HIGHLY CROSS-LINKED VITAMIN E UHMWPE TIBIAL INSERTS
K103223 · Encore Medical L.P. · OIY · Dec 21, 2010 · Orthopedic
Device Facts
Record ID
K103223
Device Name
HIGHLY CROSS-LINKED VITAMIN E UHMWPE TIBIAL INSERTS
Applicant
Encore Medical L.P.
Product Code
OIY · Orthopedic
Decision Date
Dec 21, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DJO Surgical knee devices are intended for treatment of patients who are candidates for knee arthroplasty per the indications for use. While total knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
Device Story
Highly Cross-Linked Vitamin E UHMWPE (HXL VEK) tibial insert; component of 3DKnee System. UHMWPE powder blended with alpha-tocopherol (Vitamin E), compression molded, machined, gamma-irradiated for cross-linking, and gas plasma sterilized. Used in total knee arthroplasty to restore joint mobility and reduce pain. Device provides improved wear resistance and oxidation stability compared to conventional compression molded (CM) UHMWPE. Implanted by orthopedic surgeons in clinical settings. Output is a mechanical bearing surface; clinical benefit is potential reduction in wear-related complications.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing: accelerated aging (ASTM F2003), oxidation index (ASTM F2102), Izod impact (ASTM D256), tensile testing (ASTM D638), small punch testing (ASTM F2183), posterior peel-out testing, and knee simulator wear testing (ISO 14243-1). Biocompatibility evaluated via intramuscular rabbit implant study (ISO 10993).
Technological Characteristics
Material: UHMWPE blended with alpha-tocopherol (Vitamin E). Manufacturing: Compression molding, machining, gamma irradiation cross-linking, gas plasma sterilization. Mechanical properties evaluated per ASTM D638, D256, F2183, F2003, F2102. Device is a semi-constrained cemented/uncemented tibial insert.
Indications for Use
Indicated for patients with disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (patellofemoral erosion, dysfunction, or prior patellectomy); moderate valgus, varus, or flexion deformities; or fractures unmanageable by other techniques. Also indicated for salvage of failed surgical attempts. Used with 3DKnee System for cemented or uncemented applications.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Highly Cross-Linked Vitamin E UHMWPE Tibial Insert (K091956)
Submission Summary (Full Text)
{0}------------------------------------------------
K103223
p 1/5
# Summary of Safety and Effectiveness
| Preparation Date: | December 17, 2010 | DEC 21 2010 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------|-------------|
| Applicant/Sponsor: | Encore Medical (d.b.a. DJO Surgical)<br>9800 Metric Blvd<br>Austin, TX 78758 | |
| Contact Person: | William Garzon<br>Regulatory Affairs Technician | |
| Device Name: | Highly Cross-Linked Vitamin E UHMWPE Tibial Insert | |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis (888.3560), OIY, JWH, MBH | |
## Legally Marketed Devices to Which Substantial Equivalence is Claimed:
Encore Medical L.P. - Highly Cross-Linked Vitamin E UHMWPE Tibial Insert, K091956
## Device Description:
Subject of this Traditional 510(k) Premarket Notification is a request for labeling claims for the DJO Surgical 3DKnee HXL VE Tibial Insert. The in vitro wear claim will be made for the use of the 3DKnee femoral component coupled with a 3DKnee HXL VE Insert. It is important to note that there are no new total knee components being introduced as a result of this Traditional 510(k) premarket notification.
## Indications for Use:
Total joint replacement is indicated for patients suffering from disability due to:
- degenerative, post-traumatic or rheumatoid arthritis; .
- avascular necrosis of the femoral condyle; .
- post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, . dysfunction or prior patellectomy;
- . moderate valgus, varus or flexion deformities;
- Treatment of fractures that are unmanageable using other techniques .
This device may also be indicated in the salvage of previously failed surgical attempts. This device is intended to be used with the 3DKnee System for cemented or uncemented applications.
## Intended Use:
DJO Surgical knee devices are intended for treatment of patients who are candidates for knee arthroplasty per the indications for use. While total knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.
- - -
{1}------------------------------------------------
K103223
p2/5
. .
