REPROCESSED HARMONIC SHEAR
K132566 · Sterilmed, Inc. · NLQ · Jun 14, 2014 · SU
Device Facts
| Record ID | K132566 |
| Device Name | REPROCESSED HARMONIC SHEAR |
| Applicant | Sterilmed, Inc. |
| Product Code | NLQ · SU |
| Decision Date | Jun 14, 2014 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic, Pediatric |
Indications for Use
The reprocessed harmonic shears are intended to be used for cutting soft tissue and when control of bleeding and minimal thermal injury is desired. The instruments can be used as an adjunct to or a substitute for electrosurgery, lasers, and steel scalpels in general, plastic, pediatric, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures.
Device Story
Reprocessed Harmonic Shears; used in combination with handpiece, generator, and torque wrench. Device enables surgeon to grasp, coagulate, and cut soft tissue with single instrument; coagulates vessels up to 5 mm diameter. Hand-actuated shaft with 360° rotating distal tip; ultrasonic energy delivered via hand activation or generator foot switch. Used in OR for open and endoscopic procedures. Output is mechanical ultrasonic energy for tissue effect; surgeon visualizes tissue response to guide decision-making. Benefits include reduced thermal injury and bleeding control compared to traditional electrosurgery or steel scalpels.
Clinical Evidence
Bench testing only. No clinical data. Validation included cleaning, sterilization (ISO 11135, USP <71>), biocompatibility (ISO 10993), ethylene oxide residuals (ISO 10993-7), packaging (ASTM D 4169, F 88, F 2096), shelf life (ASTM 1980), and functional performance (simulated use, fatigue, and visual inspection).
Technological Characteristics
Ultrasonic surgical instrument; 5 mm shaft diameter; 23 cm or 36 cm length. Materials and design identical to predicate. Energy source: ultrasonic generator. Sterilization: ethylene oxide. Standards: ISO 11135, ISO 10993, ASTM D 4169, ASTM F 88, ASTM F 2096, ASTM 1980.
Indications for Use
Indicated for cutting soft tissue and controlling bleeding in general, plastic, pediatric, gynecologic, urologic, thoracic, and orthopedic procedures (spine/joint space) via open or endoscopic access. Suitable for patients requiring soft tissue surgery where minimal thermal injury is desired.
Regulatory Classification
Identification
Single-use reprocessed versions of the ultrasonic devices used in surgical procedures either for fragmentation, emulsification and aspiration of soft tissue and hard tissue or for ligation of vessels (under product code “LFL”). Reprocessing validation data for this device type must be included in a 510(k) submission.. (70 FR 56911, available at https://www.govinfo.gov/content/pkg/FR-2005-09-29/pdf/05-19510.pdf).
Predicate Devices
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "STERILMED" in a simple, sans-serif font. The letters are evenly spaced and appear to be in black or a dark color against a white background. To the left of the word is a small, abstract graphic that is difficult to discern.
