K131563 · Opus Ksd, Inc. · GDW · Apr 9, 2014 · General, Plastic Surgery
Device Facts
Record ID
K131563
Device Name
SUBQ IT! BIOABSORBABLE SKIN CLOSURE SYSTEM
Applicant
Opus Ksd, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Apr 9, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SubQ It!™ Bioabsorbable Skin Closure System is intended for use in humans for abdominal, thoracic, gynecologic, orthopedic, plastic and reconstructive surgery for the subcuticular closure of skin where an absorbable tissue fastener is desired for temporary tissue approximation.
Device Story
SubQ It! Bioabsorbable Skin Closure System consists of bioabsorbable fasteners and a manual stapler; used for subcuticular skin closure. Fasteners feature two legs with barbed tips; composed of 82/18 L-lactide/Glycolide copolymer. Device is pre-loaded, single-patient use, and sterile. Operated by surgeons in clinical settings (abdominal, thoracic, gynecologic, orthopedic, plastic, reconstructive). Surgeon manually deploys fasteners to approximate tissue edges; fasteners provide temporary closure while absorbing over time. Benefits include reduced need for suture removal and potential for improved cosmetic outcomes. No clinical data; performance established via bench and porcine animal models.
Clinical Evidence
No clinical data performed. Evidence based on GLP-compliant porcine animal model studies (7cm, 2cm, and 1cm incisions) over 90 days, bench testing, shelf-life mechanical performance studies, packaging tests, and biocompatibility testing.
Indicated for humans undergoing abdominal, thoracic, gynecologic, orthopedic, plastic, or reconstructive surgery requiring subcuticular skin closure for temporary tissue approximation.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
On October 8, 2021, FDA issued a final order reclassifying surgical staplers intended for internal use (formerly regulated under the classification for "manual surgical instrument for general use") from class I (general controls) into class II (special controls) and subject to premarket review. See here:
https://www.federalregister.gov/documents/2021/10/08/2021-22041/general-and-plastic-surgery-devices-reclassification-of-certainsurgical-staplers.
As part of this reclassification, FDA also amended the existing classification for "manual surgical instrument for general use" to remove staplers and to create a separate classification regulation for surgical staplers that distinguishes between surgical staplers for internal use and external use. Under the new regulation, 21 CFR 878.4740, surgical staplers for internal use are classified into class II with special controls, while surgical staplers for external use remain in in class I, exempt from premarket notification.
{1}------------------------------------------------
June 07, 2022
Image /page/1/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Opus Ksd, Inc. Rosina Robinson, RN, MED, RAC Affiliate Principal Regulatory Consultant 62 Forest Street Suite 300 Marlborough, Massachusetts 01752
Re: K131563
Trade/Device Name: Subq It! Bioabsorbable Skin Closure System Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, QQS
Dear Rosina Robinson, RN, MED, RAC:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 09, 2014. Specifically, FDA is updating this SE Letter to reflect an administrative change corresponding to the reclassification of certain surgical staplers, as detailed in the final order published on October 8, 2021 (see here for more information: https://www.federalregister.gov/documents/2021/10/08/2021-22041/general-and-plasticsurgery-devices-reclassification-of-certain-surgical-staplers). As part of this reclassification, FDA has amended the existing classification for "manual surgical instrument for general use" to remove staplers and to create a separate classification for surgical staplers that distinguishes between surgical staplers for internal use and external use. Under the new regulation, 21 CFR 878.4740, surgical staplers for internal use are classified into class II with special controls, while surgical staplers for external use remain in class I, exempt from premarket notification. As a result of the final order, FDA has created a new product code (OOS) to better categorize surgical staplers for external use.
{2}------------------------------------------------
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Deborah Fellhauer, OHT4: Office of Surgical and Infection Control Devices, 301-796-9570, Deborah.Fellhauer@fda.hhs.gov.
