K131038 · Navilyst Medical, Inc. · LJS · May 9, 2013 · General Hospital
Device Facts
Record ID
K131038
Device Name
NMI PICC MAXIMAL BARRIER NURSING KITS
Applicant
Navilyst Medical, Inc.
Product Code
LJS · General Hospital
Decision Date
May 9, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NMI HPICC III is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for central venous pressure monitoring and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the NMI HPICC III is 6 mL/sec. The NMI PICC III is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, for central venous pressure monitoring and for power injection of contrast media. The Xcela® Hybrid PICC with PASV® Valve Technology is indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the Xcela Hybrid PICC with PASV Valve Technology is 6 mL/sec. The NMI PICC II, NMI PICC, and BSC PICC are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media.
Device Story
PICC Maximal Barrier Nursing Kit comprises a peripherally inserted central catheter (PICC) packaged with procedural aides and maximal barrier precaution devices. Designed to facilitate sterile PICC placement in clinical settings; used by healthcare professionals. Kit includes components based on CDC recommendations to minimize infection risk during insertion. Device provides venous access for therapy, sampling, and monitoring. Output is the established venous access path for clinical intervention. Benefits include standardized, sterile procedural setup to support safe catheter placement.
Clinical Evidence
Bench testing only. No clinical data presented. Performance evaluation based on risk analysis and testing per AAMI/ANSI/ISO 10993-7 (EO residuals) and AAMI/ANSI/ISO 11607-1/2 (packaging validation).
Technological Characteristics
Kit includes PICC catheter and procedural/maximal barrier components. Materials are industry-standard and well-characterized. Sterilization via ethylene oxide (EO). Packaging validated per AAMI/ANSI/ISO 11607-1/2. No electronic or software components.
Indications for Use
Indicated for patients requiring short- or long-term peripheral access to the central venous system for intravenous therapy, including fluid/medication/nutrient administration, blood sampling, central venous pressure monitoring, and power injection of contrast media.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
K131038
510(k) Premarket Notification PICC Maximal Barrier Nursing Kit
#### MAY 9 2013
Navilyst
Medical
26 Forest Street Marlborough. MA 01752 Tel 508,658.7990
www.navilystmedical.com
# 510(k) Summary for the PICC Maximal Barrier Nursing Kit
or
Date prepared: 11 April 2013
A. Sponsor Navilyst Medical, Inc. 26 Forest Street
Marlborough, MA 01752
#### B. Contact
Marion W. Gordon Sr. Project Manager Global Regulatory Affairs Phone: 508-658-7942
#### C. Device Name
Trade Name: Common/Usual name: Classification Name:
Classification Panel:
#### D. Predicate Device
Common/Usual name: Classification Name:
Premarket Notification
Lorraine M. Hanley Vice President Global Regulatory Affairs Phone: 508-494-1129
PICC Maximal Barrier Nursing Kits Peripherally Inserted Central Catheter (PICC) Catheter, Intravascular Therapeutic, short and longterm greater than 30 days 21 CFR §880.5970, Class II General Hospital Device Panel
Peripherally Inserted Central Catheter (PICC) Catheter, Intravascular Therapeutic, short and longterm greater than 30 days 21 CFR §880.5970, Class II K122882, K121089, K111906, K093366, K091261, K070002
# E. Device Description
The PICC Maximal Barrier Nursing Kit is a packaging configuration containing a specified NMI PICC, along with (1) procedural aides typically used for PICC placement and (2) maximal barrier precaution devices based upon recommendations of Center of Disease Control and Prevention (CDC).
{1}------------------------------------------------
# F. Intended Uses
#### · PICC Maximal Sterile Barrier Kit with NMI HPICC III
# NON-VALVED VERSION
The NMI HPICC III is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for central venous pressure monitoring and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the NMI HPICC III is 6 mL/sec.
### · PICC Maximal Sterile Barrier Kit with NMI PICC III
# NON-VALVED VERSION
The NMI PICC III is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids. medications and nutrients, the sampling of blood, for central venous pressure monitoring and for power injection of contrast media.
