K131457 · Shenzhen Jumper Medical Equipment Co., Ltd. · KNG · Jan 24, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K131457
Device Name
FETAL ULTRASONIC MONITOR AND ACCESSORIES
Applicant
Shenzhen Jumper Medical Equipment Co., Ltd.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Jan 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
The device is ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat. It can display values of fetal heart rate.
Device Story
Hand-held, battery-powered ultrasonic Doppler device; utilizes 2.5 MHz probe to detect fetal heart beats; processes acoustic signals to provide audio output via built-in speaker and visual display of fetal heart rate; intended for use by clinicians or patients in hospital or home settings at/after 12 weeks gestation; assists healthcare providers in monitoring fetal status; provides non-invasive fetal heart rate detection.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance requirements (IEC 61266), and biocompatibility (ISO 10993-1, -5, -10).
Technological Characteristics
Hand-held fetal Doppler; 2.5 MHz CW transducer; Ni-HM rechargeable battery powered; ABS and PP plastic construction; complies with ISO 10993 for biocompatibility; acoustic output <20 mW/cm²; standalone device.
Indications for Use
Indicated for use by a licensed physician in hospitals and homecare environments for detecting fetal heart beats at or after 12 weeks gestation.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
2 4 2014
| Shenzhen Jumper Medical Equipment Co., Ltd. | VOL 005:001 510K Summary<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | JAN |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Proposed product: Fetal Doppler | Version:A/1 | |
# VOL 05 510(k) Summary
This summary of 510(K) safety and effectiveness information is submitted As Required by requirements of SMDA and 21 CFR §807.92.
# 5.1 Administrative Information
.
| Date of Summary<br>prepared | May, 09, 2013 |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Company title:<br>Shenzhen Jumper Medical Equipment Co., Ltd<br>Company address:<br>5th Floor, Building No.34, Baoyuan Industrial<br>Zone, Xixiang Street, Baoan District, Shenzhen<br>518102, P.R. China.<br>Phone: +86-755-2669 2192<br>Fax: +86-755-2685 2025<br>Contact Person:<br>E-mail: info@jumper-medical.com |
| Submission<br>Correspondent | Shenzhen ZYTC Consulting Co., Ltd.<br>4th Floor, Jinhui Building, Nanhai BLVD,<br>Nanshan District, Shenzhen, Guangdong,<br>China.<br>Contact person: Mr. Field.Fu<br>E-Mail: cefda13485@163.com |
| Establishment registration<br>number | 3008973759 |
## 5.2 Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|----------------------------|
| Trade Name: | Fetal doppler |
| Model: | JPD-100B |
| Classification name: | Monitor, ultrasonic, fetal |
| Review Panel: | Obstetrics/Gynecology |
| Product Code: | KNG |
{1}------------------------------------------------
VOL_005:001_510K Summary Shenzhen Jumper Medical Equipment Co., Ltd. Version:A/1 Proposed product: Fetal Doppler
> Device Class: 11 Requlation Number: | 884.2660
### 5.3 Predicate Device Information
Beijing Choice Electronic Technology Co., Ltd. Sponsor: Fetal Doppler MD800 Device: 510(K) Number: | K112911
### 5.4 Device Description
JPD-100B Fetal Doppler is device prescribed by a licensed physician for use in hospitals and the homecare environment. It is a hand-held, battery powered audio Doppler device integrated with 2.5 MHz probe, used for detecting fetal heart beats. And the device is for prescription use and is intended for use at or after 12 weeks gestation.
#### 5.5 Intended Use
The device is ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat. It can display values of fetal heart rate.
#### 5.6 Indications for Use:
The Fetal Doppler JPD-100B is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats.
# 5.7 Technological characteristics of the proposed device compared to the predicate device
The proposed device and the Predicate device have the same intended use, design principle, and similar material composition. The differences do not exert adverse effect on the proposed device. An itemized comparison between the proposed device and the cited predicate devices demonstrates that the proposed device is substantially equivalent to the predicate devices. The differences and/or similarities are as follow:
- 1) In terms of intended use, differences in wording are minor and do not impact the equivalence of the statements;
- 2) Energy used: the predicate is powered by alkaline batteries, and the subject device is powered by Ni-HM rechargeable batteries, the differences in battery type do not have a significant impact on device operation;
{2}------------------------------------------------
Shenzhen Jumper Medical Equipment Co., Ltd.
