TERASON USMART3200 AND BENQ UP200ULTRASOUND SYSTEM
Applicant
Teratech Corp.
Product Code
IYN · Radiology
Decision Date
Mar 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Teratech Corporation Terason™ uSmart3200T (also known as the BenQ UP200) is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Intra-Operative (abdominal, organs and vascular), Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.
Device Story
Portable tablet-based ultrasound system (uSmart3200T/BenQ UP200) for real-time diagnostic imaging and fluid flow analysis. System comprises a tablet computer and a proprietary ultrasound engine (housed in a base compartment) that controls transducer acoustic output and processes return echoes. Data transfers via FireWire (IEEE 1394) to the tablet for image generation and display. Operated by physicians in clinical settings. Features include B-mode, M-mode, Color/Power Doppler, and spectral Doppler. Enhanced Needle Visualization (ENV) available on 15L4 transducer. System supports various transducers for diverse clinical exams. Output displayed on 11.5" touch screen; aids clinicians in diagnostic evaluation and needle/catheter guidance. Benefits include portability and high-resolution imaging for point-of-care or clinical use.
Clinical Evidence
Bench testing only. Performance evaluated via acoustic tank testing for new transducers (15L4, 16HL7, 8EC4A) per IEC 60601-2-37. Biocompatibility testing performed per ISO 10993-5 and -10. Electrical safety and EMC compliance verified per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6.
Technological Characteristics
Portable tablet (4.9 lbs) with 11.5" touch screen and integrated Li-Polymer battery. Uses custom ultrasound engine for 128-channel transmit/receive. Connectivity via FireWire. Transducers use piezoelectric elements. Patient contact materials include ABS, Silicone, and PEI. Software runs on Windows. Sterilization/reprocessing per AAMI TIR No. 12.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal, cardiac, and peripheral vascular applications. Used by qualified physicians.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K140524
Page 1 of 15
### 510(k) Summary or Statement
### Teratech Corporation
### Terason uSmart3200T and BenQ UP200 Ultrasound System
### 1. Sponsor:
Teratech Corporation 77-79 Terrace Hall Ave. Burlington, MA 01803
Contact Person: Ben Chiampa Director of Quality Assurance Telephone: 781-270-4143
Date Prepared: February 18, 2014
### 2. Device Name
Terason uSmart3200 and BenQ UP200 Ultrasound System Proprietary Name: Common / Usual Name: Diagnostic Ultrasound System Classification Name: Diagnostic Ultrasound Transducer
Ultrasonic Pulsed Doppler Imaging System (21 CFR 892.1550, 90-IYN) Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560, 90-IYO) Diagnostic Ultrasonic Transducer (21 CFR 892.1570, 90-ITX)
### 3. Predicate Device
Terason uSmart3200T Ultrasound System (K131209) Terason™ t3000 Ultrasound System (K112953) Terason™ T3200 Ultrasound System (K110020)
#### 4. Intended Use
The Teratech Corporation Terason™ uSmart3200T (also known as the BenQ UP200) is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Intra-Operative (abdominal, organs and vascular), Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.
### 5. Device Description
{1}------------------------------------------------
The Terason uSmart3200T ultrasound system is a portable tablet style full feature, general purpose, and diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. The Terason uSmart3200T Ultrasound System is equivalent to the previously cleared version of the uSmart3200T Ultrasound Systems. The modification includes the addition of 3 transducers (15L4, 16HL7, and 8EC4A) with the no change to tablet style computer form factor.
The Terason™ uSmart3200T ultrasound system was the previously cleared on the date of May 28, 2013 as described in the 510(k) submission (K131209). This system contains a proprietary ultrasound engine for controlling the acoustic output of the transducer and processing the return echoes in real time. This data are then transferred to the tablet computer over a FireWire (aka IEEE 1394) connection for further processing and generation and display of the ultrasound image.
The Terason™ uSmart3200T ultrasound tablet weighs 4.9 pounds (2.21 Kg) and has an 11.5" backlit touch screen. The tablet dimensions (8.82"(H) x 12.64"(W) x 1.25"(D)) are chosen to allow portability. A Lithium-Polymer battery (integrated into the tablet) provides 2 hours of continuous ultrasound scanning. The tablet includes a docking station (for charging) that uses a medicalgrade power supply. The ultrasound transducer connector is identical to that used in the Terason™ predicate device, the uSmart3200T. Optional accessories include a cart and printer.
