The Teratech Corporation Terason™ uSmart3200T is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.
Device Story
Portable ultrasound system; consists of tablet computer with 11.5" touch screen and proprietary ultrasound engine; engine controls acoustic output of transducer and processes return echoes in real time; data transferred via FireWire (IEEE 1394) to tablet for image generation/display; operated by physicians in clinical settings; provides real-time ultrasound imaging and fluid flow analysis; aids in clinical decision-making through visualization of internal structures and blood flow; benefits include portability and improved workflow efficiency via touch-screen interface; supports previously cleared transducers (12L5A, 5C2A, 4V2A).
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (safety), IEC 62366/60601-1-6 (usability), IEC 60601-1-2 (EMC), and IEC 60601-2-37 (ultrasonic safety) verified. Biocompatibility testing per ISO 10993-5 and 10993-10 performed for all transducers. Acoustic output measured per NEMA UD 3.
Technological Characteristics
Tablet-based ultrasound system; Windows 7 OS; custom beamformer chip for acoustic array control and signal processing; FireWire (IEEE 1394) connectivity; 11.5" backlit touch screen; Lithium-Polymer battery; supports 12L5A, 5C2A, 4V2A transducers; biocompatible patient contact materials per ISO 10993.
Indications for Use
Indicated for qualified physicians performing diagnostic ultrasound imaging or fluid flow analysis in fetal, abdominal, pediatric, small organ (thyroid, breast, testes), neonatal/adult cephalic, musculoskeletal (conventional/superficial), cardiac (adult/pediatric), and peripheral vascular applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
హ
- ---
:::
::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::::
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
ﺳﮯ m�
. I
MAY 2 4 2013
#### 510(k) Summary Teratech Corporation Terason™ uSmart3200T Ultrasound System
1. Sponsor:
Teratech Corporation 77-79 Terrace Hall Ave. Burlington, MA 01803
Contact Person: Ben Chiampa, Quality Assurance and Requlatory Affairs Telephone: 781-270-4143 Date Prepared: March 21, 2013
#### 2. Device Name
Proprietary Name: Terason™ uSmart3200T Uitrasound System Common / Usual Name: Diagnostic Ultrasound System Classification Name: Diagnostic Ultrasound Transducer
(21 CFR 892.1570. 90-ITX) Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560, 90-IYO) Diagnostic Ultrasonic Transducer (21 CFR 892.1570, 90-ITX)
#### 3. Predicate Devices
Terason™ t3000 Ultrasound System (K112953) and Terason™ t3200 Ultrasound System (K110020)
#### 4. Device Description
The Terason™ uSmarl3200T ultrasound system is equivalent to the previously cleared version of the t3200 and t3000 Ultrasound Systems described in the following 510(k) submissions (K110020 and K112953). This system contains a proprietary ultrasound engine for controlling the acoustic output of the transducer and processing the return echoes in real time. These data are then transferred to the tablet (previously laptop computer) over a FireWire (aka IEEE 1394) connection for further processing and generation and display of the ultrasound image.
The Terason™ uSmart3200T ultrasound tablet weighs 4.9 pounds (2.21 Kg) and has an 11,5" backlit touch screen. The tablet dimensions (8,82"(H) x 12,64"(W) x 1,25"(D)) are chosen to allow portability. A Lithium-Polymer battery (integrated into the tablet) provides 2 hours of continuous ultrasound scanning. The tablet includes a docking station (for charging) that uses a medical-grade power supply. The ultrasound transducer connector is identical to that used in the Terason™ predicate device, the t3200. Optional accessories include a cart and printer.
{1}------------------------------------------------
### 5. Intended Use
The Teratech Corporation Terason™ uSmart3200T is a general purpose Ultrasound System intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Musculo-skeletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.
### 6. Technology Characteristics
The design and construction of the Terason™ uSmart3200T is similar to the Terason™ t3200 Ultrasound system. These systems utilize a tablet (or laptop) computer running Windows 7 to execute the ultrasound application and a custom designed engine for control of the acoustic array and processing of the return echoes. For the uSmart3200T, the engine is housed in a compartment that is attached to the backside of the tablet computer.
The similarity and difference between the Terason™ uSmart3200T and the Terason 13200 Ultrasound System (the predicate device) include the following:
- The engines are the same with no modification in the custom beamformer chip . (as compared to earlier versions of Terason ultrasound systems) that provides for improved filtering of the return signal for wider bandwidth and better resolution across the entire image field.
- The ultrasound application software has been modified to improve the user . workflow and ease of use commensurate with a tablet application. The screen layout has been modified and the user controls have been changed for finger touch control to improve the efficiency for the targeted exam types.
Transducers: The Terason uSmart3200T and the BenQ UP200 will support 3 transducers. These transducers have been previously cleared.
- . 12L5A: Cleared in 510k submission K112953 (February 3, 2012)
- 5C2A: Cleared in 510k submission K112953 (February 3, 2012) .
- 4V2A: Cleared in 510k submission K112953 (February 3, 2012). .
