BIPOLAR FORCEPS

K131058 · Boston Endo-Surgical Technologies, Inc. · GEI · Oct 25, 2013 · General, Plastic Surgery

Device Facts

Record IDK131058
Device NameBIPOLAR FORCEPS
ApplicantBoston Endo-Surgical Technologies, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateOct 25, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

Tissue grasping and control of bleeding of small vessels by bipolar coagulation.

Device Story

Single-use, hand-held, insulated bipolar forceps; 295mm length; 190mm grasp-to-tip distance; 1.5cm bayonet offset. Titanium Nitride coated tips; 1.5mm x 1cm contact surfaces. Available in straight or 20-degree angled configurations. Used in general surgery by trained clinicians to grasp tissue and control bleeding via bipolar electrosurgical coagulation. Connects to compatible electrosurgical generators (IEC 60601 compliant) via reusable pin plug cable (1.8mm pin, 5.8mm spacing). Rated at 1000V peak; maximum 70 watts. Device provides mechanical grasping and electrical coagulation path; clinician observes tissue response to determine coagulation endpoint. Benefits include sterile, single-use convenience, eliminating reprocessing requirements associated with reusable forceps.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 (cytotoxicity, irritation, sensitization). Performance testing conducted using animal tissue of various diameters to simulate vessels at different power settings; device successfully cut and coagulated tissue. Additional testing included electrical safety, packaging validation, shipping/transportation, shelf life, and Hi-Pot testing.

Technological Characteristics

Titanium Nitride coated tips; insulated handle; 295mm length. Energy source: external electrosurgical generator (max 70W, 1000V peak). Connectivity: compatible with 1.8mm pin plug cables. Single-use, sterile, double-blister packaging with Tyvek lid. No software or electronic components.

Indications for Use

Indicated for tissue grasping and bipolar coagulation of small vessels in general surgical procedures. For use by trained personnel only.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K131058 ## 510(k) Summary for Boston Endo Surgical Technologies (division of Lacey Manufacturing) Bipolar Forceps #### SUBMITTER/510(K) HOLDER 1. Boston Endo Surgical Technologies (division of Lacey Manufacturing) 1146 Barnum Avenue Bridgeport, CT 06610 Contact Name: James Rogers. Email: JRogers@peplacey.com Telephone: (203) 336-7453 Date Revised: July 1, 2013 Prepared By: JA #### 2. DEVICE NAME | Proprietary Name: | Bipolar Forceps | |----------------------|-----------------------------------------------------| | Common/Usual Name: | Bipolar Forceps | | Classification: | Class 2, 21 CFR 878.4400 | | Classification Name: | Electrosurgical Cutting and Coagulation Accessories | | Product Code: | GEI | #### 3. PREDICATE DEVICES K032327 Modern Medical Bipolar Forceps . #### DEVICE DESCRIPTION 4. The Bipolar Forceps device is a single use hand held insulated medical device approximately 295mm in overall length featuring a 190mm distance from the user's grasp to the tip. The bayonet offset is 1.5cm. The tips are Titanium Nitride coated and have polished contact surfaces 1.5mm wide by 1cm long. The device is offered in straight and 20 degree angled tips. The Bipolar Forceps are sold as a single use sterile packaged device used in general surgical procedures. The Bipolar Forceps are limited to maximum 70 watts and shall be used only by personnel with training in electrosurgical procedures and with compatible generators compliant with IEC 60601 requirements. The Bipolar Forceps are rated at 1000V peak and can be used with a compatible reusable pin plug cable (1.8mm {1}------------------------------------------------ pin, 5.8mm spacing). The Bipolar Forceps are single use disposable devices packaged in double blisters with Tyvek lid. #### INTENDED USE 5. Tissue grasping and control of bleeding of small vessels by bipolar coagulation. #### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6. The Bipolar Forceps substantially equivalent to the Modern Medical Bipolar Forceps cleared under 510(k) K032327. The only difference between the devices is that the subject device is provided sterile and is for one time use, whereas the predicate device is reusable. The materials and design are identical. #### 7. PERFORMANCE TESTING Biocompatibility testing for the intended application of the Bipolar Forceps was performed as suggested by the International Standards Organization (ISO) 10993 Guidelines, FDA General Program Memorandum No. G95-1 and the Office of Device Evaluation (ODE) Bluebook Memorandum G95-1 "Use of ISO-10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". | Test | Conclusion | |---------------|-----------------| | Cytotoxicity | Non-cytotoxic | | Irritation | Non-irritating | | Sensitization | Non-sensitizing | The device was also subjected to performance testing after sterilization using animal tissue of various diameters to simulate vessels at different power settings. The device properly cut and coagulated the tissue. ### 8. SAFETY AND EFFICACY The subject device was subjected to testing including: - Electrical safety . - Biocompatibility . - Packaging validation . - Shipping & transportation . - . Aging Shelf Life - . Hi-Pot The testing found no issues of safety or effectiveness. {2}------------------------------------------------ ## 9. CONCLUSION Boston Endo Surgical Technologies believes that based on the indications for use, technological characteristics, and comparison to predicate devices the Bipolar Forceps has been shown to be substantially equivalent to the predicate and is safe and effective for its intended use. - . {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service October 25, 2013 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Boston Endo Surgical Technologies, Incorporated Mr. James Rogers Regulatory Affairs, Manager 1146 Barnum Avenue Bridgeport, Connecticut 06610. Re: K131058 Trade/Device Name: Bipolar Forceps Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: August 14, 2013 Received: August 16, 2013 Dear Mr. Rogers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Mr. James Rogers forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Mark Nijelkerson -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): K131058 Device Name: Bipolar Forceps Indications for Use: Tissue grasping and control of bleeding of small vessels by bipolar coagulation. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) an small of the hostin or more and the man the money of the stars and the country of the collection of the collection of the (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) pitally signed by Long H. Chen -A Long H. Chen -A 005FDA, Bu=People, cn=Long H. Chana 0.9.2342.19200300.100.1.1=13300369056 Date: 2013.11.05 06:44.29 -05:00 I for MXM .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (Division Sign-off) Division of Surgical Devices 510(k) Number: K131058 Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%