K133019 · Emtron Elektronik VE Mekanik Sanayi VE Ticaret Lim · HQF · Feb 20, 2015 · Ophthalmic
Device Facts
Record ID
K133019
Device Name
FIBERION OPHTHALMIC ENDOLASER PROBE
Applicant
Emtron Elektronik VE Mekanik Sanayi VE Ticaret Lim
Product Code
HQF · Ophthalmic
Decision Date
Feb 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Fiberion Ophthalmic Endolaser Probes are intended for use in performing ophthalmic laser treatments to deliver laser energy to the treatment area inside the eye, the illumination function is indicated for use to illuminate the interior of the eye. The probes are offered with straight or angled tips, and with a series of connectors that allows them to be used with compatible laser systems. The probes are cleared for use for the particular indications of the laser system to which they are attached.
Device Story
Fiberion Ophthalmic Endolaser Probes are fiber-optic delivery devices for ophthalmic laser photocoagulation; transmit laser energy from external laser systems to intraocular treatment sites; some models include additional fiber optics for illumination. Components include input connector, glass fiber optic, buffer coating, protective tubing, aluminum handle, and medical-grade stainless steel needle. Used in clinical settings by ophthalmologists; probes connect to compatible laser photocoagulators. Output is laser energy delivery and interior eye illumination; enables precise photocoagulation for various ophthalmic conditions. Benefits include targeted therapeutic energy delivery and visualization of the surgical field.
Clinical Evidence
Bench testing only. Testing included laser power transmission, beam divergence, beam density homogeneity, and illumination intensity/area. Laboratory testing confirmed sterility, biocompatibility, and acceptable levels of EO/ECH residues (<1.25 ug/device and <5 ug/device respectively) and endotoxins (<0.2 EU/device).
Indicated for ophthalmic laser treatments requiring delivery of laser energy to the interior of the eye and illumination of the interior of the eye. Used in conjunction with compatible ophthalmic laser systems for the indications specified by those systems.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping heads or faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20, 2015
Emtron Elektronik Ve Mekanik Sanayi Ve Ticaret Lim Dr. Mehmet Melek President Bebek Yolu Sokagi 23/3 Etiler Istanbul, 34337 TR
Re: K133019
> Trade/Device Name: Fiberion Ophthalmic Endolaser Probe Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF Dated: January 9, 2015 Received: January 12, 2015
Dear Dr. Melek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, MD Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K133019
Device Name Fiberion Ophthalmic Endolaser Probe
Indications for Use (Describe)
Fiberion Ophthalmic Endolaser Probes are in performing ophthalmic laser treatments to deliver laser energy to the treatment area inside the eye, the illumination is indicated for use to illuminate the interior of the eye. The probes are offered with straight or angled tips, and with a series of connectors that allows them to be used with compatible laser systems. The probes are cleared for use for the particular indications of the laser system to which they are attached
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary EMTRON™ Corporation Fiberion™ Ophthalmic Endolaser Probes
## Submitter's Name, Address, Telephone Number, E-mail, Contact Person and Date Prepared
EMTRON Elektronik ve Mekanik Sanayi ve Ticaret Limited Sirketi Bebek Yolu Sokaqi 23/3 Etiler 34337 lstanbul Turkey +90 (212) 257 1787 melek@emtron.com Dr. Mehmet Melek Date Prepared: February 20, 2015
### Device Information
Trade Name: Fiberion Ophthalmic Endolaser Probes Classification Name: Laser, Ophthalmic, Accessory CFR Section: 886.4390 Class II Product Code: HQF
### Predicate Devices
Fiberion Ophthalmic Endolaser Probes is substantially equivalent in intended use and/or method of operation to other currently legally marketed laser probes of IRIDEX Corporation's IRIS Medical EndoProbe (K022228).
