The OsteoMed ExtremiLOCK Ankle Plating System is intended for fixation of fractures, arthrodesis, osteotomies, and non-unions of the tibia and fibula. The ExtremiLOCK Ankle Plating System implants are intended for single use only. The 1/3 tubular plates, hook plates, screws, and washers are also intended for use in trauma, general surgery, and reconstructive procedures of bones appropriate for the size of the device. The OsteoMed ExtremiLOCK Ankle Plating System can be used for adult and pediatric patients.
Device Story
The OsteoMed ExtremiLOCK Ankle Plating System is a bone fixation system comprising plates, screws, and K-wires used by surgeons for fracture management, arthrodesis, and osteotomies. The system provides compression, non-locking, and locking hole designs to stabilize bony fractures of the distal tibia and fibula. Implants are provided in various shapes and sizes; surgical instrumentation is included to facilitate insertion, modification, and removal. The device is used in clinical settings (OR) to provide mechanical stabilization of bone segments, promoting healing and structural integrity. It is intended for single use only.
Clinical Evidence
Bench testing only. Verification evaluation included pull-out, torque, and bending tests to ensure mechanical strength. No clinical data was required to support substantial equivalence.
Technological Characteristics
Implants manufactured from Titanium (ASTM F-67) or Titanium alloy (ASTM F-136). Instrumentation made from stainless steel, anodized aluminum, and medical grade polymer. System features angulated locking screws, standard non-locking screws, and plates with compression, non-locking, and locking hole designs.
Indications for Use
Indicated for fixation of fractures, arthrodesis, osteotomies, and non-unions of the tibia and fibula in adult and pediatric patients. Also indicated for trauma, general surgery, and reconstructive procedures of bones appropriate for device size.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "OSTEOMED" in bold, black letters. The word is inside of a black oval shape. There is a registered trademark symbol in the upper right corner of the oval.
K133691 (pg 1/3)
## 510(k) Summary
FEB 2 4 2014
Submitter Information:
OsteoMed 3885 Arapaho Road Addison, Texas 75001 Phone: (972) 677-4600 Fax: (972) 677-4601
Contact Person:
Blesson Abraham
Date Prepared: November 27, 2013
Device Information:
Proprietary/Trade Name: OSTEOMED Ankle Plating System Common Name: Bone Plating System
Classification Name:
- · Regulation Number: o 21 CFR 888.3030
- o 21 CFR 888.3040
- · Regulation Name:
- o Single/multiple component metallic bone fixation appliances and accessories (Plate, Fixation, Bone)
- o Smooth or threaded metallic bone fixation fastener (Screw, Fixation, Bone: Pin, Fixation, Smooth)
- · Product Code:
- HRS o
- HWC o
Device Class: II
#### Predicate Devices:
Synthes Dynamic Compression Locking Plating System (Synthes One-Third Tubular DCL Plate), K011335 Classification Name: Single/multiple component metallic bone fixation appliances and accessories (21CFR 888.3030, Product Code HRS) Device Class: II
OsteoMed Foot Plating System, K091614 Classification Name: Single/multiple component metallic bone fixation appliances and accessories (21CFR 888.3030, Product Code HRS) Device Class: II
{2}------------------------------------------------
#### OsteoMed Super Screw 3.5mm, K954330
Classification Name: Smooth or threaded metallic bone fixation fastener (21CFR 888.3040, Product Code HWC)
#### Device Class: Il
#### Merete Locking Bone Plate System III, K120787
Classification Name: Single/multiple component metallic bone fixation appliances and accessories (21CFR 888.3030, Product Code HRS) Smooth or threaded metallic bone fixation fastener (21CFR 888.3040. Product Code HWC)
Device Class: II
#### Device Description:
The OSTEOMED ExtremiLOCK Ankle Plating System is a comprehensive ankle fracture system intended to provide solutions for various bony fractures, including simple to complex fractures of the distal tibia and fibula. The plates are provided in a variety of shapes and sizes, offering surgeons compression, non-locking and locking hole designs. The ExtremiLOCK Ankle Plating System includes angulated locking screws and standard non-locking screws as well as K-wires. Surgical instrumentation is provided to facilitate insertion, modification and/or removal of implants.
