MODIFIED FLOWGATE BALLOON GUIDE CATHETER

K131492 · Concentric Medical, Inc. · DQY · Oct 3, 2013 · Cardiovascular

Device Facts

Record IDK131492
Device NameMODIFIED FLOWGATE BALLOON GUIDE CATHETER
ApplicantConcentric Medical, Inc.
Product CodeDQY · Cardiovascular
Decision DateOct 3, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Concentric Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

Device Story

Modified FlowGate Balloon Guide Catheter is a coaxial-lumen, braid-reinforced, variable stiffness catheter. Device features distal radiopaque marker for angiographic visualization and compliant balloon for temporary vascular occlusion. Proximal end includes bifurcated luer hub for flushing, inflation, and aspiration. Used in clinical settings by physicians to facilitate insertion/guidance of intravascular catheters and as a conduit for retrieval devices in peripheral and neurovascular procedures. Accessories include dilator, hemostasis valves, sheaths, and extension tubing. Device provides temporary vascular occlusion to assist in angiographic procedures, potentially improving procedural outcomes.

Clinical Evidence

Bench testing only. No clinical data. Verification and validation included dimensional verification, simulated use in neurovascular model, balloon performance (symmetry, compliance, fatigue, burst/leakage), tensile strength, shaft flexibility, tip deflection, tip patency, kink resistance, and accessory compatibility. Biocompatibility testing (ISO 10993) confirmed non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic, and non-hemolytic properties.

Technological Characteristics

Coaxial-lumen, braid-reinforced (stainless steel) catheter. Materials: Pebax (shaft/tip), Silicone (balloon), Platinum/Iridium (marker), Polyurethane (hub), Polyolefin/Pebax (strain relief). Cyanoacrylate used for braid securement. Sterilization: Ethylene Oxide (EtO). Dimensions: 7F, 8F, or 9F. Connectivity: None (mechanical device).

