PANASIL IMPRESSION MATERIALS, INCLUDING PUTTY, BINETICS, TRAY, MONOPHASE AND CONTACT
Applicant
Kettenbach GmbH & Co. KG
Product Code
ELW · Dental
Decision Date
Sep 8, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Panasil® Impression Materials are intended to: be placed on an impression tray (or injected directly into the mouth, depending on the technique and device) and used to reproduce the structure of a patient's teeth and gums; provide models for study and for production of restorative prosthetic devices.
Device Story
Panasil® Impression Materials are addition-curing, elastomeric dental materials; available in five viscosities (putty, heavy, medium, light, x-light). Used by dental professionals in clinical settings to capture precise anatomical structures of teeth and gums. Materials are applied via impression trays or direct intraoral injection using automix cartridges or automatic dispensing systems. The material cures to form a stable, dimensionally accurate negative replica of the oral cavity. This replica is used to create physical models for study or as a foundation for fabricating prosthetic restorations (crowns, bridges, dentures, implants). Benefits include high tear strength and resistance to deformation, ensuring accurate fit of final dental prosthetics.
Clinical Evidence
Bench testing only. Evaluated per FDA Guidance for Industry: Dental Impression Materials and ISO 4823:2000/Cor 1:2004/Amd 1:2007. Testing confirmed suitability for use as dental impression materials and performance equivalent to predicate devices.
Technological Characteristics
Addition-curing elastomeric impression materials. Hydrophilic properties, high tear strength, dimensional accuracy. Available in five viscosities. Delivery systems: 1:1 50 ml automix cartridges, 5:1 362 ml foil bags, and putty jars. Complies with ISO 4823 standards.
Indications for Use
Indicated for dental patients requiring impressions for study models, restorative prosthetic devices, crown/bridgework, inlays, full/partial dentures, relines, and implant-related impressions. Used in various putty/wash, simultaneous, and functional impression techniques. Prescription use only.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Traditional Premarket Notitication [510(K)] Panasil® Impression Materials
ບບການປະເທດລ
K082560
### 510(K) Summary
SEP 0 8 2008
| | SEP 08 2008 |
|--------------------------|-------------------------------------------------------------------|
| A. Submitter Information | |
| Submitter's Name: | Kettenbach GmbH & Co. KG |
| Address: | Im Heerfeld 7 |
| | D-35713 |
| | Eschenburg, Germany |
| Phone Number: | (+49) 2774-705-58 |
| Fax Number: | (+49) 2774-705-33 |
| Contact Person: | Michaela Zinke |
| Date of Preparation: | July 10, 2008 |
| B. Device Name | |
| Trade Name: | Panasil® Impression Materials, to include: |
| | • Panasil® putty (putty, putty fast set, putty soft) |
| | • Panasil® binetics (putty fast, putty soft) |
| | • Panasil® tray (fast, soft) |
| | • Panasil® monophase |
| | • Panasil® contact (two in one, plus) |
| Common/Usual Name: | Impression Material |
| Classification Name: | Material, Impression (21 CFR 872.3660, Product Code: ELW) |
| C. Predicate Devices | |
| Trade Name: | Panasil® Impression Materials (K954281) |
| Trade Name: | Aquasil Ultra Rigid Smart Wetting® Impression Materials (K021413) |
| Trade Name: | Aquasil Ultra XLV Smart Wetting® Impression |
Kerr VPS Impression Material (K050604)
Material (K021410)
Trade Name:
. .
{1}------------------------------------------------
#### D. Device Description
Panasil® Impression Materials are addition-curing, elastomeric materials with hydrophilic properties, high tear strength, dimensional accuracy, and resistance to permanent deformation. The Panasil Impression Material family consists of five different viscosities (putty, heavy-bodied, medium-bodied, light-bodied, x-lightbodied), available in an assortment of delivery systems: traditional 1:1 50 ml automix cartridge, 5:1 362 ml foil bags for use in most automatic dispensing and mixing systems, and traditional 1:1 putty jars.
### E. Intended Use
The Panasil® Impression Materials are intended to:
- be placed on an impression tray (or injected directly into the mouth, depending on the technique and device) and used to reproduce the structure of a patient's teeth and gums;
- provide models for study and for production of restorative prosthetic devices.
### Indications for Use
Panasil putty (putty, putty fast set, putty soft) and Panasil binetics (putty fast, putty soft) are to be used as preliminary materials for:
-Two step putty/wash impression technique.
-One step putty/wash impression technique.
- -Two step putty/wash impression technique using a foil (plastic putty spacer)
- -One step putty impression technique for forming functional peripheries.
Panasil trav (fast. soft) is to be used as a heavy-bodied material for:
- -One step impression technique (simultaneous technique) using single or dual viscosities.
- -Two step impression technique using dual viscosities.
- -Functional impressions.
{2}------------------------------------------------
Panasil monophase is to be used as a medium-bodied tray or syringeable impression material for:
- -Taking impressions over fixed/removable restorations and implants
- (transferring impression posts and bridge components).
- -Functional impressions.
- -Fabricating crown and bridgework or inlays.
- -Fabricating full or partial dentures.
- -Reline impressions.
- -Use in the simultaneous mixing technique as well as the putty/wash and triple tray techniques.
- -Transferring root posts when fabricating posts and cores indirectly.
