K132037 · Mindray North America · MWI · May 16, 2014 · Cardiovascular
Device Facts
Record ID
K132037
Device Name
VITAL SIGNS MONITOR
Applicant
Mindray North America
Product Code
MWI · Cardiovascular
Decision Date
May 16, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The monitor is intended for spot-check monitoring physiologic parameters, including SpO2, PR, NIBP and TEMP, on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
Device Story
Accutorr 3 Vital Signs Monitor is a compact device for spot-check monitoring of physiological parameters: SpO2, PR, NIBP, and TEMP. It utilizes measurement accessories for data acquisition; supports SpO2 modules from Mindray, Masimo, or Nellcor. Operated by clinical physicians or medical staff in healthcare facilities. Output is displayed on the monitor for clinical assessment of patient vital signs. Device aids in basic patient monitoring needs; supports clinical decision-making through real-time parameter tracking.
Clinical Evidence
Clinical validation performed for NIBP and TEMP modules. Testing conducted per ISO 80601-2-56 (clinical thermometers) and ISO 81060-2 (automated non-invasive sphygmomanometers). No specific numerical results (e.g., mean error, SD) provided in summary; compliance with standards cited as evidence of clinical accuracy.
Technological Characteristics
Compact vital signs monitor; supports SpO2, PR, NIBP, TEMP. SpO2 modules from Mindray, Masimo, or Nellcor. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 9919 (pulse oximetry), EN 1060 series (sphygmomanometers), and ASTM E1112 (thermometers). Software life-cycle per IEC 62304. Usability engineering per IEC 62366.
Indications for Use
Indicated for spot-check monitoring of SpO2, PR, NIBP, and TEMP in adult, pediatric, and neonatal patients within healthcare facilities.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
510(k) SUMMARY
# 510(K) SUMMARY
MAY 1 6 2014
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K132037.
Submitter: 1.
> Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: (86)755-81888998 Fax: (86)755-26582680
## Contact Person:
Yanhong Bai Manager Regulatory Affairs SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. Mindray Building, Keji 12th Road South, High-tech Industrial Park, Nanshan, Shenzhen 518057, P.R. China Tel: (86) 755-81885635 Fax: (86) 755-26582680 E-mail: baiyanhong@mindray.com
## Date Prepared:
May 13, 2014
#### 2. Name of the devices: Accutorr 3 Vital Signs Monitor
## Classification
| 21 CFR 870.2300 Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
|-----------------------------------------------------------------------------|----------|
| 21 CFR 870.2700 Oximeter | Class II |
| 21 CFR 870.1130 Noninvasive blood pressure measurement system | Class II |
| 21 CFR 880.2910 Clinical electronic thermometer | Class II |
#### 3. Device Description:
{1}------------------------------------------------
The Accutorr 3 Vital Signs Monitor is a compact, easy-to-use vital signs monitor designed to satisfy basic monitoring needs. It consists of a main unit, parameters measurement accessories, peripheral equipments or accessories.
The Accutorr 3 Vital Signs Monitor is designed to monitor a fixed set of physiological parameters including Pulse Oxygen Saturation (SpO2), Pulse Rate(PR), Non-invasive Blood Pressure (NIBP) and Temperature (TEMP). This monitor provides 3 different SPO2 modules manufactured by Mindray, Masimo or Nellcor.
#### 4. Intended Use:
The monitor is intended for spot-check monitoring physiologic parameters, including SpO2, PR, NIBP and TEMP, on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
#### છે. Predicate Devices:
The Accutorr 3 Vital Signs Monitor is substantially equivalent to the predicate device VS-800 Vital Signs Monitor (K072235). They have the same intended uses, the same technological characteristics, and are comparable in key safety and effectiveness features.
#### Non-clinical Tests: 6.
The Accutorr 3 Vital Signs Monitor has been evaluated for biocompatibility, safety, parameter performance and electromagnetic compatibility, and has been designed to comply with applicable medical safety standards. This device has been tested and evaluated under the following standards:
- · ISO 14971 Medical device Application of Risk Management to Medical Devices
- · IEC 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety
- · IEC 60601-1-2 Medical electrical equipment part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility- Requirements and tests
- · IEC 60601-1-4 Medical electrical equipment Part 1-4: General requirements for safety -Collateral Standard: Programmable electrical medical systems
- . ISO 9919 Medical electrical equipment Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use
- · EN 1060-1 Specification for Non-invasive sphygmomanometers Part 1: General requirements
- · EN 1060-3 Non-invasive sphygmomanometers Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems
{2}------------------------------------------------
### 510(k) SUMMARY
- · EN 1060-4 Non-invasive sphygmomanometers-Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers
- · ANSI/AAMI SP-10 Manual, electronic, or automated sphygmomanometers
- · ASTM E1104 Standard Specification for Clinical Thermometer Probe Covers and Sheaths
- · ASTM E1112 Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
- · ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing
- · ISO 15223-1 Medical devices-Symbols to be used with medical device labels, labeling and information to be supplied
- · IEC 62304 Medical device software Software life-cycle processes
- · IEC 62366 Medical devices Application of usability engineering to medical devices
- · IEC 60601-1-1 Medical Electrical Equipment- Part 1-1: General Requirements for Safety -Collateral Standard: Safety requirements for medical electrical systems
- · EN 1041 Information supplied by the manufacturer of medical devices
#### 7. Clinical Studies
The NIBP and TEMP module of the subject device Accutorr 3 Vital Signs Monitor have been evaluated for clinical accuracy and have been designed to comply with applicable clinical These modules have been tested and evaluated under the following clinical standards. standards:
- · ISO 80601-2-56 Particular requirements for basic safety and essential performance of clinical thermometers
- · SO 81060-2 Non-Invasive Sphygmomanometers Part 2: Clinical Validation of Automated Measurement Type
#### Conclusion: 8.
The results of all testing demonstrate that the Accutorr 3 Vital Signs Monitor is as safe, as effective, and perform as well as the predicate device.
{3}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
## May 16, 2014
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. c/o Mr. Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Road South, High-Tech Industrial Park, Nanshan, Shenzhen 518057 P.R. China
Re: K132037
Trade/Device Name: Accutorr 3 Vital Signs Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI, DQA, DXN, FLL Dated: May 08, 2014 Received: May 12, 2014
Dear Mr. Bai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Yanhong Bai
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/8 description: The image shows a logo with the letters FDA. The letters are stylized and overlapping, with a double line underneath the A. The logo is black and white and appears to be a vectorized image.
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): _K132037
## Device Name: Accutorr 3 Vital Signs Monitor
## Indications for Use:
The monitor is intended for spot-check monitoring physiologic parameters, including SpO2, PR, NIBP and TEMP, on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Picture/10 description: The image shows a logo with the letters FDA in a stylized font. Above the letters, the word "Date" is written. To the right of the logo, the numbers 2014.06.16 and 04'00' are written. The image is black and white.
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.