When used with the components of the Consensus Knee System: The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems. A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis. B. Failed osteotomy or unicompartmental replacements. C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists. D. The non-porous (uncoated and coated with CoCr beads without Titanium) components may only be used with cement. E. The porous coated (CoCr beads with Titanium) components may be used with or without cement. When used as a component of the Consensus Revision Knee System: The Consensus® Revision Knee System is designed as a system and is only intended to be used with compatible components of the Consensus® Knee System. The revision knee femoral and tibial components are intended for cemented use only. The indications for use are: A. Primary intervention of rheumatoid arthritis, osteoarthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle. B. Post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy) C. Failed osteotomy or unicompartmental replacements. D. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists. E. The salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery F. Moderate valgus, varus, or flexion deformities
Device Story
VitalitE Tibial Insert and Patellar Components are orthopedic implants serving as gliding surfaces in total knee arthroplasty. The device consists of UHMWPE (GUR 1020) infused with vitamin E (a-tocopherol) to improve material properties. It functions as a bearing surface between metallic femoral and tibial components. Used in clinical settings by orthopedic surgeons during knee replacement surgery. The device is a material modification of existing Consensus Knee System inserts; it maintains the same geometry and locking mechanism as predicate inserts. Benefits include potential for reduced wear due to vitamin E stabilization. Output is a physical implant component that restores joint function and stability.
Clinical Evidence
Bench testing only. Testing included material characterization, wear testing of the worst-case tibial insert, and anterior disassembly (push-out) testing of the locking mechanism across various sizes. Biocompatibility testing was also performed.
Technological Characteristics
Material: Vitamin E (a-tocopherol) infused UHMWPE (GUR 1020). Design: Cruciate retaining or ultra-congruent/PCL substituting. Sizes: 0-6; Thickness: 10-22 mm. Patellar components: Round or oval, 7.5 mm or 10 mm thickness. Sterilization: Validated process identical to current inserts. Shelf-life: 5 years.
Indications for Use
Indicated for patients requiring primary or revision total knee arthroplasty due to rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, degenerative arthritis, polyarthritis, collagen disorders, avascular necrosis, failed osteotomy/unicompartmental replacement, or joint deformity (valgus, varus, flexion).
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
## K133919 Page 1 of 3
2014
## 2. 510(k) SUMMARY
| Sponsor Name: | Consensus Orthopedics, Inc.<br>1115 Windfield Way, Suite 100<br>El Dorado Hills, CA 95762 | JUL 1 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|
| 510(k) Contact: | Matthew M. Hull, RAC<br>Phone: (916) 355-7156/ Fax: (916) 355-7190<br>mhull@consensusortho.com | |
| Date Prepared: | 20 December 2013 | |
| Trade Name: | VitalitE Tibial Insert & Patellar Components | |
| Common Name: | Total Knee Arthroplasty System | |
| Classification Name: | Knee joint patellofemorotibial Polymer/metal/polymer semi-<br>constrained cemented prosthesis is a Class 2 device per 21 CFR<br>888.3560 (Product Code JWH/OIY) | |
| | Knee joint patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis is a Class II device per 21 CFR<br>888.3565 (Product Code MBH) | |
## Device Description:
The VitalitE Tibial Insert is a vitamin E version of the UHMWPE tibial insert that has been a component of both the Consensus Knee System (CKS) and the Consensus Revision Knee System (RKS). The new insert still serves as the gliding surface between the metallic femoral component and the metallic tibial base plate of these knee systems. The new VitalitE insert will also be available in a cruciate retaining (CR) or ultra-congruent/ PCL substituting designs in sizes 0 through 6 and in thicknesses from 10 to 22 mm. The proprietary locking mechanism design on the underside of the insert will remain unchanged. The only difference will be the vitamin E additive in the UHMWPE which is a-tocopherol. This is the exact same polymer that is used in the Consensus VitalitE Acetabular insert but with a slightly lower radiation dose. Consensus will also offer our current CKS all-poly patellae in the new vitamin E UHMWPE material. This includes both the round and oval configurations with either a 7.5 mm or 10 mm thickness.
