The OARtrac® System is specifically indicated for male prostate cancer treatment to measure photon beam therapy as an adjunct to treatment planning permitting measurement of in-vivo radiation dose received on the anterior surface of a modified Prostate Immobilization Endorectal Balloon (ERB) device to monitor and verify the surrounding organs at risk, specifically the protatic rectal interface.
Device Story
OARtrac® System provides near real-time, in-vivo, multipoint radiation-dose monitoring during prostate cancer photon beam therapy. System consists of two Plastic Scintillating Detectors (PSD) embedded on a modified Prostate Immobilization Endorectal Balloon (ERB), a Clinical Detector Unit (CDU) with a Charged Coupled Device (CCD) camera, and proprietary dose management software. ERB immobilizes prostate while positioning sensors at the rectal prostatic interface. PSDs measure light output proportional to radiation dose; fiber optic cables transmit signals to the CDU. Software processes signals to calculate dose and dose rate, displaying current and historical treatment logs. Used in clinical radiation oncology settings by physicians or trained staff. Output allows physicians to compare actual delivered dose against planned dose, facilitating real-time monitoring of organs at risk. Non-implantable design reduces risks associated with invasive dosimeters.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software validation (IEC 62304), and mechanical testing (balloon leak, burst, and tensile strength).
Technological Characteristics
System uses Plastic Scintillating Detectors (PSD) with fiber optic connectivity to a Clinical Detector Unit (CDU) containing a CCD camera. ERB is polyurethane. Energy range: 1-20 MV photons. Dose range: up to 90 Gy. Accuracy: ±6% at 95% confidence interval. Software-based dose calculation. Non-sterile. Single-use.
Indications for Use
Indicated for male patients undergoing prostate cancer treatment to measure photon beam therapy dose on the anterior surface of an endorectal balloon to monitor and verify radiation dose to the prostatic rectal interface.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Sicel Technologies, Inc., Dose Verification System (DVS) (K083055)
{1}------------------------------------------------
K141154
Page 1 of 8
JUL 2 2 2014
# 510(k) Summary
for
# OARtrac® System
#### 1. Submission Sponsor
RadiaDyne, LLC 2313 W. Sam Houston Pkwy N Suite 107 Houston Texas, 77043, USA Phone: 281.759.9600 Fax: 281.759.9609 Contact: John Isham, President & CEO
#### 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701, USA Office Phone: 512.327.9997 Fax: 512.327.9998 Contact: Stuart R. Goldman, Senior Consultant, RA/QA Email: project.management@emergogroup.com
#### 3. Date Prepared
May 2, 2014
#### 4. Device Identification
Trade/Proprietary Name: OARtrac® System Common/Usual Name: Dose radiation verification and prostate immobilizer system Classification Name: Dosimeter, ionizing radiation, implanted
Classification Regulation: 892.5050 NZT and PCT Product Code: Device Class: Class II Classification Panel: Radiology
#### 5. Legally Marketed Predicate Device(s)
Sicel Technologies, Inc., Dose Verification System (DVS) (K083055) RadiaDyne, LLC, Prostate Immobilizer Rectal Balloon (PIRB) (K132194)
#### 6. Device Description
The OARtrac® System provides radiation oncologists with near real- time, in-vivo, multipoint radiation-dose information obtained from two (2) Radiatrac® Plastic Scintillating Detectors (PSD) sensors located on the anterior surface of a modified clinically accepted OARtrac® prostate Endorectal Balloon (ERB) to monitor dose photon based radiation
{2}------------------------------------------------
therapy for prostate cancer treatment. This information allows the physician to monitor the dose at the rectal prostatic interface, compare the actual dose relative to the planned dose, and provides graphs and dose information for both the current treatment as well as a log of dose from five previous treatments. The OARtrac® System also provides dose rate during actual treatment.
The function of the OARtrac® prostate ERB is to immobilize the prostate while at the same time positioning the Radiatrac® PSD sensors so that measurement of in-vivo radiation received on the anterior surface of the balloon can be monitored to verify the radiation dose to surrounding organs at risk, specifically the protatic rectal interface. The actual verification of the dose radiation is accomplished by the other main components of the OARtrac® System, those being the Clinical Detector Unit (CDU) with its Charged Coupled Device (CCD) camera and the system's own proprietary dose management software that are further addressed in the remainder of this section.
## 7. Indication for Use Statement
The OARtrac® System is specifically indicated for male prostate cancer treatment to measure photon beam therapy as an adjunct to treatment planning permitting measurement of in-vivo radiation dose received on the anterior surface of a modified Prostate Immobilization Endorectal Balloon (ERB) device to monitor and verify the surrounding organs at risk, specifically the protatic rectal interface.
