DIGITAL FUNDUS CAMERA, MODEL CF-60DSI

K041546 · Canon, Inc. · HKI · Jun 23, 2004 · Ophthalmic

Device Facts

Record IDK041546
Device NameDIGITAL FUNDUS CAMERA, MODEL CF-60DSI
ApplicantCanon, Inc.
Product CodeHKI · Ophthalmic
Decision DateJun 23, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1120
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

DIGITAL FUNDUS CAMERA CF-60DSi is intended to be used for taking digital images of retina of human eye with a mydriatic.

Device Story

CF-60DSi is a digital fundus camera used by ophthalmologists to capture retinal images. It functions as an improved model of the CF-60UVi, featuring direct mounting for a Canon EOS digital camera without requiring an adapter. The device utilizes a light source (max 300WS) and optical components to illuminate and image the retina; focusing is achieved by aligning split lines. It offers variable powers of 60/40 degrees. The system is used in clinical settings for applications such as telemedicine and electronic filing. Captured images are viewed immediately on a monitor, facilitating clinical assessment and documentation of retinal conditions. The device requires a mydriatic for pupil dilation (minimum 5.5mm diameter required).

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and comparison of technical specifications, optical performance, and mechanical design.

Technological Characteristics

Ophthalmic camera; 60/40° field of view; 45mm working distance; split-line focusing; max 300WS light source; direct-mount EOS digital camera interface; UL60601-1 electrical safety compliance; manual/automatic filter set; external eye fixation lamp.

Indications for Use

Indicated for taking digital retinal images of the human eye using a mydriatic agent. Intended for use by ophthalmologists.

