CROSSCHECK PLATING SYSTEM

K132594 · Solana Surgical, LLC · HRS · Dec 13, 2013 · Orthopedic

Device Facts

Record IDK132594
Device NameCROSSCHECK PLATING SYSTEM
ApplicantSolana Surgical, LLC
Product CodeHRS · Orthopedic
Decision DateDec 13, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Solana Surgical LLC, CrossCheck Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet, wrist, and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use.

Device Story

CrossCheck Plating System consists of metallic bone plates and screws for orthopedic fixation; used for fractures, osteotomies, and fusions in extremities. System includes locking and non-locking screws (3mm or 3.5mm), lag screws, and cross-hole plates for compression. Implanted by surgeons in clinical settings to stabilize small bones; provides mechanical support to facilitate bone healing. Device is single-use.

Clinical Evidence

Bench testing only. Evaluated per ASTM F382 (Standard Specification and Test Method for Metallic Bone Plates) and ASTM F543-07 (Standard Specification and Test Methods for Metallic Medical Bone Screws).

Technological Characteristics

Materials: ASTM F136 Ti-6Al-4V ELI alloy. Components: Bone plates and screws (locking/non-locking, 3mm/3.5mm). Principle: Mechanical fixation. Form factor: Small bone plating system. Sterilization: Not specified.

Indications for Use

Indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones in the hand, feet, wrist, ankles, fingers, and toes. Applicable to both pediatric and adult patients.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K132594 DEC 13 2013 # 6. 510(k) Summary Summary of 510 (k) safety and effectiveness information upon which the substantial equivalence determination is based: Prepared: Applicant: August 15, 2013 Solana Surgical, LLC 6363 Poplar Ave, Suite 312 Memphis, TN 38119 Joe Clift (Contact) jclift@solanasurgical.com (901) 818-1860 Common Name: Device Trade Name: Device Classification Name: Device Classification: Reviewing Panel: Regulation Number: Product Code: Predicate Device(s): Plate, Fixation, Bone CrossCheck Plating System Single/Multiple component metallic bone fixation appliances and accessories Class II Orthopedic 21 CFR 888.3030 HRS Memometal Anchorage Bone Plate System: K083447 Integra Total Foot System K123000 #### Device Description: The Solana Surgical, LLC CrossCHECK Plating System consists of a variety of bone plates and screws to be used in fixation fractures, osteotomies, and fusions in the extremities of the ley and arm including the hand and foot. The various plate designs use either 3mm or 3.5mm screws. Screws include both locking and non-locking designs. Lag screws and cross-hole plates are included for compression. Plate designs include MTP, Lapidus, and 2-7 Hole Utility plates. All plates and screws are manufactured from ASTM F136 Ti-6A1-4V ELI alloy. The design of the Solana Surgical implant is similar to the predicate devices. No new materials or processes are used in the development of this implant. ## Indications for Use: The Solana Surgical LLC, CrossCheck Plating System is indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones in the hand, feet, wrist, and ankles, fingers and toes. The system may be used in both pediatric and adult patients. The device is intended for single use. K132594 - 510(k) Summary - Revised November 21, 2013 {1}------------------------------------------------ # Comparison to Predicate Device: Similarities of the Solana Surgical device to its predicates include these devices being: intended for single use only, intended for surgical implantation longer than 30 days, a system consisting of a series of implants, made of industry standard materials, with no new materials being introduced in the product, comparably sized, and indicated for the same uses. The technological characteristics, principles of operation and mechanisms of action are the same as the predicates. #### Summary of Device Testing: The plates and screws of the subject device were evaluated per ASTM F382 "Standard Specification and Test Method for Metallic Bone Plates" and ASTM F543-07 "Standard Specification and Test Methods for Metallic Medical Bone Screws". Examples from the predicate device systems were used for comparison. The subject device met acceptance criteria and is substantially equivalent to the predicates. # K132594 - 510(k) Summary - Revised November 21, 2013 {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service December 13, 2013 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Solana Surgical, LLC Mr. Joe Clift Senior Vice President Quality & Regulatory 6363 Poplar Avenue, Suite 312 Memphis, Tennessee 38119 Re: K132594 Trade/Device Name: CrossCheck Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: November 21, 2013 Received: November 22, 2013 Dear Mr. Clift: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register, Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - Mr. Joe Clift - - - - - - - - - - - - - - - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark NiMelkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health 1 Rail- 13-12-2018 Enclosure {4}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. 510{k} Number (if known) K132594 Device Name CrossCheck Plating System #### Indications for Use (Describe) The Solana Surgical LLC, CrossCheck Plating System is indication and fixation of frash fractures, revision procedures, joint fixsion and reconstruction of small bones in the hand, fee, wrist, and toes. The system may be used in both pediatric and adult patients. The device is intended for single use. Type of Use (Select one or both, as applicable) [2] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Stignature) # Elizabeth L. Frank -S Division of Orthopedic Devices FORM FDA 3881 (9/13) i
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...