#### 8. Substantial Equivalence Discussion
RadiaDyne has chosen two predicate devices in the 510(k) submission for its OARtrac® System, the Sicel Dose Verification System (K083035) manufactured by Sicel Technologies and cleared under Product Code NZT and §892.5050, and its own Prostate Immobilizer Rectal Balloon (PIRB) approved by the FDA through a de novo application (K132194) under Product Code PCT and §892.5720.
The following table compares the OARtrac® System to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Device | Sicel Technologies | RadiaDyne | RadiaDyne | Differences |
|---------------|-----------------------------|-----------------------------------------------|-----------------|--------------------------------------------------------------------------------------------|
| Trade Name: | Dose Verification<br>System | Prostate Immobilizer Rectal<br>Balloon (PIRB) | OARtrac® System | - |
| 510(k): | K083035 | K132194 | Pending | |
| Product Code: | NZT | PCT | NZT +<br>PCT | The OARtrac®<br>System's technology<br>is based on<br>elements of both<br>predicate device |
| Table 5-1 |
|----------------------------------------------------------------------|
| OARtrac® System vs. Predicate Devices - Similarities and Differences |
{3}------------------------------------------------
| Regulation: | §892.5050 | §892.5720 | §892.5050 +<br>§892.5720 | product codes in<br>that it acts as a<br>radiation dose<br>verification system<br>and as a prostate<br>immobilization<br>device.<br>The OARtrac®<br>System's technology<br>is based on<br>elements of both<br>predicate device<br>regulation numbers<br>in that it acts as a<br>radiation dose<br>verification system<br>and as a prostate<br>immobilization<br>device. | | | | | |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Class: | II | II | II | II | Mode of<br>Operation: | Metal oxide<br>semiconductor field-<br>effect transistor<br>(MOSFET) is implanted<br>in patient to allow for<br>dose verification of<br>their radiation<br>treatment procedure.<br>The dosimeter is<br>powered<br>telemetrically and<br>measures radiation<br>using two MOSFETs.<br>The MOSFETs are<br>hermetically sealed in<br>a biocompatible glass<br>capsule. | Polyurethane balloon is<br>designed as an immobilizer<br>to assist in positioning the<br>prostate during CT exams,<br>and during radiation<br>treatment therapy<br>procedures. | Plastic<br>scintillation<br>detectors (PSD)<br>have been<br>imbedded in the<br>center of the<br>previously<br>cleared<br>RadiaDyne PIRB<br>device to allow<br>for dose<br>verification of the<br>patient's<br>radiation<br>treatment<br>procedures. | The OARtrac®<br>System's dose<br>verification<br>technology is based<br>on fiber optics and<br>PSD sensor<br>technology, while<br>the DVS device uses<br>MOSFET technology<br>to achieve its dose<br>verification<br>function. |
| Intended Use: | The Dose Verification<br>System (DVS) is<br>intended for use in<br>radiation therapy to<br>verify treatment<br>planning and radiation<br>dose to tissue and<br>organs in or near the<br>irradiated areas of a<br>patient. | The RadiaDyne Prostate<br>Immobilizer Rectal Balloon<br>(PIRB) is intended to be<br>used for the temporary<br>positioning of the rectal<br>wall and adjacent structure<br>in the male human<br>anatomies. | The OARtrac®<br>System is<br>intended for use<br>in photon beam<br>radiation therapy<br>to monitor and<br>verify radiation<br>treatment dose<br>to the<br>surrounding<br>organs at risk<br>tissue during<br>prostate external<br>beam radiation<br>treatment. | The OARtrac®<br>System shares<br>technical elements<br>of the DVS and PIRB<br>in that it functions<br>as both a radiation<br>does verification<br>system and a<br>prostate<br>immobilizer and<br>positioning device. | Material: | Glass capsule | Polyurethane | Polyurethane | Same as above. |