Regulatory Classification

Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K041546 Section G: Summary # 510(k) Summary #### Prepared: March 31, 2004 #### Submitter: | Company Name: | Canon USA, Inc. (U.S. agent for Canon Inc.) | |------------------|---------------------------------------------| | Company Address: | One Canon Plaza<br>Lake Success, NY 11042 | | Contact Person: | Ms. Sheila Driscoll | | Phone Number: | (516) 328-5602 | | Fax Number: | (516) 328-5169 | #### Proposed Device: | Reason For 510(k): | New Model | |----------------------|---------------------------| | Manufacturer: | Canon Inc. | | Trade Name: | Canon | | Model Name: | CF-60DSi | | Classification Name: | 86HKI, Ophthalmic cameras | | FDA 510(k) #: | To be assigned | #### Predicate Device: | Manufacturer: | Canon Inc. | |----------------------|---------------------------| | Trade Name: | Canon | | Model Name: | CF-60UVi | | Classification Name: | 86HKI, Ophthalmic cameras | | FDA 510(k) #: | K946058 | CF-60DSi is intended to be used for taking digital images pictures of retina Intended Use: of human eye with a mydriatic. Description Of Device: CF-60DSi is an improved model of CF-60UVi. The DIGITAL FUNDUS Doorfpack CF-60DSi is used for taking digital images of retina of human eye without a mydriatic. Canon EOS Digital Camera is mounted with CF-60DSi, can be viewed immediately, making oundly 200 Digitiont and many different applications, such as telemedicine and electronic filing. The CF-60DSi's intended use is the same as that of CF-60UVi. However, the differences in design are as follows: - EOS digital camera can be attached to the main unit of CF-60DSi directly. But CF-60UVi . cannot be attached without an adapter. - cannot of attached while at and mount of CF-60DSi with the adapter. The . 100 angliar cambera can be attached to the sub mount of CF-60UVi with the adapter, and 35mm camera can be attached with the 35mm double adapter, and also Polaroid film unit can be attached without the adapter. - While CF-60DSi has 2 variable powers (60/40 degree), CF-60UVi has 3 variable powers (60/40/30 degree). - A data card cannot be used. • {1}------------------------------------------------ K041546 ### Section G: Summary CF-60DSi is equivalent to CF-60UVi in the following respect: - The optical components and alignment. . - The optical components and allgament. The mechanical structures of the CF-60DSi are almost same as the CF-60UVi. Please refer to . the CF-60DSi comparison table provided in this section. | | | | Table of comparison | | |-------------|-------------------------------------------------------------------------------------------|------------------------------------|------------------------------------------|------------------------------------------| | | | | CF-60UVi | CF-60DSi | | PERFORMANCE | Angle of view | | 60/40/30° | 60/40° | | | | Actual image size | $$\phi$ 29×22(on 35mm film) | Same(on monitor) | | | | | $$\phi$ 7.5×5.7(Type 1:on sensor array) | Same | | | | | $$\phi$ 4.9×3.7(Type 2:on sensor array) | No Applicable | | | | | $$\phi$ 75×57(on Polaroid film) | No Applicable | | | | Min. diameter of pupil<br>required | $$\phi$ 5.5mm(60/40/30°) | Same(60/40°) | | | | | $$\phi$ 4.0mm(with S.P switch ON in 30°) | $$\phi$ 4.0mm(with S.P switch ON in 40°) | | | | Working distance(WD) | 45mm | Same | | | | Focusing | By aligning the split lines | Same | | | | Data to be printed | Hand-written data | No Applicable | | | Eye fixation lamp | | External | Same | | | Filter set | | Automatic/Manual | Same | | | Light source for<br>photography | | Max. 300WS | Same | | | Image unit | | EOS Digital Camera(with Adapter) | EOS Digital Camera | | | | | CCD camera(with Adapter) | No Applicable | | | | | 3CCD camera(with Adapter) | No Applicable | | | | | 35mm film unit | No Applicable | | | | | Polaroid film unit | No Applicable | | | | | No Applicable | ICG digital camera(with Adapter) | | | Working range<br>Vertical<br>Forward and back<br>Right and left<br>Chin rest (vertically) | | 38mm | Same | | | | | 70mm | Same | | | | | 120mm | Same | | | | | 65mm | Same | | | | | | | | | Panning | | 30° right or left | Same | | | External dimensions | | | | | | Main unit | | W320×D560×H565mm | Same | | | Power control unit | | W225×D390×H520mm | W225×D380×H485mm | | | Weight | | | | | | Main unit | | Approx.26kg | Approx.27kg | | | Power control unit | | Approx.30kg | Approx.27kg | | | Intended use | | Taking picture of retina of human eye | Same | | | Energy | | | | | | | | | | | | used | | 1000VA | Same | | | delivered | | No Applicable | Same | | | Target population | | Ophthalmologist | Same | | | Physical safety | | UL544 | UL60601-1 | | | Compliance with standards | | UL544 | UL60601-1 | | | Biocompatibility | | No Applicable | Same | | | Labeling<br>Packaging | | Printed model name is changed | | ## Table of comparison {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 3 2004 Canon, Inc. c/o Glenn M. Luchen Underwriters Laboratories, Inc. 1285 Walt Whitman Rd. Melville, NY 11747 K041546 Re: K041540 Trade/Device Name: Digital Fundus Camera, Model CF-60DSi Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: Class II Product Code: HKI Dated: June 7, 2004 Received: June 9, 2004 Dear Mr. Luchen: We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section JTV(K) premainer in substantially equivalent (for the indications referenced above and have determined the device is substantial in interstate ferenced above and nave determined the arrest and devices marketed in interstate for use stated in the encrosule, to regally mancess promotical Device Amendments, or to commerce prior to May 20, 1970, the characted with the provisions of the Federal Food. Drug, devices that have been recuire approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approval of a previous of the Act . The and Cosment Act (Act) that to not require approval controls provisions of the Act. The You may, merelore, market the device, subject to the gents for annual registration, listing of general controls provisions of the Act include requirements misbranding and general controls provisions of the frectines and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (Scc above) inl. - Existing major regulations affecting your device can may be subject to such additional controls. Existing major regulation E may be subject to such additional controls: Extrems - 1998 In addition, FDA may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 may be found in the Code of Federal Regarations, our device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA \$ Issualitat equr device complies with other requirements of the Act that FDA has made a determination that your device complies. You must that FDA has made a delemination and regulations administered by other Federal and listing or any Federal statutes and regulations administed to: registration and listing requirements a comply with all the Act S requirements, metals), good manufacturing practice requirements as set (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing the electro (2) CFR Part 807); labellig (21 CFR Patt 820); and if applicable, the electronic forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electr torth in the quality systems (QD) regalaten (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1 {3}------------------------------------------------ Page 2 - Glenn M. Luchen This letter will allow you to begin marketing your device as described in your Section 510(k) This icter will anow your to ough finding of substantial equivalence of your device to a legally premarket notification: "The Director on on your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific and Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, A. halyi Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications Statement | 510(K)Number(if known): | K041546 | |-------------------------|---------| |-------------------------|---------| Device Name: CF-60DSi Indications for Use: (Per 21 CFR 801.109) aleations for USC. DIGITAL FUNDUS CAMERA CF-60DSi is intended to be used for taking digital images of retina of human eye with a mydriatic. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHERT PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) 1 OR Over-The-Counter Use Prescription Use (Optional Format 1-2-96) Evelyn K. Been Division Siar Division of Ophthalmic Ear. Nose and Throat 0 510(k) Number K041546
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...