{1}------------------------------------------------
## Indications for Use:
When used with the components of the Consensus Knee System:
The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems.
- A. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
- B. Failed osteotomy or unicompartmental replacements.
- C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
- D. The non-porous (uncoated and coated with CoCr beads without Titanium) components may only be used with cement.
- E. The porous coated (CoCr beads with Titanium) components may be used with or without cement.
When used as a component of the Consensus Revision Knee System:
The Consensus® Revision Knee System is designed as a system and is only intended to be used with compatible components of the Consensus® Knee System. The revision knee femoral and tibial components are intended for cemented use only.
The indications for use are:
- A. Primary intervention of rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle.
- B. Post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion. dvsfunction, or prior patellectomy)
- C. Failed osteotomy or unicompartmental replacements.
- D. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
- E. The salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery
- F. Moderate valgus, varus, or flexion deformities
#### Substantial Equivalence:
## Technological Characteristics/ Substantial Equivalence:
The new VitalitE tibial insert is the exact same design as cleared for use with the Consensus Knee System in K932837, K953443 & K110950 and also with the Consensus Revision Knee in K 100542 and is compatible with all of the metallic femoral and tibial components from those systems. The UHMWPE (GUR 1020) with a-tocopherol is the same polymer as that cleared in K 130652 for the Consensus VitalitE Acetabular Insert with the exception of the slightly lower radiation dose. The use of Vitamin E infused UHMWPE tibial insert for knee systems has been cleared for use in DJO/ Encore systems via K091956 and K103223. The UHMWPE used in the current
{2}------------------------------------------------
# K133919
Page 3 of 3
Consensus tibial inserts is GUR 1050. The VitalitE inserts are cleaned, packaged, and sterilized using the same validated processes as the current inserts and will also have a 5 year shelf-life.
## Non-Clinical Performance Data:
Non-clinical testing and analysis were provided, including bench testing, material characterization, and biocompatibility testing. Bench testing of the new insert material included wear testing of the worst case tibial insert, and anterior disassembly testing of the locking mechanism (push-out testing) on a range of various sizes and types of inserts.
All of the observed results indicate that the VitalitE Tibial Inserts and Patellas are substantially equivalent to devices currently marketed. Therefore, the devices are as safe, as effective, and perform at least as safely and effectively as legally marketed predicates.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The image is a black and white picture of the address.
July 14, 2014
Consensus Orthopedics, Incorporated Mr. Matthew M. Hull QS & RA Director 1115 Windfield Way, Suite 100 El Dorado Hills. California 95762
Re: K133919
Trade/Device Name: VitalitE Tibial Inserts & Patellar Components Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, MBH, OIY Dated: June 10, 2014 Received: June 12, 2014
Dear Mr. Hull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
Page 2 - Mr. Matthew M. Hull - .
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins
- for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 1. INDICATIONS FOR USE STATEMENT
### 510(k) Number: K133919
## Device Name: VitalitE Tibial Inserts & Patellar Components
#### Indications for Use:
When used with the components of the Consensus Knee System:
The CONSENSUS® KNEE SYSTEM Primary Knee is designed as a system and is not intended for substitution of components from other systems.
- A. Primary intervention of theumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
- B. Failed osteotomy or unicompartmental replacements.
- C. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
- D. The non-porous (uncoated and coated with CoCr beads without Titanium) components may only be used with cement.
- E. The porous coated (CoCr beads with Titanium) components may be used with or without cement.
When used as a component of the Consensus Revision Knee System:
The Consensus® Revision Knee System is designed as a system and is only intended to be used with compatible components of the Consensus® Knee System. The revision knee femoral and tibial components are intended for cemented use only.
The indications for use are:
- A. Primary intervention of rheumatoid arthritis, osteoarthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle.
- Post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction, or B. prior patellectomy )
- C. Failed osteotomy or unicompartmental replacements.
- D. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
- The salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized u at the time of surgery
- Moderate valgus, varus, or flexion deformities F.
Prescription Use X (21 CFR Part 801 Subpart D) AND/OR
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
---
Division of Orthopedic Devices
Page 1/1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.