| Indications for<br>Use: | The DVS system is<br>specifically indicated<br>for breast and<br>prostate cancer to<br>measure photon beam<br>therapy and as an<br>adjunct to treatment<br>planning to permit<br>measurement of<br>the in vivo radiation<br>dose received at the<br>tumor periphery,<br>tumor bed and/or<br>surrounding normal<br>tissues for validation<br>of the prescribed<br>dose. | The RadiaDyne Prostate<br>Immobilizer Rectal Balloon<br>is a single-use disposable,<br>inflatable, non-powered<br>positioning device intended<br>to be used for the<br>temporary positioning of<br>the rectal wall and adjacent<br>structure in the male<br>human anatomies. The<br>purpose of the device is to<br>stabilize the prostate<br>during Computed<br>Tomography (CT) exam, X-<br>ray, or Radiation Therapy<br>(RT) treatments. The<br>placement of the balloon<br>requires a Physician or a<br>Physician directed<br>healthcare professional,<br>and is performed as a<br>separate procedure outside<br>of the standard CT exam<br>and RT treatment. | The OARtrac®<br>System is<br>specifically<br>indicated for<br>male prostate<br>cancer treatment<br>to measure<br>photon beam<br>therapy as an<br>adjunct to<br>treatment<br>planning<br>permitting<br>measurement of<br>in-vivo radiation<br>dose received on<br>the anterior<br>surface of a<br>modified<br>Prostate<br>Immobilization<br>Endorectal<br>Balloon (ERB)<br>device to monitor<br>and verify the | The OARtrac®<br>System and PIRB are<br>indicated for the<br>treatment of the<br>prostate only, while<br>the Sicel DVS is<br>indicated for use on<br>the breast and<br>prostate. | Implantable: | Yes, 3-12 cm<br>maximum depth from<br>any skin surface. | No | No | The OARtrac®<br>System's ERB<br>component is not<br>implanted in the<br>body while the<br>DVS's MOSFET<br>dosimeter is. |
| Body Location: | Implanted under skin<br>in the target area. | Placed in rectum | Placed in rectum | The OARtrac®<br>System's ERB and<br>PIRB are indicated<br>for placement in the<br>rectum while the<br>DVS's MOSFET<br>dosimeter is<br>indicate for<br>placement under<br>the skin. | | | | | |
| Target Area: | Breast and Prostate | Prostate | Prostate | The OARtrac®<br>System and PIRB are<br>indicated for<br>treatment of the<br>prostate while the<br>DVS is indicated for<br>treatment of the<br>prostate and breast. | | | | | |
| Fiducial Marker: | Yes | Yes | Yes | All three devices<br>have a fiducial<br>marker that is used<br>as an aid for device<br>position and<br>location in the<br>patient. | | | | | |
| Sterile: | EtO | No | No | The OARtrac®<br>System's ERB<br>component and<br>PIRB are provided<br>non-sterile and not | | | | | |
:
:
:
.
・
:
{4}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
. . . .
______________________________________________________________________________________________________________________________________________________________________________
:
.
{5}------------------------------------------------
| | | | | intended to be<br>sterilized, while the<br>DVS's implantable<br>MOSFET dosimeter<br>is provided EtO<br>sterilized. |
|---------------------|--------------------------------|-----|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Single Use: | Yes | Yes | Yes | All three devices are<br>for single use. |
| Hard Wired: | No, powered<br>telemetrically. | NA | Yes, fiber optic<br>cable. | The OARtrac®<br>System's ERB<br>component is hard<br>wired to the robust<br>cable that is wired<br>to the clinical<br>detector unit, while<br>the DVS employs no<br>hard wiring and<br>operates<br>telemetrically. |
| Pre-Calibrated: | Yes (Cobalt-60) | NA | Yes (Cobalt-60) | Both the OARtrac®<br>System and DVS are<br>pre-calibrated using<br>Cobalt-60. |
| System<br>Software: | Yes | NA | Yes | Both the OARtrac®<br>System and DVS run<br>on their own<br>custom software<br>platforms. |
| Energy Range: | 6-18<br>MV Photon Dose | NA | 1-20<br>MV Photon Dose | Both the OARtrac®<br>System and DVS<br>have the ability to<br>measure the energy<br>range of the<br>radiation photon<br>dose administered<br>to the patient, but<br>use different<br>technologies to<br>accomplish this. PSD<br>fibers measure light<br>output in the<br>OARtrac® System,<br>while the DVS uses<br>MOSFET technology. |
| Dose Range: | 80 Gy maximum | NA | 90 Gy maximum | The capability of the<br>OARtrac® System to<br>measure a dose<br>range to the patient<br>10 Gy higher than<br>the DVS is purely a<br>function of the<br>higher cumulative<br>dose regimes used<br>in current prostate<br>radiation treatment<br>plans compared to<br>when the predicate<br>device was cleared<br>in 2008, and was<br>specified by |
.