Maximum Power Injection Flow Rate:
- 4F Single Lumen/55 cm 3.5 mL/sec .
- 5F Single Lumen/55 cm 5 mL/sec ◆
- 5F Dual Lumen/55 cm 4 mL/sec ●
- 6F Dual Lumen/55 cm 5 mL/sec .
#### VALVED VERSION
The NMI PICC III is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media.
Maximum Power Injection Flow Rate:
- . 3F Single Lumen/55 cm 1 mL/sec
- 4F Single Lumen/55 cm 3.5 mL/sec ●
- 5F Single Lumen/55 cm 5 mL/sec .
- ◆ 5F Dual Lumen/55 cm 4 mL/sec
- 6F Dual Lumen/55 cm 5 mL/sec .
- . 6F Triple Lumen/55 – 6 mL/sec
- · PICC Maximal Sterile Barrier Kit with Xcela® Hybrid with PASV® Valve Technology:
for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the Xcela Hybrid PICC with PASV Valve Technology is 6 mL/sec.
{2}------------------------------------------------
· PICC Maximal Sterile Barrier Kit with NMI PICC II; or with NMI PICC; or with BSC PICC:
for short or long-term peripheral access to the central venous system for intravenous therapy. including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media.
# G. Summary of Similarities and Differences in Technological Characteristics and Performance
## Similarities
The proposed PICC Maximal Barrier Nursing Kit contains one of the identified predicate PICCs packaged with a variety of procedural aide componentry typically used during PICC placement. The proposed PICC indications for use, technological characteristics, materials and operating principles are identical.
#### Differences
The proposed packaging configuration differs from the predicate PICC kit packaging in order to contain a selection of procedural aides used in PICC placement including those identified as maximal barrier precaution devices. All packaging is manufactured from packaging materials that are well characterized and commonly used in the medical industry.
#### H. Performance
- . No performance standards have been established under section 514 of the Food, Drug and Cosmetic Act for these devices. The performance evaluation of the PICC Maximal Barrier Nursing Kit was conducted based upon a risk analysis and included testing conducted in accordance with the following national/international standards and FDA guidance documents:
- . A AMI/ANSI/ISO 10993-7 Biological evaluation of medical devices-Part 7: Ethylene oxide sterilization residuals (2008)
- AAMI/ANSI/ISO 11607-1 Packaging for terminally sterilized medical devices-Part 1: ◆ Requirements for materials, sterile barrier systems, and packaging systems (2006)
- . A AMI/ANSI/ISO 11607-2 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes (2006)
- Deciding When to Submit a 510(k) for a Change to an Existing Device, 10 January 1997 .
- Convenience Kits, Interim Regulatory Guidance: 20 May 1997 �
- Sterilized Convenience Kits for Clinical and Surgical Use: 7 January 2002 .
- Bundling Multiple Devices or Multiple Indications in a Single Submission: 22 June 2007 ●
#### Safety and Performance Testing l.
Tests:
The successful results of the following key tests demonstrate that the proposed PICC Maximal Barrier Nursing Kit has met the pre-determined acceptance criteria applicable to the safe use of the devices.
1. Packaging Standards Testing
2. EO Sterilization Testing
#### J. Conclusion
Results of testing according to recognized standards and in consideration to the responses posed in FDA's Guidance on the CDRH Premarket Notification Review Program, 510(k) Decision Making Tree, the proposed PICC Maximal Barrier Nursing Kit is determined to be substantially equivalent to the predicate NMI PICCs.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The text is in all caps and appears to be in a sans-serif font.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
May 9, 2013
Ms. Marion W. Gordon Senior Project Manager Navilyst Medical, Incorporated 26 Forest Street MARLBOROUGH, MA 01752
Re: K131038
Trade/Device Name: PICC Maximal Barrier Nursing Kit Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: April 11, 2013 Received: April 17, 2013
Dear Ms. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Gordon
Please be advised that FDA's issuance of a substantial equivalence determination does not meanthat FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/4/Picture/8 description: The image contains a handwritten note with the word "for" at the top left. The note also includes the words "with" and what appears to be a name or word that is difficult to decipher due to the handwriting style. There are also some decorative swirls and lines around the text, adding a personal touch to the note.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital,
. Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if Known):
Device Name:
PICC Maximal Barrier Nursing Kit
with NMI PICC III
Indications for Use:
### NON-VALVED VERSION
The NMI PICC III is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, for central venous pressure monitoring and for power injection of contrast media.