Indications for Use

Indicated for patients requiring intravascular catheter insertion and guidance in peripheral and neurovascular systems, temporary vascular occlusion during angiographic procedures, and as a conduit for retrieval devices.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) Summary | Trade Name: | Modified Balloon Guide Catheter | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Percutaneous Catheter | | Classification Name: | Percutaneous Catheter, 21CFR 870.1250 Class II | | Product Code: | DQY | | Submitter: | Concentric Medical, Inc.<br>301 E. Evelyn Avenue<br>Mountain View, CA 94041<br>Tel 650-938-2100<br>Fax 650-237-5230<br>Facility Registration #2954917 | | Contact: | Rhoda Santos<br>Principal Regulatory Affairs Specialist | | Date Prepared: | September 27, 2013 | | Predicate Device: | Concentric Balloon Guide Catheter (K112404)<br>Modified Concentric Balloon Guide Catheter (K122581) | #### Device Description Like the predicate device, the Modified FlowGate™ Balloon Guide Catheter is a coaxiallumen, braid-reinforced, variable stiffness catheter designed for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. A radiopaque marker is included on the distal end for angiographic visualization. A compliant balloon is mounted on the distal end to provide temporary vascular occlusion during angiographic procedures. A bifurcated luer hub on the proximal end allows attachments for flushing, inflation and aspiration. Modified FlowGate™ Balloon Guide Catheter dimensions and maximum recommended balloon inflation volume are indicated on product label. #### Accessories The Modified FlowGate™ Balloon Guide Catheter is packaged with a Dilator, Rotating Hemostasis Valve, Tuohy Borst Valve with sideport, Peel Away Sheaths, Luer-Activated Valve, and Extension tubing. {1}------------------------------------------------ #### Indications for Use The Indications for Use are the identical to that of the predicate device and are as follows: The Concentric Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices. #### Technological Characteristics and Product Feature Comparison: The subject device, Modified FlowGate Balloon Guide Catheters is substantially equivalent to the predicate device in terms of: indications for use materials fundamental scientific technology fundamental design materials and processes for packaging and sterilization of devices A tabular comparison of the specific technological characteristics between the predicate devices and subject device is provided below. {2}------------------------------------------------ . : : . | Product Feature Comparison of Subject Device with Predicate Devices (K112404 and | | | |----------------------------------------------------------------------------------|--|--| | K122581) | | | | | Feature | Predicate Device,<br>K112404 | Results | | |---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | | | | Predicate Device,<br>K122581 | Subject Device,<br>K131492 | | | Indications for<br>Use | The Balloon Guide<br>Catheter is indicated for<br>use in facilitating the<br>insertion and guidance of<br>an intravascular catheter<br>into a selected blood<br>vessel in the peripheral<br>and neuro vascular<br>systems. The balloon<br>provides temporary<br>vascular occlusion during<br>these and other<br>angiographic procedures.<br>The Balloon Guide<br>Catheter is also indicated<br>for use as a conduit for<br>Retrieval devices. | Same with reference to<br>marketed name as<br>FlowGate™ Balloon<br>Guide Catheter | Same as predicate<br>device, K122581 | | | Device<br>Description | The Concentric Balloon<br>Guide Catheter is a<br>coaxial-lumen, braided<br>shaft, variable stiffness<br>catheter with a<br>radiopaque marker on the<br>distal end and a<br>bifurcated luer hub on the<br>proximal end. A<br>compliant balloon is<br>mounted on the distal<br>end. BGC dimensions<br>and maximum<br>recommended balloon<br>inflation volume are<br>indicated on product<br>label. If indicated on the<br>product label, a dilator is<br>provided for transition to<br>the guidewire. | Same with reference to<br>marketed name as<br>FlowGate™ Balloon<br>Guide Catheter | Same as predicate<br>device, K122581 | | | Outer/Inner<br>Jacket | Pebax | Same with revisions to<br>transition lengths and<br>material durometers | Same material durometers<br>as K122581, with<br>revisions to transition<br>lengths | | | Results | | | | | Feature | Predicate Device,<br>K112404 | Predicate Device,<br>K122581 | Subject Device,<br>K131492 | | | Distal Tip | Pebax | Same with different<br>durometer material than<br>predicate device,<br>K112404 | Same with different<br>durometer material than<br>predicate devices | | | Balloon.