## Comparable Features to Predicate Device(s):
- . similar product claims
- have the same indicated use, .
- incorporate the same design, .
- . incorporate the same materials.
- . have the same shelf life, and
- . packaged and sterilized using the same materials and processes
## Claims:
The following is a summary of the claims submitted for clearance:
#### Claim 1:
DJO Surgical HXL VEK showed no measurable oxidation during accelerated age testing per ASTM-2003
FTIR analysis was performed on samples of HXL VEK tibial inserts at three different time points: non- aged, two weeks aged, and four weeks aged. Accelerated aging was performed per ASTM F2003 (oxygen environment at 5 atmospheres at 70℃ for 14 and 28 days). The oxidation index was calculated using the method described in ASTM F2102. All three time points showed no detectable oxidation indices < 0.1). Oxidation indices were 0.0440.02, 0.04±0.00, and 0.03±0.02 respectively. As a comparison, oxidation index for Compression Molded UHMWPE (CM) at two weeks aged is 0.73±0.12.
Izod Impact testing was performed on HXL VEK test specimens per ASTM D256 in both pre and post aged material. Impact testing can indicate the relative brittleness of a material; a consequence of oxidation. Accelerated aging was performed per ASTM F2003 (oxygen environment at 5 atmospheres at 70°C for 14 days). CM was used as a control group for comparative purposes. The average impact resistance for CM before and after accelerated aging was 93.5±7.2 KJ/m² and 25.1±7.0 KJ/m² respectively while the average impact resistance for HXL VEK before and after accelerated aging was 96.2±12.3 KJ/m2 and 102.3±13.3 KJ/m2 respectively.
HXL VEK showed no reduction in impact resistance after accelerated aging whereas CM showed marked degradation in impact resistance after accelerated aging.
CM bar stock is machined into a final part and then inspected, cleaned, packaged with nitrogen flush, and sterilized via gamma irradiation.
HXL VEK is UHMWPE powder that is blended with pure liquid pharmaceutical grade alpha-tocopheral (Vitamin E), compression molded, and machined into bar stock. HXL VEK bar stock is then gamma irradiated to cross link the material. The HXL VEK bar stock is then machined into a final part and then inspected, cleaned, pre-packaged, sterilized via gas plasma, and final packaged.
Bench testing is not necessarily indicative of clinical performance.
#### Claim 2:
DJO Surgical HXL VEK maintains the mechanical strength of conventional compression molded UHMWPE (CM) during mechanical testing.
Tensile testing was performed per ASTM D638 on CM, Highly Cross-Linked (HXL), and HXL VEK material. All test specimens were accelerated aged per ASTM F2003 (oxygen environment at 5 atmospheres at 70°C for 14 days). Average yield strength for each material was 24.98±0.76 MPa, and 25.09±0.72 MPa, and 25.09±0.72 MPa respectively. There was approximately a 12% drop off in yield strength for HXL as compared to CM. However, there was no drop off in yield strength for HXL VEK as compared to CM. Maintaining yield strength is a critical parameter for knee tibial insert performance.
Page 2 of 5
---
{2}------------------------------------------------
R103223
p 3/5
Tibial insert posterior peel-out testing was performed to determine the mechanical integrity of the tibial inserts relative to the metal baseplate. A snap feature is utilized to secure the metal baseplates. Tibial insert test parts that employ identical snap features were manufactured at nominal conditions for both CM and HXL VEK material. Testing was performed to simulate a worst case peel-out phenomenon in which the thickest tibial insert is forced out of the baseplate by applying a load to the insert in the anterior direction. An acceptance criterion was that the HXL VEK inserts should exhibit at least the same amount of resistance to posterior peel-out force as CM inserts. The average resistance load to posterior peel-out for CM and HXL VEK were 169.67±18.18 lbs and 182.43±2.23 bs respectively.
CM bar stock is machined into a final part and then inspected, cleaned, packaged with nitrogen flush, and sterilized via gamma irradiation.
HXL bar stock is CM bar stock that is gamma irradiated to cross link the material. The HXL bar stock is then machined into a final part and then inspected, cleaned, sterilized via gas plasma, and final packaged.
