K133364 · Cynosure, Inc. · GEX · Jul 22, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133364
Device Name
PICOSURE WORKSTATION
Applicant
Cynosure, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PicoSure™ workstation is indicated for tattoo and benign pigmented lesion removal. The PicoSure™ workstation operating with the 3mm or 6mm handpiee and the FOCUS lens array is indicated for the treatment of acne scars in Fitzpatrick skin types I-IV.
Device Story
PicoSure™ workstation is a high-powered, Q-switched Alexandrite laser system (755-nm wavelength). System comprises console (power supply, control electronics, laser) and articulated arm for energy delivery; activated via footswitch. Used in clinical settings by trained professionals for tattoo removal, benign pigmented lesion removal, and acne scar treatment. FOCUS lens array attachment modifies beam delivery for scar treatment. Laser energy interacts with skin tissue to induce therapeutic effects, such as epidermal vacuolization and dermal remodeling, aiding in scar reduction and pigment clearance. Clinical benefits include improved appearance of acne scars and removal of tattoos/pigmented lesions.
Clinical Evidence
Two clinical studies conducted. Study 1: 17 patients (46 acne scars) treated with FOCUS lens array (mean 6.3 sessions). 70% correct identification of before/after photos by blinded evaluators; mean improvement score 1.9 (scale 0-3); 87% patient satisfaction. Study 2: Histological evaluation on 3 subjects (arm/leg). Biopsies at 0, 7, 15 days post-treatment showed immediate focal epidermal vacuolization and degenerated keratinocytes, followed by intact epidermis and mild superficial dermal lymphocytic infiltrates at days 7 and 15.
Technological Characteristics
Flashlamp-excited Q-switched Alexandrite laser; 755nm wavelength; 450ps-900ps pulse width; 0.2 J/cm2 energy per pulse; 6.37 J/cm2 max average fluence. Delivery via articulated arm with zoom (2-6mm) or fixed (2-10mm) handpieces and FOCUS lens array. Electrical safety per IEC 60601-1 and IEC 60601-1-2. Software verification and validation performed.
Indications for Use
Indicated for tattoo and benign pigmented lesion removal; treatment of acne scars in Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is the logo for the Department of Health and Human Services (HHS). The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion. Inside the circle is a symbol consisting of three stylized human profiles facing right, with flowing lines extending from the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 23, 2016
Cynosure, Inc. Ms. Connie Hoy Vice President of Regulatory Affairs 5 Carlisle Road Westford, Massachusetts, 01886
Re: K133364 Trade/Device Name: Picosure™ Workstation Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: June 15, 2014 Received: June 17, 2014
Dear Ms. Hoy:
This letter corrects our substantially equivalent letter of July 22, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
This 510(K) Summary of safety and effectiveness for the PicoSure™ workstation
is submitted in accordance with the requirements of the SMDA 1990 and following guidan concerning the organization and content of a 510(K) summary.
| Applicant: | Cynosure, Inc. | |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address: | 5 Carlisle Road<br>Westford, MA 01886<br>USA | |
| Contact Person: | Connie Hoy, SVP of Regulatory Affairs | |
| Telephone:<br>Email: | 1-781-993-2414<br>choy@cynosure, com | |
| Preparation Date: | April 4, 2016 | |
| Device Trade Name: | PicoSure™ workstation | |
| Common Name: | Laser | |
| Classification Name: | Instrument, Surgical, Powered, laser<br>79-GEX, 21 CFR 878-48 | |
| Legally Marketed Predicate<br>Device(s): | PicoSure™ workstation (K)121346<br>Hoya ConBio RevLite (K)103118 | |
| Description of the PicoSure™<br>Workstation: | The PicoSure™ workstation is a high-powered, Q-switched<br>Alexandrite system that delivers laser energy in the 755-nm<br>wavelength. The system consists of a console that houses the<br>power supply, control electronics and the laser. Laser energy<br>is delivered to the skin via an articulated arm. The laser is<br>activated using a footswitch. | |
| Intended use of the PicoSure™<br>Workstation | The PicoSure™ workstation is indicated for tattoo and benign<br>pigmented lesion removal.<br><br>The PicoSure™ workstation operating with the 3mm or 6mm<br>handpiee and the FOCUS lens array is indicated for the<br>treatment of acne scars in Fitzpatrick skin types I-IV. | |
| Performance Data: | IEC 60601-1 Medical Electrical Equipment - Part 1: General<br>Requirements for Safety | |
| | IEC 60601-1-2 Medical Electrical Equipment 1-2 General<br>Requirements for basic safety and essential performance | |
| | Steam Sterilization Test Report 673257 | |
| | Software Verification and Validation Testing Report<br>860-7012-SRV | |
{3}------------------------------------------------
Results of Clinical Study:
Two clinical studies were conducted.
