The Calypso system is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso system provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders. Implanted Beacon transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures. Permanent Beacon transponders are indicated for implantation in the body, specifically in the prostate and peri-prostatic tissue (i.e., prostatic bed), and in soft tissue to align and monitor the treatment isocenter in real time during radiation therapy.
Device Story
Permanent Beacon transponder is a small, radiopaque, echogenic, electromagnetic fiducial marker; implanted via 14G introducer needle into or near tumor/radiation target. Device consists of sealed biocompatible-glass capsule containing passive electrical circuit. Used with Calypso System (3.0+) for electromagnetic localization or radiographic systems (kV x-ray, fluoroscopy, CT) for radiographic localization. Operates as fiducial marker to ensure accurate patient positioning relative to radiation therapy isocenter. Provides continuous, real-time localization data to clinicians during radiation therapy; enables precise alignment of radiation beams to target. Benefits include improved accuracy in radiation delivery and monitoring of target position.
Clinical Evidence
No clinical data provided. Substantial equivalence based on bench testing, biocompatibility evaluation per ISO 10993, and comparison of technological characteristics to legally marketed predicate devices.
Technological Characteristics
Cylindrical fiducial marker (1.8 mm dia. x 8.5 mm length); biocompatible-glass encapsulated passive electrical circuit (copper/ferrite). Radiopaque (kV x-ray, fluoroscopy, CT) and echogenic (ultrasound). MR conditional. Sterilized via gamma irradiation. Single-use. Implanted via 14G needle.
Indications for Use
Indicated for patients requiring radiotherapy target localization. Used to mark soft tissue, specifically prostate and peri-prostatic tissue (prostatic bed), for real-time alignment and monitoring of treatment isocenter during radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Varian Medical Systems. The logo consists of the word "VARIAN" in a serif font, with a stylized "i" that extends above the other letters. Below the word "VARIAN" is the phrase "medical systems" in a smaller, sans-serif font. The logo is simple and professional, and it conveys a sense of trust and reliability.
**JUN 27 2014**
K140823
Page 1 of 5
## 510(k) Summary
Submitter information: Varian Medical Systems 2101 4th Ave., Suite 100 Seattle, WA 98121 Phone: 206-254-0600 Fax: 206-577-4597 Contact person: Lisa Levine, PhD Director, Clinical and Pre-Market Regulatory Affairs Date summary prepared: March 28, 2014 Permanent Beacon transponder or Trade name: Soft Tissue Beacon transponder, provided in the Soft Tissue Beacon Package Common name: Fiducial marker Classification name: Medical charged-particle radiation therapy system Classification number: CFR 892.5050 Class: Class II Product code: IYE Predicates: Calypso System with Beacon transponders (K060906, K080726, K123137) Gold Soft Tissue Marker (K071614) ONC Gold Seed Marker (K071673 )
{1}------------------------------------------------
#### Device description:
The Soft Tissue Beacon Transponder is a small, radiopaque, echogenic, electromagnetic fiducial marker designed for permanent implantation and intended for radiotherapy target localization to ensure accurate positioning for radiation therapy. It consists of a sealed biocompatible-glass capsule containing a small, passive electrical circuit. The Soft Tissue Beacon Transponder may be used with the Calypso System (3.0 or later) as an electromagnetic fiducial marker, or with radiographic-based systems (e.g., kV x-ray, fluoroscopy, and CT) as a radiographic fiducial marker.
Each transponder is implanted with a separate 14G introducer needle (introducer) in or near the tumor or intended radiation target. Three Soft Tissue Beacon Transponders and three single-use introducers are provided in each Soft Tissue Beacon Care Package. The device is single-use and provided sterile.
## Indications for use:
The Calypso system is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso system provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.
Implanted Beacon transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.
Permanent Beacon transponders are indicated for implantation in the body, specifically in the prostate and peri-prostatic tissue (i.e., prostatic bed), and in soft tissue to align and monitor the treatment isocenter in real time during radiation therapy.
## Substantial equivalence:
The subject device has the same intended use as the predicates. The technological characteristics of the subject device are supported by the technological characteristics of the predicates. The subject device is as safe and effective as the predicates. There are no different questions of safety or effectiveness. Thus, the subject device is substantially equivalent.
The substantial equivalence table is shown on the next page for reference.