HXL VEK is UHMWPE powder that is blended with pure liquid pharmaceutical grade alpha-tocopheral (Vitamin E), compression molded, and machined into bar stock. HXL VEK bar stock is then gamma irradiated to cross link the material. The HXL VEK bar stock is then machined into a final part and then inspected, cleaned, pre-packaged, sterilized via gas plasma, and final packaged.
Bench testing is not necessarily indicative of clinical performance.
#### Claim 3:
DJO Surgical HXL VEK maintains mechanical strength after accelerated aging.
Tensile, small punch, and Izod Impact testing were performed on HXL VEK before and after accelerated aging. There was no significant decrease in yield strength, ultimate load, or impact resistance after accelerated aging was performed per ASTM F2003 (oxygen environment at 5 atmospheres at 70°C for 14 days). Tensile testing was performed per ASTM D638. The average yield strength for HXL VEK before and after accelerated aging was 26.76+0.96 MPa and 25.09+0.72 MPa respectively. Small punch testing was performed per ASTM F2183. The average ultimate load for HXL VEK before and after accelerated aging was 80.80±4.56 N and 81.31±4.22 N respectively. Izod impact testing was performed per ASTM D256. The average impact resistance for HXL VEK before and after accelerated aging was 96.2±12.3 KJ/m² and 102.3±13.3 KJ/m² respectively.
HXL VEK is UHMWPE powder that is blended with pure liquid pharmaceutical grade alpha-tocopheral (Vitamin E), compression molded, and machined into bar stock. HXL VEK bar stock is then gamma irradiated to cross link the material. The HXL VEK bar stock is then machined into a final part and then inspected, cleaned, pre-packaged, sterilized via gas plasma, and final packaged.
Bench testing is not necessarily indicative of clinical performance.
#### Claim 4:
DJO Surgical HXL VEK tibial bearings had an average wear rate that was 57% less than that of a conventional direct compression molded UHMWPE (DCM) bearing of the same geometry in knee simulator wear testing.
Wear testing was performed on the 3D™ Knee system comparing DCM tibial inserts and machined HXL VEK tibial inserts. Each test was performed on a multi-axis, force driven knee simulator (Instron/Stanmore 4-Station Knee Simulator) per ISO 14243-1 for 5 million cycles in a 25% concentration of defined bovine caff serum. Both inserts were articulated against a 3D™ Knee CoCr femoral component of matching size. The HXL VEK inserts were aged for two weeks per ASTM F2003 (oxygen environment at 5 atmospheres at 70°C for 14 days) while the DCM inserts were not aged. Both sets of inserts employed soak controls (loaded and unloaded) to correct for fluid absorption of the polyethylene during wear testing. The test specimens were Size 6 Left Femoral Components, Size 6 Left 11mm Inserts, and Size 6 Left Tibial Baseplates. The average wear rate for the DCM and HXL VEK tibial inserts were 4.4±3.0
{3}------------------------------------------------
K103223
p 415
mg/million cycles and 1.9±1.9 mg/million cycles respectively significant (p<0.01). Even though the HXL VEK inserts were machined and aged (vs. not aged for the DCM inserts), the HXL VEK inserts still showed superior wear properties over the DCM inserts.
Note, reduction in total polyethylene wear volume or wear rate alone may not result in an improved clinical outcome as wear particle size and morphology are also critical factors in the potential for wear mediated osteolysis and associated aseptic implant loosening. Particle size and morphology were not evaluated as part of this wear claim.
DCM parts are inspected, cleaned with nitrogen flush, and sterilized via gamma irradiation,
HXL VEK is UHMWPE powder that is blended with pure liquid pharmaceutical grade alpha-tocopheral (Vitamin E). compression molded, and machined into bar stock. HXL VEK bar stock is then gamma irradiated to cross link the material. The HXL VEK bar stock is then machined into a final part and then inspected, cleaned, pre-packaged, sterilized via gas plasma, and final packaged.
Bench testing is not necessarily indicative of clinical performance.
#### Claim 5:
DJO Surgical HXL VEK material was classified as a non-irritant as compared to the control under the conditions of the muscle implantation study.