The implants of the OsteoMed ExtremiLOCK Ankle Plating System are made from Titanium (ASTM F-67) or Titanium alloy (ASTM F-136). The instrumentation is made from various grades of stainless steel, anodized aluminum, and/medical grade polymer.
#### Intended Use:
The OsteoMed ExtremiLOCK Ankle Plating System is intended for fixation of fractures, arthrodesis, osteotomies, and non-unions of the tibia and fibula. The ExtremiLOCK Ankle Plating System implants are intended for single use only.
The 1/3 tubular plates, hook plates, screws, and washers are also intended for use in trauma, general surgery, and reconstructive procedures of bones appropriate for the size of the device.
The OsteoMed ExtremiLOCK Ankle Plating System can be used for adult and pediatric patients.
#### Technological Characteristics:
The OsteoMed Extremil.OCK Ankle Plating System is recommended for fixation/reconstruction of distal tibia and fibula, or other bones appropriate for the size of the device.
ExtremiLOCK implants are manufactured from Titanium (ASTM F-67) or Titanium alloy (ASTM F-136), the same materials used in the manufacture of the predicate devices. These materials are biocompatible.
{3}------------------------------------------------
#### Performance / Clinical Data:
The OsteoMed ExtremiLOCK Ankle Plating System was compared to the Synthes One-Third Tubular DCL Plate, K011335, the OsteoMed Foot Plating System, K091614, the OstedMed Super Screw 3.5mm, K954330, and the Merete Locking Bone Plate System III, K120787. The Extremil.OCK Ankle implants underwent verification evaluation (pull-out, torque and bending tests) to ensure that the design features met the required mechanical strength criteria for their intended use. The intended use of the OsteoMed ExtremiLOCK Ankle implants is similar to the Synthes One-Third Tubular DCL Plate, the OsteoMed Foot Plating System, the OstedMed Super Screw 3.5mm, and the Merete Locking Bone Plate System III.
Performance equivalence was shown through the verification comparison to the predicate devices.
Clinical Testing is not required to support substantial equivalence.
#### Substantial Equivalence:
A design and dimensional comparison was performed to establish substantial equivalence to the legally marketed predicate devices listed in this summary. The basis of substantial equivalence for this device is based on similarities in intended use, function, performance, design, technology and operational principles to the Synthes One-Third Tubular DCL Plate (K011335), the OsteoMed Foot Plating System (K091614), the OsteoMed Super Screw 3.5mm (K954330), and similarities in intended use, material, and function to the Merete Locking Bone Plate System III (K120787).
Substantial equivalence was shown through the pullout test, and bending test to the predicate devices (the Synthes One-Third Tubular Plate, the OsteoMed Foot Plating System, and the OsteoMed Super Screw 3.5mm). The intended use, design, technology and operational principles are similar between the subject and predicates (the Synthes One-Third Tubular Plate, the OsteoMed Foot Plating System, the OsteoMed Super Screw 3.5mm, and Merete Locking Bone Plate System III), and therefore OsteoMed believes that the OSTEOMED ExtremiLOCK Ankle Plating System does not raise any new safety or effectiveness issues.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2014
OsteoMed Mr. Blesson Abraham Regulatory Affairs Specialist 3885 Arapaho Road Addison, Texas 75001
Re: K133691
Trade/Device Name: OsteoMed ExtremiLOCK Ankle Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: November 27, 2013 Received: December 2, 2013
Dear Mr. Abraham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
#### Page 2 - Mr. Blesson Abraham
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Vincent訂房@evlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: OsteoMed ExtremiLOCK Ankle Plating System
Indications for Use:
The OsteoMed ExtremiLOCK Ankle Plating System is intended for fixation of fractures, arthrodesis, osteotomies, and non-unions of the tibia and fibula. The ExtremiLOCK Ankle Plating System implants are intended for single use only.
The 1/3 tubular plates, hook plates, screws, and washers are also intended for use in trauma, general surgery, and reconstructive procedures of bones appropriate for the size of the device.
The OsteoMed ExtremiLOCK Ankle Plating System can be used for adult and pediatric patients.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth南®rank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.