HEALTHSTATS A-PULSE CASPAL + AND MC3100+ MONITORS AND MY-BP BP MANAGEMENT APPLICATION SOFTWARE
K131788 · Healthstats International Pte. , Ltd. · DXN · Mar 4, 2014 · Cardiovascular
Device Facts
Record ID
K131788
Device Name
HEALTHSTATS A-PULSE CASPAL + AND MC3100+ MONITORS AND MY-BP BP MANAGEMENT APPLICATION SOFTWARE
Applicant
Healthstats International Pte. , Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Mar 4, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The A-PULSE CASPal+ Monitor is a compact standalone monitor that combines two methods of blood pressure measurement: - Conventional oscillometric blood pressure via a brachial cuff on the upper arm, and . - Radial arterial pulse acquisition via a wrist-mounted tonometer sensor. The A-PULSE CASPal+ Monitor first measures systolic and diastolic blood pressure and pulse rate using the oscillometric method, and then acquires the radial arterial pulse waveform to derive the Central Aortic Systolic Pressure (CASP) non-invasively. The MC3100+ monitor measures systolic and diastolic blood pressure and pulse rate via the oscillometric method via a brachial cuff at the upper arm. The monitors have a non-volatile memory to store and recall measurements, and the ability to transfer measurement data via Bluetooth wireless communications to other suitably equipped electronic devices, such as a PC with HealthSTATS.my-BP PC software. my-BP PC software is used in conjunction with the HealthSTATS A-PULSE CASPal+ and MC3100+ Monitors to store and display blood pressure data obtained from the monitors. The A-PULSE CASPal+ and MC3100+ Monitors are intended for use on patients who are eighteen (18) years and older, and for the CASPal+, have a palpable radial pulse. Brachial Blood pressure, CASP and pulse rate readings obtained using the A-PULSE CASPal+ and MC3100+ Monitors are intended for use by qualified healthcare personnel as an aid to diagnosis and treatment. Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
Device Story
A-PULSE CASPal+ and MC3100+ are non-invasive blood pressure monitors. CASPal+ uses brachial cuff (oscillometric) and wrist-mounted tonometer sensor (radial arterial pulse) to derive Central Aortic Systolic Pressure (CASP). MC3100+ uses only brachial cuff. Both devices feature Bluetooth for data transfer to my-BP PC software for storage and display. Operated by healthcare personnel in clinical settings. Output provides systolic/diastolic BP, pulse rate, and CASP (CASPal+ only) to assist in cardiovascular disease management and hypertension assessment. Benefits include non-invasive CASP estimation and digital data management.
Clinical Evidence
No new clinical data submitted. Clinical testing performed on the predicate A-PULSE CASPal device was determined to be applicable to the subject devices, as they are functionally and physically identical regarding measurement technology.
Technological Characteristics
Monitors utilize oscillometric brachial cuff and radial arterial tonometry (CASPal+). Bluetooth wireless connectivity. Embedded processing unit. Biocompatible materials for skin contact. Tested for electrical safety, EMC, environmental, and mechanical strength. Software-based data management via my-BP PC application.
Indications for Use
Indicated for patients 18 years and older requiring blood pressure and pulse rate monitoring. CASPal+ model requires a palpable radial pulse. Used by qualified healthcare personnel as an aid to diagnosis and treatment.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Page 1 of 4
# 510(k) Summary of Safety and Effectiveness in accordance with 21 CFR 807.92
- . HealthSTATS International Pte. Ltd. (a) Submitted bv: 6 New Industrial Road #04-01/02/03 Singapore 536199
Phone: +65-6858 3248 Fax: +65-6858 0148
E-mail: cmting@healthstats.com.sg
- Contact Person: Position/Title:
Dr. Choon Meng TING Chairman/CEO
May 31, 2013 Date of Preparation:
Trade Name:
HealthSTATS A-PULSE CASPal®+and MC3100TM+ Monitors and my-BP™ BP Management Application Software
74 DXN, 21 CFR § 870.1130
- Common/Classification Name: System, Measurement, Blood-pressure, Non-invasive
Product Code:
Class:
Class II
- (3) Predicate Device(s):
- HealthSTATS A-PULSE CASPal® , HealthSTATS K101002 International Pte Ltd.