{2}------------------------------------------------
| Item/Characteristic | Permanent Beacon<br>Transponder with Expanded<br>Indications | Permanent Beacon<br>Transponder<br>(predicate) | Gold Soft Tissue Marker<br>(predicate) | ONC Gold Seed Marker<br>(predicate) |
|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| number | n/a | K060906, K080726, K123137 | K071614 | K071673 |
| product code | IYE | IYE | IYE | IYE |
| Intended use | Radiotherapy target<br>localization | Radiotherapy target<br>localization | Radiotherapy target<br>localization | Radiotherapy target<br>localization |
| Indications for use<br>statement<br>(bold emphasis added<br>only as an aid to the<br>reader) | The Calypso System is<br>intended for use as an adjunct<br>in treatment planning and<br>radiation therapy, to align<br>and/or monitor the patient's<br>position relative to the<br>isocenter of a radiation<br>therapy system. The Calypso<br>System provides accurate,<br>precise and continuous<br>localization of a treatment<br>isocenter by using two or<br>more Beacon transponders. | The Calypso System is<br>intended for use as an adjunct<br>in treatment planning and<br>radiation therapy, to align<br>and/or monitor the patient's<br>position relative to the<br>isocenter of a radiation<br>therapy system. The Calypso<br>System provides accurate,<br>precise and continuous<br>localization of a treatment<br>isocenter by using two or<br>more Beacon transponders. | The fiducial markers are<br>intended to be implanted into<br>the body to accurately<br>visualize and constitute the<br>reference frame for<br>stereotactic radiosurgery and<br>radiotherapy target<br>localization.<br><br>Specifically, they can be used<br>in intracranial diseases as<br>gliomas, neuromas,<br>meningiomoas, astrocytomas,<br>arteriovenous malformations,<br>and metastatic carcinomas.<br><br>Additionally, they can be used<br>in the body for treating<br>hepatic, pancreatic,<br>retroperitoneal, paraspinal,<br>skeletal, prostatic and breast<br>tumors. | The ONC Marker is indicated<br>for use to radiographically<br>mark soft tissue for future<br>therapeutic procedures. |
| | Implanted Beacon<br>transponders are indicated for<br>use to radiographically and<br>electromagnetically mark soft<br>tissue for future therapeutic<br>procedures. | Implanted Beacon<br>transponders are indicated for<br>use to radiographically and<br>electromagnetically mark soft<br>tissue for future therapeutic<br>procedures. | | |
| | | Permanent Beacon<br>transponders are indicated for<br>implantation in the body,<br>including in the prostate and | Permanent Beacon<br>transponders are indicated for<br>implantation in the prostate<br>and the peri-prostatic tissue | |
| Item/Characteristic | Permanent Beacon<br>Transponder with Expanded<br>Indications | Permanent Beacon<br>Transponder<br>(predicate) | Gold Soft Tissue Marker<br>(predicate) | ONC Gold Seed Marker<br>(predicate) |
| | the peri-prostatic tissue (i.e.,<br>prostatic bed), to align and<br>monitor the treatment<br>isocenter in real time during<br>radiation therapy. | (i.e., prostatic bed) to align<br>and monitor the treatment<br>isocenter in real time during<br>radiation therapy. | | |
| Dimensions | Cylindrical | Cylindrical | Cylindrical | Cylindrical |
| Materials | 1.8 mm dia. x 8.5 mm length<br>Biocompatible-glass-<br>encapsulated electrical<br>circuit (primarily copper<br>and ferrite) | 1.8 mm dia. x 8.5 mm length<br>Biocompatible-glass-<br>encapsulated electrical<br>circuit (primarily copper<br>and ferrite) | 1.6 mm dia. x 3 mm length<br>Gold | 1.2 mm dia. x 10 mm length<br>Gold |
| Radiographic and<br>Ultrasound imaging | Radiopaque (kV x-rays,<br>fluoroscopy, CT);<br>Echogenic (ultrasound) | Radiopaque (kV x-rays,<br>fluoroscopy, CT);<br>Echogenic (ultrasound) | Radiopaque (kV x-rays, MV<br>x-rays, CT);<br>Echogenic (ultrasound) | No details provided but known<br>to be radiopaque and<br>echogenic |
| Means of radiotherapy<br>target localization | Used as electromagnetic<br>fiducial marker with Calypso<br>system | Used as electromagnetic<br>fiducial marker with Calypso<br>system | Used as radiographic fiducial<br>marker with radiographic<br>detector | Used as radiographic fiducial<br>marker with radiographic<br>detector |
| | Used as radiographic fiducial<br>marker with radiographic<br>detector | Used as radiographic fiducial<br>marker with radiographic<br>detector | | |
| MR status | MR conditional<br>Rigid, no attachments | MR conditional<br>Rigid, no attachments | MR conditional<br>Rigid, no attachments, knurled | MR conditional<br>Rigid, no attachments |
| Introducer needle gauge | 14G | 14G | 14G | 17G |
| Introducer needles<br>available | Provided with introducer<br>needles | Provided with introducer<br>needles | Provided with introducer<br>needles | Provided with introducer<br>needles |
| Single-use | For single-use | For single-use | For single-use | For single-use |
| Sterility | Provided sterile | Provided sterile | Provided sterile | Provided sterile |
| | For permanent implantation | For permanent implantation | For permanent implantation | For permanent implantation |
| Item/Characteristic | Permanent Beacon<br>Transponder with Expanded<br>Indications | Permanent Beacon<br>Transponder<br>(predicate) | Gold Soft Tissue Marker<br>(predicate) | ONC Gold Seed Marker<br>(predicate) |
| Biocompatibility | Biocompatibility was<br>evaluated per ISO 10993 | Biocompatibility evaluated per<br>ISO 10993 | Biocompatibility evaluated per<br>ISO 10993 | No information |
| Sterilization | Sterilization: gamma<br>irradiation | Sterilization: gamma<br>irradiation | Sterilization method: EtO | Sterilization method: EtO |
Table 7-1. Substantial Equivalence Comparison Table
. .
1
.
{3}------------------------------------------------
:
:
{4}------------------------------------------------
. . .
K140823
Page 5 of 5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized symbol that resembles an eagle or bird with three curved lines representing its wings or feathers.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
June 27, 2014
Varian Medical Systems, Inc. % Lisa Levine, Ph.D. Director, Clinical and Pre-Market Regulatory Affairs 2101 4th Avenue, Suite 100 SEATTLE WA 98121
Re: K140823
Trade/Device Name: Permanent and Soft Tissue Beacon Transponder Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: March 31, 2014 Received: April 1, 2014
Dear Dr. Levine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Lisa Levine, Ph.D.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K140823
Device Name: Permanent Beacon® Transponder
Indications for Use:
The Calypso system is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso system provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.
Implanted Beacon transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.
Permanent Beacon transponders are indicated for implantation in the body, specifically in the prostate and peri-prostatic tissue (i.e., prostatic bed), and in soft tissue to align and monitor the treatment isocenter in real time during radiation therapy.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. DiHara
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k) K140823
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.