SE-601 SERIES ELECTROCARDIOGRAPH

K131503 · Edan Instruments, Inc. · DPS · Feb 25, 2014 · Cardiovascular

Device Facts

Record IDK131503
Device NameSE-601 SERIES ELECTROCARDIOGRAPH
ApplicantEdan Instruments, Inc.
Product CodeDPS · Cardiovascular
Decision DateFeb 25, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2
AttributesPediatric

Indications for Use

The intended use of SE-601 is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only.

Device Story

SE-601 series (models A, B, C) is a portable, lightweight electrocardiograph. It acquires ECG signals via body surface electrodes; features alphanumeric keyboard, one-touch operation, and internal/external printing. Device supports data archiving via USB/card reader and transmission to PC via Ethernet, Wi-Fi, or serial port. Operated by physicians or trained healthcare professionals in clinical settings. Includes SEMIP software for automatic analysis and diagnostic interpretation, heart rate variability (HRV) analysis, and two-step exercise testing. Output consists of recorded cardiograms and interpretive statements (adults only) provided on an advisory basis to assist clinicians in diagnosing heart disease. Powered by AC/DC or internal rechargeable battery.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including software testing, risk analysis, safety testing, performance testing, and environmental testing.

Technological Characteristics

Portable electrocardiograph with alphanumeric keyboard and internal/external printer. Features include rechargeable battery, AC/DC power, and connectivity via Ethernet, Wi-Fi, or serial port. Employs SEMIP software for automatic ECG analysis and diagnostic interpretation. Class II device (21 CFR 870.2340).

Indications for Use

Indicated for adult and pediatric patients requiring ECG signal acquisition via body surface electrodes. Interpretive statements and measurements are indicated for adult patients only. Intended for use by doctors and trained healthcare professionals in hospitals or healthcare facilities.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ イ ## 510(K)Summary of Safety and Effectiveness : FEB 25 2014 This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92 The 510(k) number is (if known)_ | Submitter: | Edan Instruments, Inc<br>3/F - B. Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>shekou. Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: +86755 26882220<br>Fax:+86 755 26882223 | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact person: | Cherry Sun<br>Edan Instruments, Inc. | | Proprietary Name: | SE-601 Series Electrocardiograph | | Classification<br>information: | 21 CFR 870.2340, Electrocardiograph<br>Class II | | Product code: | DPS | | Review Panel: | Cardiovascular | | Predicate Devices: | PC ECG cleared under K092010<br>Manufacturer: Edan Instruments, Inc.<br>SE-601 series Electrocardiograph cleared under K093869<br>Manufacturer: Edan Instruments, Inc. | | Device Description: | SE-601 series Smart ECG includes three models SE-601A, SE-601B<br>and SE-601C.<br>Device features include as follows: | | | ● Portable, lightweight design<br>● Easy data input and operation<br>● Alphanumeric keyboard and one-touch operation<br>● Built-in rechargeable battery, AC/DC power supply<br>● Automatic analysis and diagnostic software (SEMIP) for adults<br>and pediatrics<br>● Two-step exercise test with periodic recording<br>● Heart rate variability (HRV) analysis | {2}------------------------------------------------ Intended Use: - . Internal thermal printer and external printer - . Support external archiving: USB flash disk, card reader - . Data transmission to PC via Ethernet . WIFI or serial port This submission is to modify the indication for use (intended use) by adding the applicable pediatric population. The intended use of SE-601 is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only. ## Test Summary: The following quality assurance measures were applied to the development of the SE-601 Series Electrocardiograph: - 0 Software testing - 0 Risk analysis - 0 Safety testing - . Performance testing - . Environmental testing ## Conclusion: Verification and validation testing was done on SE-601 Series Electrocardiograph. This premarket notification submission demonstrates that the subject device SE-601 Series Electrocardiograph is substantially equivalent to the predicate device above mentioned 24 {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three wavy lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 25, 2014 Edan Instruments, Inc. Ms. Cherry Sun Certification Engineer Nanhai Road 1019# Nanshen Shenzhen, Guangdong, China 518067 Re: K131503 > Trade/Device Name: SE-601 Series Electrocardiograph Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: January 21, 2014 Received: January 28, 2014 Dear Ms. Cherry Sun: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Ms. Cherry Sun CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. 51 for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indication for Use 510(k) Number: Device Name: SE-601 Series Electrocardiograph The intended use of SE-601 is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is only intended to be used on adult patients and is offered to clinicians on an advisory basis only. Prescription Use (21 CFR Part 801 Subpart D) Over the Counter Use _ (21 CFR Part 801 Subpart C) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Or Concur for B. Zuckerman Date: 11/01/2014 14:02:25 13:39:48:05'00'
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...