- K040371 A&D Medical UA767BT Blood pressure monitor, A&D Engineering, Inc.
- Omron Health Management Software (accessory PC K061822 software to Omron HEM-780, Omron Healthcare Ltd.
Reason for Submission: New Device
- Description of Device: (4)
The HealthSTATS A-PULSE CASPal® and MC3100™ are standalone noninvasive blood pressure monitors which are designed to measure (1) conventional oscillometric blood pressure with a brachial cuff and (2)
{1}------------------------------------------------
Central Aortic Systolic Pressure (CASP) and other indices based on arterial tonometry at the radial artery of the wrist.
The CASPal®+ Monitor is comprised of four main elements:
- System embedded processing unit. �
- Integrated oscillometric blood pressure module for calibration . (identical to the BPCalibrator [listed device K041546], also cleared iin CASPal device K101002).
- . Wrist sensor module for acquistion of central aortic systolic pressure (CASP) via radial arterial tonometry at the wrist [cleared accessory and method K101002].
- Bluetooth® wireless communications module. .
The MC3100™* is a reduced feature set of the CASPal®* which only offers the oscillometric blood pressure measurement and not the CASP measurement, but is otherwise identical. my-BP™ PC software may be used with either device to store and display blood pressure and pulse rate measurements on a suitably equipped computer.
### Intended use: (5)
The A-PULSE CASPal® and MC3100™* Monitors are intended to measure systolic and diastolic blood pressure and pulse rate via the oscillometric method with cuff. The A-PULSE CASPal® utilizes a wrist sensor to obtain the radial pulse waveform to derive central aortic systolic pressure (CASP) and other waveform indices. Both monitors utilize Bluetooth wireless technology to transfer measurements to HealthSTATS my-BP™ PC Software or other suitable devices.
Cardiovascular disease remains among the leading causes of death worldwide. Hypertension assessment is one of the primary factors in the evaluation of cardiovascular disease. Assessment of CASP may improve the management of patients with elevated cardiovascular risk.
## Indications for Use:
The A-PULSE CASPal+ Monitor is a compact standalone monitor that combines two methods of blood pressure measurement:
- · Conventional oscillometric blood pressure via a brachial cuff on the upper arm, and
- . Radial arterial pulse acquisition via a wrist-mounted tonometer sensor.
The A-PULSE CASPal+ Monitor first measures systolic and diastolic blood pressure and pulse rate using the oscillometric method, and then acquires the radial arterial pulse waveform to derive the Central Aortic Systolic Pressure (CASP) non-invasively.
{2}------------------------------------------------
## Page 3 of 4
The MC3100+ monitor measures systolic and diastolic blood pressure and pulse rate via the oscillometric method via a brachial cuff at the upper arm.
The monitors have a non-volatile memory to store and recall measurements, and the ability to transfer measurement data via Bluetooth wireless communications to other suitably equipped electronic devices, such as a PC with HealthSTATS my-BP PC software.
my-BP PC software is used in conjunction with the HealthSTATS A-PULSE CASPal+ and MC3100+ Monitors to store and display blood pressure data obtained from the monitors.
The A-PULSE CASPal+ and MC3100+ Monitors are intended for use on patients who are eighteen (18) years and older, and for the CASPal+, have a palpable radial pulse.
Brachial Blood pressure, CASP and pulse rate readings obtained using the A-PULSE CASPal+ and MC3100+ Monitors are intended for use by qualified healthcare personnel as an aid to diagnosis and treatment.
### Technological Characteristics: (6)
Two monitors are offered in this 510(k), the A-PULSE CASPal®* Monitor and the MC3100™*:
- The A-PULSE CASPal®+ Monitor measures and displays blood . pressure readings with a pulse bar to display radial arterial pulsations. The device measures both oscillometric blood pressure via a brachial cuff and Central Aortic Systolic Pressure (CASP) via a wrist-mounted tonometer sensor. The A-PULSE CASPal®+ is substantially identical to the predicate A-PULSE CASPal monitor except that the CASPal®+ is capable of Bluetooth® wireless data transmission of measurements.
- . The MC3100™+ Monitor offers a reduced feature set of the CASPal®+ and measures only oscillometric blood pressure via a brachial cuff. The MC3100™+ also is capable of Bluetooth® wireless data transmission.
