be placed on an impression tray(or infected directly into the mouth, depending on the technique and device)and used to reproduce the structure of a patient's teeth and gums. provide models for study and for production of restorative prosthetic devices.
Device Story
Suflex Impression Materials are addition-curing, hydrophilic, elastomeric dental materials; available in four viscosities (putty, heavy, mono, light). Used by dental professionals to capture precise anatomical structures of teeth and gums; applied via impression trays or injected directly into the mouth. Materials facilitate creation of dental models for study and fabrication of restorative/prosthetic devices (crowns, bridges, dentures, implants). Delivery systems include 1:1 50ml automix cartridges, 5:1 362ml foil bags for automatic dispensing, and traditional putty jars. High tear strength and dimensional accuracy ensure reliable reproduction of oral structures for clinical decision-making in restorative dentistry.
Clinical Evidence
No clinical studies submitted. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization/irritation). Bench testing performed per ISO 4823 standards for mechanical properties including working time, consistency, detail reproduction, linear dimensional change, elastic recovery, and strain-in-compression.
Technological Characteristics
Addition-curing, hydrophilic, elastomeric impression material. Viscosities: putty, heavy, mono, light. Delivery: 1:1 50ml automix cartridges, 5:1 362ml foil bags, or putty jars. Complies with ISO 4823 for dental impression materials. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for dental patients requiring impressions for fixed/removable restorations, implants, crown/bridgework, inlays, full/partial dentures, relines, and functional impressions using various putty, tray, monophase, and contact techniques.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
K133527
MAR - 7 2014
Image /page/0/Picture/2 description: The image shows the word "OSSTEM" in bold, black letters. Below the word "OSSTEM" is the word "IMPLANT" in smaller, thinner, black letters. There is a degree symbol to the right of the word "OSSTEM". The background is white.
# Osstem Germany GmbH
#Mergenthaler Allee 25, 65760 Eschborn, Germany Tel. +49(0)6196 777 550 · Fax. +49 (0}6196 777 55 06 www.osstem.de, www.sinuskit.com
Osstem Germany GmbH.
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Jessica Jung
+49 6196 777 550
HIOSSEN Inc.
Patrick Lim
ELW
888 678 0001
Date: July 12, 2013
1. Company and Correspondent making the submission:
- Submitter's Name :
- Address
- Contact :
- Phone:
- Correspondent's Name:
- Address:
- Contact:
- Phone:
2. Device :
Trade or (Proprietary) Name :
Common or usual name :
Classification Name :
Suflex Impression Materials to include: Suflex putty Suflex Heavy Suflex Mono Suflex Light Impression material Material, Impression(21 CFR 872.3660.Product) 21CFR872.3640 Class II
85 Ben Fairless Dr. Fairless Hills, PA 19030
#Mergenthaler Allee 25, 65760 Eschborn, Germany
3. Predicate Device: Trade Name: Trade Name: Trade Name:
Trade Name:
Trade Name:
Panasil Impression Materials(K082560) Panasil Impression Materials(K954282) Aquasil Ultra Rigid Smart Wetting Impression Material (K021413) Aquasil Ultra XLV Smart Wetting Impression Material (K021410) Kerr VPS Impression Material(K050604)
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Osstem Implant. The logo is in black and is a stylized font. The word "OSSTEM" is on the top line and the word "IMPLANT" is on the bottom line.
# Osstem Germany GmbH.
#Mergenthaler Allec 25, 65760 Eschborn, Germany Tel. +49(0)6196 777 550 · Fax. +49 (0)6196 777 55 06 www.osstem.de. www.sinuskit.com
#### 4. Description:
Suflex Impression Materials are addition-curing, elastomeric materials with hydrophilic properties, high tear strength, dimensional accuracy, and resistance to permanent deformation. The Suflex Impression Material family consists of four different viscosities(putty, heavy, mono, light), available in an assortment of delivery systems: traditional 1:1 50ml, automix cartridge, 5:1 362ml foil bags for use in most automatic dispensing and mixing systems, and traditional 1:1 putty jars.
#### 5. Intended use
- be placed on an impression tray(or infected directly into the mouth, depending on the technique and device)and used to reproduce the structure of a patient's teeth and gums.
- provide models for study and for production of restorative prosthetic devices.
#### 6. Indication for use
Suflex Putty is to be used as panasil putty material for;
- Two step putty impression technique l
- One step putty impression technique
- Two step putty impression technique using a foil(plastic putty spacer) .
- One step putty impression technique for forming functional peripheries. -
Suflex Heavy is to be used a s panasil tray material for;
- One step impressiontechnique using sinte or dual viscosities. ।
- Two step impression technique using dual viscosities ।
- Functional impressions. -
Suflex Mono is to be used as panasil monophase material for;
- Taking impression over fixed/removable restorations and implants -(transferring impression posts and bridge components)
- -Fuctional impresssions.