### 6. Technology Characteristics
The design and construction of the Terason uSmart3200T is same to the Terason uSmart3200T Ultrasound system which was cleared in May 2013. These systems utilize a portable computer running Windows to run the ultrasound application and a custom designed engine for control of the acoustic array and processing of the return echoes. The engine is housed in a compartment that is attached to the bottom of the tablet computer.
The uSmart3200T system contains the same ultrasound engine as the predicate device Terason T3200 ultrasound system for controlling the acoustic output of the transducer and processing the return echoes in real time. These data are then transferred to the tablet computer over a FireWire connection for further processing and generation and display of the ultrasound image
The differences between the Terason uSmart3200T and the previous Terason uSmart3200T Ultrasound System (the predicate device) include the following:
- Three transducers have been added to the system. The software has been modified to . control these transducers and ensure compliance to the standards controlling acoustic and thermal power.
- . Added support for 15L4, 16HL7 and 8EC4A transducers
- Confirmed transducer id numbers and names O
- Confirmed transducer geometries and characteristic parameters o
- o Confirmed 15L4 and 16HL7 acoustic tables and added 8EC4A acoustic tables
- Added 15L4, 16HL7 and 8EC4A to the table of allowed transducers o
- Added imaging presets for 15L4, 16HL7 and 8EC4A with o
- Musculoskeletal selections same as offered on Terason t3200 B
- . 8EC4A gynecological and prostate exams
- l 15L4 same exam and preset selections as 12L5A
- l Vascular exams and presets for 15L4 and 16HL7
{2}------------------------------------------------
- Added the Vascular Application series with new exam license files .
- Added exam license files for gynecological and prostate exam types .
- . Added the Enhanced Needle Visualization (ENV) feature as a licensed option, available only on the 15L4 transducer
- . Added UI feature to optionally swap position of the Control Panel Area and the Thumbnail Area on the screen, left vs. right
- Removed TGC Display from the General tab of Setup .
- Added System Backup and Restore, initiated from the General tab of Setup ●
- Added Gesture options to the General tab of Setup .
- o Disable All Gestures
- Enable Double Tap Gesture o
- ECG options disable on General Tab of Setup .
- Added measurement configurations for Spectral Doppler and M-mode. .
- Added vessel name to the measurement configuration .
- . Added feature to modify user annotations
- In the Store/Acquire tab of Setup, added feature to enable retrospective acquisition of the . live image stream
- . Disabled selection of ECG related features on the Store/Acquire tab of Setup
- . Added DICOM communication option and features, Storage and Modality Worklist
- . Changed designation of Clinician to Operator on the Patient screen
- Added selection checkboxes to multiple selection list controls .
- . Added touch gestures for multiple selection in list controls
- Added touch gestures for freeze, play/pause, zoom and pan, depth, steering, split-� screen, spectral Doppler scale and baseline, color Doppler scale, review navigation, full screen, text and body markers
- . Added variable sector size and position for the 4V2A transducer
- Minor updates to soft key menus .
- Modified 12L5A Musculoskeletal presets to be the same selections as offered on the . Terason t3200
- . Added Musculoskeletal, Breast, Thyroid and Vascular Access exam measurements and annotations, same as offered on Terason t3200.