## Device Description
Fiberion endolaser probes are designed to be connected to ophthalmic laser photocoagulators and to transmit laser energy inside the patient's eye. One end of the probe contains a connector for attachment to the laser unit, the other, a tip section to be introduced inside the eye. Device components are an input connector, a glass fiber optic protected by a buffer coating, a protective tubing, an aluminum handle with a medical grade stainless steel needle. The illuminating endolaser probe has an additional illumination fiber optic.
## List of Fiberion Ophthalmic Endolaser Probes
| SMA-S | : Standard SMA connector, straight 0,9mm (20 Gauge) tip |
|--------------|--------------------------------------------------------------------------|
| SMA-A | : Standard SMA connector, angled 0,9mm (20 Gauge) tip |
| E-SMA-S | : Extended SMA connector, straight 0,9mm (20 Gauge) tip |
| E-SMA-A | : Extended SMA connector, angled 0,9mm (20 Gauge) tip |
| E-CYL-S | : Cylindrical connector, straight 0,9mm (20 Gauge) tip |
| E-CYL-A | : Cylindrical connector, angled 0,9mm (20 Gauge) tip |
| E-906-S | : Extended SMA 906 connector, straight 0,9mm (20 Gauge) tip |
| E-906-A | : Extended SMA 906 connector, angled 0,9mm (20 Gauge) tip |
| U-SMA-S | : SMA connector with long handle, straight 0,9mm (20Gauge) tip |
| U-SMA-A | : SMA connector with long handle, angled 0,9mm (20Gauge) tip |
| SMA-400-S | : Standard SMA connector, 400 micron fiber, straight 0,9mm (20Gauge) tip |
| SMA-400-A | : Standard SMA connector, 400 micron fiber, angled 0,9mm (20Gauge) tip |
| STA-S | : ST connector with long ferrule, straight 0,9mm (20 Gauge) tip |
| STA-A | : ST connector with long ferrule, angled 0,9mm (20 Gauge) tip |
| SMA-23-S | : Standard SMA connector, straight 0,6mm (23 Gauge) tip |
| SMA-23-A | : Standard SMA connector, angled 0,6mm (23 Gauge) tip |
| E-SMA-23-S | : Extended SMA connector, straight 0,6mm (23 Gauge) tip |
| E-SMA-23-A | : Extended SMA connector, angled 0,6mm (23 Gauge) tip |
| E-CYL-23-S | : Cylindrical connector, straight 0,6mm (23 Gauge) tip |
| E-906-23-S | : Extended SMA 906 connector, straight 0,6mm (23 Gauge) tip |
| E-906-23-A | : Extended SMA 906 connector, angled 0,6mm (23 Gauge) tip |
| U-SMA-23-S | : SMA connector with long handle, straight 0,6mm (23Gauge) tip |
| U-SMA-23-A | : SMA connector with long handle, angled 0,6mm (23Gauge) tip |
| SMA-400-23-S | : Standard SMA connector, 400 micron fiber,straight 0,6mm (23 Gauge) tip |
| STA-23-S | : ST connector with long ferrule, straight 0,6mm (23 Gauge) tip |
| STA-23-A | : ST connector with long ferrule, angled 0,6mm (23 Gauge) tip |
| SMA-25-S | : Standard SMA connector, straight 0,5mm (25 Gauge) tip |
| SMA-25-A | : Standard SMA connector, angled 0,5mm (25 Gauge) tip |
| E-SMA-25-S | : Extended SMA connector, straight 0,5mm (25 Gauge) tip |
| E-SMA-25-A | : Extended SMA connector, angled 0,5mm (25 Gauge) tip |
| E-906-25-S | : Extended SMA 906 connector, straight 0,5mm (25 Gauge) tip |
| E-906-25-A | : Extended SMA 906 connector, angled 0,5mm (25 Gauge) tip |
| U-SMA-25-S | : SMA connector with long handle, straight 0,5mm (25Gauge) tip |
| STA-25-S | : ST connector with long ferrule, straight 0,5mm (25 Gauge) tip |
| STA-25-A | : ST connector with long ferrule, angled 0,5mm (25 Gauge) tip |
| SMA-27-S | : Standard SMA connector, straight 0,4mm (27 Gauge) tip |
| E-SMA-27-S | : Extended SMA connector, straight 0,4mm (27 Gauge) tip |
| E-906-27-S | : Extended SMA 906 connector, straight 0,4mm (27 Gauge) tip |
| STA-27-S | : ST connector with long ferrule, straight 0,4mm (27 Gauge) tip |
| ILL-SMA-S | : Standard SMA connector, straight 0,9mm (20 Gauge) tip |
