The MEDINAUT Kyphoplasty System is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
Device Story
MEDINAUT Kyphoplasty System is a single-use, sterile surgical kit for percutaneous vertebral augmentation; used to reduce spinal compression fractures and restore sagittal alignment. System components include balloon catheter, bone expander syringe, needles, cannula, and cement delivery tools. Physician inserts balloon catheter into vertebral body under fluoroscopic guidance; inflates balloon to create void in cancellous bone; removes balloon; injects cleared PMMA bone cement via dispensing tools. Procedure performed in clinical/surgical setting by trained physicians. Output is restored vertebral height and stabilized fracture; benefits include reduced back pain and improved patient functional mobility.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via functional testing including tensile bond strength, fatigue strength, balloon deflation time, burst pressure (constrained/unconstrained), balloon dimension analysis, and insertion/withdrawal force. Biocompatibility testing performed per ISO 10993 standards.
Indicated for patients requiring reduction and fixation of spinal fractures or creation of a void in cancellous bone during percutaneous vertebral augmentation. Contraindications include posterior wall/pedicle instability, infection, severe bleeding, known bone cement allergies, pregnancy, >68% vertebral height loss, or anatomy unsuitable for balloon placement.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
### MEDINAUT Kyphoplasty System 510k Summary APR 0 4 2014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).
- Mar 24, 2014 1. Date:
- Applicant / Sponsor 2.
| | Company |
|--------------|----------------------------------------------------------------------------------------------------|
| Name | IMEDICOM Co., Ltd. |
| Address | #612 Hanlim Human Tower, 1-40, Guemjung-dong,<br>Gunpo-si, Gyeonggi-do, Republic of Korea, 435-824 |
| Phone<br>Fax | +82 31-479-1156<br>+82 31-479-1157 |
| Contact | Ki Ung. Choi |
## 3. Submission Contact Person
Priscilla Chung / LK Consulting Group USA, Inc.
2651 E Chapman Ave Ste 110,
Fullerton CA 92833
714-202-5789 Tel:
714-409-3357 Fax:
E-mail: info@lkconsultinggroup.com
## 4. Proposed Device Identification:
Proprietary Name - MEDINAUT Kyphoplasty System Common Name -- Inflatable Bone Tamp Classification Name - Arthroscope
Predicate Device: KyphX Inflatable Bone Tamp, K010246 & K041454 Kyphon Inflatable Bone Tamp, K981251
## 5. Proposed Device Classifications & Citations:
Classification: Class II
Product Code A:
Regulation Number: 21CFR§888.3027
Review Panel: Orthopedic
{1}------------------------------------------------
Classification: Class II Product Code B: HRX Regulation Number: 21CFR§880.1100
Review Panel: Orthopedic
## 6. Predicate Device Identification:
510(k) Number: Primary Predicate: K041454 Secondary Predicates: K010246 and K981251 KyphX® Inflatable Bone Tamp Product Name: Manufacturer's Name: Kyphon Inc. (Medtronic)
## 7. Device Description:
The MEDINAUT Kyphoplasty System is designed to reduce spinal compression fracture and restore sagittal alignment. By creating a space in the vertebral body it facilitate the insertion of bone cement through the use of the cement dispensing plunger. The benefits of MEDINAUT Kyphoplasty System are the reduction in back pain and increase of patient's functional abilities, allowing a return to the previous level of activity.
It consists of Balloon Catheter, Bone Expander Syringe, Needle Pipe, Needle Pin, Wire Pin (Troca Type, Round Type), Cannula, Expander, Spacer, Cement Pusher, Cement Filler, and Guide Wire. It is supplied sterile and for single disposable use.
The MEDINAUT Kyphoplasty System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. The system is substantially equivalent in design, function and intended use to the predicate devices.