- Fabricating crown and bridgework or inlays. .
- Fabricating full or partical dentures. .
- Reline impressions. י
- Use in the simultaneous mixing technique as well as the putty -
- Trnasferring root posts when fabricating posts and cores indirectly. .
Suflex Light is to be used as panaasil contact material for;
- Two step putty impressiontechnique. ।
- One step putty impression technique. ।
- One step impression technique using a foil(plastic putty spacer) l
- One step impression technique(simultaneous technique)using dual viscosities. .
- -Reline impressions.
- Fabricating full or partial dentures. -
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for OSSTEM IMPLANT. The word "OSSTEM" is written in a bold, sans-serif font, with the word "IMPLANT" written in a smaller font below it. There is a degree symbol to the right of the word "OSSTEM".
## Osstem Germany GmbH.
#Mergenthaler Allee 25, 65760 Eschborn, Germany Tel. +49(0)6196 777 550 - Fax. +49 (0)6196 777 55 06 www.osstem.de, www.sinuskit.com
### 7. Summary of Non-Clinical Performance Testing
Biocompatibility tests have been performed to assure biological safety in accordance with the ISO 10993 family. Tests in respect to cytotoxicity(ISO 10993-5), sensitization and mucosa irritation(ISO 10993-10) and a chemical analysis showed, that Suflex materials' biocompatibility data is comparable to other materials on the market. Additionally bench testing was performed to allow an evaluation of the mechanical properties of Sulex in comparison to already marketed products. The evaluation covers
Component colours (ISO 4823, 6.1)
- Working time (ISO 4823, 6.3)
- | Compatibility with gypsum (ISO 4823, 6.4)
- Consistency(ISO 4823, 6.4)
- Detail reproduction(ISO 4823, 6.4)
- Linear dimensional change(ISO 4823, 6.4)
- Elastic recovery(ISO 4823, 6.4)
- Strain-in-compression(ISO 4823, 6.4)
The results indicate adequate consistency results compared to leading Impression materials.
### 8. Summarv of clinical testing
No clinical studies are submitted
#### 9. Conclusions
Osstem Germany GmbH believes that the Suflex Impression material is substantially equivalent to the currently legally marketed products. It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or safety risks.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of a human figure with outstretched arms, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2014
Osstem Germany GmbH C/O Mr. Patrick Lim QA/RA Manager HIOSSEN, Inc. 85 Ben Fairless Drive Fairless Hills, PA 19030
Re: K133527
Trade/Device Name: Suflex Impression Materials Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: November 27, 2013 Received: December 11, 2013
Dear Mr. Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicaiDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image shows the name "Erin F. Keith-S" in a stylized font. The name is written in black ink and is easily readable. The letters are bold and have a slightly rounded appearance. The overall impression is that of a professional and well-designed logo or nameplate.
Erin I. Keith, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for OSSTEM IMPLANT. The word "OSSTEM" is in a bold, sans-serif font, with the letters slightly overlapping each other. The word "IMPLANT" is in a smaller, sans-serif font and is located directly below "OSSTEM". There is a degree symbol to the right of the word "OSSTEM".
### Osstem Germany GmbH.
#Mergenthaler Allec 25, 65760 Eschborn, Germany Tel. +49(0)6196 777 550 - Fax. +49 (0)6196 777 55 06 www.osstem.de. www.sinuskit.com
510(k) Number K _133527
Device Name : Suflex impression materials
Indication for use :
Suflex Putty is to be used as panasil putty material for;
- Two step putty impression technique 。
- One step putty impression technique -
- Two step putty impression technique using a foil(plastic putty spacer) l
- One step putty impression technique for forming functional peripheries. .
Suflex Heavy is to be used a s panasil tray material for;
- One step impression technique using single or dual viscosities. ,
- Two step impression technique using dual viscosities .
- Functional impressions. -
Suflex Mono is to be used as panasil monophase material for;
- Taking impression over fixed/removable restorations and implants . (transferring impression posts and bridge components)
- Functional impressions. .
- -Fabricating crown and bridgework or inlays.
- Fabricating full or partial dentures. 1
- Reline impressions. ।
- Use in the simultaneous mixing technique as well as the putty .
- Transferring root posts when fabricating posts and cores indirectly. -
Suflex Light is to be used as panasil contact material for;
- Two step putty impression technique. .
- One step putty impression technique. -
- One step impression technique using a foil(plastic putty spacer) -
- One step impression technique (simultaneous technique) using dual viscosities. -
- Reline impressions. -
- Fabricating full or partial dentures. .
Prescription Use X (Per 21CFR801 Subpart D)
Over-The-Counter Use OR (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runner -5
FDA 2024.09.06
07:56:58-05'00'
1 / 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.