the new device and the predicate device has same application mode of garments/cuffs.<br>Any difference in technological characteristics does not raise any new safety and effectiveness issues. The conclusion drawn from the testing (Electrical Safety test, Pressure testing, Inflation/deflation time, Air leakage test, Burst testing, and so on) is that the device is substantially equivalent to the predicate device. | |
.
:
.
{5}------------------------------------------------
Dalian Labtek Science & Development Co., Ltd.
# 2-2-2-2, No.4 Zhengren Street, Shahekou District, Dalian, China 116021
4. Discussion of Substantial Equivalence to the Predicate Device
Veinoflow SCD system is a non-invasive medical device that delivers air to the garments and sequentially pressures them. Accordingly, the Veinoflow SCD system is intrinsically safe. Apart from the intrinsic safety of the physiological mechanism of the Veinoflow SCD system, the product components are designed to minimize potential risks to patients during product use. In particular, the product is equipped with the following safety features:
1) Audio and visual alarms are activated if inflation garments pressure either exceeds or fails to achieve recommended levels.
2 Relevant contraindications, numerous warning proper use and maintenance, are contained in the instruction manual.
3 The product labeling indicates that the device is restricted to sale by or on the order of a physician.
4 Software validation and other safety features.
The Veinoflow SCD, Model LBTK-M-I 5001, has passed biocompatibility safety, electricity safety and electromagnetic compatibility safety testing according to:
ISO 10993-1: 2009: Evaluation and testing within a risk management process
ISO 10993-5: 2009: Tests for cytotoxicity: In vitro methods;
ISO 10993-10: 2010: Tests for Irritation and Sensitization;
IEC 60601-1: 2005: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
Moreover, the garments of the device and the whole system passed bench testing of: Pressure testing, Inflation/deflation time, Anti-stretch test of garment materials, Anti-stretch test of air chamber welding point, Air leakage test, Burst testing of garment, Bladder fatiguing test, and Velcro adhesion test.
Veinoflow SCD system, Model LBTK-M-1 5001 has the same technological characteristics, intend use, design, function, composition and mode of operation to the predicate device.
The conclusion drawn from the bench testing is that the device is substantially equivalent to the predicate device. Furthermore, the device complies with the recognized standards and performs its intended tasks as well as the legally marketed predicate device.
୧-୧
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WQ66-G609 Silver Spring, MD 20993-0002
August 16, 2013
Dalian Labtek Science & Development Co., Ltd. c/o Doris Dong Consultant Shanghai CV Technology Co., Ltd. Room 1706, No. 128 Songle Rd., Songjiang Area Shanghai, China 201600
Re: K123830
Trade/Device Name: Veinoflow SCD, Model LBTK-M-1 5001 Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: June 27, 2013 Received: July 16, 2013
Dear Ms. Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
Page 2 - Ms. Dong
device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillerman
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Section 4 Indications for Use
510(k) Number (if known): K123830
Device Name: Veinoflow SCD, Model LBTK-M-I 5001
## Indications for Use:
Veinoflow SCD system, Model LBTK-M-1 5001 is a system to prevent DVT (Deep Vein Thrombosis) by improving the blood velocity of patients. LBTK-M-I 5001 is indicated for Circulation Enhancement, Deep Vein Thrombosis Prophylaxis, Edema - Chronic, Extremity Pain Incident to Trauma or Surgery, Leg ulcers, Venous Stasis / Venous Insufficiency.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
MA. Lillehemen
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.