BARD RITECATH INTERMITTENT CATHETER

K133470 · C.R. Bard, Inc. · EZD · Apr 4, 2014 · Gastroenterology, Urology

Device Facts

Record IDK133470
Device NameBARD RITECATH INTERMITTENT CATHETER
ApplicantC.R. Bard, Inc.
Product CodeEZD · Gastroenterology, Urology
Decision DateApr 4, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Bard RiteCath Intermittent Urinary Catheter is intended for use by male and female patients for draining urine from the bladder.

Device Story

The Bard RiteCath Intermittent Urinary Catheter is a single-use, biocompatible PVC device designed for intermittent bladder drainage. It features a funnel, a shaft with two staggered eyelets, and a tip available in straight or coude configurations. The device is inserted through the urethra into the bladder, allowing urine to drain through the eyelets and shaft into the funnel. It is available in multiple French sizes (8–18 Fr.) and lengths. The device is ethylene oxide sterilized and intended for use by patients or clinicians to manage urinary retention or bladder emptying.

Clinical Evidence

Bench testing only. Functional performance testing conducted per BS EN 1616:1997 + A1:1999. Biocompatibility testing conducted per ISO 10993-1:2009 and FDA Bluebook Memorandum G95-1.

Technological Characteristics

Material: Polyvinyl chloride (PVC). Design: Funnel, shaft, staggered eyelets, straight or coude tip. Sizes: 8–18 Fr. Sterilization: Ethylene oxide per ANSI/AAMI/ISO 11135-1:2007. Single-use.

Indications for Use

Indicated for male and female patients requiring bladder drainage.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Summary Traditional 510(k) | Bard RiteCath Intermittent Catheter ## Bard Medical Division C.R. Bard. Inc. 8195 Industrial Blvd. Covington, GA 30014 # BARD | MEDICAL ## 510(k) Summary In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Bard RiteCath Intermittent Catheter 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005. Sponsor: BARD Medical Division C. R. BARD, Inc. 8195 Industrial Blvd. Covington, GA 30014 Establishment Registration Number: 1018233 Contact: Michele Davis, RAC Regulatory Affairs Project Manager Bard Medical Division Tel: 770-784-6274 Fax: 770-385-4706 Date: April 2, 2014 Subject Device: Trade Name: Bard® RiteCath™ Intermittent Urinary Catheter Common Name: Urological Catheter Classification Name: Urological catheter and accessories Regulation: 21 CFR 876.5130 Classification: II Primary Product Code: EZD Secondary Product Codes: EZC Legally marketed device to which substantial equivalence is claimed: Coloplast A/S Self Cath Catheter, K100878 . #### Device Description The Bard RiteCath Intermittent Urinary Catheter is a biocompatible, polyvinyl chloride (PVC) catheter used to drain urine from the bladder. The catheter consists of a funnel, shaft with two staggered eyelets and a tip. The tip is available in a straight or coude configuration. The tip of the catheter passes through the urethra into the bladder to allow urine to drain into the eyelets and then through the catheter shaft, exiting through the funnel. The catheter will be offered in multiple French sizes (8 – 18 Fr.), lengths (6" and two tip designs (straight and coude). The product is ethylene oxide sterilized (per ANSI/AAMI/ISO 11135-1:2007, Sterilization of health care products – ethylene oxide – Part 1: Requirements for development, validation ond routine control of a sterilized process for medical devices). The catheter is for single use. {1}------------------------------------------------ #### Intended Use The Bard RiteCath Intermittent Urinary Catheter is intended for use by male and female patients for draining urine from the bladder. ## Technological Characteristics The Bard RiteCath Intermittent Urinary Catheter has the same technological characteristics as the predicate device. The subject and predicate device are made from the same catheter material, polyvinyl chloride (PVC) and have the same catheter design consisting of color-coded funnel, shaft, staggered eyelets, and straight or coude tip. The subject and predicate device are available in multiple French sizes and lengths. ## Performance Data Nonclinical functional performance testing was performed per BS EN 1616: 1997 + A1:1999, Sterile urethral catheters for single use. Nonclinical biocompatibility testing was conducted in accordance with ISO 10993-1:2009, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and FDA Bluebook Memorandum G95-1, Use of International Standard ISO 10993 "Biological Evaluation of Medical Devices Part 1: Evaluation of Testing." #### Substantial Equivalence The Bard RiteCath Intermittent Urinary Catheter has the same design features and is indicated for the same use as the predicate device, Coloplast Self Cath Catheter, K100878. The subject device is substantially equivalent to the legally, marketed predicate device and nonclinical test data demonstrates that the subject device is safe and effective. {2}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 4, 2014 C. R. Bard, Inc. Michele Davis, RAC Regulatory Affairs Project Manager 8195 Industrial Blvd. Covington, GA 30014 K133470 Re: > Trade/Device Name: Bard RiteCath Intermittent Urinary Catheter Regulation Number: 21 CFR& 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD, EZC Dated: March 4, 2013 Received: March 6, 2014 Dear Michele Davis, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {3}------------------------------------------------ Page 2 - Michele Davis, RAC You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, BenjaminFisher-S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use Traditional 510(k) | Bard RiteCath Intermittent Catheter # Section 4 # Indications for Use Statement 510(k) Number: K133470 Bard RiteCath Intermittent Urinary Catheter Device Name: INDICATIONS FOR USE: The Bard RiteCath Intermittent Urinary Catheter is intended for use by male and female patients for draining urine from the bladder. Prescription Use X (Part 21 CFR 801 Subpart D) and/or Over-the-Counter Use (21 CFR 807 Subpart C) -Counter Use 307 Subpart C) ☐ (Please do not write below this line – continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/4/Picture/15 description: The image shows the word "Benjamin" on the top left corner. Below it, the date "2014.04" is written. There are some graphical elements on the right side of the image, but they are not clear enough to describe. Section 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...