- 7. Table of Similarities and Differences Compared to the Predicate Devices
Terason uSmart3200T System and Transducers Comparison and Discussion (Transducers 8EC4A, 16HL7, and 15L4) and previously cleared 510(k) K131209 transducers (12L5A, 5C2A, and 4V2A)
### Terason uSmart3200T Tablet Computer
| Subject Device Model | Comparable Predicate<br>Device | Comparable Predicate<br>Devices |
|----------------------|---------------------------------|---------------------------------|
| Terason uSmart3200T | Terason uSmart3200T<br>K1301209 | Terason t3000<br>K112953 |
| (This Submission) | | |
{3}------------------------------------------------
| Intended Use | Diagnostic Ultrasound | Diagnostic Ultrasound | Diagnostic Ultrasound |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | imaging or fluid flow | imaging or fluid flow | imaging or fluid flow |
| | analysis of the human | analysis of the human | analysis of the human |
| | body | body | body |
| Indication for Use | Fetal, Abdominal, Intra-<br>operative (Spec.),<br>Pediatric, Small Organ<br>(Thyroid, Breast, Testes,<br>etc.), Neonatal Cephalic,<br>Adult Cephalic, Trans-<br>rectal, Trans-vaginal,<br>Musculo-skel.<br>(Convent.), Musculo-<br>skel. (Superfic), Cardiac<br>Adult, Cardiac Pediatric,<br>Peripheral vessel | Fetal, Abdominal,<br>Pediatric, Small Organ<br>(Thyroid, Breast, Testes,<br>etc.), Neonatal Cephalic,<br>Adult Cephalic, Musculo-<br>skel. (Convent.),<br>Musculo-skel. (Superfic),<br>Cardiac Adult, Cardiac<br>Pediatric, Peripheral<br>vessel | Fetal, Abdominal, Intra-<br>operative (Spec.), Intra-<br>operative (Neuro),<br>Laparoscopic, Pediatric,<br>Small Organ (Thyroid,<br>Breast, Testes, etc.),<br>Neonatal Cephalic,<br>Adult Cephalic, Trans-<br>rectal, Trans-vaginal,<br>Musculo-skel.<br>(Convent.), Musculo-<br>skel. (Superficial),<br>Cardiac Adult, Cardiac<br>Pediatric, Peripheral<br>vessel |
| Transducer Types | Linear Array<br>Curved Array<br>Phased Array<br>Endocavity - curved<br>array<br>Hockey Stick - Linear | Linear Array<br>Curved Array<br>Phased Array<br>Endocavity - curved<br>array | Linear Array<br>Curved Array<br>Phased Array<br>Endocavity - curved<br>array<br>Hockey Stick - Linear<br>Bi-plane -Linear/curved |
| Acoustic Output and | Display Features for | Display Features for | Display Features for |
| FDA Limits | High Outputs | High Outputs | High Outputs |
| Global Maximum<br>Outputs/Worst Case<br>Setting | ISPTA3: 652.9 mW/cm²<br>(4V2A)<br>TI Type: TIC (15L4)<br>TI Value: 5.8 (15L4)<br>MI: 1.78 (8EC4A)<br>IPA3@MI Max: 1029<br>W/cm² (15L4) | ISPTA.3: 652.9 mW/cm²<br>(4V2A)<br>TI Type: TIB (4V2A)<br>TI Value: 5.6 (4V2A)<br>MI: 1.74 (4V2A)<br>IPA3@MI Max: 350.3<br>W/cm² (4V2A) | ISPTA.3: 678.1 mW/cm²<br>(12HL7)<br>TI Type: TIB (4V2A)<br>TI Value: 5.6 (4V2A)<br>MI: 1.5 (4V2A)<br>IPA3@MI Max: 227.3<br>W/cm² (4V2A) |
| Modes of Operation | B-Mode Grayscale | B-Mode Grayscale | B-Mode Grayscale |
| | Imaging | Imaging | Imaging |
| | Tissue Harmonic | Tissue Harmonic | Tissue Harmonic |
| | Imaging | Imaging | Imaging |
| | M-Mode (motion) | M-Mode (motion) | M-Mode (motion) |
| | Anatomical M-Mode | Anatomical M-Mode | Anatomical M-Mode |
| | Color M-Mode | Color M-Mode | Color M-Mode |
| | Color Power Doppler | Color Power Doppler | Color Power Doppler |
| | Velocity Color Doppler | Velocity Color Doppler | Velocity Color Doppler |
| | Duplex/Triplex - Doppler | Duplex/Triplex - Doppler | Duplex/Triplex - |
| | imaging | imaging | Doppler imaging |
| | Pulsed Wave (PW) | Pulsed Wave (PW) | Pulsed Wave (PW) |
| | Doppler | Doppler | Doppler |
| | TeraVision II | TeraVision II | TeraVision I |
:
・
t
.
