K133544 · Palodex Group OY · OAS · Mar 26, 2014 · Radiology
Device Facts
Record ID
K133544
Device Name
OP300
Applicant
Palodex Group OY
Product Code
OAS · Radiology
Decision Date
Mar 26, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examination of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as x-ray projection images of an examined volume for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.
Device Story
OP300 is an extra-oral dental X-ray system producing 2D panoramic, cephalometric, and TMJ images, and 3D cone beam CT volumes. System uses CMOS flat panel detectors to capture X-ray projections during rotation around patient head. Reconstruction software (FBP or ART) processes projections into 3D views. Operated by qualified healthcare professionals in clinical settings. Modified version introduces medium-size CMOS sensor (12x12), expanded FOV options (up to 130x150mm), and modified beam-limiting aperture plate. Includes low-radiation dose mode for follow-up, adjusting voxel size as input parameter. GUI updated for new FOV selections. Output viewed via 3D software to assist dental diagnosis and treatment planning.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing comparing image quality of the modified device against the predicate using anthropomorphic phantoms. Testing confirmed no significant differences in image quality and successful validation of software modifications.
Indicated for dental radiographic examination of teeth, jaw, and TMJ areas in patients requiring 2D panoramic, cephalometric, or 3D cone beam computed tomography imaging. For use by qualified healthcare professionals only.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
**MAR 2 6 2014**
Page 1 of 4
K133544
# 510(k) Summary
Date: October 25th, 2013 Revised: 4th Febuary, 2014
## Manufacturer:
PaloDEx Group Ov Nahkelantie 160 Tuusula, Finland 04300
Tel: +358 10 270 2000 Fax: +358 9 701 5263
Contact person: Mr. Terho Turkumäki, Tel +358 50 320 9113
# Trade Name:
OP300
## Common Name:
Dental panoramic, cephalometric and cone beam computed tomography x-ray device
## Classification Name:
x-ray, tomography, computed, dental (21 CFR § 892.1750, product code OAS)
#### Description:
The Orthopantomograph OP300 is an extra oral source dental x-ray device that is softwarecontrolled which produces conventional digital 2D panoramic, cephalometric and TMJ x-ray images as well as digital x-ray projection images taken during cone beam rotations around a patient's head. The projection images are reconstructed to be viewed in 3D by a 3D viewing software.
#### Indication for Use:
The OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examination of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as x-ray projection images of an examined volume for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.
#### Intended Use:
The intended use of the OP300 is for dental radiographic examination of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as x-ray projection images of an examined volume for the reconstruction of a 3D view.
{1}------------------------------------------------
# Summary of Technological Characteristics:
OP300 is substantially equivalent in design, composition and function to the current OP300 unit.
| | Concept | OP300 (K122018) | OP300 (Modified) |
|----|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| 1. | Indications for use | The OP300 dental panoramic,<br>cephalometric and cone beam<br>computed tomography x-ray device is<br>intended for dental radiographic<br>examination of teeth, jaw and TMJ<br>areas by producing conventional 2D x-<br>ray images as well as x-ray projection<br>images of an examined volume for the<br>reconstruction of a 3D view. The<br>device is operated and used by<br>qualified healthcare professionals. | Same |
| 2. | Imaging modes | Panoramic, Cephalometric, TMJ, 3D | Same |
| 3. | X-ray source | 3D mode: 90kV<br>Pan mode: 57-90 kV<br>Ceph mode: 60-90 kV<br>kV accuracy: +/-5kV<br>mA range: 3.2-16 mA<br>3D power mode: pulsed | Same |
| 4. | Focal spot | 0.5mm | Same |
| 5. | Image detector(s) | CMOS Flat Panel<br>+ CMOS for pan/ceph imaging | Same |
| 6. | 3D imaging technique | Reconstruction from 2D images | Same |
| 7. | 3D's Field Of View | 61 x 41 mm<br>61 x 78 mm | 50 x 50 mm<br>61 x 78 mm<br>78 x 78 mm<br>78 x 150 mm<br>130 x 150 mm |
| 8. | 3D's total viewing angle | 200 degrees | Same |
| 9. | Pixel size | CMOS flat panel for 3D: 200 μm<br>CMOS for panoramic imaging: 100 µm | Same |
| 10 | Voxel size | 80-350 µm | 80-600 µm |
| 11 | Reconstruction Software | Filtered Back Projection (FBP) or<br>Algebraic Reconstruction Technique<br>(ART) | Same |
| 12 | 3D's effective exposure time | 2 - 20 sec | Same |
| 13 | 3D Reconstruction Time | 1-3 min | Same |
| 14 | Patient's Position | Standing and wheelchair | Same |
| 15 | System footprint | H161-241cm x D1390cm x W97-193<br>cm | Same |
| 16 | Weight | Pan/3D 205 kg | Same |
{2}------------------------------------------------
The differences between the Predicate Device OP300 (K122018), and the modified device are the following:
New panel and Field of Views: The small size 3D CMOS flat panel sensor (10x7) used in the Predicate Device OP300 (K122018) is replaced with a medium size CMOS flat panel sensor (12x12), This brings new bigger Field of View (FOV) sizes for the user. The new 3D programs to support the new FOV -sizes are: 50 x 50 mm. 61 x 78 mm (as in the Predicative Device), 78 x 78 mm. 78 x 150 mm and 130 x 150 mm.
