ENDOSCOPE POSITION MARKING PROBE,ENDOSCOPE POSITION DETECTING UNIT

K134026 · Olympus Medical Systems Corp. · PGU · Mar 7, 2014 · Gastroenterology, Urology

Device Facts

Record IDK134026
Device NameENDOSCOPE POSITION MARKING PROBE,ENDOSCOPE POSITION DETECTING UNIT
ApplicantOlympus Medical Systems Corp.
Product CodePGU · Gastroenterology, Urology
Decision DateMar 7, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

This instrument has been designed to be used with Olympus endoscope systems for the detection and display of the shape of an endoscope once inserted into the patient. This instrument has been designed to be used with an Olympus endoscope and the endoscope position detecting unit for detection and display of the location and/or shape of the inserted endoscope.

Device Story

System comprises UPD-3 Endoscope Position Detecting Unit and MAJ-1878 Endoscope Position Marking Probe. Device detects magnetic fields generated by magnetic coils integrated into endoscopes or the MAJ-1878 probe inserted into the endoscope's instrument channel. System computes position coordinates of coils via arithmetic processing; displays endoscope shape/location on monitor. Used in clinical settings during colonoscopy and enteroscopy procedures. Provides real-time visualization of endoscope configuration to assist clinicians in navigation and positioning. Benefits include improved procedural guidance and potential reduction in patient discomfort during complex endoscopic maneuvers.

Clinical Evidence

No clinical data. Evidence consists of bench testing using a small intestine model, risk analysis per ISO 14971:2007, software validation per FDA guidance, and biocompatibility testing per G95-1.

Technological Characteristics

Magnetic field detection system. Components: UPD-3 unit and MAJ-1878 probe with integrated magnetic coils. Standards applied: IEC 60601-1, IEC 60601-1-1, IEC 60601-2-18, IEC 60601-1-2, ISO 14971, IEEE Std C95.1. Software classified as Minor Level of Concern.

