K250083 · Nu-Rise, SA · NZT · Oct 1, 2025 · Radiology
Device Facts
Record ID
K250083
Device Name
PRO-DOSE System
Applicant
Nu-Rise, SA
Product Code
NZT · Radiology
Decision Date
Oct 1, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The PRO-DOSE System is intended to provide dosimetry or detect in real-time the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of HDR brachytherapy after-loaders involving Iridium-192 sources.
Device Story
PRO-DOSE System provides real-time in vivo dosimetry during HDR brachytherapy. Input: radiation dose/rate detected by three single-use, pre-calibrated fiber optic probes inserted into patient via standard brachytherapy needles/catheters. Processing: Reading Device captures signals; MiniPC running kiosk-mode GUI processes data and displays real-time dose/rate measurements. Output: real-time monitoring display and PDF Dosimetry Reports. Used in hospitals/clinics by trained personnel. Device does not control radiation delivery; output used by clinicians to verify treatment accuracy. Benefits: enables immediate monitoring of dose delivery, ensuring treatment alignment with planned parameters.
Clinical Evidence
No clinical or animal studies conducted. Evidence based on bench testing, including EMC/electrical safety (IEC 60601-1/1-2), mechanical integrity, dosimetric characterization (accuracy, linearity, repeatability, stability, calibration), phantom testing, human factors validation (IEC 62366), software validation (IEC 62304), and cybersecurity assessment.
Technological Characteristics
Fiber optic sensing technology; single-use probes; MiniPC-based GUI; Ethernet connectivity. Complies with IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366. Shelf-life testing per ASTM F1980-21; transit validation per ASTM D4332-22 and D4169-22.
Indications for Use
Indicated for real-time dosimetry and verification of radiation dose output during HDR brachytherapy procedures involving Iridium-192 sources in clinical settings.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
MOSFET System Portable Dosimeter Dose Verification System, Model TD-RD- 90 (K092285)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 1, 2025
Nu-Rise, SA
% Keira Jessop
Regulatory Affairs Consultant
AlvaMed
935 Great Plain Avenue
Suite 166
Needham, Massachusetts 02492
Re: K250083
Trade/Device Name: PRO-DOSE System
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: NZT, IYE
Dated: January 10, 2025
Received: January 13, 2025
Dear Keira Jessop:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250083 - Keira Jessop
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250083 - Keira Jessop
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250083 | |
| Device Name PRO-DOSE System | |
| Indications for Use (Describe) The PRO-DOSE System is intended to provide dosimetry or detect in real-time the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of HDR brachytherapy after-loaders involving Iridium-192 sources. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250083
Page 1 of 6
510(k) Summary
Name and Address of Submitter
NU-RISE SA
Univ. Aveiro Incubator
PCI – Creative Science Park
Via do Conhecimento, Edifício
Central
3830-352 Ilhavo, Portugal
Joana Melo
+351 919010803
Correspondent/Primary Contact Person
Keira Jessop, Regulatory Affairs Consultant
AlvaMed, Inc.
935 Great Plain Avenue, Suite 166
Needham, MA 02492
Submission Information
Date Summary Prepared: September 30, 2025
| Subject Device: | Trade/Device Name: | PRO-DOSE System |
| --- | --- | --- |
| | Manufacturer: | NU-RISE |
| | Common Name: | |
| | Regulation Number: | 892.5050 |
| | Regulation Name: | Medical charged- particle radiation therapy system |
| | Regulation Class | II |
| | Product Code: | NZT, IYE |
| | Review Panel: | Radiology |
| Predicate Device: | Clearance: | K092285 |
| --- | --- | --- |
| | Trade/Device Name: | MOSFET System Portable Dosimeter Dose Verification System, Model TD-RD- 90 |
| | Manufacturer: | Best Medical Canada, Ltd. |
| | Regulation Number: | 892.5050 |
| | Regulation Name: | Medical charged- particle radiation therapy system |
| | Regulation Class | II |
| | Product Code: | IYE |
{5}
Page 2 of 6
Reference Device: Clearance: K182395
Trade/Device Name: OARtrac System
Manufacturer: AngioDynamics (formerly RadiaDyne)
Regulation Number: 892.5050
Regulation Name: Medical charged- particle radiation therapy system
Regulation Class: II
Product Code: NZT
## Valid Predicate Discussion
MOSFET System Portable Dosimeter Dose Verification System, Model TD-RD- 90 was selected as the valid predicate device and the OARtrac System was selected as the reference device to support the 510(k) submission because they received FDA clearance using well-established methods. Both the predicate and reference device meet the expected predicate performance due to no reports of unexpected injury, deaths, or malfunctions associated with the device. After conducting a search on the Manufacturer and User Facility Device Experience (MAUDE) Database, Medical Device Reporting (MDR), MedSun Reports Database, Medical Device Safety and CBER Safety & Availability (Biologics) and Recall Database websites, it was found that there are no known unmitigated use-related or design safety issues and the predicate device have not been subject to a design-related recall. Design related recalls were identified in the FDA Database for the reference device. The reason for recall was that the device may result in readings outside of the expected accuracy range. All the recalls were terminated on October 26, 2020.