## 8. Indication for use:
The MEDINAUT Kyphoplasty System is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
## 9. Bench Test Data:
The MEDINAUT Kyphoplasty System complies with the internal quality control procedure of IMEDICOM Co., Ltd. and the characteristics are compared with Kyphon Inflatable Bone Tamp.
| Test | Test Criteria | Test Result | |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | MÉDINAUT<br>Kyphoplasty system | Kyphon<br>Inflatable Bone Tamp |
| Tensile Bond<br>Strength | Outer shaft and Hub with the<br>1mm/sec<br>Test report#; IMT-TBST12-01 | IBE-15; 20.85kgf | K13A; 14.76kgf |
| Fatigue<br>strength | No burst and leakage at 15 cycles<br>of maximum pressure limit<br>400psi<br>Test report#; IMT-FST12-01 | IBE-10; Pass<br>IBE-15; Pass<br>IBE-20; Pass | K13A; Pass<br>K09A; Pass<br>K08A; Pass |
| Test | Test Criteria | Test Result | |
| | | MEDINAUT<br>Kyphoplasty system | Kyphon<br>Inflatable Bone Tamp |
| Balloon<br>deflation time | Compare the performance<br>(seconds)<br>Test report:IMT-BDT12-01 | IBE-10 2ml; 0.50 sec. | K13A 2ml; 0.27 sec. |
| | | 3ml; 0.76 sec. | 4ml; 0.62 sec. |
| | | IBE-15 3ml; 0.51 sec. | K09A 2ml; 0.28 sec. |
| | | 5ml; 1.16 sec. | 4ml; 0.62ml |
| | | IBE-20 3ml; 0.47 sec. | K08A 2ml; 0.31 sec. |
| | | 5ml; 1.12 sec.<br>7ml; 1.29 sec. | 4ml; 0.54 sec.<br>6ml; 0.69 sec. |
| Burst<br>pressure<br>constrained | 400 psi at the Constrained jig<br>for 30 seconds<br>Test report: IMT-BPC12-08 | IBE-10; 3.5ml Pass<br>IBE-15; 5.5ml Pass<br>IBE-20; 7.5ml Pass | K09A; 5.5ml Pass<br>K08A; 7.5ml Pass |
| Burst strength<br>unconstraint | Measure the burst inflation<br>pressure<br>Test report#: IMT-BST12-01 | IBE-10; 250psi<br>IBE-15; 207psi<br>IBE-20; 183psi | K09A; 243psi<br>K08A; 237psi |
| Balloon<br>dimension<br>before and<br>after the<br>inflation | Initial balloon length<br>IBE-10; 10mm<br>IBE-15; 15mm<br>IBE-20; 20mm<br>Test report#: IMT-IDT11-02 | Balloon Diameter after<br>inflation<br>IBE-10, 3ml inflation;<br>14.1mm<br>IBE-15, 5ml inflation;<br>17.1mm<br>IBE-20, 7ml inflation;<br>19.3mm<br>Balloon Length after<br>inflation<br>IBE-10, 3ml inflation;<br>16.3mm<br>IBE-15, 5ml inflation;<br>22.2mm<br>IBE-20, 7ml inflation;<br>29.4mm | Balloon Diameter after<br>inflation<br>K09A; 4ml inflation;<br>15.2mm<br>K08A, 6ml inflation;<br>16.8mm<br>Balloon Length after<br>inflation<br>K09A, 4ml inflation;<br>20.1mm<br>K08A, 6ml inflation;<br>29.2mm |
| Insertion and<br>withdrawal<br>force | Measure the insertion and<br>withdrawal force with the<br>0.01m/s of test velocity.<br>Test report#: IMT-IFWF12-01 | Insertion Force<br>IBE-10; MEAN 1.24N<br>IBE-15; MEAN 1.65N<br>IBE-20; MEAN 2.11N<br>Withdrawal Force<br>IBE-10; MEAN 1.11N<br>IBE-15; MEAN 1.42N<br>IBE-20; MEAN 1.79N | Insertion Force<br>K09A; MEAN 1.08N<br>K08A; MEAN 1.49N<br>Withdrawal Force<br>K09A; MEAN 0.90N<br>K08A; MEAN 1.26N |
{2}------------------------------------------------
··
:
.
> :
.
...