:
:
{4}------------------------------------------------
| | Postprocessing | Postprocessing | Postprocessing |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PW Doppler | Available for all<br>transducers<br>Triplex Mode<br>B-Mode and PW<br>Doppler<br>High PRF | Available for all<br>transducers<br>Triplex Mode<br>B-Mode and PW<br>Doppler<br>High PRF | Available for all<br>transducers<br>Triplex Mode<br>B-Mode and PW<br>Doppler<br>High PRF |
| Transducer Frequency | 2.0 - 15.0 MHz | 2.0 - 15.0 MHz | 2.0 - 10.0 MHz |
| #Transmit Channels | 128 Channels | 128 Channels | 128 Channels |
| # Receive Channels | 128 Channels | 128 Channels | 128 Channels |
| Acoustic Output<br>Measurement Standard | NEMA UD 2-2004<br>NEMA UD 3-2004 | NEMA UD 2-2004<br>NEMA UD 3-2004 | NEMA UD 2-2004<br>NEMA UD 3-2004 |
| DICOM | DICOM 3.0 Structured<br>Reporting, Worklist -<br>Image Viewer | DICOM 3.0 Structured<br>Reporting, Worklist -<br>Image Viewer | DICOM 3.0 Worklist -<br>Image Viewer |
| Product Safety<br>Certification | IEC60601-1<br>IEC60601-1-2<br>IEC60601-1-6<br>IEC60601-2-37 | IEC60601-1<br>IEC60601-1-2<br>IEC60601-1-6<br>IEC60601-2-37 | IEC60601-1<br>IEC60601-1-2<br>IEC60601-1-4<br>IEC60601-2-37 |
| EMC | IEC60601-1-2<br>CISPR11 Class B | IEC60601-1-2<br>CISPR11 Class B | IEC60601-1-2<br>CISPR11 Class B |
| System<br>Characteristics | uSmart3200T:<br>tablet computer<br>weighs 4.9 lbs (2.21 Kg)<br>11.5" backlit touch<br>screen.<br>Tablet dimensions<br>(8.82"(H) x 12.64"(W) x<br>1.25"(D)).<br>A Lithium-Polymer<br>battery (integrated into<br>the tablet) provides 2<br>hours of continuous<br>ultrasound scanning<br>Docking station (for<br>charging) that uses a<br>medical-grade power<br>supply<br>Data transferred to the<br>tablet computer over a<br>FireWire (aka IEEE<br>1394) | uSmart3200T:<br>tablet computer<br>weighs 4.9 lbs (2.21 Kg)<br>11.5" backlit touch<br>screen.<br>Tablet dimensions<br>(8.82"(H) x 12.64"(W) x<br>1.25"(D)).<br>A Lithium-Polymer<br>battery (integrated into<br>the tablet) provides 2<br>hours of continuous<br>ultrasound scanning<br>Docking station (for<br>charging) that uses a<br>medical-grade power<br>supply<br>Data transferred to the<br>tablet computer over a<br>FireWire (aka IEEE<br>1394) | Terason T3000:<br>Laptop Computer<br>weighs: 8.0 pounds (3.6<br>kg), LED backlit display<br>and keyboard, Laptop<br>dimensions (H): 2.5",<br>(W): 13.8", (D): 11.0"<br>A Lithium-Polymer<br>battery, (integrated into<br>the Laptop) provides 2<br>hours of continuous<br>ultrasound scanning.<br>Uses a medical-grade<br>power supply<br>Data transferred to the<br>laptop computer over a<br>FireWire (aka IEEE<br>1394) |
:
:
:
:
.
.
.
{5}------------------------------------------------
There are no hardware design changes to the system (ultrasound engine and tablet computer). The hardware design released at the time original submission (K131209) already had all the necessary functions and features to support the changes in software and the additional 3 transducers. The hardware revisions of the ultrasound engine and tablet computer remain unchanged.
## 8EC4A Transducer
| Key Features | Subject Device Model<br>Terason 8EC4A<br>Transducer | Comparable<br>Predicate Device<br>Terason 8EC4A<br>Transducer | Same or Different |
|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification | ITX | ITX | Same |
| 510(k) Number | K1XXXXX | K112953 | n/a |
| Indications for Use | The transducer is<br>intended to be used<br>with a conventional<br>ultrasound system<br>(Terason uSmart3200T)<br>to image trans-rectal<br>and trans-vaginal. | The transducer is<br>intended to be used<br>with a conventional<br>ultrasound system<br>(Terason t3000) to<br>image trans-rectal and<br>trans-vaginal. | Same. The proposed<br>transducer and the<br>predicate transducer<br>have the identical claim<br>of imaging similar<br>regions in the human<br>body. |
| Acoustic Array<br>Technology: | Piezoelectric elements | Piezoelectric elements | Same.<br>Effectiveness: Both<br>arrays allow focused<br> |
| Acoustic Array Style: | Endocavity Curved<br>Array | Endocavity Curved<br>Array | transmission and<br>reception of ultrasound<br>energy to enhance<br>image quality within the<br>region of interest. |
| Acoustic Array<br>Characteristics:<br>Element count...<br>Center frequency...<br>Element size (pitch x<br>elevation)...<br>Elevation focus... | 128<br>6.5<br>0.205mm X 5.0mm<br>30mm | 128<br>6.5<br>0.205mm X 5.0mm<br>30mm | Same: element count.<br>Acoustic characteristics<br>have met safety<br>guidelines of IEC60601-<br>2-37. Safety and<br>effectiveness<br>unchanged from<br>predicate. |
| Acoustic Array | The transducer imaging<br>performance has been<br>evaluated in an acoustic<br>tank. | The transducer<br>performance has been<br>evaluated in the<br>previous 510(k) filing<br>(K112953). | Same; The 8EC4A uses<br>a same acoustic array<br>as the predicate device<br>and therefore has same<br>acoustic characteristics.<br>To ensure proper safety<br>guidelines are met,<br>acoustic testing was<br>performed per the<br>IEC60601-2-37 |
{6}------------------------------------------------
| Patient Contact<br>Material | ABS Silicone 747 | ABS Silicone 747 | Same |
|-----------------------------|------------------|------------------|------|
| | | | |
## Discussion:
The 8EC4A uses a same acoustic array as the predicate device and therefore has same acoustic characteristics. To ensure proper safety guidelines are met, acoustic testing was performed per the IEC60601-2-37 standard.