Maximum Power Injection Flow Rate:
- 4F Single Lumen/55 cm 3.5 mL/sec .
- 5F Single Lumen/55 cm 5 mL/sec .
- 5F Dual Lumen/55 cm 4 mL/sec .
- 6F Dual Lumen/55 cm 5 mL/sec .
#### VALVED VERSION
The NMI PICC III with PASV Vale Technology is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media.
Maximum Power Injection Flow Rate:
- 3F Single Lumen/55 cm 1 mL/sec ●
- 4F Single Lumen/55 cm 3.5 mL/sec .
- 5F Single Lumen/55 cm 5 mL/sec .
- 5F Dual Lumen/55 cm 4 mL/sec .
- 6F Dual Lumen/55 cm 5 mL/sec .
- 6F Triple Lumen/55 6 mL/sec .
Prescription Use (21 CFR 801 Subpart D) And/Or
×
AND/OR Over-The-Counter Use: (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Kathleen E. Fitzgerald 200 Concurrence of CDRH, Office of Device Evaluation (ODE)
> (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
{6}------------------------------------------------
### Indications for Use
510(k) Number (if Known):
Device Name:
PICC Maximal Barrier Nursing Kit
NMI HPICC III with i
Indications for Use:
NON-VALVED VERSION
The NMI HPICC III is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for central venous pressure monitoring and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the NMI HPICC III is 6 mL/sec.
Prescription Use (21 CFR 801 Subpart D) × And/Or
AND/OR Over-The-Counter Use: (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Kathle 300.100.1-1=0010093027. Fitzger
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices
510(k) Number: K131038
4-2
D
{7}------------------------------------------------
510(k) Premarket Notification PICC Maximal Barrier Nursing Kit
# Indications for Use
510(k) Number (if Known):
Device Name:
PICC Maximal Barrier Nursing Kit
K131038
Xcela® Hybrid PICC with PASV® Valve Technology with
Indications for Use:
for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the Xcela Hybrid PICC with PASV Valve Technology is 6 ml /sec.
Prescription Use (21 CFR 801 Subpart D)
"
区 And/Or AND/OR Over-The-Counter Use: (21 CFR 801 Subpart C)
□
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/7/Picture/14 description: The image shows the letters FDA in a stylized, outlined font. The letters are interconnected and have a three-dimensional appearance due to the multiple outlines. The overall design is simple and recognizable, likely representing the Food and Drug Administration.
Richard C. Chapman 2013.05.07 12:39:02 -04'00'
(Division Sign-Off) (Division Sign-Cor)
Division of Anesthesiology. General Hospital Division of Anesian Devices
510(k) Number: K131038
{8}------------------------------------------------
# 510(k) Premarket Notification PICC Maximal Barrier Nursing Kit
# Indications for Use
| 510(k) Number (if Known): | |
|---------------------------|------------------------------------------------------------|
| Device Name: | PICC Maximal Barrier Nursing Kit<br>with NMI PICC II<br>Or |
| | PICC Maximal Barrier Nursing Kit<br>with NMI PICC<br>Or |
| | PICC Maximal Barrier Nursing Kit<br>with BSC PICC |
Indications for Use:
for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media.
Prescription Use (21 CFR 801 Subpart D) ಸ And/Or AND/OR Over-The-Counter Use: (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kathleen E
Fitzgerald
Digitally signed by Kathleen E. Fitzgerald
DN: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People,
0.9.2342.19200300.100.1.1-0010093027,
cn=Kathleen E. Fitzgerald
Date: 2013.05.09 15:27:31 -04'00'
(Division Sign-Off) (Division of Anesthesiology, General Hospital Intection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.