VOL_005:001_510K Summary
Version:A/1
Proposed product: Fetal Doppler
3) The working frequency of transducer:
- . 2MHz for the predicate device
- 2.5MHz for the subject. .
- 3MHz another model on the market, JPD-100S fetal Doppler . (Shenzhen Jumper, K110124) ,
2.5MHz falls in the range of 2MHZ to 3MHz. thus, it is believed to be similar to products on the market;
### 4) Materials:
- · patient contact materials of predicate are unknown,
- · materials for subject device are either
- o the same as the aforementioned Doppler JPD-100S, K110124 (for ABS),
- o Complied with ISO 10993-5 and -10. (for PP);
- 5) Acoustic output: The output is different, but the value of both the predicate and the subject are less than 20 mW/cm², complies with sub-section 2.1.2 of FDA Guidance --- Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers.
## 5.8 Brief discussion of the nonclinical tests
JPD-100B Fetal Doppler conforms to the following standards:
- � IEC 60601-1, Medical electrical equipment Part 1: General requirements for basic safety and essential performance, 2005, CORR. 1:2006 + CORR.2. 2007:
- � IEC 60601-1-2:2007 Medical electrical equipment -- Part 1-2: General requirements for safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests;
- � IEC 61266:1994 Ultrasonic - Hand-held probe Doppler fetal heartbeat detectors -Performance requirements and methods of measurement and reportina:
- & ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process;
- & ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for In Vitro cytotoxicity;
- & ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
{3}------------------------------------------------
Shenzhen Jumper Medical Equipment Co., Ltd.
VOL_005:001_510K Summary
Proposed product: Fetal Doppler Version:A/1
## 5.9 Brief discussion of clinical tests
Not applicable.
# 5.10 Other information (such as required by FDA guidance)
No other information.
#### 5.11 Conclusions
The subject device JPD-100B Fetal Doppler is substantially equivalent to MD800 Fetal Doppler whose 510(k) number is K112911.
#### Page 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### January 24, 2014
Shenzhen Jumper Medical Equipment Co., Ltd. % Field Fu Shenzhen ZYTC Consulting Co., Ltd. 4th Floor, Jinhui Building, Nanhai BLVD, Nanshan District Shenzhen, Guangdong 518052 China
Re: K131457
> Trade/Device Name: Fetal Doppler JPD-100B Regulation Number: 21 CFR§ 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: December 6, 2013 Received: December 23, 2013
Dear Field Fu,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
#### Page 2 - Field Fu
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
in meetings or teleconferences. Requires attention to detail. Requires confident
Sincerely yours.
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
VOL_004:001_Indications for Use Shenzhen Jumper Medical Equipment Co., Ltd. Proposed product: Fetal Doppler Version:A/1
# VOL 04 Indications for Use
510(k) Number (if known): K131457
Device Name: Fetal Doppler JPD-100B
Indications for Use:
..............................................................................................................................................................................
The Fetal Doppler JPD-100B is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats.
Prescription Use _ AND/OR 风 (Part 21 CFR 801 Subpart D)
Over-The-Counter Use : (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner 2014.01.24 14:29:02 =05'00'
Page 1 of 2
{7}------------------------------------------------
# Diagnostic Ultrasound Indications For Use
Device Name: Fetal Doppler JPD-100B Transducer: 2.5MHz CW inherent of the main unit Intended Use: Detect fetal heart beats as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|-------------------------|-------------------|---|-----|-----|------------------|----------|---------------------|
| General<br>(Track 1 | Specific<br>(Tracks 1 & | B | M | PWD | CWD | Color<br>Doppler | Combined | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | N | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| Fetal<br>Imaging<br>& Other | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| Cardiac | Cardiac | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| Peripheral<br>Vessel | Peripheral | | | | | | | |
| | Peripheral | | | | | | | |
ട * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner 2014.01.24 14:34:11 -05'00' Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.