| ILL-SMA-A | : Standard SMA connector, angled 0,9mm (20 Gauge) tip |
| ILL-E-SMA-S | : Extended SMA connector, straight 0,9mm (20 Gauge)tip |
| ILL-E-SMA-A | : Extended SMA connector, angled 0,9mm (20 Gauge) tip |
| ILL-E-906-S | : Extended SMA 906 connector, straight 0,9mm (20Gauge) tip |
| ILL-E-906-A | : Extended SMA 906 connector, angled 0,9mm (20Gauge) tip |
| ILL-U-SMA-S | : SMA connector with long handle, straight 0,9mm (20Gauge) tip |
| ILL-U-SMA-A | : SMA connector with long handle, angled 0,9mm (20Gauge) tip |
| ILL-STA-S | : ST connector with long ferrule, straight 0,9mm (20 Gauge) tip |
| ILL-STA-A | : ST connector with long ferrule, angled 0.9mm (20 Gauge) tip |
{6}------------------------------------------------
#### Intended Use
Fiberion Ophthalmic Endolaser Probes are intended for use in performing ophthalmic laser treatments to deliver laser energy to the treatment area inside the eye, the illumination function is indicated for use to illuminate the interior of the eye. The probes are offered with straight or angled tips, and with a series of connectors that allows them to be used with compatible laser systems. The probes are cleared for use for the particular indications of the laser system to which they are attached.
{7}------------------------------------------------
## Substantial equivalence
| Application for 510K | Substantial equivalence to |
|-----------------------------------------|-----------------------------------------|
| Fiberion Ophthalmic Endolaser Probes | IRIDEX Corporation's IRIS Medical |
| | EndoProbe (K022228) |
| Light transmission for photocoagulation | Light transmission for photocoagulation |
| Aluminum handpiece | Aluminum handpiece |
| Stainless steel needle | Stainless steel needle |
| Glass optical fiber | Glass optical fiber |
| Thermoplastic rubber jacket | Vinyl jacket |
| ETO sterilized. | ETO sterilized. |
## Differences between the Fiberion Ophthalmic Endolaser Probes and the predicate device
Some terminology differences exist in the intended use description due to trademark issues, but both devices' intended use is exactly the same.
Some models of the predicate device are not available in the Fiberion line. Fiberion probes offer a range of connectors that have been validated with the ophthalmic lasers listed in the instructions for use. The probes have been used internationally for years without safety or laser compatibility problems.
The protective tubing which does not come in contact with the patient is of a different material. Fiberion probes use thermoplastic rubber meeting FDA food grade requirements.
Hence, none of these differences are cause for safety or effectiveness concerns.
# Bench testing
Fiberion Ophthalmic Endolaser Probes and the predicate device were tested for laser power transmission, beam divergence, beam density homogeneity, and additionally for illuminating models, illumination intensity and illumination area diameter; and were found to be equivalent. Fiberion Ophthalmic Endolaser Probes were tested for safe delivery of the intended output power.
# Laboratory testing
Fiberion Ophthalmic Endolaser Probes were tested for sterility and their sections in contact with the patient for biocompatibility, EO-ECH residues and endotoxins and were found to fulfil the requirements. EO levels were determined to be < 1.25 micrograms per device, ECH levels < 5 micrograms per device, and Endotoxin levels < 0.2 Endotoxin units per device.
# Conclusion
Fiberion Ophthalmic Endolaser Probes shares similar indications for use, materials, and similar performance characteristics as, and thus are substantially equivalent to. IRIDEX Corporation's IRIS Medical EndoProbe (K022228).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.