QRmaster-H/Revo is a dedicated X-ray imaging device that acquires a 360-degree rotational X-ray sequence of images for use as diagnostic support in radiology of the dento-maxillo-facial complex and in the field of maxillofacial surgery. The device accomplishes this task by reconstructing a three-dimensional matrix of the examined volume, producing two-dimensional views of this volume, and displaying both 2D images and 3D renderings. This technique is known as cone beam computed tomography, or CBCT. All devices are sold by or on the order of a physician. They are not for use by the general public or over-the-counter.
Device Story
QRmaster-H/Revo is an extraoral 2D panoramic and 3D CBCT X-ray imaging system for dental professionals. Device captures 360-degree rotational X-ray sequences using a cone-shaped beam projected onto a single amorphous silicon flat panel detector. System utilizes three field-of-view (FOV) options, positioned via panoramic scout, 2D scout, or four laser beams. Firmware operates the device via an attached control panel; image acquisition software reconstructs 3D volumes from transmission data. Output consists of 2D images and 3D renderings stored in DICOM format for export to third-party viewing stations (PCs). Clinicians use these images for diagnostic support in dento-maxillo-facial radiology and maxillofacial surgery. Device benefits include targeted 3D visualization of oral structures to aid surgical planning and diagnosis.
Clinical Evidence
Bench testing only. Clinical images were evaluated by a professor at Asahi University School of Dentistry and deemed diagnostically relevant and reliable. Submission included sample 2D and 3D images using test phantoms. Software verification and validation, EMC, and electrical safety testing (60601 series) were performed.
Technological Characteristics
High-frequency rotating anode X-ray source; amorphous silicon flat panel detector (Hamamatsu) with CsI scintillator. Supports 2D panoramic and 3D CBCT imaging. Connectivity via DICOM export to third-party workstations. Electrical safety per 60601; laser safety per 60825-1. Software-based reconstruction of 3D matrix from 360-degree rotational X-ray data.
Indications for Use
Indicated for diagnostic support in radiology of the dento-maxillo-facial complex and maxillofacial surgery. Intended for use by dental professionals on patients requiring 2D panoramic or 3D CBCT imaging of dentition, jaws, and oral structures.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 4. 2014
Telesystems Co., Ltd. % Mr. Claude Berthoin President Denterprise International, Inc. 110 East Granada Blvd. ORMOND BEACH FL 32176
Re: K133797
Trade/Device Name: Omaster-H/Revo Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: August 1, 2014 Received: August 4, 2014
Dear Mr. Berthoin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Smh.p)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K133797
Device Name QRmaster-H/Revo
Indications for Use (Describe)
QRmaster-H/Revo is a dedicated X-ray imaging device that acquires a 360-degree rotational X-ray sequence of images for use as diagnostic support in radiology of the dento-maxillo-facial complex and in the field of maxillofacial surgery. The device accomplishes this task by reconstructing a three-dimensional matrix of the examined volume, producing two-dimensional views of this volume, and displaying both 2D images and 3D renderings. This technique is known as cone beam computed tomography, or CBCT.
All devices are sold by or on the order of a physician. They are not for use by the general public or over-the-counter.
