K103711 · Sirona Dental Systems GmbH · OAS · Mar 22, 2011 · Radiology
Device Facts
Record ID
K103711
Device Name
ORTHOPHOS XG 3D / CEPH
Applicant
Sirona Dental Systems GmbH
Product Code
OAS · Radiology
Decision Date
Mar 22, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The ORTHOPHOS XG 3D/ Ceph is intended to produce two dimensional images and three dimensional volume reconstructions, including partial volumes and selected projections of the dentomaxillofacial areas, for use in planning and diagnostic support. Image acquisition, modes include panoramic X-ray, cephalometric X-ray, specialized tomographic X-ray, and cone beam tomography X-ray. The system can also acquire carpal exposures. The system includes an X-ray source, flat panel X-ray detector, positioning devices, as well as interactive 3D reconstruction, processing, and archiving software.
Device Story
Dental computed tomography system; produces 2D (panoramic/cephalometric) and 3D (cone beam) images of teeth, jaw, and head. Device uses X-ray source and combi-sensor (line sensor for 2D, flat panel for 3D) mounted on motor-driven rotatable ring. Patient positioned via bite block, forehead/temple supports, and four class 1 laser localizers. Operator uses control panel for mode/program selection and height adjustment; exposure triggered via handhold push-button. PC software reconstructs 3D volumes and processes 2D images; data conveyed to SIDEXIS database for clinician review. Supports planning and diagnostic decision-making in dental/maxillofacial clinical settings. Benefits include integrated multi-modal imaging capabilities in single device.
Clinical Evidence
No clinical data. Bench testing only; system-level testing performed using test phantoms to evaluate exposure equality against predicate devices (ORTHOPHOS XG Plus DS and GALILEOS).
Technological Characteristics
Dental X-ray system; 2D line sensors and 3D flat panel detector. Conical X-ray beam. Class 1 laser localizers. Biocompatible patient-contact materials (ISO 10993-1). PC-based software for 3D reconstruction and 2D processing. Networked connectivity to SIDEXIS database. Height-adjustable support stand with rotatable ring.
Indications for Use
Indicated for patients requiring dentomaxillofacial imaging for diagnostic support and treatment planning, including panoramic, cephalometric, specialized tomographic, cone beam CT, and carpal exposures.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
## 510(k) Summary for Sirona Dental Systems ORTHOPHOS XG 3D/ Ceph
MAR 2 2 2011
### ー SPONSOR
Sirona Dental Systems GmbH Fabrikstrasse 31 D-64625 Bensheim Germany Contact Person: Fritz Kolle 49 6251 16 32 94 Telephone: Date Prepared: December 14, 2010
### રપ DEVICE NAME
| Proprietary Name: | ORTHOPHOS XG 3D / Ceph |
|----------------------|-------------------------------------|
| Common/Usual Name: | x-ray, tomography, computed, dental |
| Classification Name: | Computed tomography x-ray system |
#### 3 PREDICATE DEVICES
Sirona GALILEOS (K060892) and Sirona ORTHOPHOS XG"]48 DS (K033073)
### INTENDED USE ব
The ORTHOPHOS XG 3D/ Ceph is intended to produce two dimensional images and three dimensional volume reconstructions, including partial volumes and selected projections of the dentomaxillofacial areas, for use in planning and diagnostic support. Image acquisition, modes include panoramic X-ray, cephalometric X-ray, specialized tomographic X-ray, and cone beam tomography X-ray. The system can also acquire carpal exposures. The system includes an X-ray source, flat panel X-ray detector, positioning devices, as well as interactive 3D reconstruction, processing, and archiving software.
"
"
{2}------------------------------------------------
### ട് Device Description and Function
The ORTHOPHOS XG 3D/ Ceph is a dental computed tomography system intended to produce X-rays for obtaining two dimensional images (panoramic and cephalometric) and three dimensional volume reconstructions of the teeth, jaw, and the head area, which includes dentomaxillofacial areas, for use in planning and diagnostic support. In 3D mode the OP XG 3D device generates a conical x-ray beam that rotates round the patient´s head within a certain angle.
The device comprises a combi sensor with 2D panoramic and cephalometric sensors and a flat panel sensor for 3D volume exposure. Five volume regions are defined by the geometry of the ORTHOPHOS XG 3D/ Ceph. Four class I laser beam light localizers serve for postioning the patient's head that may be fixed through bite block and adjustable forehead and temple supports.
From the obtained exposures in the 3D mode a three dimensional image is reconstructed and can be viewed as well as panoramic/cephalometric images. The constructed 3D volume and simulated projection exposures as well as panoramic/cephalometric data are conveyed to SIDEXIS and stored in the SIDEXIS data base.
An operator control panel allows height adjustment, selection of mode and program and indicates machine states.
A separate handhold push-button serves for exposure release.
An optional remote control is available.
Sirona Dental Systems 510(k) ORTHOPHOS XG 3D/ Ceph
Page 2 of 6
{3}------------------------------------------------
### 5.1 Scientific Concept
The underlying scientific concept is combining two exposure technologies, panoramic including cephalometric, (2D), and volumetric (3D), in one device. Volumetric exposures are obtained by cone-beam technology.
### 5.2 Physical and Performance Characteristics
# 5.2.1 Design
The ORTHOPHOS XG 3D/ Ceph comprises of a support stand to which a height adjustable sled is attached. The sled carries the patient fixation, the operator control panel (easy pad) and the motor driven rotatable ring. The X-ray source and a combi sensor are fixed to the ring. The combi sensor houses a panoramic and cephalometric line sensor (2D) and a flat panel sensor for volumetric exposures (3D). In 3D mode the OP XG 3D device generates a conical x-ray beam that rotates round the patient's head within a certain angle.
