K110124 · Shenzhen Jumper Medical Equipment Co., Ltd. · KNG · Feb 4, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K110124
Device Name
FETAL DOPPLER JPD-100S
Applicant
Shenzhen Jumper Medical Equipment Co., Ltd.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Feb 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Fetal Doppler JPD-100S is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats.
Device Story
JPD-100S is a hand-held, battery-powered audio Doppler device; utilizes 3MHz continuous wave (CW) ultrasound probe to detect fetal heart beats. Operated by licensed physicians in hospitals, clinics, or private offices. Device transforms reflected ultrasonic signals into audible sound; healthcare provider listens to output to confirm fetal heart rate. Provides non-invasive monitoring of fetal status during pregnancy (12+ weeks). Benefits include simple, portable assessment of fetal well-being in clinical settings.
Clinical Evidence
Bench testing only. Validated against IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 61266 (ultrasonic performance), and ISO 10993-5/10 (biocompatibility). Acoustic output testing confirmed compliance with standard criteria for 3MHz transducer and Isata levels.
Technological Characteristics
Hand-held, battery-powered audio Doppler. 3MHz CW transducer. Materials evaluated for biocompatibility per ISO 10993-5 and ISO 10993-10. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 61266. Standalone device.
Indications for Use
Indicated for detection of fetal heart beats in pregnant patients at 12+ weeks gestation. Prescription use only in hospitals, clinics, and private offices.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
Predicate Devices
Contec Pocket Fetal Doppler Baby Sound A (K082480)
{0}------------------------------------------------
. . .. ..
.
K110124
# Attachment II 510(k) Summary
As Required by CFR 870.92
| Preparing Date | November 23, 2010 | FEB - 4 2011 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Sponsor | Shenzhen Jumper Medical Equipment Co., Ltd | |
| | 5th Floor, Building No.34, Baoyuan Industrial Zone, Xixiang Street, | |
| | Baoan District, Shenzhen, Guangdong 518067, P.R. China | |
| | Boley He Quality Manager | |
| | Tel: +86-755-2669 6279 | |
| | Fax: +86-755-2685 2678 | |
| | Email: rd@jumper-medical.com | |
| Submission<br>Correspondent | Ms. Diana Hong / Mr. Lee Fu | |
| | Shanghai Mid-Link Business Consulting Co., Ltd | |
| | P.O. BOX 237-023 | |
| | Shanghai, 200237, China | |
| | Tel: +86-21-64264467 | |
| | Fax: 240-238-7587 | |
| | Email: diana.hong@mid-link.net | |
| Proposed Product | | |
| Trade Name | Fetal Doppler | |
| Model | JPD-100S | |
| Product Code: | KNG | |
| Regulation Number: | 21 CFR 884.2660 | |
| Device Class: | Class II | |
| Submission Purpose: | New Device | |
| Submission Type | Traditional 510(k) | |
| Predicate Device: | | |
| Device Name | Contec Pocket Fetal Doppler Baby Sound A | |
| K Number | K082480 | |
| Manufacturer | Contec Medical Systems Co., LTD | |
| Intended Use | The Fetal Doppler JPD-100S is a hand-held, battery powered audio | |
| | Doppler device used for detecting fetal heart beats. | |
| Device Description | JPD-100S Fetal Doppler is a prescription from licensed physician in<br>hospitals, clinics and private offices. It is a hand-held, battery powered<br>audio Doppler device integrated with 3MHz probe, used for detecting<br>fetal heart beats. And the device is normally applied to pregnancy for<br>more than twelve weeks. | |
| Test Conclusion | Laboratory testing was conducted to validate and verify that the<br>proposed devices met all design specifications, including electrical<br>safety, EMC, specification per following standards. | |
| | IEC60601-1(1988; Amendment 1, 1991-11, Amendment 2, 1995):<br>Medical electrical equipment -Part 1: General requirements for basic<br>safety and essential performance<br>IEC60601-1-2 (2001): Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential performance -<br>Collateral standard: Electromagnetic compatibility -Requirements and<br>tests<br>IEC 61266 (1994 ): Ultrasonic - Hand-held probe Doppler fetal<br>heartbeat detectors -Performance requirements and methods of<br>measurement and reporting<br>ISO 10993-5 (2009): Biological evaluation of medical devices -- Part<br>5: Tests for In Vitro cytotoxicity<br>ISO 10993-10 (2002; Amd. 1:2006): Biological evaluation of medical<br>devices - Part 10: Tests for irritation and delayed-type hypersensitivity<br>Results of these tests demonstrate compliance to the requirements of<br>all consensus standards. | |
| Comparison Summary | The proposed device shares same classification, intended use, similar<br>technical specifications, same safety specifications and skin-contact<br>material with the predicate device. The main difference includes<br>battery, operation and storage conditions, transducer frequency and<br>Isata, but the difference is considered to have no effect on<br>effectiveness and safe. The SE analysis of the difference is provided as<br>following.<br>The batteries of proposed and predicate device are different. But both<br>of them comply with the requirements of electrical safety and EMC of<br>IEC 60601-1 and IEC 60601-1-2. Therefore, this difference is | |
.