Table 1: Valid Predicate Device
| Valid Predicate Device | A - Well established methods | B - Meets or exceeds expected predicate performance | C - Unmitigated use-related or design-related safety issues | D – Associated design-related recall |
| --- | --- | --- | --- | --- |
| MOSFET System Portable Dosimeter Dose Verification System, Model TD-RD- 90 (Predicate Device) | Used relevant methods that were published in the public domain. | History of Safe use, established due to duration of device on the market | No known unmitigated use-related or design related safety issues | No design-related recall identified |
{6}
Page 3 of 6
| OARtrac System
(Reference Device) | Used relevant methods that were published in the public domain. | History of safe use established due to no reports of unexpected injury, deaths, or malfunctions associated with the device | No known unmitigated use related or design-related safety issues | Design-related recalls identified in the FDA Database |
| --- | --- | --- | --- | --- |
## Device Description
The PRO-DOSE System is intended to provide dosimetry or detect in real-time the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of HDR brachytherapy after-loaders involving Iridium-192 sources.
It is designed to measure the total dose and dose rate administered to patients during radiation therapy treatments. It enables real-time, in vivo monitoring throughout the treatment and generates a Dosimetry Report upon completion.
The Reading Device allows simultaneous dosimetry measurements in up to three independent positions by using three probes.
The System can be operated via an Software Graphical User Interface (GUI) that runs on the MiniPC, which functions in kiosk mode to ensure secure and exclusive use. The Software is pivotal in controlling the Reading Device, processing the signal and displaying measurements. As treatment progresses, real-time dose and dose rate measurements are provided. It also allows for the export of a Dosimetry Report as PDF files for documentation and review. The Probes allow instant detection and quantification of radiation dose. These probes are strategically placed at predefined locations by the user to ensure optimal monitoring throughout the treatment. The probes are single-use and pre-calibrated. Due to their small size and water equivalence, these probes can be inserted into the patient body using the same techniques through needles used in the HDR brachytherapy treatments.
The real-time radiation dose data is intended for use by qualified and trained personnel who recognize radiation hazards associated with the use of radiation therapy and radiology equipment. The system does not interact with the radiation source and is solely designed to monitor and report the accuracy of the treatment delivery. It does not deliver, plan, simulate, or control the radiation delivery. The trained and qualified user is responsible for the interpretation of the information.
The current version of PRO-DOSE System is specifically suitable for monitoring High Dose Rate (HDR) Brachytherapy treatments involving Iridium-192 sources.
## Indications for Use
The PRO-DOSE System is intended to provide dosimetry or detect in real-time the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of HDR brachytherapy after-loaders involving Iridium-192 sources.