Panasil contact (two in one, plus) is to be used as a syringeable impression material for:
- -Two step putty/wash impression technique.
- -One step putty/wash impression technique.
- -One step impression technique using a foil (plastic putty spacer)
- -One step impression technique (simultaneous technique) using dual viscosities
- -Reline impressions.
- -Fabricating full or partial dentures.
### F. Technological Characteristics Summary
The technological characteristics of Panasil® Impression Materials are substantially equivalent to the predicate device technological characteristics. Panasil Impression Materials (Panasil putty, Panasil binetics, Panasil tray, Panasil monophase, Panasil contact) and the predicate devices are addition-curing, elastomeric materials designed and manufactured for use as dental impression materials.
{3}------------------------------------------------
#### G. Performance Data
No performance standards have been established for this type of device. Panasil® Impression Materials have been evaluated in accordance with the applicable criteria established in Guidance for Industry and FDA Staff: Dental Impression Materials – Premarket Notification (FOD#2203, 8/17/1998) and ISO 4823 (Dentistry – Elastomeric impression materials):2000/Cor 1:2004/Amd 1:2007. The results of device performance testing demonstrated that Panasil Impression Materials (Panasil putty, Panasil binetics, Panasil tray, Panasil monophase, Panasil contact) are suitable for use as dental impression materials. Panasil Impression Materials (Panasil putty, Panasil binetics, Panasil tray, Panasil monophase, Panasil contact) have been designed and manufactured to perform in a manner substantially equivalent to that of the predicate devices.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.
Public Health Service
SEP 0 8 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kettenbach GmbH & Company KG C/O Mr. Stefan Preiss Responsible Third Party Official TÜV SÜD America, Incorporated 1775 Old Highway 8 NW, Suite 104 New Brighton, Minnesota 55112-1891
Re: K082560
Trade/Device Name: Panasil® Putty (Putty, Putty Fast Set, Putty Soft), Panasil® Binetics (Putty Fast, Putty Soft), Panasil® Tray (Fast, Soft), Panasil® Monophase, Panasil® Contact (Two in One, Plus) Regulation Number: 872.3660 Regulation Name: Partially Fabricated Denture Kit Regulatory Class: II Product Code: ELW Dated: September 2, 2008 Received: September 4, 2008
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Preiss
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CI'R Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Thhammelb. Lerdiner frell
Chiu S. Lin. Ph. D Division Director Division of Anesthesiology, Gencral Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): _ KO22560
Device Name: Panasil® putty (putty, putty fast set, putty soft),
Indications for Use:
Panasil putty (putty, putty fast set, putty soft) is to be used as a preliminary material for:
- -Two step putty/wash impression technique.
- -One step putty/wash impression technique.
- -Two step putty/wash impression technique using a foil (plastic putty spacer)
- -One step putty impression technique for forming functional peripheries.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
| AND/OR | |
| Over-The-Counter Use | |
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | 1082560 |
|----------------|---------|
|----------------|---------|
Page 1 of 5
{7}------------------------------------------------
510(k) Number (if known): \$0825 60
Device Name: Panasil® binetics (putty fast, putty soft)
Indications for Use:
Panasil binetics (putty fast, putty soft) is to be used as a preliminary material for:
-Two step putty/wash impression technique.
-One step putty/wash impression technique.
-Two step putty/wash impression technique using a foil (plastic putty spacer)
- Two step putty/waon improberhnique for forming functional peripheries.
Prescription Use __ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sina Ruom
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 2 of 5
510(k) Number:
K082560
Page 17 of 510
{8}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Panasil® tray (fast, soft)
Indications for Use:
### Panasil tray (fast, soft) is to be used as a heavy-bodied material for:
- -One step impression technique (simultaneous technique) using single or dual viscosities.
- -Two step impression technique using dual viscosities.
- -Functional impressions.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices Page 3 of 5
510(k) Number:
Kattanhach GmhH & C.n KG
Page 18 of 510
{9}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Panasil® monophase
Indications for Use:
### Panasil monophase is to be used as a medium-bodied tray or syringeable impression material for:
-Taking impressions over fixed/removable restorations and implants
(transferring impression posts and bridge components).
-Functional impressions.
- -Fabricating crown and bridgework or inlays.
-Fabricating full or partial dentures.
-Reline impressions.
-Use in the simultaneous mixing technique as well as the putty/wash and triple tray techniques.
-Transferring root posts when fabricating posts and cores indirectly.
Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runar
Division Sign-Off) )ivision of Anesthesiology, General Hospital nfection Control, Dental Devices
Page 4 of 5
10(k) Number: K082560
{10}------------------------------------------------
510(k) Number (if known): _KOKAS (eC)
Device Name: Panasil® contact (two in one, plus)
Indications for Use:
Panasil contact (two in one, plus) is to be used as a syringeable impression material for:
- -Two step putty/wash impression technique.
- -One step putty/wash impression technique.
- -One step impression technique using a foil (plastic putty spacer)
- -One step impression technique (simultaneous technique) using dual viscosities
- -Reline impressions.
- -Fabricating full or partial dentures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| Division Sign-Off) | |
|----------------------------------------------|---------|
| Division of Anesthesiology, General Hospital | |
| nfection Control, Dental Devices | |
| 510(k) Number: | K082560 |
Page 5of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.