Indications for Use

Indicated for patients requiring colonoscopy and enteroscopy. Contraindicated in patients with critical active implantable devices (e.g., pacemakers, implantable cardioverter defibrillators) and pregnant women.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ENDOSCOPE POSITION MARKING PROBE ENDOSCOPE POSITION DETECTING UNIT Olympus Medical Systems, Corp. Traditional 510(k) Notification December 27, 2013 MAR - 7 2014 K134026 Page 1 of 5 # 510(k) SUMMARY UPD-3 ENDOSCOPE POSITION DETECTING UNIT MAJ-1878 ENDOSCOPE POSITION MARKING PROBE ### 1. General Information - OLYMPUS MEDICAL SYSTEMS CORP. . Applicant: 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047 - B Official Daphney Germain-Kolawole Regulatory Affairs Project Manager Correspondent: Olympus Corporation of the Americas 3500 Corporate Parkway . PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-5691 FAX: 484-896-7128 Email: daphney.germain-kolawole@olympus.com - Manufacturer: SHIRAKAWA OLYMPUS CO., LTD. 1 3-1.Aza Ookamivama Ooaza Odakura, Nishigo-Mura Nishirakawa-Gun Fukushima, JAPAN 961-8061 Registration Number: 3002808148 AIZU OLYMPUS CO., LTD. 500 Aza Muranishi Ooaza-lidera Monden-cho Aizuwakamatsu-Shi Fukushima,JAPAN 965-8520 Registration Number: 9610595 - II. Device Identification <MAJ-1878> - 1 Device Trade Name: MAJ-1878 - . ENDOSCOPE POSITION MARKING PROBE Common Name: - 미 Regulation Number: 876.1500 Section 5 510(k) Summary Page 1 of 5 {1}------------------------------------------------ Traditional 510(k) Notification December 27, 2013 ### ENDOSCOPE POSITION MARKING PROBE ENDOSCOPE POSITION DETECTING UNIT Olympus Medical Systems, Corp. ■ Regulation Name: Endoscope and Accessories - 이트 Regulatory Class: ll - Classification Panel: Gastroenterology and urology - 트 Product Code: PGU FDA FDF - II. Device Identification <UPD-3> - 파 UPD-3 Device Trade Name: - 트 Common Name: ENDOSCOPE POSITION DETECTING UNIT - E Regulation Number: 876.1500 - 해 Regulation Name: Endoscope and Accessories - 8 Regulatory Class: ll - 를 Classification Panel: Gastroenterology and urology - 를 Product Code: PGU FDA FDF ### Predicate Device Information III. | MAJ-1878 | | | |---------------------------------------------------|----------------------------------|------------| | Model name | Applicant | 510(k) No. | | Probe XB01-657-6 | Olympus Optical Co., Ltd. | K002749 | | UPD-3 | | | | Model name | Applicant | 510(k) No. | | ENDOSCOPE POSITION<br>DETECTING UNIT<br>UPD-Y0003 | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K103312 | {2}------------------------------------------------ ### Device Description s ### < MAJ-1878&UPD-3 > The Endoscope Position Detecting Unit (UPD-3) has been designed to be used with Olympus endoscope systems and Endoscope Position Marking Probe (MAJ-1878) for the detection and display of the shape of an endoscope once inserted into the patient. The UPD-3 detects the magnetic field generated by magnetic coils built directly into the endoscope or the MAJ-1878 inserted into the endoscope's instrument channel. The position coordinate of each coil is computed by carrying out arithmetic processing and is displayed on the UPD-3 monitor. The system is used in patients who require colonoscopy and enteroscopy except patients with a critical active implantable device such as a pacemaker and an implantable cardioverter defibrillator or pregnant women. No modifications were made to the technology, performance, and specifications for the subject UPD-3 in comparison to the predicate device (K103312). ## V. Indications for Use ### < MAJ-1878 > This instrument has been designed to be used with an Olympus endoscope and the endoscope position detecting unit for detection and display of the location and/or shape of the inserted endoscope. ### <UPD-3> This instrument has been designed to be used with Olympus endoscope systems for the detection and display of the shape of an endoscope once inserted into the patient. ### <Patient population> Patients who require colonoscopy and enteroscopy except as follows: Patients with a critical active implantable device such as a pacemaker and an implantable cardioverter defibrillator Pregnant women ### VI. Comparison of Technological Characteristics ### < MAJ-1878 > The addition of the device to be used in the small intestine does not affect the safety and effectiveness. The following design specifications have been modified and do not affect the safety and effectiveness of the device. ### -Materials -Dimensions -Applicable method of reprocessing Section 5 510(k) Summary Page 3 of 5 {3}------------------------------------------------ ENDOSCOPE POSITION MARKING PROBE ENDOSCOPE POSITION DETECTING UNIT Olympus Medical Systems, Corp. Traditional 510(k) Notification December 27, 2013 ### <UPD-3> There is no technological difference between the subject device and the predicate device: however, the modification on the subject device's software has been made to include additional minor software features, and it does not affect the safety and effectiveness. ### VII. Summary of non-clinical testing The differences of technological characteristics between the predicate device and the subject device are confirmed that they are substantially equivalent through the following tests and standards. · Bench test of small intestine model. · Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment. The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The device software is considered a Minor Level of Concern. Biocompatibility testing is performed in accordance with the FDA Guidance," Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices, Blue Book Memo, G95-1". The validation test on the reprocessing has been perfomed in accordance with the FDA quidance "Labelling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance - April 1996". The FDA quidance "Draft Guidance for Industry and FDA Staff -Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling – May 2011" was also taken into consideration. The following standards have been applied to the MAJ-1878&UPD-3: - · IEC 60601-1 - · IEC 60601-1-1 - · IEC 60601-2-18 - · IEC 60601-1-2 - · ISO 14971 - · IEEE Std C95.1(only UPD-3) {4}------------------------------------------------ ENDOSCOPE POSITION MARKING PROBE ENDOSCOPE POSITION DETECTING UNIT Olympus Medical Systems, Corp. Traditional 510(k) Notification December 27, 2013 ### VIII. Conclusion When compared to the predicate device, the MAJ-1878 & UPD-3 do not incorporate any significant changes in intended use, method of operation, or design that could affect the safety or effectiveness of the device. > Section 5 510(k) Summary Page 5 of 5 {5}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wings, representing health, services, and people. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 7, 2014 OLYMPUS MEDICAL SYSTEMS CORP. % Daphney Germain-Kolawole Regulatory Affairs Project Manager Olympus Corporation of the Americas, Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610 Re: K134026 > Trade/Device Name: ENDOSCOPE POSITION MARKING PROBE and ENDOSCOPE POSITION DETECTING UNIT Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: PGU, FDA, FDF Dated: December 27, 2013 Received: December 30, 2013 Dear Daphney Germain-Kolawole, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Image /page/6/Picture/6 description: The image shows the name "Benjamin Fisher-S" in a bold, sans-serif font. The letters "FDA" are stacked on top of each other in the middle of the name. The letters "FDA" are outlined in black and white. Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) ### K134026 ### Device Name ENDOSCOPE POSITION DETECTING UNIT ### Indications for Use (Describe) ### <UPD-3> This instrument has been designed to be used with Olympus endoscope systems for the detection and display of the shape of an endoscope once inserted into the patient. ### <Patient population> Patients who require colonoscopy and enteroscopy except as follows: Patients with a critical active implantable device such as a pacemaker and an implantable cardioverter defibrillator Pregnant women Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) 43 1 20 11 12 Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. 3 THE FOR FOR FOR FOR FOR FOR THE CONTRACT CART CONTRACTOR PRODUCTION لقارير الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموق . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . * Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) FORM FDA 3881 (9/13) PSC Publishing Services (JQ) | 443-6740 Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. . {8}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. ### 510(k) Number (if known) ### K134026 ### Device Name ENDOSCOPE POSITION MARKING PROBE ### Indications for Use (Describe) This instrument has been designed to be used with an Olympus endoscope and the endoscope position detecting unit for detection and display of the location and/or shape of the inserted endoscope. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) , Over-The Counter Use (21 CFR 801 Subpant C) # Please do not write below this line - continue on a separate page if needed. Comments of the States of Children Comments of Top FDA JISE ONLY - Portugues of Program Comments 1992 15 17 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/8/Picture/15 description: The image contains the name "Benjamin" in bold font on the top left. Below the name is the date "2014.03.07". The right side of the image contains some illegible text that appears to be a signature. The text ends with "05'00'". FORM FDA 3881 (9/13)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...