:
{6}------------------------------------------------
| Dose Per<br>Fraction: | 150-250 cGY | NA | 1-2000 cGy<br>At dose rates of 1<br>cGy/s and above | RadiaDyne in the design and development of their device.<br>Both the OARtrac® System and DVS have the ability to measure the radiation dose per fraction administered to the patient, but use different technologies (PSD fibers vs. MOSFET technology) to accomplish this, which accounts for the differences in the range of the values shown. |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------|----|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Time to Read<br>Dosimeter: | Up to 10 minutes<br>following radiation<br>therapy. Optimal<br>read time is 2 to 3<br>minutes following<br>irradiation. | NA | 20 seconds | Both the OARtrac® System and DVS have the ability to read and store the radiation dose administered to the patient, but use different technologies (PSD fibers vs. MOSFET technology) to accomplish this which accounts for the differences in the range of the values shown. Since the OARtrac® System uses PSD fibers that are hardwired to the CDU and its camera, readings are updating every 20 seconds during the treatment. This compares to the DVS where the radiation exposure time values are first stored in the implanted dosimeter. Once the radiation is delivered to the patient, the dose information can then be retrieved with the aid of the hand hold reader |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーー
:
.
{7}------------------------------------------------
| | | | | wand. |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dose<br>Accuracy: | <5.5% (2σ) up to 20 Gy<br><6.5% (2σ) up to 74 Gy*<br>Dosimeter axis is<br>placed roughly parallel<br>to the body axis.<br>Actual lot accuracy is<br>specified in the Lot<br>Calibration Certificate.<br>*Accuracy decreases<br>slightly above 74 Gy. | NA | +- 6% at 95%<br>confidence<br>interval | The radiation dose<br>accuracy for the<br>OARtrac® System<br>was obtained using<br>controlled phantom<br>data, and is similar<br>to accuracy<br>reported for the<br>DVS. |
# 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the OARtrac® System and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, RadiaDyne completed a number of tests. The OARtrac® System meets all the requirements for overall design, sterilization, biocompatibility, package shelf-life, electrical safety and EMC, which confirms that the output meets the design inputs and specifications for the device.
The OARtrac® System passed all the testing in accordance with national and international standards shown below to support substantial equivalence of the subject device to the predicates for which substantial equivalence is being claimed.
- Biocompatibility Testing per ISO 10993-1 (Parts 5, 10 and 11) .
- Electrical Safety per IEC 60601-1
- � EMC per IEC 60601-1-2
- Software Verifications and Validation per IEC 62304 .
- . Bioburden per ISO 11137-1
- Package Shelf-Life per ASTM F1090-07
- Device Risk Analysis per ISO 14971
- Dose Range Verification Testing
- Ship Testing Calibration Testing �
- . Balloon Leak Test
- . Balloon Tensile Strength Test '
- . Balloon Burst Strength Test
- . Balloon Stopper Resistance Test
## 10. Clinical Performance Data
There was no human clinical testing required to support the OARtrac® System as the indications for use is equivalent to the predicate devices. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the OARtrac® System.
## 11. Statement of Substantial Equivalence
{8}------------------------------------------------
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared device, or has the same intended use and different technological characteristics, but it can be demonstrated that the new device is substantially equivalent to the predicate device, and that the new device does not raise any questions regarding its safety and effectiveness when compared to the predicate device.
The OARtrac® System functions as both a radiation does verification system and as a prostate immobilizer and positioning device in one complete system, whereas the predicate devices would have to be used in combination to have the same intended use on a male subject. In addition, the OARtrac® System's ERB component is not implanted, thus reducing the risk of complications such as infection, implant migration, implant placement, and adverse reactions to implants by the patient, etc.
Therefore, based on the substantial equivalence analysis described above, the OARtrac® System, as designed, developed and manufactured for RadiaDyne, is determined to be substantially equivalent to the Dose Verification System (K083035) manufactured by Sicel Technologies, and to the Prostate Immobilizer Rectal Balloon (K132194) designed, developed and manufactured for RadiaDyne.
{9}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol featuring three abstract shapes resembling human figures or flowing lines, arranged in a stacked formation.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66.GoG9 Silver Spring, MD 20993-0002
July 22, 2014
RadiaDyne, LLC % Mr. Stuart Goldman Senior Consultant Emergo Group 816 Congress Avenue, Suite 1400 · AUSTIN TX 78701
K141154 Re: Trade/Device Name: OARtrac® System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: NZT and PCT Dated: May 2, 2014 Received: May 5, 2014
Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
Page 2 - Mr. Stuart Goldman
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510{k) Number (if known) Not Assigned K141154
Device Name OARtrac® System
Indications for Use (Describe)
The OARtrace System is specifically indicated for male prostate cancer treatment to measure photon beam therapy as an adjunct to treatment planning measurement of in-vivo radiation dose received on the anterior surface of a modified Prostate Immobilization Endorectal Balloon (ERB) device to monitor and verify the surrounding organs at risk, specifically the protatic rectal interface.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.