Intramuscular implant testing was performed per ISO 10993 on three healthy adult New Zealand White rabbits. Compression molded UHMWPE (CM) material was used as the control article to compare to HXL VEK. Five HXL VEK sites and five CM sites were implanted for each rabbit. The surgical sites were closed, and the animals were observed daily for 13 weeks, After 13 weeks, the rabbits were euthanized and the HXL VEK and CM specimens were explanted at necropsy. All tissues were fixed in 10% neutral buffered formalin. Hematoxyiin and eosin (H&E) stained sections of the test and control implant sites were prepared from all animals. A veterinary pathologist microscopically evaluated the H&E stained tissue sections of each implant site. There were no gross abnormalities. The average imitant scores for CM and HXL VEK were 5.7 and 5.1 respectively. The Irritant Ranking Score is defined as: Test Group Average - Control Group Average. In this case, 5.1 - 5.7 = - 0.6. According to the interpretation of the Irritant Ranking Score, a non-irritant is classified as having a range of 0.0 - 2.9. Therefore, with an Irritant Ranking Score of -0.6. HXL VEK is classified as a non-irritant. Note, inflammatory response to device wear particles has not been evaluated.
CM bar stock is machined into a final part and then inspected, cleaned, packaged with nitrogen flush, and sterilized via gamma irradiation.
HXL VEK is UHMWPE powder that is blended with pure liquid pharmaceutical grade alpha-tocopheral (Vitamin E), compression molded, and machined into bar stock. HXL VEK bar stock is then gamma irradiated to cross link the material. The HXL VEK bar stock is then machined into a final part and then inspected, cleaned, pre-packaged, sterilized via gas plasma, and final packaged.
Per the ISO 10993-1 standard, all permanently implanted medical devices should be evaluated for irritation potential as part of the biocompatibility assessment.
Animal testing is not necessarily indicative of clinical performance.
### Non-Clinical Testing:
The following non-clinical laboratory testing was performed to determine substantial equivalence: mechanical material characterization (Tensile, Small Punch, Izod Impact, and crack propagation), physical and chemical characterization ( Oxidation Index, Compressive Modulus, Poisson's Ratio, Surface Roughness, Density, Onset Melting Temperature, Peak Melting Temperature, Delta H, Degree of Crystallinity, Crosslink Density, Swell Ratio, Molecular Weight, Polydispersity Index,
1
{4}------------------------------------------------
K103.223
pg 515
Lamallae Thickness, Free Radical Concentration, Vitamin E Concentration, Vitamin E Consolidation, Vitamin E Elution/Extraction, Trans-vinylene Index), tibial insert peel-out strength, wear testing, and biocompatibility. All testing has demonstrated the device is substantially equivalent to the predicate devices.
# Clinical Testing:
None provided as a basis for substantial equivalence.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
Encore Medical, L.P. % Mr. William Garzon Regulatory Affairs Technician 9800 Metric Boulevard Austin, Texas 78758
DEC 2 1 2010
Re: K103223
Trade/Device Name: Highly Cross-Linked Vitamin E UHMWPE Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: OIY, JWH, MBH Dated: October 29, 2010 Received: November 4. 2010
Dear Mr. Garzon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. | You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
..............................................................................................................................................................................
{6}------------------------------------------------
## Page 2 - Mr. William Garzon
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
DEC 2 1
510(k) Number (if known):
Device Name: Highly Cross-Linked Vitamin E UHMWPE Tibial Insert
GEC 2 1 2010
# Indications for Use:
Joint replacement is indicated for patients suffering from disability due to:
- degenerative, post-traumatic or rheumatoid arthritis; .
- . avascular necrosis of the femoral condyle;
- post-traumatic loss of joint configuration, particularly when there is patellofemoral . erosion, dysfunction or prior patellectomy;
- . moderate valgus, varus or flexion deformities;
- Treatment of fractures that are unmanageable using other techniques .
This device may also be indicated in the salvage of previously failed surgical attempts. This device is intended to be used with the 3DKnee System for cemented or uncemented applications.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for m. Melkeron
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103223
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.