<br>Material | Silicone | Same with different<br>material durometer than<br>predicate device,<br>K112404 | Same with same material<br>durometer as predicate<br>device, K112404 | | | Braid | Stainless Steel | Same | Same | | | Braid distal end<br>securement | PET heat shrink | Acrylic (Acrylated<br>Urethane) | Cyanoacrylate, with<br>different viscosity than<br>Cyanoacrylate used on<br>predicate devices | | | Marker Band | Platinum/Iridium | Same | Same | | | Catheter Hub | Polyurethane | Same | Same | | | Strain Relief | Polyolefin | Same | Additional Pebax strain<br>relief included with<br>subject device | | | Labeled Shaft<br>Inner / Outer<br>Diameter<br>(nominal) | Inner Diameter:<br>0.059" (7F), 0.078" (8F)<br>or 0.085" (9F)<br><br>Outer Diameter:<br>0.088 "(7F), 0.104 "(8F)<br>or 0.116" (9F) | Inner Diameter:<br>0.084in<br><br>Outer Diameter:<br>0.106in | Same as predicate<br>device, K122581 | | | Maximum outer<br>diameter along<br>effective length | 0.0945"(7F),<br>0.1080"(8F) or<br>0.1205" (9F) | 0.1080in | Same as predicate<br>device, K122581 | | | Effective<br>Lengths | 80cm or 95cm | 90cm or 100cm | 85cm and 95cm | | | Distal Tip<br>length | 1.5mm | 2mm | 0.75mm | | | Radiopaque<br>Marker Length<br>(nominal) | 0.0400in | 0.0200in | Same as predicate<br>device, K122581 | | | | Results | | | | | Feature | Predicate Device,<br>K112404 | Predicate Device,<br>K122581 | Subject Device,<br>K131492 | | | Accessory<br>Devices<br>Provided | Dilator | Dilator<br>Peel Away Sheath<br>Rotating Hemostatic<br>Valve<br>Tuohy Borst Valve<br>with Sideport<br>Luer Activated Valve | Same as predicate<br>device, K122581 with<br>addition of Extension<br>Tubing | | | Packaging<br>Materials and<br>Configuration | Polyethylene Tube and<br>HDPE Packaging Card<br>Twek/PE/PET Pouch | Polyethylene Tube and<br>HDPE Packaging Card<br>Tyvek/LLDPE Pouch | Same as predicate<br>device, K122581 | | | Sterilization<br>Method | Eto | Same | Same | | | How Supplied | Sterile, Single Use | Same | Same | | . {3}------------------------------------------------ ## Product Feature Comparison of Subject Device with Predicate Devices (K112404 and K122581) {4}------------------------------------------------ ## Product Feature Comparison of Subject Device with Predicate Devices (K112404 and K122581) #### Risk Assessment Risk assessment of the modifications has been conducted in accordance with EN ISO 14971:2012. Concentric Medical, Inc. has determined the modifications to the predicate devices raise no new questions of safety or effectiveness. Results of verification and validation testing have demonstrated the Modified FlowGate Balloon Guide Catheters are substantially equivalent to the predicate devices. Furthermore, the modifications did not result in any new failure modes nor were there any changes to existing failure modes. {5}------------------------------------------------ #### Testing Summary The results of verification and validation conducted on the subject device demonstrate that it performs as designed, is suitable for its intended use and is substantially equivalent to the predicate devices. Specifically, the following tests were performed and successfully evaluated on the subject device and accessories: Dimensional Verification: the ability of the device to meet predetermined dimensional requirements Simulated Use Testing: the ability of the device and accessories to be used per procedural instructions outlined in the Instructions for Use in a neurovascular model Balloon Testing: the symmetry and compliance of the distal balloon; balloon inflation and deflation rates; balloon fatigue strength; constrained balloon burst and leakage Tensile Testing: the mechanical integrity of the device under tensile loads Distal Shaft Flexibility Testing: the force to flex the distal shaft of the device Soft Tip Deflection Testing: the force to deflect the distal edge of the soft tip Tip Patency Testing: the ability of the device distal tip to maintain patency under aspiration Kink Resistance: the ability of the device shaft to resist kinking Hub Luer Fitting Testing: compatibility of the Extension Tubing accessory with the hub luer fitting Extension Tubing Aspiration: the ability to aspirate using the Extension Tubing Accessory Packaging Verification: the ability of the packaging to protect the finished device following climatic conditioning and distribution simulation Shelf life testing (Product and Packaging) {6}------------------------------------------------ The subject device uses the same materials and processes as the predicate devices (K112404 and K122581). Biocompatibility tests were performed on the predicate devices (K112404 and K122581); results for all tests met the acceptance criteria and apply to the subject device. In addition, some biocompatibility screening tests were also performed on the subject device. The biocompatibility tests conducted on the predicate device K122581 and the subject device are summarized in the tables below: | ISO 10993<br>Standard | Test | Results | Conclusion | |-----------------------|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------| | ISO 10993-10 | Sensitization<br>Kligman Maximization<br>Test - ISO | The USP 0.9% Sodium Chloride<br>for injection (NaCl) and<br>Cottonseed Oil (CSO) extracts of<br>the test article showed no<br>evidence of causing delayed<br>dermal contact sensitization. | Non-sensitizer | | ISO 10993-5 | Cytotoxicity<br>L929 MEM Elution Test | No biological activity (Grade 0) was<br>observed in the L929 mammalian cells<br>at 48 hours post exposure to the test<br>article extract. The observed cellular<br>response obtained from the positive<br>control article extract (Grade 4) and<br>negative control article extract (Grade<br>0) confirmed the suitability of the test<br>system. | Non-cytotoxic | | ISO 10993-10 | Intracutaneous Reactivity<br>Intracutaneous Injection<br>Test | The sites injected with the USP 0.9%<br>Sodium Chloride for Injection (NaCl)<br>and Cottonseed Oil (CSO) extracts of<br>the test article did not show a<br>significantly greater biological<br>reaction than the sites injected with<br>the control extracts. The difference