Sincerely,
## Deborah A. Fellhauer -S
Deborah Fellhauer RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure in profile, with three faces overlapping to suggest community and connection. The overall design is simple and professional, reflecting the department's role in public health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 9, 2014
Opus KSD Incorporated % Ms. Rosina Robinson, RN, Med, RAC Aptiv Solutions 62 Forest Street, Suite 300 Marlborough, Massachusetts 01752
Re: K131563
Trade/Device Name: SubQ It!" Bioabsorbable Skin Closure System Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: February 20, 2014 Received: February 21, 2014
Dear Ms. Robinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: SubQ It!™ Bioabsorbable Skin Closure System
Indications for Use:
The SubQ It!™ Bioabsorbable Skin Closure System is intended for use in humans for abdominal, thoracic, gynecologic, orthopedic, plastic and reconstructive surgery for the subcuticular closure of skin where an absorbable tissue fastener is desired for temporary tissue approximation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{6}------------------------------------------------
### 510(k) Summarv for the Opus KSD, Inc. SubQ It!™ Bioabsorbable Skin Closure System (per 21CFR 807.92)
#### 1. SUBMITTER/510(K) HOLDER
Opus KSD, Inc. 210 Worcester Rd Peacham, VT 05862 Contact Person: Charles H. Rogers, C.E.O. Telephone: +1.781.929.7406
May 21, 2013 Date Prepared:
#### 2. DEVICE NAME
| Proprietary Name: | SubQ It!™ Bioabsorbable Skin Closure System |
|----------------------|---------------------------------------------|
| Common/Usual Name: | Skin staples and stapler |
| Classification Name: | Staple, implantable |
#### 3. PREDICATE DEVICES
Incisive Surgical, INSORB™ Absorbable Staple (Multiple 510(k)'s e.g. K024117, K033602, K061784, K090159, K120373)
#### 4. DEVICE DESCRIPTION
The SubQ It!" Bioabsorbable Skin Closure System comprises bioabsorbable fasteners made of 82/18 L-lactide/Glycolide copolymer, a well-established material for resorbable medical devices. The fasteners have two legs with barbed tips (4.3mm long, 0.88mm diameter and spaced 2.75mm center-to-center). The fasteners are inserted using the SubQ It!" Stapler.
#### 5. INDICATION FOR USE/INTENDED USE
The SubQ It!™ Bioabsorbable Skin Closure System is intended for use in humans for abdominal, thoracic, gynecologic, orthopedic, plastic and reconstructive surgery for the subcuticular closure of skin where an absorbable tissue fastener is desired for temporary tissue approximation.
{7}------------------------------------------------
#### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE 6. PREDICATE DEVICE/S
The SubQ It!" Bioabsorbable Skin Closure System is substantially equivalent to the INSORB™ Absorbable Staple when used with the INSORB Subcuticular Skin Stapler manufactured by Incisive Surgical, Inc. The intended use is the same, the fasteners are similar in size and material, with similar strength and mass. Both the new device and the predicate are sterilized by gamma radiation and pre-loaded into a manual stapler, packaged for single patient use.
#### SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL 7. EQUIVALENCE
The SubQ It!" Bioabsorbable Skin Closure System was tested side-by-side with the INSORB™ device in 7cm long incisions in a GLP compliant porcine animal model for 90 days. Data show no significant differences between the SubO It!" and the predicate device.
#### 8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
Clinical testing has not been performed.
#### SUMMARY OF OTHER INFORMATION 9.
The SubQ It!™ Bioabsorbable Skin Closure System was also tested in 1 cm and 2 cm incisions in a GLP compliant porcine animal model, subjected to bench tests, shelflife study of mechanical performance, and packaging tests. These studies demonstrate the safety and effectiveness of the SubQ It!" Bioabsorbable Skin Closure System for its intended use. Additionally, the finished, sterilized product met requirements of biocompatibility tests.
### 10. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
Based on descriptive information and test results, Opus KSD, Inc. has determined that the SubQ It!" Bioabsorbable Skin Closure System is substantially equivalent to the INSORB™ Absorbable Staple manufactured by Incisive Surgical, Inc., that differences between the two products are minor, and raise no new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.