Type of Use (Select one or both, as applicable)
|× Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 807 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| 510(k) Owner: | Telesystems Co., Ltd.<br>2-8-19, Ebisunishi, Naniwa-Ku<br>Osaka, Japan |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Yoshimitsu Kobayashi<br>General Manager<br>Phone: +81-6-6631-0925<br>Fax: +81-6-6631-5926 |
| Date Prepared: | December 10, 2013 |
| 510(k) Preparer: | Denterprise International, Inc.<br>110 E. Granada Blvd., Suite 207<br>Ormond Beach, FL 32176<br>386-672-0450 (Voice)<br>386-672-4402 (Fax) |
| Primary Contact: | John Gillespy, VP Regulatory Affairs (john@510kfda.com) |
| Secondary Contact: | Claude Berthoin, President (claude@510kfda.com)<br>Holly Layne, Assistant (holly@510kfda.com) |
| Trade Name: | QRmaster-H/Revo |
| Common Name: | Cone Beam Computed Tomography (CBCT) X-Ray System |
| Classification Name: | Dental Computed Tomography X-Ray System (21 CFR 892.1750, Product<br>Code OAS) |
| Predicate Devices: | Hyperion (K113497); Cefla Dental/MyRay, Italy<br>Orthophos XG (K103711); Sirona Dental, Germany<br>NewTom VG (K072357); Cefla Dental/QR, Italy |
| Device Description: | QRmaster-H/Revo is an extraoral 2D (panoramic) and 3D (CBCT) x-ray<br>imaging system for use by dental professionals. The device is designed to produce cross-<br>sectional images of the body by computer reconstruction of x-ray transmission data from |
e body by lput the same axial plane taken at different angles. To accomplish this it uses a cone-shaped xray beam projected on a flat panel detector to capture a rapid series of 2D images using the standard narrow beam technique. Such images from a single 360° scan are reconstructed with special software to be viewed on 3D viewing stations (PCs) that are not part of the system.
The system makes use of a single flat panel detector for both panoramic and cone beam images. Three field of views (FOVs) provide targeted regions of interest that can be positioned by panoramic scout, two-dimensional scout, or four positioning laser beams.
{5}------------------------------------------------
QRmaster-H/Revo includes firmware to operate the device through a control panel attached to the main device. The device also includes image acquisition software that stores 3D images in standard DICOM format for export to third party viewing or image management software not part of the system. Users are expected to have or acquire such image management software for communication both with the device's image acquisition software and with third party 3D-viewing software. These latter applications are readily available in the domestic marketplace.
- Indications For Use: ORmaster-H/Revo is a dedicated X-ray imaging device that acquires a 360-degree rotational X-ray sequence of images for use as diagnostic support in radiology of the dento-maxillo-facial complex and in the field of maxillofacial surgery. The device accomplishes this task by reconstructing a three-dimensional matrix of the examined volume, producing two-dimensional views of this volume, and displaying both 2D images and 3D renderings. This technique is known as cone beam computed tomography, or CBCT.
- Comparison to Predicates QRmaster-H/Revo is intended for 2D panoramic diagnostic examination of dentition, jaws, and oral structures, the same as all three predicates. The device is also intended for 3D tomographic examination of dento-maxillo-facial areas and to support maxillofacial surgery, the same as two predicate devices with CBCT technology (Orthophos XG and NewTom VG).
Table 5 on the next page compares technological features between the subject and predicate devices. It shows the QRmaster-H/Revo and predicates share the same fundamental scientific technology. All are dedicated high-end x-ray imaging systems for dentistry.
More specifically, ORmaster-H/Revo is a PAN/CBCT x-ray system. All three predicates share the subject's 2D PAN technology, whereas two of the three share its 3D CBCT technology (Orthophos XG and NewTom VG). Two predicates also include CEPH technology for orthodontists (Hyperion and Orthophos XG), a feature not included in the subject device.
All devices feature similar x-ray generation technology, while QRmaster-H/Revo shares the same single-panel x-ray detector technology as NewTom VG.
Scan performance is continuous for both ORmaster-H/Revo and Hyperion. Exposure times are higher for the subject device, but measured effective exposure is equal to or lower than the predicates (as discussed in Section 12, Substantial Equivalence Discussion).
None of the differences in technological features between the subject and predicate devices raise new issues of safety or effectiveness.