The following provides additional details of the three transducers, presented in this submission, that were previously cleared.
- . 12L5A: equivalent indications for use, frequency settings, shape of transducer head and needle guide/software brackets. Same manufacturer, same acoustic array and patient contact materials.
- 5C2A: equivalent indications for use, frequency settings, and needle guide bracket / . software. Same manufacturer, same shape, same acoustic array and palient contact materials.
- 4V2A: equivalent indications for use, frequency settings, and needle guide bracket . / software. Same manufacturer, same shape, same acoustic array and palient contact materials.
{2}------------------------------------------------
### B1. Non Clinical Tests
The Terason™ uSmart3200T system has been tested for compliance to the following standards (with the corresponding report referenced for each standard).
- · IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Safety.
- o Intertek Test Record Number 100825075BOX-001.
- IEC 62366, Medical Devices: Application of usability engineering to medical ● devices.
- o Intertek Project: 100825075BOX-004.
- IEC60601-1-6, Medical Electrical Equipment Part 1-6: General . requirements for safety- Collateral standard: Usability o Intertek Project: 100825075BOX-003.
- . IEC 60601-1-2:2007, Medical Electrical Equipment - Part 1-2; General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and tests
- o IEC60601-1-2 Intertek Test Record Number, 100933162BOX-017.
- . IEC 60601-2-37 / EN60601-2-37 Medical Electrical Equipment Part 2: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment.
- o Transducer Model 5C2A: Intertek Report Number IEC60601-2-37 uSmart3200T 5C2A: 100825075BOX-006
- o Transducer Model 12L5A: Intertek Report Number IEC60601-2-37 uSmart3200T 12L5A: 100825075BOX-007
- o Transducer Model 4V2A: Intertek Report Number IEC60601-2-37 uSmart3200T 4V2A: 100825075BOX-005.
- . NEMA UD 3 Acoustic Output Display Terason uSmart3200T Ultrasound System User Guide (16-3301).
- Biocompatibility Tests, ISO 10993 Part 5 and Part 10 .
- o Biocompatibility reports for all transducers.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 18, 2013
TeraTech Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street NW BUFFALO MN 55313
Re: K131209
Trade/Device Name: Terason® uSmart3200T and BenQ UP200 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, and ITX Dated: May 15. 2013 Received: May 16, 2013
Dear Mr. Job:
This letter corrects our substantially equivalent letter of May 24, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the Terason " uSmart3200T and BenQ UP200 Ultrasound System, as described in your premarket notification:
Transducer Model Number
| 12L5A |
|-------|
| 5C2A |
| 4V2A |
{4}------------------------------------------------
Page 2 - Mr. Mark Job
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
If you have any questions regarding the content of this letter. please contact Robert Ochs. Ph.D. at (301) 796-6661.
Sincerely yours.
FDA
for
Janine M. Morris Director, Division of Radiological Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K131209
## Device Name: _Terason uSmart3200T and BenQ UP200 Ultrasound System
Indications for Use: Diagnostic ultrasound imaging system or fluid flow analysis of the human body as follows:
The Teratech Corporation Terason™ uSmart3200T (also known as the BenQ UP200) is a The Teration ourporation Teracem intended for use by a qualified physician for evaluation by yenoral porpood of acconnelysis of the human body. Specific clinical applications and exam types include: Fetal, Abdominal, Pediatrics, Small Organ (Thyroid, Breast, Testes); exam types inolade. Total Aboundai, Reletal (Conventional and Superficial); Cardiac (Adult & Pediatric); Peripheral Vascular.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k)________________________________________________________________________________________________________________________________________________________________________
Page 1 of 5
{6}------------------------------------------------
### 510(k) Number (if known): K131209
| Indications For Use: Diagnostic ultrasound imaging system or fluid flow analysis of the human body as follows:<br>Clinical Application | | Mode of Operation | | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------------|---|-----|-----|----------------|-----------------|--------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | Otherc |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetalh | N | N | N | N | N | N | N |
| | Abdominalc: | N | N | N | N | N | N | N |
| | Intra-operative (Spec.)d,e | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatricc: | N | N | N | N | N | N | N |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | N | N | N | N | N | N | N |
| | Neonatal Cephalicc: | N | N | N | N | N | N | N |
| | Adult Cephalicc: | N | N | N | N | N | N | N |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginalg: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)c: | N | N | N | N | N | N | N |
| | Musculo-skel. (Superfic)c: | N | N | N | N | N | N | N |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | N | N | N | N |
| | Cardiac Pediatric | N | N | N | N | N | N | N |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vesselc: | N | N | N | N | N | N | N |
| | Other (Specify) | | | | | | | |
Device Name: _Terason uSmart3200T and BenQ UP200 Ultrasound System
ound imaging system or fluid flow analysis of the human body as follows:
Vessel N= new indication; P= previously cleared by FDA; E= added under Appendix E
No new Indication; P= previously cleared by I DA, L= addod Sharin =
ª Includes Color Doppler (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler.