{7}
Page 4 of 6
Summary of Technological Characteristics
| | Subject Device | Predicate Device | Reference Device |
| --- | --- | --- | --- |
| Trade Name | PRO-DOSE System | Portable Dosimeter Dose Verification System, Model TD-RD-90 (MOSFET System) | OARtrac System |
| 510(k) Number | K250083 | K092285 | K182395 |
| Indications for Use | The PRO-DOSE System is intended to provide dosimetry or detect in real-time the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of HDR brachytherapy after-loaders involving Iridium-192 sources. | To provide dosimetry or to detect the dose delivered to a patient during therapeutic radiation procedures. The system is intended for the verification of the output of radiation producing devices. | The OARtrac System with patient specific, reusable up to five times, precalibrated PSD sensors are intended for use during cancer treatments to measure photon and electron radiation therapy as an adjunct to treatment planning permitting measurement and validation of radiation dose received by the patient to the targeted area of their body. The system is indicated for use with electron energy when adhered to the skin with or without a bolus, and for photon energy when adhered to the skin with a minimum 3mm bolus or inserted into the rectum for measurements at the rectal wall during cancer treatment via a specifically designed OARtrac endorectal balloon device. |
| Regulation Number | 892.5050 | 892.5050 | 892.5050 |
| Regulation Name | Medical charged- particle radiation therapy system | Medical charged- particle radiation therapy system | Medical charged- particle radiation therapy system |
| Device Class | II | II | II |
| Product Code | NZT, IYE | IYE | NZT |
| Technology | Fiber Optic | Semiconductor | Fiber Optic |
| Dose Monitoring | Continuous Real Time | Total dose after irradiation or updates every 20 | - |
{8}
Page 5 of 6
| | Subject Device | Predicate Device | Reference Device |
| --- | --- | --- | --- |
| | | seconds | |
| How Provided and Use | Clean, Single Use | Clean, Reusable | Clean, Reusable |
| Location of Use | Within the body in conjunction with catheters | Adhered to skin, or within the body in conjunction with catheters | Adhered to skin, or within rectal balloon accessory device |
| Accessories | Probes are connected to Reading Device, which is connected to software running on a miniPC | Probes are connected to Reading Unit which displays the measurements, data can be transferred to a computer via USB | Rectal balloon and Connectors |
| Use Setting(s) & Environment | For use during brachytherapy at Hospital or Radiation Therapy Clinic | For use during radiation beam therapy or brachytherapy at Hospital or Radiation Therapy Clinic | For use during radiation beam therapy or brachytherapy at Hospital or Radiation Therapy Clinic |
The difference in the indications between the subject and predicate device do not affect the safety and effectiveness. Differences in the two indications are related to the time period in which the dose amount is provided and the radiation source of which the output is being verified. The subject device is able to provide the dose and dose rate during the procedure in real time while the predicate provides information only about total dose after irradiation or updates every 20 seconds. The subject device is also specific to verifying the output of HDR brachytherapy after-loaders involving Iridium-192 sources, which is a subset of the predicate's indication of "radiation producing sources. These differences are due to the device technology or specific testing conducted and test set-up, but the intended use of the device remains the same.
With regards to the technology between the subject and predicate device, there are minor differences, but they do not affect the safety and effectiveness of the device. The subject device is inserted into standard brachytherapy needles or catheters, while the predicate device can either be inserted in the same manner or adhered to the skin. The subject device probes are single use, while the predicate probes are reusable until their useful life. The subject device uses fiber optic technology, while the predicate uses semiconductor technology. It should be noted that the reference device uses fiber optic technology. In the predicate device, the dosimetry records can be transferred to a computer through a USB connection, while the subject device is connected to software running on a computer through an ethernet cable. The fiber optic technology and connection to software allow for real time dosimetry readings, which is not available with the predicate device. The intended use of both devices is dosimetry readings, and any technological differences do not impact this. None of the technology differences represent new technology.
{9}
Page 6 of 6
# Performance Testing Summary
## Summary of Non-Clinical Functional and Performance Testing
The following tests were performed to evaluate the performance and safety of the device.
- EMC/Electrical Safety Software Validation per IEC60601-1, IEC 60601-1-2, and FDA Guidance document, Electromagnetic Compatibility (EMC) of Medical Devices
- Conditioning and Transit Validation per ASTM ASTM D4332-22 and ASTM D4169-22
- Shelf-life Testing for probes per ASTM F1980-21
- Cycle Testing to determine use life of the Reading device
- Operating Ranges Testing to confirm device operates within labeled humidity and temperature conditions
- Mechanical Integrity for:
- Handling and bending of probe
- Probe integrity with needle fixer
- Physical and Dosimetric Bench Characterization
- Accuracy
- Linearity
- Repeatability
- Stability
- Calibration
- Angular/Spatial Response
- Phantom Testing
- Ensure device meets performance specifications in representative simulated use phantoms
- Cleaning Validation to demonstrate effectiveness of cleaning and that device functionality is maintained
- Human Factors Validation per IEC 62366 and FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices
- Software Validation per IEC 62304 and FDA guidance document, Content of Premarket Submissions for Device Software Functions
- Cybersecurity Assessment per FDA guidance document, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions
## Summary of Animal and Clinical Data
No animal or clinical studies were conducted to demonstrate safety of the device.
## Conclusion
Based on the information presented in this Premarket Notification, NU-RISE concludes that PRO-DOSE is substantially equivalent to the predicate based on the same intended use. Any differences in technological characteristics between the subject and predicate device have demonstrated continued performance and do not raise issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.