{3}------------------------------------------------
#### 10. Comparison to the Predicate Device
The MEDINAUT Kyphoplasty System has the same device characteristics as the predicate device, The KyphX Inflatable Bone Tamp System; intended use, materials, design and use concept, sterilization, etc. Only the technical specifications - shape, diameter and length, tension and pull out force - are slightly different.
| Product Name | MEDINAUT Kyphoplasty System | KyphX Inflatable Bone Tamp |
|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | N/A | K010246, K041454 and K981251 |
| Manufacturer | IMEDICOM Co., Ltd. | Kyphon Inc. (Medtronic) |
| Indication for use | The MEDINAUT Kyphoplasty<br>System is intended to be used<br>for the reduction and fixation of<br>fractures and/or creation of a void<br>in cancellous bone in the spine.<br>This includes use during<br>percutaneous vertebral augmentation.<br>This system is to be used with<br>cleared spinal PMMA bone cements<br>indicated for percutancous vertebral<br>augmentation, such as kyphoplasty. | KyphX Inflatable Bone Tamps are<br>intended to be used as conventional<br>bone tamps for the reduction of<br>fracture and/or creation of a void in<br>cancellous bone in the<br>spine(including use during balloon<br>kyphoplasty with KyphX HV-RTM<br>Bone Cement), hand, tibia, radius<br>and calcaneus. |
| Articles<br>Max. Inflation Volume<br>Initial Length(mm)<br>Inflated Length(mm)<br>Inflated Diameter(mm) | IBE-10, IBE-15, IBE-20<br>3cc, 5cc, 7cc<br>10, 15, 20<br>16, 22, 34<br>14, 17, 19 | K08A, K09A, K13A<br>4cc, 4cc, 6cc<br>10, 15, 20<br>17.3, 19.7, 28.2<br>15.6, 14.3, 15.8 |
| Accessories kit | Needle Pipe,<br>Needle Pin,<br>Wire Pin,<br>Cannula,<br>Expander,<br>Spacer,<br>Cement Pusher,<br>Cement Filler and Guide Wire | Bone Access Needle,<br>Introducer System,<br>Precision Drill,<br>Cannula,<br>Expander,<br>Spacer,<br>Cement Pusher,<br>Cement Filler and Guide Wire |
| Composition of Material<br>Balloon Tip<br>Radiopaque Marker<br>Expander Syringe Body<br>Accessory Kit | Thermoplastic Polyurethane<br>Platinum<br>Polycarbonate and ABS<br>Stainless Steel and ABS | Thermoplastic Polyurethane<br>Platinum<br>Polycarbonate and ABS<br>Stainless Steel and ABS |
| Packaging | Pouch, Tyvek Blister Tray,<br>Cardboard Box | Pouch, Tyvek Blister Tray,<br>Cardboard Box |
| Sterilization | Gamma Sterilization | Gamma Sterilization |
| Biocompatibility | Meets ISO 10993 | Meets ISO 10993 |
#### ・11. Testing and Biocompatibility
Testing performed including functional testing, such as insertion and withdrawal force, puncture force, burst and simulated use which demonstrated that the MEDINAUT Kyphoplasty System is equivalent to the predicates in specifications and performance characteristics. Biocompatibility testing confirmed that the device meets the applicable requirements of the FDA Blue Book Memorandum #95-1 entitled Use of International Standards ISO 10993 Biological Evaluation of Medical Devices Part -1: Evaluation and Testing, and is biocompatible.
{4}------------------------------------------------
#### 12. Contraindications
Instability of posterior wall and/or pedicles
Infection
Severe bleeding
Known allergies to bone cement
Pregnancy
Fractures in which more than 68% of vertebral height is lost
Should not be used if vertebral dimensions or fracture pattern do not allow safe placement and inflation of the balloon
Instability of posterior wall and/ or pedicles
Any known severe allergy to contrast media
#### 13. Warnings
- For a transpedicular approach, if the pedicle is not large enough or stable enough to . withstand the procedure, pedicle fracture may occur.
- Complications that may occur during a parapedicular approach include pneumothorax and . bleeding
- Avoid contract between the balloon and bone cement .
- The balloon component of the Vertebral Balloon may fail due to bone splinters and/or . surgical tool contact
- Do not inflate the balloon until it has been fully deployed in the vertebral body. Inflating the . balloon prior to full deployment may result in premature balloon failure due to contact between the balloon and the access cannula.
- Do not use this product after the expiration date printed on the package. The device may not . be safe or effective beyond its expiration date.
- Do not pressurize more than 350PSI. .