The 8EC4A consists of same patient contact material as the predicate device. To ensure proper safety guidelines are met, biocompatibility tests were run on the patient contact materials
### Transducer 16HL7
| Key Features | Subject Device Model<br>Terason 16HL7<br>Transducer | Comparable Predicate<br>Device<br>Terason 16HL7<br>Transducer | Same or Different |
|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification | ITX | ITX | Same |
| 510(k) Number | K1XXXXX | K110020 | n/a |
| Indications for Use | The transducer is<br>intended to be used with<br>a conventional<br>ultrasound system<br>(Terason uSmart3200T)<br>to image intra-operative,<br>small parts, muskulo-<br>skel and peripheral<br>vascular regions. | The transducer is<br>intended to be used<br>with a conventional<br>ultrasound system<br>(Terason t3200) to<br>image intra-operative,<br>small parts, muskulo-<br>skel and peripheral<br>vascular regions. | Same. The proposed<br>transducer and the<br>predicate transducer<br>have the identical<br>claim of imaging<br>similar regions in the<br>human body. |
| Acoustic Array<br>Technology | Piezoelectric elements | Piezoelectric elements | Same |
| Transducer Style | Hockey-Stick Linear<br>Array | Hockey-Stick Linear<br>Array | Same. |
| Acoustic Array<br>Characteristics:<br>Element count...<br>Center frequency...<br>Element size (pitch x<br>elevation)...<br>Elevation focus... | 128<br>9.2 MHz<br>0.2mm X 3.5mm<br>12mm | 128<br>9.2 MHz<br>0.2mm X 3.5mm<br>12mm | Same. |
| Acoustic Output and<br>Device Settings | The transducer imaging<br>performance has been<br>evaluated in an acoustic<br>tank. | The transducer<br>performance has been<br>evaluated in the<br>previous 510(k) filing<br>(K110020) | Same. |
{7}------------------------------------------------
| Patient Contact<br>Material | Silicone PEI | Silicone PEI | Same. |
|-----------------------------|--------------|--------------|-------|
### Discussion:
There are no differences between this device and the predicate device used in this comparison. The transducer has been added to the uSmart3200T Ultrasound system.
Based on the identical indications for use, technological characteristics and performance testing, Teratech Corporation, Inc. believes the Terason 16HL7 transducer is substantially equivalent to the Terason 16HL7 transducer (K110020) with respect to safety and effectiveness.
