The PM-50 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate monitor. It operates only on battery power using existing PM-50 disposable and reusable sensors labeled for patients ranging from neonates to adults. The VS-800 Vital Signs Monitor is used to monitor physiologic parameters including SpO2, PR and NIBP, and to measure Temperature parameter on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for transport or home use.
Device Story
PM-50: Portable, battery-powered pulse oximeter; uses two-wavelength (red/infrared) pulsatile system to measure SpO2 and pulse rate via finger sensor; displays values on LCD. VS-800: Configurable vital signs monitor; measures SpO2, PR, NIBP, and temperature; powered by AC mains or internal battery (lead-acid/lithium); features LCD/LED display, menu-driven interface, and optional thermal recorder. Used in healthcare facilities by clinicians; provides audible/visual alarms for parameter violations. Both devices utilize existing sensor technology to provide real-time physiological data for clinical decision-making.
Clinical Evidence
Clinical study conducted according to ISO9919 to validate SpO2 performance following probe modification. Bench testing included design change analysis, biocompatibility, performance, and safety/EMC testing. Results demonstrated devices meet design specifications and perform equivalently to predicates.
Technological Characteristics
PM-50: Portable, battery-powered. VS-800: AC/battery-powered (lead-acid/lithium). Both use two-wavelength (red/infrared) optical sensing for SpO2. VS-800 includes NIBP and temperature modules. Connectivity via patient cables; optional thermal recorder. Software-based parameter processing and alarm management.
Indications for Use
Indicated for monitoring SpO2, pulse rate, NIBP, and temperature in adult, pediatric, and neonatal patients in healthcare facilities. Not for transport or home use.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
OCT 1 9 2007
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:
## Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
- Contact Person: .
Li Dongling Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
- Date Prepared: .
July 31, 2007
## Name of the device:
| Trade/Proprietary Name | PM-50 Pulse Oximeter | VS-800 Vital Signs Monitor |
|------------------------|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Pulse Oximeter | Vital Signs Monitor |
| Classification | 21 CFR 870.2700, Oximeter, Pulse, 74 DQA<br>21 CFR 870.2710, Ear Oximeter, Pulse, 74 DQA | 21 CFR 870.2700, Oximeter, Pulse, 74 DQA<br>21 CFR 870.2710, Ear Oximeter, Pulse, 74 DQA<br>21 CFR 870.2300, Cardiac monitor (including cardiotachometer and rate alarm), 74 MWI<br>21 CFR 870.1130, Non-Invasive blood pressure measurement System ,74 DXN<br>21 CFR 880.2910, Clinical Electronic Thermometer, 80 FLL |
| Predicate Device | K061442, PM-50 Pulse Oximeter, cleared on Sep 5, 2006 | K063055, VS-800 Vital Signs Monitor, cleared on Dec 21, 2006 |
{1}------------------------------------------------
#### Device Description:
PM-50 Pulse Oximeter:
The PM-50 is a flexible, portable, battery powered Pulse Oximeter. The PM-50 Pulse Oximeter acquires the physiological signals - oxygen saturation (SpO2) and pulse rate (PR). The signals are converted into digital data and processed, and the SpO2 and pulse rate values are calculated and displayed on LCD screen.
PM-50 uses a two-wavelength pulsatile system - red and infrared light - to obtain SpO2 based on the different light absorption of oxygenated and reduced hemoglobin. The light source in the finger sensor emits red and infrared light, which are partially absorbed and modulated by the arterial blood pulsation at the sensor site. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The electronic signals are sent to the oximeter and processed by the oximeter's circuitry. Thereafter, the SpO2 and pulse rate are obtained and indicated on the LCD screen.
VS-800 Vital Signs Monitor:
The VS-800 Vital Signs Monitor is a prescription device intended for use by health care professionals. And the device is capable of operation from an external AC mains powers source or an internal battery including rechargeable lead-acid battery and lithium battery. The device uses the same or similar technology and materials as the predicate devices, see Legally Marketed Predicate Devices listed above.
The VS-800 Vital Signs Monitor is a configurable monitor with options selected by customer preference. Device's options including module configuration and language setting are configured at the time the monitor is manufactured. Options may be upgrade via upgrade port by the manufacturer. The monitor also provides customer with the convenient operating control and human-machine interface (HMI). All of the patient cable connections are located on the monitor. The LCD and LED display patient information and the menu provides single control operations of all main functions. Operator can adjust parameter alarm settings that give audible and visual indication when a violation occurs. The VS-800 provides option for printing information by a thermal recorder.
This monitor has the following parameters measurement functions:
- V SpO2 measurement: pulse oxygen saturation (SpO2), pulse rate (PR), and SpO2 plethysmogram.
- イ NIBP measurement: systolic pressure (S), diastolic pressure (D), mean pressure (M), and pulse rate (PR).
- V Rectal / oral / axillary TEMP measurement: temperature (TEMP).
{2}------------------------------------------------
## Intended Use:
The PM-50 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate monitor. It operates only on battery power using existing PM-50 disposable and reusable sensors labeled for patients ranging from neonates to adults.
The VS-800 Vital Signs Monitor is used to monitor physiologic parameters including SpO2, PR and NIBP, and to measure Temperature parameter on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for transport or home use.
# Comparison of Technological Characteristics:
The PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor employ the same functional technology as the predicate devices.
### Testing:
The device modification only involved the addition of SpO2 probes to PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor. Bench testing and clinical study was conducted based on the modification to verify and validate that the PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor met all design specifications and were substantially equivalent to predicate devices.
The following quality assurance measures were applied to the devices:
- Design Change and Modification Influence Analysis
- Biocompatibility Testing
- Performance Testing
- Safety & Environmental &EMC Testing
- Clinical Study according to ISO9919
### Conclusion:
The conclusions drawn from clinical and bench testing of the PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor demonstrate that the devices is as safe, as effective, and performs as well as the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged in a circle around the symbol. The caduceus symbol is composed of four abstract human figures, each with their head bowed. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 9 2007
Shenzhen Mindray Bio-Medical Electronics Company, Limited C/O Ms. Susan D. Goldstein-Falk Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
Re: K072235
Trade/Device Name: PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: September 13, 2007 Received: September 19, 2007
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D.
Chiu Lin, Ph.D Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor Indications For Use:
The PM-50 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate monitor. It operates only on battery power using existing PM-50 disposable and reusable sensors labeled for patients ranging from neonates to adults.
The VS-800 Vital Signs Monitor is used to monitor physiologic parameters including SpO2, PR and NIBP, and to measure Temperature parameter on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for transport or home use.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mull Whil
Control, Dental Devices
Page of
Number: K072235
nesthesiology, General Hospital.
0032
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.