EMBOL-X INTRODUCER SHEATH

K140398 · Edwards Lifesciences, LLC · DYB · Mar 19, 2014 · Cardiovascular

Device Facts

Record IDK140398
Device NameEMBOL-X INTRODUCER SHEATH
ApplicantEdwards Lifesciences, LLC
Product CodeDYB · Cardiovascular
Decision DateMar 19, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The EMBOL-X Introducer Sheath is indicated for use in procedures requiring the introduction of EMBOL-X Intra-Aortic Filters.

Device Story

17 Fr, sterile, single-use catheter introducer; facilitates introduction of EMBOL-X Intra-Aortic Filters. Features single lumen housing, removable snap-lock obturator, internal valve to prevent backbleeding, and flexible suture flange for vessel attachment. Includes orientation markings and venting system (vent grooves, holes, central lumen) to assist air removal. Used by clinicians in surgical procedures. Device provides access path for filter deployment; ensures secure vessel interface and air management during insertion.

Clinical Evidence

Bench testing only. Testing included leak testing, tensile testing of bond joints, obturator performance (vent time, latch/insertion/removal forces), filter compatibility (deployment/retraction forces), and suture flange flex testing. Design validated via surgeon evaluation and bench-top study.

Technological Characteristics

17 Fr catheter introducer; constructed of flexible and non-flexible polymeric materials. Features internal valve, snap-lock obturator, and integrated venting system. Single-use, sterile, non-pyrogenic. No electronic or software components.

Indications for Use

Indicated for use in procedures requiring the introduction of EMBOL-X Intra-Aortic Filters.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Edwards Lifesciences LLC Special 510(k) Premarket Notification EMBOL-X Introducer Sheath . K140398, 510(k) Summary page 1 of 2 | 510(k) Summary | | | |----------------------|--------------------------------------------------------------------------------------------------------------|--------------| | Submitter: | Edwards Lifesciences LLC | MAR 1 9 2014 | | Contact Person: | Luke Meidell, Regulatory Affairs Associate III<br>12050 Lone Peak Pkwy<br>Draper, UT 84020<br>(801) 565-6212 | | | Date Prepared: | February 17, 2014 | | | Trade Name: | EMBOL-X Introducer Sheath | | | Classification Name: | Catheter Introducer<br>21 CFR Part 870.1340, Product Code DYB, Class II | | | Predicate Device: | K123714, EMBOL-X Introducer Sheath | | ## Device Description: Edwards Lifesciences' EMBOL-X Introducer Sheath is a 17 Fr, sterile, non-pyrogenic, single-use introducer made of flexible and non-flexible polymeric materials. It consists of a single lumen housing and a removable, snap-lock obturator. The housing has a flexible suture flange for attachment to the vessel and orientation markings on the shaft and suture flange to assist in the placement of the EMBOL-X Introducer Sheath. Inside the housing is an internal valve to prevent backbleeding. The housing also has suture loops on the proximal end of the Introducer housing, and vent grooves, vent holes, and a central lumen in the obturator that leads to the vent plug to assist in venting air from the Introducer. The EMBOL-X Introducer Sheath is intended for use in procedures requiring the introduction of EMBOL-X Intra-Aortic Filters. #### Indications For Use: The Indications for Use for the EMBOL-X Introducer Sheath is as follows: The EMBOL-X Introducer Sheath is indicated for use in procedures requiring the introduction of EMBOL-X Intra-Aortic Filters. {1}------------------------------------------------ ## Comparative Analysis: The subject device has the same intended use and technological characteristics (i.e., design, material, chemical composition) as the predicate device. The subject EMBOL-X Introducer Sheath is comparable to the predicate device in fundamental scientific technology, materials, principles of operation, and functional performance evaluations. No new issues of safety or effectiveness have been raised as a result of the specification change or shelf life increase. ## Functional/Safety Testing: The following functional and accelerated aging bench testing was performed to support the specification change and the shelf life increase. All data met acceptance criteria, demonstrating that the device is as safe and effective as the predicate device. - . Leak Testing - Leak without device inserted, leak with obturator inserted, leak with filter inserted. - Tensile Testing Introducer bond joints, obturator vent plug bond joint. . - . Additional Obturator Testing - Vent time, latch compression force, insertion force, removal and retention force. - . Filter Compatibility Testing - Insertion force, deployment capability, deployment force, retraction force, removal and retention force. - Additional Introducer Testing Suture flange flex . - . Design Validation - Design was validated through surgeon evaluation and a bench top study . #### Conclusion: Based on the Indications for Use, comparative analysis, and functional/safety testing results, the EMBOL-X Introducer Sheath is substantially equivalent to the predicate device and demonstrates no new issues of safety and effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-0002 March 19, 2014 Edwards Lifesciences c/o Luke Meidell Regulatory Affairs Associate III 12050 Lone Peak Pkwy Draper, Utah 84020 K140398 Re: > Trade/Device Name: Embol-X Introducer Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: February 17, 2014 Received: February 18, 2014 Dear Mr. Meidell. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {3}------------------------------------------------ Page 2 - Luke Meidell the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, MA.S.L.Lieman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K140398, Indications for Use Page 1 of 1 # Statement of Indications for Use # Indications for Use 510(k) Number (if known): K140398 Device Name: EMBOL-X Introducer Sheath The EMBOL-X Introducer Sheath is indicated for use in procedures requiring the introduction of EMBOL-X Intra-Aortic Filters Prescription Use X (Per 21 CFR 801.109) OR Over-The-Counter Use_ 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office Of Device Evaluation (ODE) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - M. L. Lieberman Page B-2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...