One study assessed facial acne scars. In 17 patients who completed the study, a total of 46 acne scars underwent a mean of 6.3 treatments with either the 3mm or 6mm handpiece, each operated with the FOCUS lens array. Before and after photographs were evaluated by three blinded evaluators, who were able to identify correctly the before and after images in 70% of the 17 subjects. The mean improvement score (scale 0-3) was 1.9. The non-blinded treating physician reported a 100% overall satisfaction rate, scoring the improvement in 47% of subjects as "satisfied" and 53% as "extremely satisfied." Subjects reported a satisfaction rate of 87% (53% "satisfied" and 27% "extremely satisfied").
A second study was conducted on normal skin on the arm of 2 subjects and the leg of a 3rd subject to evaluate the tissue response to treatment of the FOCUS lens array. Three patients were treated with the 6mm lens with FOCUS array at 0.71J/cm2 and with the 3mm lens with FOCUS array at 2.83J/cm2. Biopsies were performed in untreated skin and in treated skin immediately post treatment, 7 days post treatment and 15 days post treatment. Treatment induced immediate focal epidermal vacuolization and small foci of degenerated keratinocytes. At days 7 and 15, skin exhibited intact epidermis with mild, superficial dermal lymphocytic infiltrates in peri-vascular pattern.
| Technical<br>Comparison: | Specifications | | |
|----------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| | PicoSure™ Workstation<br>(current submission) | PicoSure™ Workstation<br>(previously cleared) | RevLite |
| Laser Type | Flashlamp excited Q-<br>switched alexandrite | Flashlamp excited Q-<br>switched alexandrite | Flashlamp excited Q-<br>switched Nd:YAG |
| Wavelength | 755nm | 755nm | 532 nm, 585 nm 650 nm, &<br>1064 nm |
| Energy per pulse | 0.2 J/cm2 | 0.2 J/cm2 | 0.85 J/cm2 |
| Maximum Average<br>Fluence | 6.37 J/cm² | 6.37 J/cm² | 12 J/cm² (1064nm)<br>5 J/cm² (532nm)<br>10 J/cm² (585nm)<br>6 J/cm²<br>(650 nm)<br>1.2 J/cm² (532Lite) |
| Repetition Rate | Single pulse, or 1, 2.5, 5,<br>or 10 pulse(s) per second<br>(Hz) | Single pulse, or 1,2,5, or<br>10 pulse(s) per second<br>(Hz) | Single & double pulse, 1, 2,<br>5, & 10 Hz pulses per<br>second |
| Pulse Width | 450ps-900ps | 450ps-900ps | 5-20 ns |
| Spot Sizes | Zoom 2-6 mm,<br>Fixed 2, 3, 4, 6, 8, 10 mm<br>With FOCUS lens array:<br>3mm 6mm | Zoom 2-6 mm,<br>Fixed 2, 3, 4, 6, 8, 10 mm | Fixed 2 - 8 mm (varies by<br>wavelength) |
{4}------------------------------------------------
Conclusion:
The PicoSure Workstation is substantially equivalent to other existing laser systems in commercial distribution for use in Dermatology and Plastic Surgery.
.
{5}------------------------------------------------
concerning the organization and content of a 510(K) summary.