{6}------------------------------------------------
| Devices | QRmaster-H/Revo<br>(Subject) | Hyperion | Orthophos XG | NewTom VG |
|------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------|------------------------------------------------------|
| 510(k) Number | NA | K113497 | K103711 | K072357 |
| Manufacturer | Telesystems<br>(Japan) | Cefla Dental/MyRay<br>(Italy) | Sirona Dental<br>(Germany) | Cefla Dental/QR<br>(Italy) |
| Device Photo | Image: QRmaster-H/Revo | Image: Hyperion | Image: Orthophos XG | Image: NewTom VG |
| Common Name | PAN & CBCT<br>X-Ray System | PAN & CEPH<br>X-Ray System | PAN, CEPH &<br>CBCT<br>X-Ray<br>System | PAN & CBCT<br>X-Ray System |
| X-Ray Generator | | | | |
| X-Ray Source | High Frequency,<br>Rotating Anode | High Frequency, Rotating<br>Anode | High Frequency,<br>Rotating Anode | High Frequency,<br>Rotating Anode |
| Shape of X-Ray<br>Beam | Fan-shaped x-ray<br>beam and cone<br>beam for<br>tomography x-ray | Fan-shaped x-ray beam | Fan-shaped x-ray<br>beam and cone<br>beam for<br>tomography x-ray | Cone beam for<br>tomography x-ray |
| X-Ray Detector | | | | |
| Single Image<br>Detector? | YES | NO | NO | YES |
| Image Detector<br>(PAN) | Amorphous silicon<br>flat panel detector<br>(Hamamatsu) | Digital CCD linear detector,<br>repluggable for CEPH<br>exposure | Digital CCD linear<br>detector,<br>repluggable for<br>PAN exposure | Amorphous silicon<br>flat panel detector<br>(Varian) |
| Image Detector<br>(CBCT) | Amorphous silicon<br>flat panel detector<br>(Hamamatsu) | NA | Digital FPD with<br>CMOS technology,<br>integrated for 3D<br>exposure technique | Amorphous silicon<br>flat panel detector<br>(Varian) |
| Scintillator<br>(CBCT) | Csl | NA | Not documented | Csl |
| Scan Performance | | | | |
| Type of X-Ray<br>Emission | Continuous for PAN<br>and CBCT exams | Continuous for PAN and<br>CEPH exams | Not documented | Pulsed for CBCT<br>exams |
| Exposure Time<br>(Standard Pan) | 12 s, continued<br>exposure | 9.3 s, continued exposure | 14.1 s | NA |
| Exposure Time<br>(CBCT) | 18 s, effective<br>radiation time | NA | 3.2-5.0 s, effective<br>radiation time | 3.6 s - 5.4 s,<br>effective radiation<br>time |
| FOV (CBCT -<br>Voxel Size - O X H<br>cm) | 5.0 X 4.5 cm<br>9.5 X 4.5 cm<br>9.5 X 9.0 cm | 7 cm (H.R. Zoom 4")<br>11 cm (6" detector)<br>15 cm (9" detector) | 8.0 X 8.0 cm | 8.0 X 10.0 cm |
# Table 5 - Comparison of Technological Features
{7}------------------------------------------------
- Performance Data: Clinical images were examined by a qualified professor at Asahi University School of Dentistry and found to be diagnostically relevant and reliable. Sample 2D and 3D images utilizing test phantoms are also included in the submission.
- Biocompatibility Assessment: Biocompatibility testing is not warranted since surfaces in contact with users are covered with single-use protective barriers during use.
- Software and Risk Analysis: A software level of concern analysis that included a device risk analysis in conformance with ISO 14971 was performed. Results of that analysis determined the residual risks inherent in the use of QRmaster-H/Revo are "acceptable." Verification and validation of the software included tests that confirmed its functionality.
- EMC and Electrical Safety Testing: Independent laboratory and in-house testing was performed to demonstrate conformance with recognized electrical (60601 et al) and laser device (60825-1) standards.
- FDA Guidance Compliance: The subject device and this petition comply with the following FDA guidance documents: (a) Pediatric Information for X-Ray Imaging Device Premarket Notifications, draft guidance issued on May 10, 2012; (b) Guidance for the Submission of 510(k)s for Solid State X-Ray Imaging Devices, issue on August 6, 1999; (c) Laser Products – Conformance with IEC 60825-1 and IEC 60601-2-22, issued on June 24, 2007; and (d) Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005.
- Conclusions: Both intended use and fundamental scientific technology are the same in the subject and predicate devices, and any differences in technological features do not raise new issues of safety or effectiveness. A comparison of such features, combined with performance data that includes clinical evaluation of device images, a software and device risk analysis, and electromagnetic compatibility as well as electrical and laser safety tests support the conclusions that ORmaster-H/Revo is safe for its intended use and is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.