1
## II. SUMMARY AND CERTIFICATION
| A. | 510(k) Summary | K132566 | | JUN 14 2014 |
|----|------------------------|-----------------------------------------------------------------------------------------------------------------|--|-------------|
| | Submitter: | Sterilmed, Inc. | | |
| | Contact Person: | Jason Skramsted<br>11400 73rd Avenue North<br>Maple Grove, MN 55369<br>Phone: 763-488-3483<br>Fax: 763-488-4491 | | |
| | Date Prepared: | 14 August 2013 | | |
| | Trade Name: | Reprocessed Harmonic Shears | | |
| | Regulation Name: | Scalpel, Ultrasonic, Reprocessed | | |
| | Device Classification: | Unclassified | | |
| | Product Code: | NLQ | | |
| Predicate<br>Devices: | The reprocessed harmonic shears are substantially equivalent to the Ethicon Harmonic Shears (K121550,<br>K120729 and K060245). |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description: | The harmonic shears are used in combination with a handpiece, generator and torque wrench and are intended<br>to be used for cutting and coagulation of soft tissue during laparoscopic and open procedures. The<br>instruments allow the surgeon to grasp, coagulate and cut soft tissue with one instrument. The instruments<br>allow for the coagulation of vessels up to and including 5 mm in diameter. The instruments are hand-actuated<br>with a shaft and distal tip that can be rotated 360°. The ultrasonic energy is delivered by hand activation or<br>with an optional generator foot switch. |
| Intended Use: | The reprocessed harmonic shears are intended to be used for cutting soft tissue and when control of bleeding<br>and minimal thermal injury is desired. The instruments can be used as an adjunct to or a substitute for<br>electrosurgery, lasers, and steel scalpels in general, plastic, pediatric, gynecologic, urologic, thoracic,<br>exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures. |
| Technological<br>Characteristics: | The reprocessed harmonic shears are identical to the predicate devices in design, materials of construction,<br>and intended use. There are no changes to the clinical applications, patient population, performance<br>specifications, or method of operation. |
| Functional and<br>Safety Testing: | Representative samples of reprocessed harmonic shears were tested to demonstrate appropriate functional<br>characteristics. Process validation testing was performed to validate the cleaning and sterilization procedures<br>as well as device packaging. In addition, the manufacturing process includes visual and validated functional<br>testing of all products produced. |
| Summary of<br>Non-clinical<br>Tests<br>Conducted: | Specific non-clinical tests performed included: cleaning validation, sterilization validation (ISO 11135, USP<br><71>), biocompatibility testing (ISO10993), ethylene oxide residual testing (ISO 10993-7), packaging<br>validation (ASTM D 4169, ASTM F 88, ASTM F 2096), and shelf life validation (ASTM 1980). In addition,<br>validation of functional performance (bench testing) was performed through simulated use, visual inspection.<br>fatigue testing, and function testing. Performance testing shows the reprocessed harmonic shears to perform<br>as originally intended. |
| Conclusion: | Sterilmed concludes that the reprocessed harmonic shears are safe, effective, and substantially equivalent to<br>the predicate devices, Ethicon Harmonic Shears (K121550, K120729, and K060245), as described in this<br>premarket notification submission. |
ر
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 14, 2014
SterilMed % Mr. Jason Skramsted Regulatory Affairs Specialist 11400 73td Avenue, North Maple Grove, Minnesota 55369
Re: K132566
Trade/Device Name: Reprocessed Harmonic Shears (See enclosed list) Regulatory Class: Unclassified Product Code: NLQ Dated: May 20, 2014 Received: May 21, 2014
Dear Mr. Skramsted:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Régister.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Mr. Jason Skramsted
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 3 – Mr. Jason Skramsted
| Model Number | Device Name | Shaft<br>Diameter | Shaft<br>Length |
|--------------|-----------------------------|-------------------|-----------------|
| HAR23M | Reprocessed Harmonic Shears | 5 mm | 23 cm |
| HAR36M | Reprocessed Harmonic Shears | 5 mm | 36 cm |
| ACE23E | Reprocessed Harmonic Shears | 5 mm | 23 cm |
| ACE36E | Reprocessed Harmonic Shears | 5 mm | 36 cm |
| N/A | Reprocessed Torque Wrench | N/A | N/A |
: 上一篇:
:
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "STERILMED" in a simple, sans-serif font. To the left of the word is a stylized graphic, possibly representing a medical symbol or logo. The text and graphic are in black against a white background, creating a clear contrast.
## Indications for Use
510(k) Number (if known): K132566
Device Name: Reprocessed Harmonic Shears
Indications for Use:
The reprocessed harmonic shears are intended to be used for cutting soft tissue and when control of bleeding and minimal thermal injury is desired. The instruments can be used as un udjunct to or a substitute for clectrosurgery, lasers, and steel scalpels in general, plastic, pediatic, gyncoologic, thoracic, exposure to orthopedic structures (such as spine and joint space) and other open and endoscopic procedures.
Prescription Use X (Part 2) CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## ua C. N