{1}------------------------------------------------
イ・・・・
11-2
{2}------------------------------------------------
considered has no effect on effectiveness and safety.
The operation and storage conditions of the proposed device have slight differences with those of the predicate device. But these conditions are considered to be able to cover the normal working and storage environment. In addition, these conditions are clearly on the proposed labeling, to remind the user handle and operate with these conditions. Therefore, this difference is considered to have no effect on effectiveness and safe.
The transducer frequency of proposed device and predicate device is different. But the proposed transducer was tested for its acoustic output, and the result demonstrates that it complies with the standard criteria. In addition, 3MHz transducer of fetal doppler is very common in clinical use, such as, 3MHz normal transducer of Contec Pocket Fetal Doppler SONOLINE A and SONOLINE B (k082480). Therefore, this difference is considered to have no effect on effectiveness and safe.
The Isata of proposed device and predicate device is different. But the proposed transducer was tested for its acoustic output, and the result demonstrates that it complies with the standard criteria. In addition, transducer of fetal doppler with Isata <5 mW/cm² is very common in clinical use, such as, CMS-150-T2 and CMS-150-T5 transducer of Contec Pocket Fetal Doppler SONOLINE A and SONOLINE B (k082480). Therefore, this difference is considered to have no effect on effectiveness and safe.
SE Determination
The proposed device, Fetal Doppler JPD-100S, is substantially equivalent (SE) to the predicate device Contec Pocket Fetal Doppler Baby Sound A (K082480).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - U.S.A." is arranged in a circular pattern around the bird-like figure. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Shenzhen Jumper Medical Equipment Co., Ltd % Mr. Marc M. Mouser Engineering Leader/FDA Office Coordinator Underwriters Laboratories, Inc. 2600 N.W. Lake Road CAMAS WA 98607-8542
FEB - 4 2011
Re: K110124
Trade/Device Name: Fetal Doppler JPD-100s Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasound monitor and accessories Regulatory Class: II Product Code: KNG Dated: December 17, 2010 Received: January 18, 2011
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Fetal Doppler JPD-100s, as described in your premarket notification:
#### Transducer Model Number
### 3.0MHz CW
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
and the country of the county of
{4}------------------------------------------------
Page 2 - Mr. Mouser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Lauren Hefner at (301) 796-6881.
Sincerely Yours,
Amharl. D. O'Rhum for
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
## Section II Indications for Use Statement
510(k) Number: Device Name:
Fetal Doppier JPD-100S
Indications for Use:
The Fetal Doppler JPD-100S is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats.
Prescription Use > X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
---
(Division Sign Off)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K110124
Page 1 of 2
{6}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
Device Name: Fetal Doppler JPD-100S
Transducer: 3.0MHz CW inherent of the main unit
Intended Use: Detect fetal heart beats as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|--|
| General<br>(Track 1 only) | Specific (Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | N | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (Specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (Specify) | | | | | | | | |
N= new indication: P= previously cleared by FDA; E= added under Appendix E Comments:
(PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
_Page 2 of 2 II-2
Division of Radiological Devices Office of In Vitro Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K110621
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.