Four class 1 laser beam light localizers serve for positioning the patient's head that may be fixed through bite block and adjustable forehead and temple supports.
The exposure area is defined by the geometry of the ORTHOPHOS XG 3D/ Ceph. A control panel allows the user to select the exposure modes and the exposure factors, view the machine status information, control the height adjustment and turn on the laser indicator.
An optional remote control is available.
The PC software reconstructs the three dimensional image as well as processing panoramic/ cephalometric images. The constructed 3D volume and simulated projection exposures as well as panoramic/ cephalometric data are conveyed to SIDEXIS and stored in the SIDEXIS data base.
{4}------------------------------------------------
# 5.2.2 Material Used
Materials that come into patent contact intentionally are biocompatible and evaluated according to ISO 10993-1: 2003, "Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process".
# 5.2.3 Physical Properties
Not applicable.
### SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS ರಿ
The ORTHOPHOS XG 3D/Ceph is a further development of the ORTHOPHOS XGPlus DS / Ceph (K033073) 1 and expands the device capabilities with the 3D function of GALILEOS (K060892).
In 2D mode the ORTHOPHOS XG 3D /Ceph provides the same 2D operating principles and programs as the predicate device ORTHOPHOS XGRUS DS/Ceph.
In 3D mode the ORTHOPHOS XG 3D /Ceph provides an adapted program-subset of the GALILEOS and the same 3D operating principles as the predicate device GALILEOS.
The modified operator panel of the ORTHOPHOS XGP™s DS/Ceph now incorporates the 3D functionality in addition.
In the ORTHOPHOS XG 3D/Ceph a solid state flat panel image receptor replaces the image intensifier of the GALILEOS and is combined in a common housing with image receptors of the ORTHOPHOS XG°l45 DS / Ceph. The appropriate sensors are moved in the correct positions automatically according to mode and program selected.
ORTHOPHOS XG 3D/Ceph provides the selection of five volume regions whereas the GALILEOS provides these regions together in one volume.
<sup>1</sup> It includes the ORTHOPHOS XG 5DS / Ceph which provides a subset of programs.
{5}------------------------------------------------
The X-ray generator of the ORTHOPHOS XGP™ DS / Ceph has been adapted in order to be capable of 3D scans and to have similar technical characteristics as the GALILEOS generator.
ORTHOPHOS XG 3D/Ceph has a smaller collimator opening in 3D mode compared with GALILEOS (smaller field of view and five volume regions).
An additional class I laser light beam localizer has been added to ORTHOPHOS XGPUS DS / Ceph that in conjunction with another one marks the Field of View in 3D mode.
The PC software performs same functions and algorithms as with the ORTHOPHOS XG 105 DS / Ceph in the 2D mode and GALILEOS in 3D mode. For the 3D mode the software has been adapted to the smaller field of view and the selection for five volume regions.
ORTHOPHOS XG 3D/Ceph offers a calculated panoramic view in combination with slices orthogonal to the panoramic curve ('transversal slices') as predicate GALILEOS.
ORTHOPHOS XG 3D/Ceph offers same functionality in viewing slices, projections and volume views.
### 7 NONCLINICAL TESTING
The ORTHOPHOS XG 3D/ Ceph system functions have been tested in a system test.
For both modes, 2D and 3D, exposures have been tested utilizing test phantoms. The tests evaluate the equality of exposures of proposed ORTHOPHOS XG 3D/ Ceph and predicate devices ORTHOPHOS XG" " DS/Ceph and GALILEOS.
For 3D mode additional tests has been performed taking into account FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, Document issued on: August 6, 1999"
{6}------------------------------------------------
### 8 CLINICAL TESTING
Clinical tests have not been performed.
### த CONCLUSION
Based on a comparison of intended use, indications, construction materials, principal of operations, features and technical data, the Sirona Dental ORTHOPHOS XG 3D /Ceph is safe and effective to perform its intended use as well as substantially equivalent to the Predicate Devices.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Fritz Kolle Primary Contact Sirona Dental System GmbH Fabrikstrasse 3, D-64625 Bensheim GERMANY
MAR 2 2 2011
Re: K103711
Trade/Device Name: Orthophos XG 3D/ Ceph Regulation Number: 21 CFR 872.1800 · Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: December 14, 2010 Received: December 20, 2010
Dear Mr. Kolle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{8}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary S Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
510(k) Number (if known):
K.103711
ORTHOPHOS XG 3D/ Ceph Device Name:
Indications for Use:
The ORTHOPHOS XG 3D/ Ceph is intended to produce two dimensional images and three dimensional volume reconstructions, including partial volumes and selected projections of the dentomaxillofacial areas, for use in planning and diagnostic support. Image acquisition, modes include panoramic X-ray, cephalometric X-ray, specialized tomographic X-ray, and cone beam tomography X-ray. The system can also acquire carpal exposures. The system includes an X-ray source, flat panel X-ray detector, positioning devices, as well as interactive 3D reconstruction, processing, and archiving software.
Prescription Use (Part 21 CFR 801 Subpart D) OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary Scott
Vivision Sign-On)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K103711
Sirona Dental Systems 510(k) ORTHOPHOS XG 3D/ Ceph
November 10, 2010
Page viii
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.