Classic™ intermittent catheters are intended for use in male, female and paediatric patients when regular assisted bladder drainage has been instructed by a physician. Classic™ intermittent catheters are indicated for use by individuals whose ability to void the bladder fully by natural bodily functions is impaired.
Device Story
Classic™ Intermittent Catheter is a sterile, single-use PVC device for bladder drainage. Consists of a tube bonded to a drainage funnel; features a round polished tip and two large eyelets for efficient drainage. Includes a sliding tube grip to minimize user contact with the invasive tube, reducing infection risk. Available in various French sizes (6-20) and lengths (Male 400mm, Female 160mm, Paediatric 240mm) with straight or coudé tip options. Used by patients or caregivers in clinical or home settings to drain urine when natural voiding is impaired. Output is the physical drainage of urine into a collection bag or receptacle. Benefits include improved bladder management and reduced risk of urinary tract infection via the sliding grip mechanism.
Clinical Evidence
Bench testing only. Evaluated per BS EN 1616:1997 and BS EN 1618:1997. Tests included visual/dimensional inspection, flow rate, tensile strength, force to connect/disconnect, leak testing, accelerated aging, EO/ECH residuals, and biocompatibility (cytotoxicity, irritation, sensitization per ISO 10993). All samples passed performance criteria.
Technological Characteristics
Materials: PVC (tube and funnel). Design: Round polished tip, two drainage eyelets, sliding tube grip. Dimensions: French sizes 6-20; lengths 160mm-400mm. Connectivity: Compatible with standard drainage bags. Sterilization: Ethylene Oxide (EO). No software or electronic components.
Indications for Use
Indicated for male, female, and pediatric patients requiring assisted bladder drainage due to impaired natural voiding function.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
K132619 Page I of 4
Image /page/0/Picture/1 description: The image shows the logo for Flexicare. Below the name is the phrase "TOTAL QUALITY - TOTAL CARE". The address "Cynon Valley Business Park, Mountain Ash" is printed below the phrase.
Mid Glamorgan, CF45 4ER UK. Tel: +44 (0) 1443 474647 Fax: +44 (0)1443 474222 Customer Services: +44 (0;1443 471580 Purchasing Fax: +44 (0)1443 476332 Email. enquiries@flexicare.com www.llexicare.com
OCT 1 1 2013
## 510(k) Summary
Trade name
C.F.R Section Product Code:
Manufacturer:
Regulatory Affairs Contact:
Telephone:
Date Summary Prepared:
Common Name:
Classification:
Predicative Devices
Description:
Classic ™ Intermittent Catheter.
21 CFR 876.5130. Urological catheter and accessories EZD
Flexicare Medical Limited Cynon Valley Business Park Mountain Ash. Mid. Glam. CF45 4ER. United Kingdom
Christopher Watkins Quality Regulatory & Technical Director Flexicare Medical Limited Cynon Valley Business Park Mountain Ash, Mid. Glam. CF45 4ER. United Kingdom
00 44 1443 471 593
September 2013
Intermittent catheter
Class 2
- BARD "Interglide" which was cleared for marketing by 510(k) � No K951260.
- Coloplast "Self Cath" which was cleared for marketing by . 510k No K100878
Classic ™ Intermittent Catheter is a PVC plastic device that is intended to be inserted through urethra and into bladder of patient to allow drainage of urine. The Classic Intermittent catheter in pouch consists of a tube which is bonded onto a drainage funnel.
Classic™ Intermittent Catheter has a round polished finish tip to provide maximum comfort for user. Each catheter features two large evelets in order to drain urine efficiently. Drainage eyelets cut-outs are smooth and rounded to minimize the risk of injury/discomfort when in use.
Classic ™ Intermittent Catheters are available in: Adult Male. Available in coude tip or straight tip
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Flexicare. The logo is black and white and features a stylized "F" followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" are the words "TOTAL QUALITY . TOTAL CARE" in a smaller font.
Adult Female. Available in straight tip Paediatric. Available in straight tip
| Coudé tip | Straight tip |
|-----------|--------------|
|-----------|--------------|
Sizes are as follows:
| French size | ID (mm) | OD (mm) | Total length |
|-------------|---------|---------|-------------------|
| 6 | 1.1 | 2.0 | |
| 8 | 1.7 | 2.7 | Male: 400mm |
| 10 | 2.3 | 3.3 | |
| 12 | 2.8 | 4.0 | Female: 160mm |
| 14 | 3.0 | 4.7 | |
| 16 | 3.6 | 5.3 | Paediatric: 240mm |
| 18 | 4.2 | 6.0 | |
| 20 | 4.7 | 6.7 | |
Intended Use:
Classic ™ intermittent catheters are intended for use in male, female and paediatric patients when regular assisted bladder drainage has been instructed by a physician. Classic ™ intermittent catheters are indicated for use by individuals whose ability to void the bladder fully by natural bodily functions is impaired.
Substantial Equivalence
The Classic Intermittent Catheter has the same intended use as predicate devices:
- BARD "Interglide", which was cleared for marketing by . 510k No K951260.
- Coloplast "Self Cath" which was cleared for marketing by . 510k No K100878.
All three intermittent cathelers are single patient use devices. All three devices are not life supporting or life sustaining.
Only the Catheter Tube has invasive patient contact. Patient Contact - Invasive - Limited Exposure (A) Classification: Mucosal Membranes (B) | Limited (A).
All three intermittent catheters do not use software/ not electronically driven. All three intermittent catheters are supplied sterile in individual peel pouches.
All three intermittent catheters are able to be used with accessories such as drainage bags, which are industry standard devices used with catheters.