Mv-BP™ PC Software is a software application which, when loaded on a computer suitably equipped with Bluetooth® functionality, is able to acquire measurement data from CASPal® and MC3100 ™* for subsequent storage and display.
#### Non-Clinical Tests Submitted: (1) (b)
The A-PULSE CASPal®/MC3100™* Monitors have been tested to meet applicable standards for medical device electrical safety, electromagnetic
{3}------------------------------------------------
### Page 4 of 4
compatibility, environmental (temperature and humidity), and mechanical strength (shock and vibration).
Materials utilized in skin contact surfaces are in accessories cleared for the same intended use. The materials met biocompatibility requirements.
The embedded software of the A-PULSE CASPar® /MC3100™* Monitors has been verified to requirements and validated to meet intended use. HealthSTATS mv-BP™ PC Software has been verified to meet specified requirements for the storage and display of retrospective vital blood pressure and pulse rate data.
System level risk, hazard, and failure mode analysis has been performed on both hardware and software, and residual risks were determined to be acceptable.
### Clinical Tests Submitted: (2)
Clinical testing performed on the cleared predicate A-PULSE CASPal device, which is functionally and physically identical to the A-PULSE CASPal® device except for the addition of the Bluetooth® wireless data transmission, was determined to be applicable to the CASPal®* and MC3100 ** monitors. Therefore no new clinical data was submitted.
### (3) Conclusions from Tests:
As described in (b)(1) and (b)(2) above, the testing demonstrates that the A-PULSE CASPal® /MC3100 ** Monitors and my-BP™ PC Software are as safe and effective as, and function in a manner equivalent to the referenced predicate devices.
{4}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Image /page/4/Picture/3 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed within a circular border, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle.
March 4, 2014
Healthstats International Pte. Ltd. Stephen Gorski S65 W35739 Piper Road Eagle, WI 53119 US
Re: K131788
> Trade/Device Name: Trade/Device Name: HealthSTATS A-PULSE CASPal®+ and MC3100 TM+ Monitors and my-BPTM BP management application software
Regulation Number: 21 CFR 870.1130 Regulation Name: System, Management, Blood-Pressure, Non-Invasive Regulatory Class: Class II Product Code: DXN Dated: January 22, 2014 Received: January 27, 2014
Dear Stephen Gorski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good connolo provisions of the reling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statues and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); the recordinations, more 801); medical device reporting of medical device-related adverse labeling (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems
{5}------------------------------------------------
Page 2 - Stephen Gorski
(QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
FDA
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
. . . . . . . .
# Indications for Use
510(k) Number (if known):
HealthSTATS A-PULSE CASPal®+ and MC3100™+ Monitors and Device Name: my-BPTM BP Management Application Software
Indications for use:
The A-PULSE CASPal+ Monitor is a compact standalone monitor that combines two methods of blood pressure measurement:
- Conventional oscillometric blood pressure via a brachial cuff on the upper arm, and .
- Radial arterial pulse acquisition via a wrist-mounted tonometer sensor.
The A-PULSE CASPal+ Monitor first measures systolic and diastolic blood pressure and pulse rate using the oscillometric method, and then acquires the radial arterial pulse waveform to derive the Central Aortic Systolic Pressure (CASP) non-invasively.
The MC3100+ monitor measures systolic and diastolic blood pressure and pulse rate via the oscillometric method via a brachial cuff at the upper arm.
The monitors have a non-volatile memory to store and recall measurements, and the ability to transfer measurement data via Bluetooth wireless communications to other suitably equipped electronic devices, such as a PC with HealthSTATS.my-BP PC software.
my-BP PC software is used in conjunction with the HealthSTATS A-PULSE CASPal+ and MC3100+ Monitors to store and display blood pressure data obtained from the monitors.
The A-PULSE CASPal+ and MC3100+ Monitors are intended for use on patients who are eighteen (18) years and older, and for the CASPal+, have a palpable radial pulse.
Brachial Blood pressure, CASP and pulse rate readings obtained using the A-PULSE CASPal+ and MC3100+ Monitors are intended for use by qualified healthcare personnel as an aid to diagnosis and treatment.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
Prescription Use × (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Date:
2014:03.04
15:57:04 -05'00'
for Bram Zuckerman
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.