B+M; B+PWD; B+CD; B+DPD; B+PD.
C Harmonic Imaging (HI)
Includes ultrasound guidance for placement of needles, catheters.
· Abdominal, thoracic and peripheral vessel.
· Abdominal, thoracic and penpheral vosso.
I Includes ultrasound guidance for placement of needles, cathers of follolo dovelopment
Includes ultrasound guidance for placement of nocerso, of fortility monitoring of form
Includes guidance of amniocentesis, infertility monitoring of follicle development.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
r a come
{7}------------------------------------------------
### 510(k) Number (if known): K131209
## Device Name: _Terason uSmart3200T and BenQ UP200 - 5C2A Transducer
Indications For Use: Diagnostic ultrasound imaging system or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | Otherc |
|-----------------------------|--------------------------------------------------|-------------------|----|-----|-----|----------------|-----------------|--------|
| General<br>(Track Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetalh | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Abdominali: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Intra-operative (Spec.)d,e | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatricd: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | | | | | | | |
| | Neonatal Cephalicd: | | | | | | | |
| | Adult Cephalicd: | | | | | | | |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginalg: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)d: | N | N | N | | N | N | N |
| | Musculo-skel. (Superfic)d: | N | N | N | | N | N | N |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | | N | N | N |
| | Cardiac Pediatric | N | N | N | | N | N | N |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesseld: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Ne new noneation, i = provided on on one Doppler (DPD), and (non-directional) Power Doppier.
b +M; B+PWD; B+CD; B+DPD; B+PD.
° Harmonic Imaging (HI)
9 Includes ultrasound guidance for placement of needles, catheters.
9 Abdominal, thoracic and peripheral vessel.
Includes ultrasound guidance for placement of needles, catheters, cryosurgery, and brachytherapy
Includes ultrasound guidance of transvaginal biopsy, infertility monitoring of follicle development.
Includes guidance of amniocentesis, infertility monitoring of follicle development.
Additional Comments: P1: uses previously cleared under K112953
| Prescription Use | x |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)_ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
n a come
{8}------------------------------------------------
### 510(k) Number (if known): K131209
# Device Name: _ Terason uSmart3200T and BenQ UP200 - 12L5A Transducer
Indications For Use: Diagnostic ultrasound imaging system or fluid ilow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | Otherc |
|-----------------------------|--------------------------------------------------|-------------------|----|-----|-----|----------------|-----------------|--------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetalh | | | | | | | |
| | Abdominald: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Intra-operative (Spec.)d,e | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatricc: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Neonatal Cephalicc: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Adult Cephalicd: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginalg: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)d: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Musculo-skel. (Superfic)d: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vesseld: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
1 Includes Color Doppler (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler.
b B+M; B+PWD; B+CD; B+DPD; B+PD.
° Harmonic Imaging (HI)
Includes ultrasound guidance for placement of needles, catheters.
· Abdominal, thoracic and peripheral vessel.
'Includes ultrasound guidance for placement of needles, catheters, cryosurgery, and brachytherapy
9 Includes ultrasound guidance of transvaginal biopsy, infertility monitoring of follicle development.
" Includes guidance of amniocentesis, infertility monitoring of follicle development.
Additional Comments: P : uses previously cleared under K112953
| Prescription Use | x |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)_
{9}------------------------------------------------
### 510(k) Number (if known): K131209
## Device Name: __ Terason uSmart3200T and BenQ UP200 - 4V2A Transducer
Indications For Use: Diagnostic ultrasound imaging system or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | Otherc |
|-----------------------------|--------------------------------------------------|-------------------|----|-----|-----|----------------|-----------------|--------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppa | Comb.<br>Modesb | |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetalb | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Abdominald: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Intra-operative (Spec.)d,e | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatricc: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Small Organ (Thyroid,<br>Breast, Testes, etc.)d: | | | | | | | |
| | Neonatal Cephalicc: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Adult Cephalicc: | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Trans-rectalf: | | | | | | | |
| | Trans-vaginalg: | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.)d: | | | | | | | |
| | Musculo-skel. (Superfic)d: | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Cardiac Pediatric | P1 | P1 | P1 | | P1 | P1 | P1 |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vesseld: | | | | | | | |
| Vessel | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
No now nonomic (CD), Directional Power Doppler (DPD), and (non-directional) Power Doppler.
b B+M; B+PWD; B+CD; B+DPD; B+PD.
° Harmonic Imaging (HI)
Includes ultrasound guidance for placement of needles, catheters.
9 Abdominal, thoracic and peripheral vessel.
1 Includes ultrasound guidance for placement of needles, catheters, cryosurgery, and brachytherapy
® Includes ultrasound guidance of transvaginal biopsy, infertility monitoring of follicle development.
AND/OR
" Includes guidance of amniocentesis, infertility monitoring of follicle development.
Additional Comments: P : uses previously cleared under K1 12953
Prescription Use __x (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
mint
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)_K131209
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.