### 15L4 Transducer
| Key Features | Subject Device Model<br>Terason 15L4<br>Transducer | Comparable<br>Predicate Device<br>Terason 15L4<br>Transducer | Same or Different |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K140524 | K110020 | n/a |
| Classification | ITX | ITX | Same |
| Indications for Use | The transducer is<br>intended to be used<br>with a conventional<br>ultrasound system<br>(Terason<br>uSmart3200T) to image<br>abdomen, small parts,<br>musculo-skel and<br>peripheral vascular<br>regions. | The transducer is<br>intended to be used<br>with a conventional<br>ultrasound system<br>(Terason T3200) to<br>image abdomen, small<br>parts, musculo-skel<br>and peripheral<br>vascular regions. | Same: The proposed<br>transducer and the<br>predicate transducer<br>have the identical claim<br>of imaging similar<br>regions in the human<br>body. |
| Acoustic Array<br>Technology: | Piezoelectric elements | Piezoelectric elements | Same: Safety<br>requirements of<br>IEC60601 are equally<br>met.<br>Effectiveness: Both<br>arrays allow focused<br>transmission and<br>reception of ultrasound<br>energy to enhance<br>image quality within the<br>region of interest. |
| Acoustic Array Style: | Linear Array | Linear Array | |
{8}------------------------------------------------
| Acoustic Array<br>Characteristics:<br>Element count...<br>Center frequency...<br>Element size (pitch x<br>elevation)...<br>Elevation focus... | 128 elements<br>7.5 MHz<br>0.3mm X 4.0mm<br>20mm | 128 elements<br>7.5 MHz<br>0.3mm X 4.0mm<br>20mm | Same: element count.<br>Acoustic characteristics<br>have met safety<br>guidelines of IEC60601-<br>2-37. |
|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Acoustic Output and<br>Device Settings | The transducer imaging<br>performance has been<br>evaluated in an<br>acoustic tank. | The transducer<br>performance has been<br>evaluated in the<br>previous 510(k) filing<br>(K110020). | Same; acoustic output<br>safety guidelines. Safety<br>is not compromised.<br>Effectiveness equal. |
| Patient Contact Material | The transducer uses<br>Silicone SIM R1001 as<br>a contact material. | The transducer uses<br>Silicone SIM R1001 as<br>a contact material. | Same. The 15L4<br>transducer consists of<br>same patient contact<br>materials as the<br>predicate device. The<br>safety of each device<br>with respect to<br>biocompatibility is<br>equivalent. |
### Discussion:
The 15L4 uses a same acoustic array than the predicate device and therefore has same acoustic characteristics. To ensure proper safety guidelines are met, acoustic testing was performed per the IEC60601-2-37 standard.
The 15L4 transducer consists of same patient contact materials as the predicate device.
Based on the identical indications for use, technological characteristics and performance testing, Teratech Corporation, Inc. believes the Terason 15L4 transducer is substantially equivalent to the Terason 15L4 transducer (K110020) with respect to safety and effectiveness.
### Transducers previously cleared in submission K131209
### 12L5A Transducer
| Key Features | Subject Device Model | Comparable Predicate<br>Device | Same or Different |
|--------------|----------------------|--------------------------------|-------------------|
| | Terason 12L5A | Terason 12L5A | |
{9}------------------------------------------------
| | Transducer | Transducer | |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Classification | ITX | ITX | Same |
| 510(k) Number | KXXXXXX | K131209 | n/a |
| Indications for<br>Use | The transducer is intended<br>to be used with a<br>conventional ultrasound<br>system (Terason<br>uSmart3200T) to image<br>Fetal, Abdominal,<br>Pediatric, Small Organ,<br>Cephalic, Musculo-skel,<br>Cardiac and Peripheral<br>vessel. | The transducer is intended<br>to be used with a<br>conventional ultrasound<br>system (Terason<br>uSmart3200T) to image<br>Fetal, Abdominal,<br>Pediatric, Small Organ,<br>Cephalic, Musculo-skel,<br>Cardiac, and Peripheral<br>vessel. | Same. The proposed<br>transducer and the<br>predicate transducer have<br>the identical claim of<br>imaging similar regions in<br>the human body. |
| Acoustic Array<br>Technology: | Piezoelectric elements | Piezoelectric elements | Same. |
| Acoustic Array<br>Style: | Linear Array | Linear Array | Regarding Safety: Both<br>arrays allow focused<br>transmission and<br>reception of ultrasound<br>energy to enhance image<br>quality within the region of<br>interest. |
| Acoustic Array<br>Characteristics: | | | Same in elevation<br>Safety and effectiveness<br>unchanged from predicate |
| Element count... | 128 | 128 | |
| Center<br>frequency... | 7.5 | 7.5 | |
| Element size<br>(pitch x<br>elevation)... | 0.3mm X 4.0mm | 0.3mm X 4.0mm | |
| Elevation<br>focus... | 19mm | 19mm | |
| Acoustic Array | The transducer imaging<br>performance has been<br>evaluated in an acoustic<br>tank. | The transducer<br>performance has been<br>evaluated in the previous<br>510(k) filing (K131209). | Same. As the predicate<br>device and therefore has<br>same acoustic<br>characteristics. To ensure<br>proper safety guidelines<br>are met, acoustic testing |
| | | | was performed per the<br>IEC60601-2-37 standard. |
| Patient Contact<br>Material | Silicone ABS | Silicone ABS | Same. The 12L5A<br>transducer consists of<br>same patient contact<br>materials as the predicate<br>device. |
.
.
{10}------------------------------------------------
## 5C2A Transducers
| Key Features | Subject Device Model | Comparable<br>Predicate Device | Same or Different |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification | ITX | ITX | Same…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.