of<br>the overall mean score between the<br>test article and the control article was<br>0.0. | Non-irritant | | ISO 10993-11 | Systemic Toxicity (Acute)<br>Systemic Injection Test | The 0.9% Sodium Chloride for<br>Injection (NaCl) and Cottonseed Oil<br>(CSO) extracts of the test article did<br>not induce a significantly greater<br>biological reaction than the control<br>extracts, when tested in Swiss Albino<br>mice. | No acute<br>systemic<br>toxicity | Biocompatibility Tests, Results, Conclusions (Predicate device, K122581) {7}------------------------------------------------ | ISO 10993<br>Standard | Test | Results | Conclusion | |------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------| | ISO 10993-11<br>USP 34 <151> | Systemic Toxicity<br>Rabbit Pyrogen Test,<br>Material Mediated | No rabbit injected with the USP 0.9%<br>Sodium Chloride for Injection (NaCl)<br>extract of the test article showed an<br>individual rise in temperature of 0.5°C<br>or more above the baseline<br>temperature. | Non-pyrogenic | | ASTM F756 | Hemocompatibility<br>Hemolysis - Rabbit Blood<br>- ASTM Direct and<br>Indirect Contact | The test article exhibited 0.17%<br>hemolysis above the level of<br>hemolysis exhibited by the negative<br>control via the direct method and<br>0.12% hemolysis above the level of<br>hemolysis exhibited by the negative<br>control via the indirect method. | Non-hemolytic | | ISO 10993-4 | Hemocompatibility<br>Complement<br>Activation Assay | The plasma exposed to test article<br>exhibited no statistically significant<br>increase in C3a when compared to<br>the untreated control and the negative<br>control after 90 minute exposure.<br>The plasma exposed to test article<br>exhibited a statistically significant<br>increase in SC5b-9 when compared to<br>the untreated control and the negative<br>control after 90 minute exposure. | Minimal<br>complement<br>activation<br>potential | | ISO 10993-4 | Hemocompatibility<br>Thrombogenicity Study in<br>Dogs | No significant thrombosis with a Grade<br>of 0 (a very small clot is acceptable)<br>was observed in 1 out of 2 test sites<br>and 1 out of 2 control sites. Minimal<br>thrombosis with a Grade of 1<br>(thrombus found at one location) was<br>observed in 1 out of 2 test sites and 1<br>out of 2 control sites. | No significant<br>thrombosis | ## Biocompatibility Tests Results Conclusions (Predicate device, K122581), continued : . : : , {8}------------------------------------------------ | ISO 10993<br>Standard | Test | Results | Conclusion | |-----------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------| | ISO 10993-5 | Cytotoxicity<br>L929 MEM Elution Test | No cytotoxicity or cell lysis was noted<br>in any of the test wells. No pH shift<br>was observed at 48 hours. The<br>reagent control, negative control and<br>the positive control performed as<br>anticipated. | Non-cytotoxic | | ISO 10993-4 | Hemocompatibility<br>ASTM Hemolysis (Direct<br>contact and extract test) | The hemolytic index for the test article<br>in direct contact with blood was 0.0%<br>and the hemolytic index for the test<br>article extract was 0.4% | Non-hemolytic | | ISO 10993-18 | Fourier Transform<br>Infrared (FTIR) Scan<br>(+ control sample) | The distal 1mm white tip was scanned<br>and compared to the control sample.<br>The samples had a match of 97.90%. | Materials for<br>test article and<br>control sample<br>are equivalent. | | ISO 10993-18 | Physicochemical Tests<br>Plastics (USP <661>) | Non-volatile residue: 1 mg<br>Residue on ignition: < 1 mg<br>Heavy metal: < 1 ppm<br>Buffering capacity: < 1.0 ml | Met USP <661><br>requirements | ・ . 11. 11. . . ## Biocompatibility Screening Tests. Results, Conclusions (Subject device, K131492) : The biocompatibility results demonstrate that the subject device is biocompatible. : {9}------------------------------------------------ ## Summary of Substantial Equivalence The subject device, Modified FlowGate Balloon Guide Catheter, is substantially equivalent to the predicate devices with regard to device design, materials, intended use, and patient population. The conclusions drawn from the risk assessments, verification and validation testing conducted demonstrate that the subject device performs as designed, is as safe, as effective, and performs as well as or better than the predicate devices. {10}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002 ## October 3, 2013 Concentric Medical, Inc. % Ms. Rhoda Santos Principal Regulatory Affairs Specialist 301 East Evelyn Avenue Mountain View, CA 94041 Re: K131492 Trade/Device Name: Modified FlowGate™ Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutancous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 21, 2013 Received: August 22, 2013 Dear Ms. Santos: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {11}------------------------------------------------ Page 2 - Ms. Rhoda Santos device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Joyce M. Whang -S Victor Krauthamer, Ph.D. for Acting Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {12}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K131492 Device Name: Modified FlowGate™ Balloon Guide Catheter Indications For Use: The Concentric Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices. Prescription Use كي__________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) Joyce M. Whang -S Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...