- Do not put contrast media more than the specified Maximum volume: 10mm(3cc), . 15mm(5cc), 20mm(7cc)
- Inflating the vertebral balloon beyond the maximum inflation volume may cause the . balloon to rupture before reaching the maximum inflation pressure.
- Inflating the vertebral balloon beyond the maximum inflation pressure may cause the . balloon to rupture before reaching the maximum inflation volume.
#### Precautions 14.
- · It is important to read the instructions for use these precautions prior to device operation.
- · Use the Balloon Catheter prior to the use by date noted on the package.
- · Do not use damaged products. Before use, inspect the Balloon Catheter and packaging to verify that no damage has occurred.
- · Prior to use, the Balloon Catheter should be examined to verify functionality and ensure that its size is suitable for the specific procedure for which it is to be used.
- · Do not use this product if you have not been properly trained. The Balloon Catheter should only be used by physicians who are trained in the techniques of bone tanip use. Physicians using the devices should be familiar with the physiology and pathology of the selected anatomy.
- · The Balloon Catheter should be manipulated only while under fluoroscopic observation with radiographic equipment that provides high quality images like C-arm.
- The Balloon Catheter should only be inflated using an inflation syringe having a 20ml volume capacity.
{5}------------------------------------------------
- · Only inflate the Balloon Catheter with liquid contrast medium a 60% solution is recommended.
- · Follow manufacturer's instructions for contrast medium indications, usage and cautions.
- · Do not use air or other gas to inflate the Balloon Catheter.
- · The inflatable component of the Balloon Catheter may fail due to contact with bone splinters, bone cement and/or surgical tools.
- · The inflation characteristics of the Balloon Catheter are altered by inflation inside bone.
- · Do not re-sterilize and/or reuse. The Balloon Catheter is for single use only.
- · Reconditioning, refurbishing, repair, modification, or resterilization of the device to enable further use is expressly prohibited.
#### 15. Mandatory Performance Standard
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices.
#### 16. Voluntary Standards
.
The MEDINAUT Kyphoplasty System was developed under the auspices of the following Standards.
| Standard | Application |
|--------------------|--------------------------------------------------------------------------------------|
| ISO 10993-1[2009] | Guidance on the selection of biocompatibility testing |
| ISO 10993-4[2002] | Applied to the biological evaluation tests for interactions<br>with blood |
| ISO 10993-5[2009] | Applied to the biological evaluation tests for In Vitro<br>cytotoxicity |
| ISO 10993-10[2010] | Applied to the intracutaneous reactivity test & skin<br>sensitization test |
| ISO 10993-11[2006] | Applied to tests for systemic toxicity |
| ISO 11137-1:2006 | |
| ISO 11137-2: 2012 | |
| ISO 11137-3:2006 | Applied to sterilization validation |
| ISO 11607-1:2006 | |
| ISO 11607-2: 2006 | |
| ISO 14971:2007 | Applied to risk management |
| ISO 9001:2008 | Applied to the certification of the quality management system<br>of Medicom Co., Ltd |
| ISO 13485:2003 | |
| ASTM F 1980-02 | Applied to accelerated aging testing |
| ISO 11737-1:2006 | |
| ISO 11737-2: 2009 | Applied to sterilization validation |
| ISO 11737-3:2004 | |
| EN 556-1:2001 | |
#### 17. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification IMEDICOM Co., Ltd. submits that the MEDINAUT Kyphoplasty System in this submittal is substantially equivalent to the predicate System, the KyphX Inflatable Bone Tamp System, as described within this submittal.
{6}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with flowing lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2014
IMEDICOM Co., Ltd. % Ms. Priscilla Chung Consultant/Official Correspondent LK Consulting Group USA, Incorporated 2651 East Chapman Avenue, Suite 110 Fullerton, California 92833
Re: K133669
Trade/Device Name: MEDINAUT Kyphoplasty System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: February 11, 2014 Received: February 18, 2014
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
Page 2 - Ms. Priscilla Chung
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
K133669 - Page 1 of 1
## Indications for Use
510(k) Number (if known): K 133669
Device Name: MEDINAUT Kyphoplasty System
Indications for Use: The MEDINAUT Kyphoplasty System is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. This system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
# Laurence D. Coyne -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K133669
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.