Modified beam-limiting device is introduced to support the new FOV -sizes. Sliding aperture plate has been modified to consist of 3 apertures to allow the usage of new FOVs, instead of 1 aperture in the Predicative Device. Control principle remains the same as the predicate device.
The GUI software that is operated through the touch screen control panel has been modified to incorporate the changes to the imaging programs. Each new FOV -size has a new selection button in the updated GUI.
Following the ALARA (As Low As Reasonably Achievable) principle, a very low radiation dose mode for 3D imaging is introduced. The low radiation dose mode can be used for example in treatment follow-up cases where the most accurate image quality is not needed. As the radiation dose is lower, the voxel size needs to be bigger (lower resolution) in order to keep the dose per voxel in adequate level. Voxel size is an input parameter for the 3D reconstruction calculation.
Design verification and validation has been performed to ensure the safety and effectiveness of the device.
#### Performance (bench) Test Data:
In- house Performance (bench) testing has been conducted to compare the image quality and the sensor performance of the proposed OP300 and the predicate OP300 according to the recommendations in the "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices". As a result of the performance test it was concluded that there is no significant differences in image quality between the proposed and the predicate device.
The modified OP300 has been tested against FDA Recognized Consensus Standards (IEC60601-1:1988, IEC60601-1-2:2001, IEC 60601-1-3:1994, IEC60601-1-4:1996, IEC 60601-2-7:1998, IEC 60601-2-28:1993 and IEC 60601-2-32:1994). The standards used are the same standards as with the Predicate Device.
Images of an anthropomorphic phantom image for the 3D imaging mode were evaluated for the OP300 device and the predicate to demonstrate the modified device was capable of producing images without severe defects.
The modified OP300 has been successfully verified and validated to ensure the safety and effectiveness of the device.
{3}------------------------------------------------
K133544
Page 4 of 4
## Clinical Test Data:
Sample clinical images of patients were not used to support substantial equivalence of the OP300 device because the primary change to the imaging chain was a change in the detector size (other detector characteristics remained the same); the bench performance testing demonstrated that the new detector options had equivalent laboratory image quality performance to the predicate.
#### Conclusion:
Based upon the similar technological/performance characteristics to the predicate device and the successful validation of the OP300 software, the clinical performance of the OP300 is deemed to be substantially equivalent to the predicate devices.
The OP300 has been successfully verified and validated to ensure the safety and effectiveness of the device.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 26, 2014
PaloDex Group Oy % Mr. Terho Turkumäki OA&RA Manager Nahkelantie 160 O4300 Tuusula FINLAND
Re: K133544 Trade/Device Name: OP300 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: February 20, 2014 Received: February 24, 2014
Dear Mr. Turkumäki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Mr. Turkumäki
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.lda.gov/Medicall.cevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K133544
Device Name: OP300
Indications for Use:
The OP300 dental panoramic, cephalometric and cone beam computed tomography x-ray device is intended for dental radiographic examination of teeth, jaw and TMJ areas by producing conventional 2D x-ray images as well as x-ray projection images of an examined volume for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$(m,h,7)\$
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> K | 33544 510(k) =======================================================================================================================================================================
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.