All three intermittent catheters are designed for the same intended use in the same intended conditions.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Flexicare. The logo has a stylized letter F on the left side, followed by the word "lexicare" in a bold, sans-serif font. Below the word "lexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in a smaller font.
All three intermittent catheters consist of 2-3 components made from injection molded, sheet & extruded plastics.
From companson testing it was determined that all invasive components of the three devices are manufactured from PVC. The connector of the Classic 74 Intermittent Catheler and the Colopiast Self Cath are also manufactured from PVC, and have the same colour connector for each size.
The BARD Interglide catheter's connector is manufactured from TPE, however, TPE has similar characteristics to PVC (easily moulded, coloured, flexible, provides secure push fit connections, bonds easily with adhesives) The Bard Interglide size 14 connector was green like the FML and Colopfast samples, however, other sizes were not available to compare colours.
All samples ODAD's were very similar, within a range of ±0.2mm for each size. Full ID/OD details are recorded in TR079/12 Test report. (Section 12).
| Sample | Total length |
|----------------------|--------------|
| FML male | 400mm |
| FML female | 160mm |
| FML paediatric | 240mm |
| Coloplast male | 420mm |
| Coloplast female | Un-available |
| Coloplast paediatric | 240mm |
| BARD male | 355mm |
| BARD female | Un-available |
| BARD paediatric | Un-available |
Summary of Testing: "Classic" intermittent catheter has been evaluated in accordance with: BS EN 1616 - 1997 - Sterile Urethral Catheters for single use BS EN 1618 - 1997 - Catheters other than intravascular catheters
Tests Performed:
| Test | Standard? / In-House? | Outcome? |
|---------------------------------------------------------------------------------|---------------------------------------------|------------------|
| Visual inspection | In-House/ EN 1616:1997 - 4.2.<br>ASTM F1886 | All samples pass |
| Dimensional inspection | In-House/ EN 1616:1997 - 4.3 | All samples pass |
| Flow rate | ASTM F623-99 | All samples pass |
| Tensile testing - catheter to connector | In-House | All samples pass |
| Tube hardness | In-House | All samples pass |
| Force to connect/ disconnect catheter<br>connector to leg bag stepped connector | In-House | All samples pass |
| Strength of catheter | EN 1616:1997 - Annex A | All samples pass |
| Leak testing - catheter tube to connector | EN 1618:1997 - Annex C | All samples pass |
| Accelerated age testing - then all above | ASTM F1980, ASTM F88, ASTM<br>F1929 | All samples pass |
| EO & ECH residuals | ISO 10993-7:2008 | All samples pass |
| Cytotoxicity, Irritation, sensitization | BS EN ISO 10993-10:2010 | All samples pass |
| | BS EN ISO 10993-5:2009 | All samples pass |
| Colour comparison | BS EN ISO 8836:2009 | All samples pass |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Flexicare. The logo is black and white and features a stylized letter "f" with two horizontal lines above it. To the right of the "f" is the word "lexicare" in a sans-serif font. Below the word "lexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in a smaller font.
All Samples passed the performance testing when tested against methods and criteria from both In-House test methods and relevant BS EN standards. The results of this testing which are included in TR079/12 show that the Classic™ Intermittent Catheters pass all performance tests and perform at least as well as both marketed predicate devices. TR079/12 can be found in section 12 of this document.
Although very similar in design and function, there are some differences between the Flexicare Intermittent catheter and its predicate devices listed within this submission.
The Flexicare Classic™ Intermittent Catheter is un-coated whilst the predicate BARD "Interglide" Intermittent Catheter has a Hydrophilic coating which aids insertion when submerged in water. However, users of un-coated intermittent catheters can use a lubricant (not supplied) if required which will aid insertion without needing a sterile water source.
The Flexicare Classic™ Intermittent Catheter also differs from the BARD "Interglide" and the Coloplast "Sell-Cath" predicates as is includes a sliding tube grip, which is not present on the predicate devices. This prevents the user having to touch the invasive tube of the catheter during insertion, reducing the risk of urinary tract infection.
The Flexicare Classic™ Intermittent Cathers from the predicate devices in available lengths, being available in a French size 20.
The Flexicare Classic™ Intermittent Catheter & Coloplast "Self-Cath" drainage funnels are Manufactured from PVC. This differs from the drainage funnel of the BARD "Interglide" sample which ls manufactured from TPE. This difference is not critical as both materials are flexible, able to be Colored and have an elastic nature, allowing a push fit connection to a drainage bag connector.
Signature of contact Person:
Chris Watkins
ewaxreno
20-09-2013
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
October 11, 2013
FLEXICARE MEDICAL LTD. % Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K132619
> Trade/Device Name: Classic™ Intermittent Catheter Regulation Number: 21 CFR\$ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD Dated: September 27, 2013 Received: September 30. 2013
Dear Mark Job.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enacunent date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mark Job
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "Flexicare". The logo is black and white and features a stylized "F" with two curved lines above it. The word "Flexicare" is written in a bold, sans-serif font to the right of the "F". Below the word "Flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in a smaller font.
## Indications for Use
510(k) Number (if known): K 132619
Page 1 of 1
Device Name:
Classic ™ Intermittent Catheter.
Indications for Use:
Classic™ intermittent catheters are intended for use in Mate, Female and Paediatic patients when regular assisted bladder drainage has been instructed by a physician. Classic™ intermittent catheters are indicated for use by individuals whose ablity to void the bladder fully by natural bodily functions is impaired.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Benjamin R. Fisher -S 2013.10.11 13:12:10 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.