CALYPSO SYSTEM

K123137 · Varian Medical Systems · IYE · Dec 5, 2012 · Radiology

Device Facts

Record IDK123137
Device NameCALYPSO SYSTEM
ApplicantVarian Medical Systems
Product CodeIYE · Radiology
Decision DateDec 5, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders. The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces. Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures. Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy. Surface Beacon® transponders are indicated for temporary external placement on the skin, to monitor respiratory motion and other patient motion in real time during radiation therapy.

Device Story

Calypso System provides continuous, objective 3D localization of radiation therapy targets; utilizes non-ionizing electromagnetic signals from implanted or surface-applied Beacon transponders; infrared optical system tracks array and transponders. System aligns patient to therapy isocenter; monitors target position in real-time. Dynamic Edge Gating feature interfaces with external radiation therapy systems to trigger beam-hold when target moves outside defined tracking limits; automatically resumes beam when target returns to range. Used in clinical radiation therapy environments; operated by trained personnel. Output provides real-time motion data to clinicians; enables precise beam delivery; minimizes radiation exposure to healthy tissue by gating beam based on target position.

Clinical Evidence

Bench testing only. Software, hardware, and system-level verification and validation testing performed to demonstrate that the system meets intended use requirements and safety standards.

Technological Characteristics

Electromagnetic and infrared optical sensing; Beacon transponders (implanted/surface); console and tracking station computer hardware; interface for external radiation therapy gating. Electrical safety: EN60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007); EMC safety: EN60601-1-2 ed. 2.0.

Indications for Use

Indicated for patients undergoing radiation therapy requiring target localization and motion monitoring. Permanent Beacon transponders indicated for prostate and peri-prostatic tissue implantation. Surface Beacon transponders indicated for temporary skin placement to monitor respiratory and patient motion.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K123137: 510(K) SUMMARY # DEC 0 5 2012 This 510(k) Summary is provided in accordance with 21 CFR 807.92. | Date of preparation: | 04 October 2012 | |------------------------|--------------------------------------------------------------------------------------------------------------------------| | Revised Date: | 05 November 2012 | | Submitter information: | Varian Medical Systems<br>2101 Fourth Avenue, Suite 500<br>Seattle, WA 98121<br>Phone: 206-254-0600<br>Fax: 206-254-0606 | | Contact: | Marcia A Page<br>Director Quality Assurance and Regulatory Affairs | | Device trade name: | Calypso® System<br>Calypso® System with Dynamic Edge™ Gating | | Common name: | Patient localization system | | Classification name: | Medical charged-particle radiation therapy system | | Classification: | CFR 892.5050<br>Class II<br>Product code: IYE | | Predicate devices: | Calypso System (K060906, K102373, K112768) | Device Description: The Calypso System utilizes non-ionizing electromagnetic and optical technology to provide accurate, objective, and continuous localization of a treatment target for patient alignment and target position monitoring during radiation therapy. Use of the Calypso System for target localization is based on the system's detection of non-ionizing electromagnetic signals from passive markers, called Beacon® transponders. The Beacon transponders are implanted in or near the treatment target or placed externally on the surface of a patient. {1}------------------------------------------------ Implanted Beacon transponders when used with the Calypso System enable objective measurement of the location of the target in 3 dimensions. The system operator uses this information to align the patient's treatment target to the isocenter of the therapy system prior to treatment. This information can also be used to monitor (track) the position of the treatment target during radiation therapy treatment. When using Surface Beacon transponders this information can be used to monitor respiratory motion and other patient motion in real time during radiotherapy. When correlated to the patient's treatment target motion the system provides objective measurement of the location of the target in 3 dimensions. The Dynamic Edge Gating feature enables the Calypso System to connect with radiation therapy systems configured with gating capabilities via an interface to external systems. With this feature added to the Calypso System, a beam hold is signaled to the radiation therapy system when the treatment target position has moved outside the defined tracking limits. The radiation therapy system is signaled to remove the beam-hold upon the target's return to a position inside the defined tracking limits. ## Indications for Use: The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders. The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces. Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures. Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy. Surface Beacon® transponders are indicated for temporary external placement on the skin, to monitor respiratory motion and other patient motion in real time during radiation therapy. {2}------------------------------------------------ Technological Characteristics – See device comparison table below | Feature and/or<br>Specification of the<br>Calypso System | Cleared Calypso System<br>Feature/Specification | Modified Calypso<br>System Feature/<br>Specification | |------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | Infrared optical system | Cameras and infrared targets<br>on array and Beacon<br>Transponders | No change | | Monitoring | Real Time Monitor patient<br>motion | No change | | Tracking | Real Time Tracking Patient<br>and Respiratory motion | No change | | Gating | Interface to external systems | No change | | Records | Display and Record motion | No change | | Computer hardware and<br>software | Computer systems (console<br>and tracking station) to<br>control device functions and<br>provide for user interface | No change | | Electromagnetic<br>technology (array) | Used to track motion during<br>therapeutic procedures | No change | | Compatibility with the<br>environment and other<br>devices | Operates in a radiation<br>therapy system environment | No change | | Electrical safety<br>mechanical safety | EN60601-1: 1990 / A1:1993 /<br>A2:1995<br>EN 60601-1-1: 2001 | EN60601-1 EN<br>60601-1: 2005 +<br>CORR. 1 (2006) +<br>CORR. 2 (2007) | | EMC Safety | EN60601-1-2 ed. 2.0 | No Change | ## Summary of Performance Testing: Results of software, hardware and system verification and validations testing demonstrate that the Calypso System satisfies the intended use as described above. . {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines extending from its body, representing the department's commitment to health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol. The logo is simple, yet recognizable, and represents the department's role in providing essential services to the American people. #### DEPARTMENT OF HEALTH & HUMAN SERVICES #### Public Health Service December 5. 2012 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Ms. Marcia Page Director Quality Assurance and Regulatory Affairs Varian Medical Systems 2101 Fourth Avenue South, Suite 500 SEATTLE, WA 98121 Re: K123137 Trade/Device Name: Calypso® System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: October 4, 2012 Received: October 5, 2012 Dear Ms. Page: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Michael D. O'Hara Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use ### 510(k) Number (if known): K123137 Device Name: Calypso® System Indications for Use: The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and/or monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders. The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces. Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures. Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy. Surface Beacon® transponders are indicated for temporary external placement on the skin. to monitor respiratory motion and other patient motion in real time during radiation therapy. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR - Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Michael D. O'Hara (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510(k)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...