| Applicant: | Cynosure, Inc. | |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address: | 5 Carlisle Road<br>Westford, MA 01886<br>USA | |
| Contact Person: | Connie Hoy, SVP of Regulatory Affairs | |
| Telephone:<br>Email: | 1-781-993-2414<br>choy@cynosure, com | |
| Preparation Date: | April 4, 2016 | |
| Device Trade Name: | PicoSure™ workstation | |
| Common Name: | Laser | |
| Classification Name: | Instrument, Surgical, Powered, laser<br>79-GEX, 21 CFR 878-48 | |
| Legally Marketed<br>Device(s): | Predicate PicoSure™ workstation (K)121346<br>Hoya ConBio RevLite (K)103118 | |
| Description of the PicoSure™<br>Workstation: | The PicoSure™ workstation is a high-powered, Q-switched<br>Alexandrite system that delivers laser energy in the 755-nm<br>wavelength. The system consists of a console that houses the<br>power supply, control electronics and the laser. Laser energy<br>is delivered to the skin via an articulated arm. The laser is<br>activated using a footswitch. | |
| Intended use of the PicoSure™<br>Workstation | The PicoSure™ workstation is indicated for tattoo and benign<br>pigmented lesion removal. | |
| | The PicoSure™ workstation operating with the 3mm or 6mm<br>handpiee and the FOCUS lens array is indicated for the<br>treatment of acne scars in Fitzpatrick skin types I-IV. | |
| Performance Data: | IEC 60601-1 Medical Electrical Equipment - Part 1: General<br>Requirements for Safety | |
| | IEC 60601-1-2 Medical Electrical Equipment 1-2 General<br>Requirements for basic safety and essential performance | |
| | Steam Sterilization Test Report 673257 | |
| | Software Verification and Validation Testing Report<br>860-7012-SRV | |
{6}------------------------------------------------
Results of Clinical Study:
Two clinical studies were conducted.
One study assessed facial acne scars. In 17 patients who completed the study, a total of 46 acne scars underwent a mean of 6.3 treatments with either the 3mm or 6mm handpiece, each operated with the FOCUS lens array. Before and after photographs were evaluated by three blinded evaluators, who were able to identify correctly the before and after images in 70% of the 17 subiects. The mean improvement score (scale 0-3) was 1.9. The non-blinded treating physician reported a 100% overall satisfaction rate, scoring the improvement in 47% of subjects as "satisfied" and 53% as "extremely satisfied." Subjects reported a satisfaction rate of 87% (53% "satisfied" and 27% "extremely satisfied").
A second study was conducted on normal skin on the arm of 2 subjects and the leg of a 3rd subject to evaluate the tissue response to treatment of the FOCUS lens array. Three patients were treated with the 6mm lens with FOCUS array at 0.71J/cm2 and with the 3mm lens with FOCUS array at 2.83J/cm2. Biopsies were performed in untreated skin and in treated skin immediately post treatment, 7 days post treatment and 15 days post treatment. Treatment induced immediate focal epidermal vacuolization and small foci of degenerated keratinocytes. At days 7 and 15, skin exhibited intact epidermis with mild, superficial dermal lymphocytic infiltrates in peri-vascular pattern.
| Technical<br>Comparison: | Specifications | | | | |
|----------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------|--|--|
| | PicoSure™ Workstation<br>(current submission) | PicoSure™ Workstation<br>(previously cleared) | RevLite | | |
| Laser Type | Flashlamp excited Q-<br>switched alexandrite | Flashlamp excited Q-<br>switched alexandrite | Flashlamp excited Q-<br>switched Nd:YAG | | |
| Wavelength | 755nm | 755nm | 532 nm, 585 nm 650 nm, &<br>1064 nm | | |
| Energy per pulse | 0.2 J/cm2 | 0.2 J/cm2 | 0.85 J/cm2 | | |
| Maximum Average<br>Fluence | 6.37 J/cm² | 6.37 J/cm² | 12 J/cm² (1064nm)<br>5 J/cm² (532nm) | | |
| | | | 10 J/cm² (585nm) | | |
| | | | 6 J/cm²<br>(650 nm) | | |
| | | | 1.2 J/cm² (532Lite) | | |
| Repetition Rate | Single pulse, or 1, 2.5, 5,<br>or 10 pulse(s) per second<br>(Hz) | Single pulse, or 1,2,5, or<br>10 pulse(s) per second<br>(Hz) | Single & double pulse, 1, 2,<br>5, & 10 Hz pulses per<br>second | | |
| Pulse Width | 450ps-900ps | 450ps-900ps | 5-20 ns | | |
| Spot Sizes | Zoom 2-6 mm,<br>Fixed 2, 3, 4, 6, 8, 10 mm<br>With FOCUS lens array:<br>3mm,6mm | Zoom 2-6 mm,<br>Fixed 2, 3, 4, 6, 8, 10 mm | Fixed 2 - 8 mm (varies by<br>wavelength) | | |
{7}------------------------------------------------
Conclusion:
The PicoSure Workstation is substantially equivalent to other existing laser systems in commercial distribution for use in Dermatology and Plastic Surgery.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.